Calibration
Ensure that the AED equipped with its magnetic base is in working order, putting
the AED status sensor on its base requires a OK AED for calibration purposes.
Unplug then plug again the AED status sensor from its
socket. The LED will blink red/green during three seconds.
During this time, you must put the AED Status Sensor on
its base.
Check the AED status sensor LED :
o If the LED is blinking orange, calibration has
failed. (Sensor was not put onto the AED
into the 3 first seconds) In that case, pull the
AED from its base and restart the calibration
procedure.
o If the status LED is orange, then
after about a minute maximum
green, procedure has succeeded,
the sensor is now operational.
Note :
In order to calibrate the sensor, the AED must show an operational status.
The AED status sensor calibration is to be done at installation or if changing
the AED. Once sensor is calibrated, you can remove it from its support,
Calibration settings are memorized in the AED status sensor, even when
power is lost.