Mettler Toledo Test Procedure for Moisture Analyzer Sensitivity Testing Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo Test Procedure for Moisture Analyzer Sensitivity Testing Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Routine Testing
SOP for Periodic Sensitivity Tests
of Moisture Analyzers
Title: SOP for Periodic Sensitivity Tests of Moisture Analyzers
Document No.:
Author: METTLER TOLEDO Laboratory & Weighing Technologies
Number of Pages: 4
Controls
Replacement Document: N/A
Reason for Revision: N/A
Release Date: 11. O c t. 2 011
Release for Routine Operation
Reviewed by:
Date:
Signature:
Approved by:
Date:
Signature:
2 of 4
SOP for Periodic Sensitivity Test
METTLER TOLEDO
Routine Testing of Moisture Analyzers
General
To test the sensitivity of the weighing unit of a moisture analyzer, a large test weight (between 50% and 100%
of capacity) is recommended.
Basic Rules for Handling Moisture Analyzers
•Beforeusingamoistureanalyzer,makesuretheinstrumentwasleftonpowerforasufficientperiodoftime
(mentioned in the user manual).
•Makesurethemoistureanalyzerisleveled.
•Minimizeenvironmentalinfluences,e.g.airdraft,vibrationsordirectsunlight.
Basic Rules for Handling Weights
Important to know
•Onlyanexternalweightwithcalibrationcertificatecanmaketheweighingunitofamoistureanalyzeratrace-
able piece of equipment.
•Weightsshouldalwaysbeplacedgentlyonthesamplepanholderandputbackimmediatelyintheirstorage
placeafteruse.Theymaybeparkedonacleanpieceofpaper.
•Weights(sincetheyarealsopartofmeasuringequipment)needtobere-calibratedatspecifiedintervals
(ISO 9001).
•Anydamage,whichmighthaveaffectedthevalueoftheweight,shouldtriggeranimmediatere-calibration.
METTLER TOLEDO’s calibration services will give advice on this.
How to Store Weights
• Weightsshouldbestoredintheiroriginalbox.
• Weightsshouldbestoredatroomtemperature,sincetemperaturedifferencesbetweenweightsandtheir
surrounding could lead to measurement errors.
• Weightsthathavenotbeenstoredatthesametemperatureneedacclimatizationbeforeuse.
How to Move Weights
• Weightsshouldonlybehandledwithappropriatetoolssuchastweezers,forks,handlesorgloves(see
METTLER TOLEDO’s accessories for weights).
• Thesetoolsshouldbeexclusivelyusedfortransferringweights,duetopossiblecontamination.
3 of 4
SOP for Periodic Sensitivity Test
METTLER TOLEDO
1)
– Value within the warning limit: no action is necessary.
–Valuebetweenthewarningandcontrollimitarewithinthetolerancebutmustbekeptundersurveillance.
2)
– Value within the control limit, see
1)
– Value beyond the control limit show that weighing process is no longer under control and immediate action is therefore required.
Sensitivity Test
Preparation
•Beforethetestisperformed,thetestweightsmustbeacclimatizedtotheambienttemperatureofthemoisture
analyzer.
•Whenthetestisperformed,theheatingunitofthemoistureanalyzershouldbeatroomtemperature.
Test Procedure HX and HS Models
•Starttestmodeandfollowtheinstructions.
Test Procedure other Moisture Analyzers
•Removethesamplepan.
•Tarethebalance.
•Placethetestweight(s)onthesamplepanholder.
•Readthestablevaluefromthedisplayandnoteit.
Evaluation
•Evaluatewhetherthenotedvalueexceedsthedefined“WarningLimit.
1)
•Evaluatewhetherthenotedvalueexceedsthedefined“ControlLimit.
2)
Deviation
Warning Limit
1)
(where defined)
• Ifthewarninglimitisexceeded,repeatthetest.
• Ifthewarninglimitisexceededagain,reportthatthewarninglimitwasnotmettothelaboratorysupervisoror
the person responsible for the moisture analyzer.
• Levelthemoistureanalyzer,performadjustmentwithbuilt-in(ifapplicable)orexternalweightandrepeatthe
test.
• Ifthewarninglimitisstillexceeded,reporttheproblemtothelaboratorysupervisororthepersonresponsible
of the moisture analyzer. Optionally, contact METTLER TOLEDO’s service organization for advice.
Control Limit
2)
• Ifthecontrollimitisexceeded,reporttheproblemtothelaboratorysupervisororthepersonresponsibleofthe
moisture analyzer.
• Markthemoistureanalyzeras“outofcontrollimits”.
• ContactMETTLERTOLEDOserviceorganizationforadvice.
Mettler-Toledo AG
Laboratory & Weighing Technologies
CH-8606Greifensee
Tel. +41 44 944 22 11
Fax.+41449443170
Subjecttotechnicalchanges
©11/2011Mettler-ToledoAG
PrintedinSwitzerland30031488
GlobalMarComSwitzerland
RecommendedTestWeights,WarningandControlLimits
Recommendation is based on 0.1% weighing accuracy and safety factor of 2. The absolute tolerances depend
on the mass of the test weight.
Various weighing parameters can contribute to the measurement uncertainty of a weighing result. With the
exceptionoftherepeatabilitytest,allcontrollimitsaresetto1/2oftheweighingaccuracyinordertohave
a security reserve accounting for any other influences such as eccentricity and linearity on the result (control
limit = test weight * weighing accuracy / 2)
Moisture Analyzer HX/HS HR/HG HB/MJ
Test weight 100 g 50 g 20 g
Warning Limit ± 25 mg ± 12.5 mg ± 5 mg
ControlLimit ± 50 mg ± 25 mg ± 10 mg
For more information