Invacare HomeFill HF2ECE9, HomeFill HF2ECE6 User manual

  • Hello! I am an AI chatbot trained to assist you with the Invacare HomeFill HF2ECE9 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Invacare®HomeFill®System
HF2ECE9,HF2ECE6
ENOxygenConservingCylinders
UserManual................................3
FRBouteilledeconservationd’oxygène
Manueld'utilisation..........................23
DESauerstoffAufbewahrungs-Zylinder
Gebrauchsanweisung.........................45
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuturereference.
©2013Invacare®Corporation
Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithout
priorwrittenpermissionfromInvacare.Trademarksareidentifiedby™and®.Alltrademarksare
ownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted.
WD-40and3–IN-ONEareregisteredtrademarksoftheWD-40Company.
Contents
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuture
reference.
1General......................................4
1.1IntendedUse..............................4
1.2Features.................................4
2Safety........................................5
2.1Symbols..................................5
2.2GeneralGuidelines..........................5
2.3Disposal.................................7
3Setup........................................8
3.1ReceivingInspection.........................8
3.2CylinderPrefillInspection.....................8
3.3TheHydrostaticTestingDate..................9
3.4ExternalExamination........................9
4Usage........................................11
4.1Connecting/DisconnectingtheCylinderto/from
theCompressor...........................11
ConnectingtheCylindertothe
Compressor............................11
DisconnectingtheCylinderfromthe
Compressor............................13
4.2CylinderOperation.........................14
5Maintenance..................................17
5.1Maintenance..............................17
5.2ReplacementPartsAccessories..............17
6Troubleshooting...............................18
6.1Troubleshooting...........................18
7TechnicalData................................19
7.1TypicalProductParameters...................19
HomeFill®System
1General
1.1IntendedUse
Thiscylinderisintendedtostore,regulate,anddeliversupplemental
oxygentopatientswhomayhavedifficultyextractingoxygenfrom
theairtheybreathe.Itisforpatientswhowouldnormallyreceivethe
oxygenviaanasalcannula.ThedevicedeliversUSP93±3%oxygenat
anumberofdifferentflowsettings.Itisintendedtobeusedasan
oxygendeliverydeviceandmayincreasethedryingoutofairways.
1.2Features
4
1176148-A
Safety
2Safety
2.1Symbols
Signalwordsareusedinthismanualandapplytohazardsorunsafe
practiceswhichcouldresultinpersonalinjuryorpropertydamage.
Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious
injury.
WARNING!
Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious
injury.
CAUTION!
Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage
orminorinjuryorboth.
Givesusefultips,recommendationsandinformationfor
efficient,trouble-freeuse.
ThisproductcomplieswithDirective93/42/EEC
concerningmedicaldevices.
ThelaunchdateofthisproductisstatedintheEC
declarationofconformity.
2.2GeneralGuidelines
WARNING!
Thissectioncontainsimportantinformationforthe
safeoperationanduseofthisproduct.
DONOTUSETHISPRODUCTORANYAVAILABLE
OPTIONALEQUIPMENTWITHOUTFIRST
COMPLETELYREADINGANDUNDERSTANDING
THESEINSTRUCTIONSANDANYADDITIONAL
INSTRUCTIONALMATERIALSUCHASUSER
MANUALS,SERVICEMANUALSORINSTRUCTION
SHEETSSUPPLIEDWITHTHISPRODUCTOR
OPTIONALEQUIPMENT.IFYOUAREUNABLE
TOUNDERSTANDTHEWARNINGS,CAUTIONS
ORINSTRUCTIONS,CONTACTAHEALTHCARE
PROFESSIONAL,DEALERORTECHNICAL
PERSONNELBEFOREATTEMPTINGTOUSETHIS
EQUIPMENT-OTHERWISE,INJURYORDAMAGE
MAYOCCUR.
GENERALGUIDELINEScontainsimportant
informationforthesafeoperationanduseofthis
product.
ACCESSORIESWARNING
Invacareproductsarespecificallydesignedand
manufacturedforuseinconjunctionwithInvacare
accessories.Accessoriesdesignedbyother
manufacturershavenotbeentestedbyInvacareand
arenotrecommendedforusewithInvacareproducts.
1176148-A5
HomeFill®System
DANGER!
Thisproductisnotintendedasalife-sustainingor
life-supportingdevice.
DONOTSMOKEwhileusingthisdevice.Keepall
matches,lightedcigarettesorothersourcesofignition
outoftheroominwhichthisproductislocatedand
awayfromwhereoxygenisbeingdelivered.
NOSMOKINGsignsshouldbeprominentlydisplayed.
Textilesandothermaterialsthatnormallywouldnot
burnareeasilyignitedandburnwithgreatintensityin
oxygenenrichedair.Failuretoobservethiswarning
canresultinseverefire,propertydamageandcause
physicalinjuryordeath.
Theinformationcontainedinthisdocumentissubject
tochangewithoutnotice.
OperatingInformation
CylinderMUSTbeemptiedofgaspriortoshippingbycourieror
postalservice.
OnlypersonnelinstructedandtrainedintheuseoftheHomeFill
ConservingCylindershouldoperatethisproduct.
Productcontainsmagnetic,ferrousmaterialthatmayaffectthe
resultsofanMRI.
Besureallconnectionsaretightandleakfree.
DONOTusealiquidleakdetectortotestforleaks.
DONOTautoclave.
DONOTgassterilizewithethyleneoxide.
DONOTcleanwitharomatichydrocarbons,including,butnot
limitedtobenzeneandtoluene.
DONOTattempttorepairproduct.
DONOTimmerseproductinanykindofliquid.
Storeproductinacleanareawhennotinuse.
OnlyuseInvacareCorporationcarryingbag.
NEVERsmokeinanareawhereoxygenisbeingadministered.
ALWAYSconfirmprescribeddosebeforeadministeringtopatient
andmonitoronafrequentbasis.
DONOTuseifdirtorcontaminantsarepresentonoraround
cylinder,valve,productorconnectingdevices.
NEVERusenearanytypeofflameorflammable/explosivesubstances,
vapoursoratmosphere.
NOOXYGENisdeliveredwhenthepointerisalignedwith“O”.
DONOTblocktheoutletfittingorkinkthecannulawhenthe
productisinuse,asthismaydamagetheunit.
Theuseofthisdeviceislimitedtotheoxygenpatient.Cylinders
MUSTbeusedonlybytheoxygenpatientandarenottobe
distributedtoanyotherindividualforanypurpose.
ThisproductisNOTtobeusedbypatientswhobreathethrough
theirmouths.
DONOTallowoilfromyourhandsorothersourcestocome
intocontactwiththecompressorcouplingorbottlefittings.These
substancesmaybecomeflammableinthepresenceofoxygenand
causeinjury.Avoidtouchingthebottlefittingwheneverpossible.
Exampleofitemstoavoidarehandcreamsorlotions,cookingoils,
suntanoilsorsunblocksandsimilaroilyproducts.Ifyouusethese
61176148-A
Safety
typesofproductsInvacaresuggestsyouwashyourhandspriorto
usingtheHomeFillunit.Ifthebottlefittingshouldcomeincontact
withsuchsubstancesitshouldbecleanedoffwithadryclothprior
toconnectingittothecompressor.
NEVERoilorlubricatethecompressorcouplerorbottlefittings.
NEVERuseanypenetratingoilorlubricantsuchasWD-40®or
3-IN-ONE®.Ifconnectiondifficultyisexperienced,verifythatthe
bottlefittingisbeinginsertedstraightdownintothecompressor
coupler.Ifcontinueddifficultyisexperienced,contactyoursupplier
forassistance.
DONOTuseahumidifierwiththisproduct.
DONOTallowcylinderstotiporfall.Securegascylinderssothey
cannotfall.Ensurecylinderisalwaysinuprightposition.
DONOTstorecylindersnearsourcesofheatorflame.
Thisproductisequippedwithareliefvalve.Ifyouhearaloudhissing
orpoppingsoundcomingfromtheproduct,discontinueuseof
product,turntheflowselectortothe“O”position,andcontactyour
dealer/HomeCareequipmentprovider.
DONOTusewhilesleepingwithoutconsultingyourhealthcare
provider.
Thisproductisdesignedtooperatewithastandardorhighflow
duallumen,adultoxygennasalcannulawithamaximumlengthof
2.1m(7ft).
Thecannulaisforsinglepatientuseonly.
CannulaMUSTbeinstalledonoutletconnectionpriortouse.
DONOTplacetheunitunderclothingwhileinuse.Whenthis
productisinuseasmallamountofoxygenisvented.Wearingthe
unitunderclothingmaysaturatefabricswithoxygenandcausethem
toburnrapidlyifexposedtosparksorflame.Itmaytakeseveral
hoursforoxygenlevelsinfabricstoreturntonormal.
DONOTuseHomeFillcompressorunit,tofillcylinderoutdoors
oranywherebelow13°C(55°F).
NEVERattempttofillcylinderwithanythingotherthantheHomeFill
compressor.
2.3Disposal
Thisproducthasbeensuppliedfromanenvironmentallyaware
manufacturerthatcomplieswiththeWasteElectricalandElectronic
Equipment(WEEE)Directive2002/96/CE.Thisproductmaycontain
substancesthatcouldbeharmfultotheenvironmentifdisposedofin
places(landfills)thatarenotappropriateaccordingtolegislation.
Followlocalgoverningordinancesandrecyclingplansregarding
disposalofthedeviceorcomponentsnormallyusedinoperation.
Thedevicedoesnotgeneratewasteorresidueinoperation.DO
NOTdisposeofthedeviceinthenormalwastestream.Any
accessoriesnotpartofthedeviceMUSTbehandledinaccordance
withtheindividualproductmarkingfordisposal.
1176148-A7
HomeFill®System
3Setup
3.1ReceivingInspection
Removeproductfrompackageandinspectfordamage.Ifthereisany
damage,DONOTuse.Contactyourdealer/HomeCareequipment
provider.
3.2CylinderPrefillInspection
RefertotheHomeFillowner’smanualforcompressor
operatinginstructions.
DANGER!
Improperfillingorabuseofthiscylinder,orfailureto
followthesesafetyinstructionsmaycauseserious
bodilyinjuryordeath.Highpressureoxidizinggas
vigorouslyacceleratescombustion.
DONOTalterormodifycylinderorrelated
components.
DONOTuseoilorgrease.Keepcylinder,cylinder
valve,allequipmentandconnectionsfreeofoiland
greasetoavoidviolentignitionresultinginserious
bodilyinjuryordeath.
WARNING!
Formedicalapplications,useonlyasdirectedbya
licensedpractitioner.Uninterrupteduseofhigh
concentrationsofoxygenoveralongduration,
withoutmonitoringitseffectonoxygencontentof
arterialblood,maybeharmful.
WARNING!
OxygencylindersfilledbytheHomeFillshallbeused
forpersonaluseonly.“Nottobefilledforresaleor
usebyprofessionalusers”.
ThiscylinderistobeusedwiththeHomeFillsystem.
Thiscylindercontains93±3%oxygenproducedby
anoxygenconcentratorusingthepressureswing
adsorptionprocess.
DONOThandlecylinderorusecontentsuntil
youareprofessionallytrained,includingemergency
procedures.
UsethiscylinderinaccordancewiththeHomeFill
owner’smanual.Openvalveslowly,Closevalveafter
eachuseandwhenempty.Securecylinderduring
storageanduse.Nosmokingincylinderarea.Keep
awayfromheat,flameorspark.Keepoutofthereach
ofchildren.DONOTdrop.
FillingofcylinderisperformedbyInvacareHomeFill
oxygensystemonly.
81176148-A
Setup
WARNING!
Cylindersthathavebeenrefinishedand/orexposed
toelevatedtemperaturesMUSTbehydrostatically
testedbeforefilling.Cylindertemperatureshouldnot
exceed52°C(125°F).Aluminumcylinderssubjectto
theactionoffireorheatedtemperaturesinexcessof
177°C(350°F)MUSTbewithdrawnfromservice
andcondemnedbytrainedpersonnel.DONOT
usecausticpaintstrippers.Valveandsafetyrelief
devicesMUSTberemovedandreplacedbytrained
andauthorizedpersonnel.DONOTalterorchange
valveassembly.
AllcylindersMUSTbeinspectedbeforeattemptingto
fill;otherwise,injuryordamagemayoccur.
NEVERsmokeinanareawhereoxygenisbeing
administered.
NEVERusenearanytypeofflameor
flammable/explosivesubstances,vapoursor
atmosphere.
3.3TheHydrostaticTestingDate
AluminumcylindersMUSTundergotestingeveryfiveyears.
WARNING!
DONOTfillcylindersthathavenotbeentested
inthepastfiveyears.Otherwise,seriouspersonal
injurymayresult.Contactyourdealer/HomeCare
equipmentproviderforreplacement.
3.4ExternalExamination
1.Examinetheoutsideofthecylinderforthefollowingconditions,
andreplacethecylinderiftheyexist:
Dentsordings
Arcburns
Oilorgrease
Anyothersignsofdamagethatmightcauseacylindertobe
unacceptableorunsafeforuse.
2.Examinethecylinderforevidenceoffireorthermaldamage.
evidenceincludescharringorblisteringofthepaint,orother
protectivecoatingorheatsensitiveindicator.Iffireorthermal
damageisfound,replacethecylinder.
3.Inspectthecylinderfillportforthefollowing:
Debris,oilorgrease
Noticeablesignsofdamage
Signsofcorrosioninsidethevalve
Signsofexcessiveheatorfiredamage
Signsofpressureleaks
DONOTusethecylinderifanyoftheseitemsare
found.Contactyourdealer/HomeCareequipment
providerforreplacement.
1176148-A9
HomeFill®System
101176148-A
Usage
4Usage
4.1Connecting/DisconnectingtheCylinder
to/fromtheCompressor
ConnectingtheCylindertotheCompressor
WARNING!
NEVERusetoolsofanykindtoconnect/disconnect
thecylinderandthecompressor.Otherwise,severe
injuryand/ordamagemayoccur.
DONOTdropoxygencylinders.Usetwohandswhen
handling/transportingoxygencylinders.Otherwise,
injuryordamagemayoccur.
1.Examinethecylindergauge.Ifthecylinderpressureislessthan
1500psig(10342kPa),proceedtoSTEP2.
Ifthecylinderpressureisgreaterthan1500psig(10342
kPa),DONOTattempttotopoffthiscylinder.Itmay
notfill.
2.Settheconserverflowdialonthecylinderto“O”.
3.Slowlyopentheshutoffvalvefully(counterclockwise).
4.Removethecylinderandconnectorfillportcovers.
CAUTION!
DONOTconnecttheoutlet(cannula)connection
totheHomeFillunit,otherwisetheconserving
cylinderwillnotworkproperly.
5.MomentarilypushDOWNontheouterring(sleeve)ofthe
connectorfillportuntilGREENdotsarevisibletoresetthe
connector.
Iftheouterring(sleeve)isintheUPposition(GREEN
dotsnotvisible),theconnectorfillportwillnotbe
abletoacceptthecylinderfillport.PushingDOWN
momentarilywillresettheconnectorfillport(GREEN
dotsvisible)toacceptthecylinderfillport.
1176148-A
11
HomeFill®System
12
1176148-A
Usage
Fillportcoversnotshownforclarity.
6.Graspthecylinder/regulatorassemblyintheareabehindthe
cylindergauge.
7.Positionthecylinderinthecompressorcradle.
8.Alignthecylinderfillportwiththeconnectorfillport.
9.PullUPontheouterring(sleeve)oftheconnectorfillportwhile
pushingDOWNonthecylinder/regulatorassemblytocouple
thecylinderfillportintotheconnectorfillport.
Thecylinderisproperlyconnectedwhenanaudible
“click”isheard.
DisconnectingtheCylinderfromtheCompressor
WARNING!
NEVERusetoolsofanykindtoconnect/disconnect
thecylinderandthecompressor.Otherwise,severe
injuryand/ordamagemayoccur.
1.PressthecompressorpowerswitchtotheOff(O)position.
2.Graspthecylinder/regulatorassemblyintheareabehindthe
cylindergauge.
1176148-A13
HomeFill®System
3.Withtheotherhand,grasptheouterring(sleeve)ofthe
connectorfillportandpushDOWN.
4.Liftuponthecylinder/regulatorassem-
blytoremovefromtheconnectorfillport.
5.Whenthecylinderfillportisdisconnectedfromtheconnector
fillport,releasetheouterring(sleeve)oftheconnectorfillport
andusetwohandstoremovethecylinderfromthecompressor
cradle.
WARNING!
Thefillportcoverontheconnectorfillportand
thecylinderfillportMUSTbereplacedafter
fillingandwhenevernotinuse.Ifeitherofthe
fillportcoversaremissingcontactyourhealthcare
providerbeforeusing.
6.Placethefillportcoversontotheconnectorfillportandcylinder
fillport.
7.Examinethecylindergaugetoensurecylinderisfull(needle
pointingintoGREENarea).Ifthecylinderisnotfull,perform
thefollowing:
RepeatSTEPS1–9ofConnectingtheCylindertothe
Compressor,page11
.
Ifcylinderisnotfullaftersecondattempttofill,contact
theprovider.
8.Slowlyclosetheshutoffvalvefully(clockwise).
4.2CylinderOperation
WARNING!
NEVERsmokeinanareawhereoxygenisbeing
administered.
NEVERusenearanytypeofflameor
flammable/explosivesubstances,vapoursor
atmosphere.
1.Priortoeachuse,inspectproductforvisibledamage.Referto
3.4ExternalExamination,page9
.DONOTuseifanydamage
isfound.
14
1176148-A
Usage
Ifforanyreason,anylabelbecomesillegibleorlost,
contactyourdealer/HomeCareequipmentprovider.
2.Inspectthecylindercontentsgaugetoensurethepressureisat
least500psig(3447kPa).Ifbelow500psig(3447kPa),thenrefill
thecylinderusingtheHomeFillsystem.
WARNING!
Whenpointeroncontentspressuregaugeenters
redzone,itisrecommendedtorefillthecylinder
ontheInvacareHomeFillCompressor.
3.Positionthecylindersotheoxygencylindervalveoutletis
pointingawayfromtheuserandanyotherperson(s).
WARNING!
DONOTdirectflowofoxygenatanyperson,or
flammablematerialwhenadjustingtheflowdial.
4.Besuretheflowdialisinthe“O”position.Align“O”withthe
indicatorwindow.
5.Attachaduallumenadultoxygennasalcannulawithamaximum
lengthof2.1m(7feet)totheproduct’stwooutletconnections
accordingtothecannulamanufacturer’sinstructions.
Itdoesnotmatterwhichtubeisinstalledonwhich
outletconnection.
DONOTusepediatric,lowflownasalcannulasor
oxygenmaskswiththisproduct.
6.Placethecannulaoveryourearsandpositiontheprongsinyour
noseasinstructedbyyourHealthCareproviderorcannula
manufacturer.
DANGER!
RiskofDeath
Thecannulacancauseinjuryifimproperlypositioned
andsecured.
ThecannulaMUSTberoutedandsecured
properly.
DONOTpositionthecannulaaroundtheneck.
Ensurethepatientcanmovefreelywhilewearing
thecannula.
Avoidpositioningcannulasacrossareasofhigh
traffic.
7.Ensuretherearenokinksortwistsinthecannulatubing.
8.SlowlyturntheshutoffvalveknobintheONdirection
(counterclockwise).
Thiswillallowcylinderpressuretotravelthroughthe
regulator.
9.Usetheswitchtoselectoneofthefollowingmodes:
ContinuousFlow
ConservingPulse
1176148-A15
HomeFill®System
Theregulatorislabeled“CONTINUOUS”and“PULSE”.
Thefirstthreeflows(0.25,0.50and0.75L/min)arein
continuousmodeatalltimes,regardlessoftheswitch
position.
10.Turnflowdialuntiltheindicatorwindowisalignedwiththe
prescribedsetting.
Whentherearetwonumbersdisplayedintheindicator
window,thelargernumberisthePULSEflowsettingand
thesmallernumberistheCONTINUOUSflowsetting.
WARNING!
NOOXYGENisdeliveredbetweensettings.
Toavoidinjurytopatient,ALWAYSconfirm
prescribedsettingbeforeadministeringtopatient
andmonitorflowonafrequentbasis.
OnlyusetheInvacareCorporationcarryingbag.
11.Breathethroughthenoseandfeelapulseofoxygenasyouinhale.
Whenonapulsesetting,thereisflow,orapulse,onlyat
thebeginningofeachbreath.Ifyoudonotfeelthepulse
atthebeginningofeachbreath,checkthesetting.
Becauseeachpatient’sbreathingpatternisdifferentand
theenvironmentvaries,itmaybedifficulttofeelsome
lowsettingpulses.
Ifoperatingthecylinderbelow1.7ºC(35ºF)usethe2
L/Mincontinuoussettingonly.
Whentheunitisfirstswitchedfrompulsetocontinuous
flow,asingleinhalationisrequiredtoinitiatethe
continuousflow.
12.Whenfinishedusingthecylinder,performthefollowingsteps:
a.Turntheflowdialtothe“O”position.
b.Turntheshutoffvalveclockwisetoturnofftheregulator.
161176148-A
Maintenance
5Maintenance
5.1Maintenance
CAUTION!
DONOTusecleaningsolutionsonthecoupleror
fillportfitting.
DONOTimmerseproductinanykindofliquid.
AllrepairsMUSTbedonebyInvacareCorporationor
inanInvacareapprovedrepaircenter.
1.Asrequiredwhendirty,cleanexterioroftheproductwitha
dry,lintfreeclothonly.
2.Storeproductinacleanareafreefromgrease,oil,andother
sourcesofcontamination.
5.2ReplacementPartsAccessories
HF2POST6BAG
HF2POST9BAG
1176148-A17
HomeFill®System
6Troubleshooting
6.1Troubleshooting
IfConservingCylinderfailstofunction,consulttheTroubleshootingguide.Ifproblemcannotbecorrected,consultyourDealer/HomeCare
equipmentprovider.
Problem
PossibleCauseSolution
Noflow.1.Regulatorin“O”position.
2.Cylinderempty.
3.Dialnotsetatincrement.
4.Conservernotsensingbreath.
1.Settoprescribedsetting.
2.Replace/fillcylinder.
3.Setdialsoindicatorpointstoasetting.
4.Checkpositionofcannulainnose.Donot
breathethroughmouth.
5.Checktoseeifthecannulaiskinked.
181176148-A
TechnicalData
7TechnicalData
7.1TypicalProductParameters
Flows
PulseVolumes(L)
EquivalentFlowRateMinimumNominalMaximum
1/4L/minsetting:0.2130.2500.288
1/2L/minsetting:0.4250.5000.575
3/4L/minsetting:0.6380.7500.863
0.4L/minsetting:0.3400.4000.460
0.5L/minsetting:0.4250.5000.575
1L/minsetting:0.8501.0001.150
1.5L/minsetting:1.2751.5001.725
2L/minsetting:1.7002.0002.300
2.5L/minsetting:2.1252.5002.875
3.5L/minsetting:2.9753.5004.025
ContinuousFlow
Setting
4.5L/minsetting:3.8254.5005.175
1176148-A19
HomeFill®System
1,1.5,2,2.5,3,4,5,6
Conserving(Savings)Ratio,
BolusSizeandAccuracy
@20BPM
PulseVolumes(L)
EquivalentFlowRateMinimumNominalMaximum
1L/minsetting:0.0140.01650.019
1.5L/minsetting:0.0210.02480.029
2L/minsetting:0.0290.03300.038
2.5L/minsetting:0.0360.04130.047
3L/minsetting:0.0430.04950.057
4L/minsetting:0.0560.06600.076
5L/minsetting:0.0710.08250.095
6L/minsetting:0.0840.09900.114
Sensitivity0.130to0.150in(0.330to0.381cm)H
2
O
“ON”time<125ms
PulseSetting
Delivery
“OFF”time<300ms
Pressures
InletPressureRange
2068-14479kPa(300-2100psig)
MaxFillPressure14479kPa(2100psig)
InternalRegulatedPressure
345kPa(50psig)
SafetyBurstDisk
Burstsat20.823.2Mpaat22°C(3025–3360psigat72°F)
201176148-A
/