Aqua Lung EBIS-4500 User manual

Type
User manual

This manual is also suitable for

User's Manual
Rev. 04/17
EBIS - 3000 & 4500
(Emergency Breathing Ination System)
2EBIS - 3000 & 4500 User's Manual
COPYRIGHT NOTICE
This user's manual is copyrighted, all rights reserved. It may not, in whole or in
part, be copied, photocopied, reproduced, translated, or reduced to any electronic
medium or machine readable form without prior consent in writing from Aqua
Lung America, Inc.
©2017 Aqua Lung International
EBIS - 3000 & 4500 Users Manual PN 103689
You can contact a Technical Advisor via e-mail at:
grubin@aqualung.com
TRADEMARK NOTICE
Aqua Lung® is a registered trademark of
Aqua Lung America, Inc.
Warnings, Cautions and Notes:
Pay special attention to information provided in Warnings, Cautions
and Notes that are accompanied by one of these symbols:
A NOTE is used to emphasize important points, tips and reminders.
A WARNING indicates a procedure or situation that, if not avoided,
could result in serious injury or death to the user.
A CAUTION indicates any situation or technique that could
cause damage to the product and could subsequently result in
injury to the user.
3
General Precautions and Warnings ............................................................ 4
Product Description ...................................................................................... 6
Preparation and Setup .................................................................................. 7
General Filling Procedures ...................................................................... 7
Filling Methods ........................................................................................ 9
Filling with a SCUBA Fill Adapter ............................................................ 9
Filling with a Compressor Fill Adapter ................................................... 12
Filling with a HP Fill Adapter ................................................................. 15
Additional Filling Options ....................................................................... 17
Vest Conguration ................................................................................. 18
Pre-Issue Inspection ................................................................................... 20
Pre-Dive Inspection .................................................................................... 23
Post-Dive Inspection .................................................................................. 24
Care and Maintenance ................................................................................ 24
Inspection and Service ............................................................................... 26
Warranty Information ................................................................................. 26
Technical Specications ............................................................................ 27
EBIS Accessories ....................................................................................... 29
Notes ............................................................................................................ 30
TABLE OF CONTENTS
4EBIS - 3000 & 4500 User's Manual
WARNING: Before using EBIS, it is important to receive in-water training
which simulates proper use of this system. Use of the EBIS without proper
training is dangerous and can result in serious injury or death.
WARNING: Visual inspection and factor y prescribed ser vice for the EBIS must be
performed annually by a factory trained service technician. Repair, service and
visual inspection must not be attempted by untrained or unqualied personnel.
WARNING: DO NOT attempt to overll the EBIS beyond its maximum
ll pressure at 70ºF/ 21ºC. Doing so may seriously weaken the safety
plug assembly and cylinder, resulting in serious injury or death.
WARNING: The EBIS is designated compatible for use only with normal,
atmospheric, compressed air (21% oxygen and 79% nitrogen by volume).
DO NOT attempt to ll with other gases, including pure oxygen or air which
has been enriched with oxygen exceeding 21% in content. Failure to obser ve
this warning may result in serious injury or death due to re and explosion
or the serious deterioration and failure of the equipment.
WARNING: DO NOT ll or use the EBIS if it has been exposed to extreme
heat exceeding 250º F / 121º C or open ame. Instead, discharge the cylinder
completely and return it to a factory trained service technician for inspection
and possible hydrostatic re-testing.
GENERAL PRECAUTIONS AND WARNINGS
WARNING: DO NOT apply any type of aerosol spray to the EBIS. Doing
so may cause permanent damage to plastic components.
WARNING: Before using the EBIS, you must have successfully received
training and certication in the technique of SCUBA diving from a Military
or government operated diving school (or any recognized certication
agency). Use of SCUBA equipment by uncertied or untrained persons
is dangerous and can result in serious injury or death.
WARNING: Due to its limited air volume, EBIS is intended to be used for
making emergency ascents from a depth no greater than 45 ft / 13.5 m,
minor buoyancy adjustments or emergency otation during maritime
diving or surface operations.
5
WARNING: It is important to ensure that the EBIS is always pressurized
whenever it is submerged in order to prevent the entrance of water into
the system. If the system has been completely emptied of air underwater,
it is important to return the EBIS as soon as possible to a factory trained
service technician for visual inspection and any necessary service before
attempting to rell it.
WARNING: It is important to ll the EBIS only with dry, ltered air with a
water vapor content that does not exceed - 65°F / - 53ºC dew point. Excess
water vapor in the air can cause ice to form inside the EBIS and interfere
with the operation of the system at colder temperatures.
NOTE:
When instructed to remove, unscrew or loosen a part, turn the part
counter-clockwise.
When instructed to install, screw, or tighten a part, turn the part clockwise.
NOTE:
When instructed to depressurize the system, close the handwheel and
depress the second stage purge button for 5 seconds and release for 5 seconds.
Repeat this procedure until the system is depressurized.
NOTE:
When instructed to “OPEN” the handwheel, turn the handwheel
counter-clockwise. The red indicator ring will not be visible in the
handwheel window, this indicates the system in the ONposition.
When instructed to CLOSE the handwheel, turn the handwheel clockwise.
The red indicator ring will be visible in the handwheel window, this indicates
the system is in the“OFFposition.
NOTE:
When instructed to disconnect or reconnect the inator hose to the
QD tting, ensure the system is off and depressurized.
Disconnect: Depress the quick disconnect (QD) button on the QD tting to
remove the inator hose.
Reconnect: Insert the inator hose into the QD tting, making sure the quick
disconnect (QD) button “Pops Up”. Ensure the inator hose is rmly attached
prior to pressurizing the system.
WARNING: DO NOT apply any type of petroleum-based lubricant, such
as household oil or motor oil to any part of the EBIS. The EBIS does
not require any lubrication under normal circumstances, except during
annual service performed by a factory trained service technician.
6EBIS - 3000 & 4500 User's Manual
PRODUCT DESCRIPTION
The Emergency Breathing Ination System (EBIS) is specically designed for
use on a variety of plate carriers used for maritime diving and surface operations.
A light weight cylinder with integrated regulator provides a source of breathable
air for emergency situations that may occur in maritime environments. A low
prole inator w/ QD hose is incorporated into the plate carriers bladder system,
allowing the operator to make minor buoyancy adjustments underwater and
achieve emergency buoyancy on the surface.
Figure 1
EBIS Component Identication
Braided MP Hose
First Stage
On / Off Knob
Second Stage
Fill Port
Aluminum Cylinder
or
Carbon Fiber Composite Cylinder
Safety Plug Assy
360º Dual Swivel
Dial Pressure Gauge
(Located on backside)
QD Fitting
Inator Hose
Inator
Mouthpiece
Cover
7
The purpose of this manual is to familiarize personnel with the correct setup,
lling, inspection and maintenance of the EBIS.
The EBIS is packaged fully assembled, ready to use after it has been lled with
air and congured to a vest. Before using, it is very important to carefully read
and understand the procedures outlined in this manual.
PREPARATION AND SETUP
General Filling Procedures
NOTE: The average duration of air supply listed in the Technical Specications
section of this manual is based upon the maximum ll pressure as marked
on the cylinder. It is strongly recommended that the cylinder be lled to its
maximum ll pressure (cold ll), in order to provide maximum breathing volume.
1. Before attempting to ll the EBIS, ensure that the ll adapter and rst stage
are completely dry – especially in the area surrounding the ll port.
2. The EBIS is congured with a 3000 PSI (207 BAR) aluminum cylinder or
4500 PSI (310 BAR) carbon ber composite cylinder. Examine the cylinder
markings to determine the maximum ll pressure. Aluminum cylinders must
have a current visual inspection. Carbon ber composite cylinders must have
a current visual inspection and hydrostatic test date (Fig. 2).
NOTE:
Aluminum cylinders do not require hydrostatic re-testing. Carbon
ber composite cylinders require hydrostatic re-testing every 5 years (Fig. 2).
Month & Year
of Manufacture
(Hydro Date)
Maximum Fill
Pressure
Figure 2
Maximum Fill Pressure
Month & Year of Manufacture (Hydro Date)
8EBIS - 3000 & 4500 User's Manual
WARNING: DO NOT attempt to ll the EBIS if the cylinder markings indicate
that it is assembled with a non standard cylinder rated for a different ll
pressure (See Fig. 2). Doing so may result in rupture or explosion in the
event of re or overlling. Instead, immediately return the unit to a factory
trained service technician and do not use under any circumstances.
Figure 5
4. Unscrew the ll port cap from the
ll port (Fig. 5).
5. Closely inspect the ll port opening
to ensure that no debris, residue or
moisture is present.
CAUTION: If moisture is found to be present inside the ll port opening, this
indicates that water may have entered the rst stage and cylinder. DO NOT
ll or attempt to use the EBIS until it has received a complete inspection
and any required service by a factory trained service technician.
Figure 3
"OFF"
POSITION
Red Indicator
Ring Visible
3. Check to ensure the handwheel is closed and the system is depressurized (Fig.
3). Disconnect the inator hose from the QD tting (Fig. 4). Put the ination system to
the side, it will be reassembled during the Vest Conguration section of the manual.
Figure 4
QD Fitting
Button
WARNING: DO NOT attempt to loosen or remove any components of
the EBIS other than instructed. Doing so could cause a dangerous
malfunction which could result in serious injury or death.
9
Filling Methods
Filling with a SCUBA Fill Adapter
NOTE:
The EBIS system does not include a SCUBA ll adapter. This adapter
(PN 108325) may be purchased separately (Fig. 6).
1. Supply (SCUBA) cylinder with SCUBA ll adapter (PN 108325).
2. Compressor with compressor ll adapter (PN M100656).
3. Compressor with HP ll adapter (PN 102865).
Figure 6
O-Ring
CAUTION: The dial pressure gauge is for reference only. Use the ll
system gauge to indicate accurate cylinder pressure.
CAUTION: Do not attempt to ll the EBIS directly from a supply cylinder
unless you have received the necessary training and authorization to
do so. If done incorrectly, this procedure poses certain hazards which
may cause severe injury or death.
WARNING: DO NOT attempt to use the SCUBA ll adapter to ll the EBIS
from an air supply where the regulated pressure exceeds 3000 PSI (207
BAR). Doing so could cause a dangerous malfunction to the ll adapter or
weaken the cylinder and safety plug, resulting in serious injury or death.
10 EBIS - 3000 & 4500 User's Manual
6. While supporting the EBIS with one
hand, place the yoke of the ll adapter
over the cylinder valve to align the inlet
tting ush against the valve o-ring.
Screw the ll adapter yoke screw into
the small dimple on the backside of
the cylinder valve until finger-tight
(Fig. 8).
7. Close the bleed valve screw on the
ll adapter (Fig. 8).
Figure 8
1. Using a calibrated pressure gauge, check the supply cylinder to ensure that
it contains 3000 PSI (207 BAR). It is very important to ensure that the EBIS is
lled to its total capacity but not overlled.
3. Screw the threaded male tting on
the ll adapter into the ll port. DO
NOT apply a wrench or otherwise
overtighten the ll adapter (Fig. 7).
4. Loosen the ll adapter yoke screw
as needed so that the dust cap can be
removed from the inlet tting.
5. Inspect the supply cylinder o-ring
to make sure it is present, in good
condition and seated evenly in the
supply cylinder valve.
2. Remove the protector cap from the threaded male tting of the ll adapter.
Inspect the ll adapter to ensure that the o-ring is present and seated evenly at
the base of the threads. Check the threads of the male tting, making sure they
are clean and not damaged.
Figure 7
Bleed Valve Screw
11
NOTE:
The EBIS will ll with the handwheel in the “OFFposition (red
indicator ring visible). In this position the dial pressure gauge will not
register pressure. This is not recommended.
8. While holding the rst stage and
ll adapter secure, slowly open the
handwheel to pressurize the system
(Fig. 9).
NOTE:
Always ll the cylinder as slowly as possible (do not exceed 300 PSI / 20.7
BAR per minute) by turning the handwheel of the supply valve slowly to control the
rate of ll. Rapid lling will generate heat and will result in an incomplete ll after the
cylinder cools. If the cylinder is warm to the touch afterward, the ll rate was too rapid.
Figure 9
"ON"
POSITION
Red Indicator
Ring Not Visible
9. While supporting the cylinder with one hand, very slowly open the handwheel
on the supply cylinder to begin lling (do not exceed a ll rate of 300 PSI / 20.7
BAR per minute). Make sure the dial pressure gauge shows that the cylinder
is lling properly.
10. When air can no longer be heard owing from the supply cylinder into the
EBIS, completely open the supply cylinder valve. Verify the cylinder has been
lled to its maximum capacity as marked.
11. While holding the rst stage and ll adapter secure, close the handwheel on the
supply cylinder. Close the handwheel on the EBIS and depressurize the system.
12. Open the bleed valve screw on the ll adapter to relieve the line pressure.
13. While supporting the EBIS cylinder with one hand, loosen the yoke screw
on the ll adapter and remove it from the supply cylinder valve.
14. Unscrew the ll adapter from the ll port. Replace the dust cap over the
inlet tting and tighten the yoke screw nger-tight. Replace protector cap on the
threaded male tting of the ll adapter.
15. Return the ll adapter to its storage location.
16. Check that the o-ring on the ll port cap is present, not damaged and seated
evenly at the base of the cap. Thread the ll port cap back into the ll port until
nger-tight.
12 EBIS - 3000 & 4500 User's Manual
Filling with a Compressor Fill Adapter
NOTE:
The EBIS system does not include a compressor ll adapter. This
adapter (PN M100656) may be purchased separately (Fig. 10).
O-Ring
O-Ring
Figure 10
CAUTION: Do not attempt to ll the EBIS directly from a compressed
air lling station unless you have received the necessary training and
authorization to do so. If done incorrectly, this procedure poses certain
hazards which may cause severe injury or death.
CAUTION: The dial pressure gauge is for reference only. Use the ll
system gauge to indicate accurate cylinder pressure.
WARNING: DO NOT attempt to use the compressor ll adapter to ll the
EBIS from a compressor or air supply where the regulated pressure exceeds
3000 PSI (207 BAR). Doing so could cause a dangerous malfunction to the
ll adapter or weaken the cylinder and safety plug, resulting in serious
injury or death.
13
Figure 11
Bleed Valve Screw
1. Remove the protector cap from the threaded male tting of the ll adapter. Inspect
the ll adapter to ensure that the o-ring is present, not damaged and seated evenly
at the base of the threads. Check the threads of the male tting, making sure they
are clean and not damaged. Inspect the o-ring in the groove on the ll block end
of the adapter, making sure it is present, not damaged and seated evenly.
2. Screw the threaded male tting on the ll adapter into the ll port of the rst stage
until nger-tight. DO NOT apply a wrench or otherwise overtighten the ll adapter.
3. Loosen the yoke screw on the compressor ll yoke as needed so that the
yoke can be placed over the block of the ll adapter.
4. While supporting the EBIS with one
hand, place the yoke over the block of
the ll adapter to align the inlet tting
ush against the ll block o-ring. Screw
the yoke screw into the small dimple on
the backside of the ll adapter block
until nger-tight (Fig. 11).
5. Close the bleed valve screw on the
compressor ll yoke (Fig. 11).
NOTE:
The EBIS will ll with the handwheel in the “OFFposition (red
indicator ring visible). In this position the dial pressure gauge will not
register PSI. This is not recommended.
Figure 12
6. While holding the rst stage and
ll adapter secure, slowly open the
handwheel to pressurize the system
(Fig. 12).
"ON"
POSITION
Red Indicator
Ring Not Visible
14 EBIS - 3000 & 4500 User's Manual
NOTE:
Always ll the cylinder as slowly as possible (do not exceed 300 PSI
/ 20.7 BAR per minute) by turning the handwheel of the supply valve slowly
to control the rate of ll. Rapid lling will generate heat and will result in an
incomplete ll after the cylinder cools. If the cylinder is warm to the touch
afterward, the ll rate was too rapid.
7. While supporting the cylinder with one hand, very slowly open the
compressor valve to begin lling (do not exceed a ll rate of 300 PSI /
20.7 BAR per minute). Make sure the dial pressure gauge shows that the
cylinder is lling properly.
8. When air can no longer be heard owing from the compressor ll hose into
the EBIS, completely open the compressor valve. Verify the cylinder has been
lled to its maximum capacity as marked.
9. While holding the rst stage and ll adapter secure, close the handwheel
on the compressor valve. Close the handwheel on the EBIS and depressurize
the system.
10. Open the bleed valve screw on the compressor ll yoke to relieve the line
pressure.
11. While holding the EBIS cylinder, loosen the yoke screw on the compressor
ll yoke and remove it from the ll adapter. Return the compressor ll yoke to
its storage location.
12. Unscrew the ll adapter from the ll port. Replace protector cap on the
threaded male tting of the ll adapter.
13. Return the ll adapter to its storage location.
14. Check that the o-ring on the ll port cap is present, not damaged and seated
evenly at the base of the cap. Thread the ll port cap back into the ll port until
nger-tight.
15
Filling with a HP Fill Adapter
NOTE:
The EBIS system does not include a HP ll adapter. This adapter
(PN 102865) may be purchased separately (Fig. 13).
O-Ring
CAUTION: Do not attempt to ll the EBIS directly from a compressed
air lling station unless you have received the necessary training and
authorization to do so. If done incorrectly, this procedure poses certain
hazards which may cause severe injury or death.
CAUTION: The dial pressure gauge is for reference only. Use the ll
system gauge to indicate accurate cylinder pressure.
Figure 13
WARNING: DO NOT attempt to use the HP ll adapter to ll the EBIS
from a compressor or air supply where the regulated pressure exceeds
the cylinders maximum ll pressure. Doing so could cause a dangerous
malfunction to the ll adapter or weaken the cylinder and safety plug,
resulting in serious injury or death.
16 EBIS - 3000 & 4500 User's Manual
1. Remove the protector cap from the threaded male tting on the ll adapter.
Inspect the ll adapter to ensure that the o-ring is present, not damaged and
seated evenly at the base of the threads. Check the threads of the male tting,
making sure they are clean and not damaged.
Figure 14
2. Insert the threaded male tting of the
ll adapter into the ll port on the rst
stage. Screw the adapter nut into the
ll port until nger-tight. DO NOT apply
a wrench or otherwise overtighten the
ll adapter (Fig. 14).
3. Close the bleed valve screw on the
ll adapter (Fig. 14).
NOTE:
The EBIS will ll with the handwheel in the “OFFposition (red
indicator ring visible). In this position the dial pressure gauge will not
register pressure. This is not recommended.
Figure 15
4. While holding the rst stage and
ll adapter secure, slowly open the
handwheel to pressurize the system
(Fig. 15).
Bleed Valve
Screw
Adapter
Nut
"ON"
POSITION
Red Indicator
Ring Not Visible
17
5. While supporting the cylinder with one hand, very slowly open the
compressor valve to begin filling (do not exceed a fill rate of 300 PSI /
20.7 BAR per minute). Make sure the dial pressure gauge shows that the
cylinder is filling properly.
6. When air can no longer be heard owing from the compressor ll hose into
the EBIS, completely open the compressor valve. Verify the cylinder has been
lled to its maximum capacity as marked.
7. While holding the rst stage and ll adapter secure, close the handwheel
on the compressor valve. Close the handwheel on the EBIS and depressurize
the system.
8. Open the bleed valve screw on the ll adapter to relieve the line pressure.
9. While supporting the EBIS cylinder with one hand, unscrew the adapter nut
and remove it from the ll port. Replace protector cap on the threaded male
tting of the ll adapter.
10. Return the ll adapter to its storage location.
11. Check that the o-ring on the ll port cap is present, not damaged and seated
evenly at the base of the cap. Thread the ll port cap back into the ll port until
nger-tight.
NOTE:
Always ll the cylinder as slowly as possible (do not exceed 300 PSI
/ 20.7 BAR per minute) by turning the handwheel of the supply valve slowly
to control the rate of ll. Rapid lling will generate heat and will result in an
incomplete ll after the cylinder cools. If the cylinder is warm to the touch
afterward, the ll rate was too rapid.
Additional Filling Options
Several lling options for the EBIS - 3000 are available through Capewell /
Aerial Systems LLC. For additional information please contact:
Capewell / Aerial LLC
4298 JEB Stuart Highway
Meadow of Dan, Va 24120
Phone: 276-952-2006
Website: capewellaerialsystems.com
18 EBIS - 3000 & 4500 User's Manual
Vest Conguration
1. Disconnect the upper and lower inator tubes from the “T” connector on the
vest (Fig. 16). Remove the upper inator tube from the hose retainer (Fig. 17).
Figure 16 Figure 17
Upper Tube
Lower Tube
3. Insert the inator hose through the
bottom and out the top of the hose
retainer. Insert the upper inator tube
through the top and out the bottom of
the hose retainer (Fig. 19).
Figure 19
Figure 18
Hose Retainer
2. Check the handwheel is closed and
the system depressurized. Disconnect
the inator hose from the QD tting
(Fig. 18) if this was not done during
the Preparation and Set-up section
of the manual.
"OFF"
Position
19
4. Connect both inator tubes (upper and
lower) to the barb ttings (top and bottom)
on the inator mechanism. Secure the
inator tubes in place using the supplied
clamp straps. Trim the excess clamp strap
using diagonal pliers (Fig. 20).
Figure 20
5. Reconnect the inator hose to the QD tting. Insert the EBIS cylinder into the
pocket on the rear, left side of the vest with the dual swivel face up. Connect
the mouthpiece cover with a clamp strap to a user preferred location on the
vest. Trim the excess clamp strap using diagonal pliers. Fit the second stage
mouthpiece into the mouthpiece cover (Fig. 21).
Figure 21
NOTE:
The inator button can face inward or outward, orientation is entirely
up to user preference. Facing the inator button inward will prevent the inator
button from being pressed accidentally.
20 EBIS - 3000 & 4500 User's Manual
PRE-ISSUE INSPECTION
Before each use, the system must be given a thorough visual inspection and
functional test. NEVER use an EBIS which shows signs of damage, leakage
or substandard performance until it has received inspection and service from
a factory trained service technician.
2. Ensure the medium pressure hose is securely connected to the second stage. Inspect
the hose and hose ttings to ensure that there is no blistering, cuts, or corrosion present.
3. Verify the inator hose is securely connected to the inator mechanism and the heat
shrink seal is in place and not damaged. Check the inator hose is rmly attached to
the QD tting. Inspect the entire length of the inator hose, along with the upper and
lower inator tubes to ensure there is no blistering, cuts or damage present.
4. Visually inspect the entire system for any external damage, such as dents, gouges,
cracks or external corrosion.
5. Conrm the handwheel is closed and the system depressurized. Inspect the
dial pressure gauge to ensure that it reads “0 PSI / BAR” (Fig. 23).
WARNING: DO NOT issue an EBIS if the safety wire is missing or damaged.
An inspection must be performed and the safety wire replaced prior to issue.
1. Carefully inspect the first stage to
make sure all external components (dial
pressure gauge, dual swivel, safety plug,
etc) are properly secured. Ensure the
safety wire is rmly attached and in good
condition (Fig. 22).
Safety Wire
Figure 22
Figure 23
0 PSI / BAR 0 PSI / BAR
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32

Aqua Lung EBIS-4500 User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI