  • Hello! I am an AI chatbot trained to assist you with the HTA Instrumentation Equinox-COND-101C User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
USER’S GUIDE
TABLE TOP
CONDUCTIVITY METER
Model : Equinox-COND-101
Doc no: HTAIPL/MAN/MET-02
Page 1/4
HTA INSTRUMENTATION PVT LTD.
NO.73, Ramachandra Agrahara, Near T. R. Mills,
Chamarajpet, Bangalore560018, India
TEL: 080-26749750, 26759253 FAX: 080-26740681,
E-mail: [email protected], Web: www.htaipl.com
HALLMARK OF EXCELLENCE
Registration No.
SI910799IN
JAS-ANZ
NQA
QSR
ISO 9001-2000
CERTIFIED
Doc no: HTAIPL/MAN/MET-02
Page 4/4
GUARANTEE :
The instrument is guaranteed for a period of ONE YEAR from the date of
supply against any manufactured defects. The repair will be carried out at our
service centers by our authorized personals only. THE GUARANTEE is valid
only if the product is used in accordance with the instructions provided in this
manual.
Accessories :
Complete Kit consisting of Conductivity Meter, Conductivity Electrode of
Glass body and of cell constant = 1, Electrode Stand with clip, One Bottle
(200 ml) Conductivity Buffer Standard of 1400mS/cm, Power Cord, Small
calibration screw driver.
Ordering Information :
Order No. : Item
Equinox- COND 101C : Complete Kit consisting of Conductivity
Meter, Conductivity Electrode of Glass
body and of cell constant = 1, Electrode
Stand with clip, One Bottle (200 ml)
Conductivity Buffer Standard of
1400mS/cm
Accessories :
Equinox-COND 101 : Consists only Conductivity meter,
Electrode stand with clip
Equinox- COND 101/01E : Conductivity Electrode of Glass Body
and K=1
Equinox- COND 101/02E : Conductivity Electrode of Glass Body
and K=0.1(for low conductivity
measurement less than 200 ms/cm)
Equinox- COND 101/01S : Electrode stand with Clip
Equinox- COND 101/01B : Conductivity Buffer Standard of 1400 mS/cm
Equinox- COND 101/02B : Conductivity Buffer Standard of 14000 mS/cm
General :
Digital Table top conductivity Meter is accurate & reliable Instrument for
use in process measurement parameters. It gives direct reading of the
conductivity of the solution under test in ms/ms on a 3 ½ digit panel meter
attached to the main instrument.
OPERATION :
1. MEASUREMENT :
a. All power & Input connections are given from the back panel of the
instrument.
b. All controls & functions are provided in the Front Panel of the unit for
calibration & monitoring purposes.
c. Connect the power main cord on to the 3 pin socket.
d. Connect the conductivity cell terminations to the pair of socket in the rear
of the unit.
e. Connect the temperature probe to the socket in the rear of the unit.
f. Keep the selection switch to the desired range
g. The instrument is now ready to read the conductivity of the solution.
CALIBRATION :
4Keep the Selector Switch in 2000 ms range.
4Dip the Electrode in 1400 ms Standard solution, display should read
1400 ms if not adjust the trim pot marked GAIN on the front panel so that
display read 1400 ms.
4Now the instrument is ready for taking measurement.
4Temperature probe is used only for compensation purpose. Temperature
value can't be viewed.
FEATURES :
Bright Red LED : Better Visibility even in 'Hard - to See' areas.
Auto Temp
Compensation(ATC) : Built-In
Accuracy : Good accuracy levels to match the Industry
Standard.
Calibration : 1 Point calibration for better stability and
accurate readings.
Electrode Holder : Provided, for added safety to the electrode.
SPECIFICATIONS :
Measurement Range : 0.0 to 199.9ms
0 to 1999 ms
0.0 to 19.99 ms
0 to 199.9 ms
Accuracy : 1% of F.s. ( Excluding Probe Error)
Resolution : 0.1 ms
1 ms
0.01 ms
0.1 ms
Temp. compensation : Auto 0 – 60°C
Range Selection : Manually Using Rotor Switch
Power supply : 230 VAC 1 50 Hz
Dimensions : 230 (W)×180(D)×90(H)mm