3M Retrofit CO Monitor Kit W-2808/37027(AAD), 1 EA/Case, Calibration Kit 529-04-49, 1 EA/Case, Portable Compressed Air Filter and Reg Panel 256-02-00, 50 cfm, 4 outlets, w/Carbon Monoxide Filtration and Monitor 1 EA/Case, Portable Compressed Air Filter and Regulator Panel 256-02-01, 100 cfm, 8 outlets,, Span Gas Cylinder 529-05-19, 1 EA/Case, Zero Gas Cylinder 529-05-18, 1 EA/Case Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the 3M Retrofit CO Monitor Kit W-2808/37027(AAD), 1 EA/Case Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
GENERAL SAFETY INFORMATION
Intended Use:
This 3M
TM
Calibration Kit is designed to allow the 3M
TM
Carbon Monoxide Monitor to be
zeroed and spanned to insure accurate readings of CO in the air flowing through 3M
TM
Portable Air Purification Panels.
USE INSTRUCTIONS AND LIMITATIONS
Important:
Before use, the user must read and understand these User Instructions and the
3M
TM
Carbon Monoxide Monitor User Instructions. Keep for reference.
Use For:
Calibration of 3M carbon monoxide monitor.
Do Not Use For:
Any other purposes other than calibration of 3M carbon monoxide monitor.
Use Instructions:
The calibration kit 529-04-49 is furnished with a non-adjustable regulator to ease the
calibrating procedure.
CAUTION: To avoid damage to the monitor’s sensor, the regulator provided in this kit must be
used. Before attempting to use this kit, be certain that the components are connected as shown
in figure 1 and read the section on “Calibration and Adjustment” in the User Instruction furnished
with the CO monitor. It will save time and conserve calibration gas if the monitor is “warmed up”
and the calibration adjustment controls have been located on the monitor prior to screwing the
regulator onto the test gas cylinder.
CALIBRATION PROCEDURE
Note: Have all tubing and equipment ready for performing the following calibration procedure
before screwing bottle of test gas onto regulator. Gas will flow out of cylinder when regulator
is attached to the test gas cylinder.
Zeroing the Monitor
1. Remove the protective cap (if supplied) from the ZERO AIR cylinder outlet.
2. Connect the plastic tubing provided in the kit to the plastic fitting as shown in the figure 1,
and connect the opposite end of the tubing securely to the barbed fitting on the regulator.
3. Connect the monitor’s sample tube to the proper end of the above tubing/regulator assembly.
4. Locate ZERO adjustment potentiometer and verify monitor has had adequate warm-up time
(see 3M
TM
CO Monitor User Instructions).
5. Screw regulator valve onto the ZERO AIR cylinder outlet, making sure not to “cross thread”
the regulator or cylinder outlet port, and tighten regulator firmly.
6. Observe display on monitor. Reading should move to zero (00) after approximately one (1)
minute. If display does NOT read zero (00), adjust ZERO potentiometer such that display
now reads zero (00).
7. When the above steps have been completed, unscrew regulator valve from the ZERO AIR
cylinder and replace plastic protective cap.
3 Calibration Kit
529-04-49
User Instructions
For use with 3M
TM
Carbon Monoxide Monitor
Exported by/Exporté par/
Exportado por:
3M Occupational Health and
Environmental Safety Division
3M Center, Building 235-2W-70
P.O. Box 33010
St. Paul, MN 55133-3010
Printed in the U. S. A.
Imported by :
3M Occupational Health and
Environmental Safety Division
3M Canada Company
P.O. Box 5757
London, Ontario N6A 4T1
Printed in the U. S. A. ©3M 2000 38-9017-2791-7
Use of equipment described in these User Instructions must be in accordance with applicable
health and safety standards or pursuant to the recommendations of an industrial hygienist.
Do not use with parts or accessories other than those supplied by 3M as specified in these User
Instructions. Failure to do so may reduce respirator effectiveness and result in sickness or death.
W WARNING
All copy and graphics
print black.
Output at 100%
Ensure that the Title Block
is facing out on the front
side of the folded insert.
ARTWORK INFORMATION
38 #: 38-9017-2791-7
34 #: 34-7049-0697-2
Author: KA
(3845 - Zip 52)
Date: 12-5-00
Trim Size: 11˝ x 8.5˝
Fold Size: 3.667˝ x 4.25˝
NIOSH: None
Scale=1 inch
3M APPROVALS:
N/A 3M Packaging Preview, St. Paul
11-29-00 3M Electronic Routing, St. Paul
_______ 3M Final Approval, St. Paul
FINAL APPROVAL NOTE:
Please review carefully.
Once released to you (the client) or
your agents, liability for its contents
by Kuschner Associates, Inc. is
restricted to correction of this file
only, and not for the cost of reprinting
or loss or damage to product from
packaging specified.
3M’s respirator products help provide protection against certain airborne
hazards. Misuse may result in sickness or death. For proper use, see
supervisor or User Instructions, or call 3M in U.S.A., 1-800-243-4630.
In Canada, call Technical Service at 1-800-267-4414.
W WARNING
Spanning the Monitor
1. Remove the protective plastic cap (if supplied) from the SPAN GAS cylinder, noting the
concentration of carbon monoxide printed on the cylinder.
2. Connect the plastic tubing provided in the kit to the plastic fitting as shown in the figure 1,
and connect the opposite end of the tubing securely to the barbed fitting on the regulator.
3. Connect the monitor’s sample tube to the proper end of the above tubing/regulator assembly.
4. Locate SPAN adjustment potentiometer and verify monitor has had adequate warm-up
time (see CO Monitor's User Instructions).
5. Screw regulator valve onto the SPAN GAS cylinder outlet, making sure not to “cross
thread” the regulator or cylinder outlet port, and tighten regulator firmly.
6. Observe the display on the monitor. The reading will increase, and then stabilize. As
soon as reading is stable (approximately one minute), compare the value indicated on the
display, with the value noted on the SPAN GAS cylinder. If the display does NOT read the
same value as noted on the SPAN GAS cylinder, adjust the SPAN potentiometer until the
reading indicated on the display matches the concentration of carbon monoxide printed
on the SPAN GAS calibration gas cylinder.
7. When the above steps have been completed, unscrew regulator valve from SPAN GAS
cylinder and replace the protective plastic cap.
8. Disconnect the associated calibration test tubing from the monitor and reconnect
monitor’s sample tube to sampling supply on filtration system.
Place all calibration kit components back into the protective plastic case to prevent damage
or loss to components.
Failure to reconnect monitor to breathing air system will void appropriate standards
requiring a carbon monoxide monitor and may result in sickness or death.
W WARNING
SPECIFICATIONS
Span Gas 95 ppm (parts per million)
FOR MORE INFORMATION and assistance on 3M occupational health and environmental
safety products, contact your local 3M representative or call 3M OH&ESD Technical Service
toll free in U.S.A., 1-800-243-4630. In Canada, call Technical Service at 1-800-267-4414.
Fig. 1
Regulator/hose
barb assembly
Slide plastic fitting from kit onto kit
supplied tube. Attach other end of
fitting to monitor’s sampling tube.
3M monitor
Monitor’s sample tube
Calibration gas
(zero and span)
38-9017-2791-7_529-04-49CalKit 12/5/00 10:45 AM Page 1
/