WIKA 73 tag:model:74 tag:model:74-8xx tag:model:75 Operating instructions

Type
Operating instructions

This manual is also suitable for

GB
P
Operating instructions
Manual de instruções
Examples/Exemplos
Gas actuated thermometers
Termómetro à pressão de gás
also per directive
também conforme a directiva
94/9/EG (ATEX)
GB
P
2
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers
Operating Instructions Gas acutated thermometers Page 3 - 28
Manual de Instruções Termómetro à pressão de gás
Página 29 - 39
GB
Contents
1. Safety instructions 4
2. Description 4
3. Installation conditions and installation 4
4. Troubleshooting 7
5. Indicator check 7
6. Indicator correction 8
7. Switch contacts 8
8. Explosion protection 11
9. Maintenance and servicing / cleaning 13
10. Repairs 13
11. Disposal 13
Enclosure 1: Declaration of conformity for Model 73 and 74 14
Enclosure 2: Declaration of conformity for Model 76 15
Enclosure 3: Declaration of conformity for Models 73 and 74
with inductive contacts Model 831 16
Enclosure 4: EC-type examination certificate (Ex approval for
gases) for slot-type initiators types SJ (WIKA-Model 831) 17
Enclosure 5: EC-type examination certificate (Ex approval for
gases) for SN-sensors types SJ (WIKA-Model 831-SN/S1N) 20
Enclosure 6: EC-type examination certificate (Ex approval for
dust) for SN-proximity sensors types SJ (WIKA-Model 831
and 831-SN/S1N) 24
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 3
Contents
Information
This symbol provides you with information, notes and tips.
Warning!
This symbol warns you against actions that can cause injury
to people or damage to the instrument.
GB
1. Safety instructions ... 3. Installation conditions, installation
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 4
1. Safety instructions
The appropriate national safety regulations (i.e. VDE 0100 /
EN 60 079-14 / 837-2) must be observed when installing,
commissioning and operating these devices.
Do not work on gauges with switch contacts unless the power has
been isolated
Serious injury and/or damage can occur if the appropriate regulations
are not observed.
Only appropriately qualified personnel should work on these instru-
ments.
2. Description
Gas actuated thermometers consist of a stem, a capillary and a case containing
the bourdon tube element. These components are connected to form a single
system. The complete measuring system is filled with an inert gas under
pressure.
Any temperature variation causes a change in the internal pressure of the
stem, leading to a deflection of the bourdon tube. A mechanical linkage
(movement) transmits this deflection to the pointer. Variations in the ambient
temperature acting on the case are compensated for by a bimetal element
mounted between the movement and the bourdon tube.
WIKA gas actuated thermometers are available for temperature ranges from
-80 ... +700 °C with an accuracy complying with Class 1 of EN 13 190.
3. Installation conditions and installation
Before installing the probe, check whether the probe material used (speci-
fied in the delivery note) is chemically resistant/neutral to the medium being
measured. This also applies to thermowells.
Ensure that all the necessary accessories for the mounting method ordered
have been supplied e.g. instrument mounting brackets, clamp straps for
panel mounting, contact adjustment buttons, etc..
The accessories will be either on the instrument or supplied in a separate
bag, which in most cases will be attached to the thermometer.
GB
3. Installation conditions and installation
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 5
3.1 Installation conditions
If possible, the entire length of the stem should be exposed to the tempera-
ture to be measured, but, if not, it should at least be the length of the active
part (active length), which corresponds to the length of the gas expansion
vessel.
In pipelines or other measuring points the temperature probe should be
directed as far towards the flow as possible.
Heat conduction errors occur when the size, volume or area of the medium
to be measured is very small, so that the temperature probe becomes
noticeable as a thermal mass. Heat conduction errors may also occur
when the insertion length is not sufficient, or when the instrument mounting
fittings are attached to a good heat conductor (e.g. metal plates) and the
temperature difference between them and the medium to be measured is
very high.
The indicator case must be mounted free of shock and vibration.
If necessary, the indicator can be isolated from the measuring point e.g. by
a flexible capillary between the two with the thermometer mounted on a
suitable instrument mounting bracket.
If this is not possible, the following limits must not be exceeded:
Dry gauges: Frequency range < 150 Hz
Acceleration < 0.7 g (7 m/s
2
)
Liquid-filled gauges: Frequency range < 150 Hz
Acceleration < 4 g (40 m/s
2
)
The liquid filling must be checked on a regular basis.
The liquid level must not drop below 75 % of the gauge diameter.
High shock and vibration leads to indication errors, increased wear in the
transmission mechanism and to fractures at welded and soldered joints.
In thermometers with integral contacts, the switches may chatter as a result
of vibrations, which can lead to increased wear of the contacts and a shift in
the contact set-point. By using magnetic snap-action contacts or slow-acting
relays, the effects of shocks and vibrations on contact thermometers can be
reduced, within certain limits.
3.2 Installation
When mounting a gauge with a screw fitting the sealing torque must not be
applied through the case or terminal box but only with a suitable tool using the
spanner flats provided for this purpose on the square connector shaft.
GB
3. Installation conditions and installation
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 6
When mounting a rotatable and inclinable gas-actuated thermometer, specific
instructions must be followed. In order to set the indicator to the desired
position, the following steps must be taken:
1. The lock nut or union nut must be loosened at the process connection.
2. The hexagon bolts and slotted screws at the swivel joint must be loosened.
3. Position the indicator as required, tighten the hexagon bolts and slotted
screws, and finally tighten the lock nut or union nut firmly.
If a thermowell is used, the stem must not be allowed to touch the bottom
of the thermowell.
Capillaries must be kink protected. The minimum bending radius for
capillaries should not be less than 6 mm. Any kinks or discontinuities in the
capillary may result in the failure of the instrument.
If the probe is installed in a location which is exposed to shock and
vibration, it is essential that the capillary is coiled in several loops, and free
of tension between the last attachment point and the probe. Any excess
length should also be coiled into loops, as cutting the cable would make the
instrument unserviceable.
Any welding or soldering of the capillary, as well as any permanent fastening
of the capillary, must be avoided, as this can severely damage the capillary,
and impair the operation of the instrument.
Installation with
spanner
loosen
Make sure to loosen the screws
on the opposite side as well!
GB
3. Installation conditions, installation ... 5. Indicator check
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 7
3.3 Ambient conditions
Unless the ingress protection of the thermometer's case is specifically identi-
fied in the order confirmation, the instrument must be protected from humid air
and other aggressive atmospheres.
The ambient temperature at the indicator case should be between 0 ... +40 °C in
order to ensure the best possible measuring accuracy. Higher or lower ambient
temperatures can lead to indication errors.
IP Ingress protection (EN 60 529 / IEC 529)
IP 66
IP 65 for thermometers with switch contacts
4. Troubleshooting
Gas-actuated thermometers, both with and without contacts, are
maintenance-free instruments according to their basic design. As measuring
instruments, their measuring accuracy should be checked at application-
specific intervals.
The wear condition of probes exposed to a permanent thermal stress (even
if this stress is very low), of electro-mechanical contacts and of capillaries
exposed to vibratory stress must be checked from time to time.
If any visible damage is found, the instrument must be replaced.
5. Indicator check
Indicator checks should only be carried out in comparison with a more
accurate instrument or, if possible, with a calibrated instrument. The tempera-
ture during the check must remain constant. Fluctuating temperatures can lead
to reading errors caused by the different response times of the probes.
Before checking the indicators of thermometers without thermowells, a waiting
time of at least 5 minutes with the stem‘s full length properly inserted is
required to allow temperature equalisation.
Temperature checks using thermometers with probes assembled with
thermowells can, in many cases, only result in a reference temperature at
the instrument under test, due to permanent heat dissipation caused by the
thermowell. In temperature applications fitted with static checking thermo-
meters (e.g. pipeline systems), permanent offsets in reference temperatures
can also be caused by the heat transfer path.
GB
5. Indicator check ... 7. Switch contacts
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 8
Ambient temperatures around the indicator housing which differ substantially
from room temperature can lead to steady indication errors with constant
ambient temperatures and to varying indication errors with fluctuating ambient
temperatures.
Permanent indication errors caused by the ageing of the measuring system
can be ignored, as they only account for a fraction of the indication accuracy.
6. Indicator correction
Any interference with or modification to the instrument will invali-
date the warranty!
Indicator corrections on gas-actuated thermometers with contacts may only
be carried out by the manufacturer or in adequately equipped workshops by
qualified persons.
The micro-adjustment mechanism built into the pointer should only be
operated using a screwdriver, and only if the thermometer has been
damaged due to improper handling, severe shocks, during transport, etc.
When correcting an indication error using the micro-adjustment mechanism,
a calibrated thermometer must be used for comparison.
The thermometer can be opened by turning the bayonet-lock bezel of the
case counterclockwise using a strap wrench.
7. Switch contacts
Gas-actuated thermometers can be equipped with magnetic snap-action,
sliding, inductive or electronic contacts (see data sheet AC 08.01).
In most cases, gas-actuated thermometers with contacts are used as
two-position controllers for on-off control or as three-position controllers
for on-off-on control with adjustable dead-band. In general thermometers‘
contacts can only carry the operating current of an interposing control relay.
The contact mechanism consists of the red set pointer and the measured-
value pointer. The red set pointer can be adjusted using the supplied setting
tool. The measured-value pointer in a magnetic snap-action contact mecha-
nism consists of a pair of drag pointers, which are held against the red set
pointer by springs and is opened and closed at the set value by means of an
operating pin on the measured-value pointer.
GB
7. Switch contacts
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 9
Electrical connection should only be carried out by qualified electricians.
The switches are terminated on screw terminals within the terminal box.
Conductor cross-section max. 1.5 mm
2
.
The terminal assignment is stated on the connection plate of the thermometer.
To adjust red set pointers
The red set pointers for the switch contacts are adjustable via the adjustment
lock on the window using an adjustment key (included in delivery; on standard
gauges this is found on the outside edge of the junction box).
The red set pointers for the switch contacts are adjustable over the full range
of the instrument. Switching points should be set in the range between 10 %
and 90 % of the full scale, to ensure switching accuracy and the long life of the
measuring mechanism.
7.1 Magnetic snap-action contact Model 821 and sliding contact Model 811
Magnetic snap-action or sliding contacts are control switches which make or
break connected electric circuits as the contact arm, moved by the measured-
value pointer, reaches the appointed set-point.
Surface oxidation at the contact points may lead to malfunction, which can
result in contact-arcing, particularly in the case of intrinsically safe circuits (low
voltages and currents), and in the case of relatively high contact loads.
Since the indication must not be affected by the contact mechanism, actuating
forces, and therefore also the possible contact rating, are low.
Adjustment
key removable
Red set pointers
Adjustment lock
GB
7. Switch contacts
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 10
Sliding contacts are not suitable for liquid-filled instruments!
Magnetic snap-action or sliding contacts are not intrinsically safe and therefore
they are not suitable for applications in potentially explosive atmospheres!
Details see data sheet AC 08.01.
7.2 Inductive switch sensor Model 831
Inductive contacts are non-contact inductive proximity sensors, which make
or break when a control flag, linked to the measured-value pointer, moves
adjacent to or disengages from the control head (slot-type initiator). Changes
in the signal level will then actuate a control unit (switch amplifier).
Details see data sheet AC 08.01.
7.3 Electronic contact Model 830 E
Direct switching of low-level signals, which are typically used in conjunction
with a PLC, is possible using these inductive contacts, complete with
integrated amplifiers, factory-fitted within the instrument.
The usual advantages of inductive contacts are also valid for this system, such
as highly-secure contact operation, zero wear due to non-contact operation
and virtually no adverse effect on the measuring system, thus ensuring the
accuracy of the indication.
No additional control unit is required.
Electronic contact Model 830 E is not intrinsically safe and therefore not
suitable for applications in potentially explosive atmospheres!
Details see data sheet AC 08.01.
7.4 Electromagnetic compatibility
EMC per EN 60 947-5-2.
The instruments should be protected against strong electromagnetic fields.
7.5 Contact functions
The contact function of the respective switch contact is identified by the index
1 or 2 after the model number.
1 = Contact makes when the pointer approaches the set point in clockwise direction
2 = Contact breaks when the pointer approaches the set point in clockwise direction
GB
8. Explosion protection
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 11
a) When used in accordance with the intended use, these instruments have no potential
source of ignition and therefore do not fall within the scope of the Directive 94/9/EC (ATEX).
Source: ATEX 94/9/EC Chap. 1 Art. 1
b) When used in accordance with the intended use these instruments have no active
source of ignition and therefore do not fall within the scope of the Directive 94/9/EC (ATEX).
Source: ATEX 94/9/EC Chap. 1 Art. 1
Use of mechanical thermometer
models specified overleaf
Thermometer
with contact?
Glass
window?
Glass
window?
Inductive
contact ?
Plastic
window
NS 100 ?
Plastic
window
NS 100 ?
Use in
potentially
explosive
atmospheres
not possible
Use without Ex-specific
marking in Zones 1 and 2
(for gases, ignition groups IIA,
IIB, IIC) and Zone 21 or 22
(for dusts)
Manufacturer´s conformity
certivicate according to
Directive 94/9/EC (ATEX)
no
yes
no
no no
no
no
yes
yes
yes
yes
yes
see a)
see b)
8. Explosion protection
GB
8. Explosion protection
Circuit Sensor type Model designation EC-type exami-
nation certificate
WIKA-
Model
(s. Ex-certific.) Fa. Pepperl & Fuchs
Model 1
standard KFD2-SR2-Ex1 PTB 00 ATEX 2080 904.31
standard KFD2-SR2-Ex2 PTB 00 ATEX 2080 904.32
Model 2
standard KFA6-SR2-Ex1 PTB 00 ATEX 2081 904.28
standard KFA6-SR2-Ex2 PTB 00 ATEX 2081 904.29
SN-sensors KFD2-SH-Ex1 PTB 00 ATEX 2042 904.33
SN-sensors KFA6-SH-Ex1 PTB 00 ATEX 2043 904.30
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 12
The following gas-actuated instrument versions always require a certificate of
conformity issued by the manufacturer in accordance with the Directive
94/9/EC (ATEX), when operated in potentially explosive atmospheres:
Combi-Thermometer with RTD (R7380, R7381)
Combi-Thermometer with thermocouple (A7390, R7391)
8.1 Marking of instruments intended for use in potentially explosive
atmospheres
WIKA Model
Year of manufacture
Lodged document number (approval number): 8000550764
8.2 EC-type examination certificates for inductive contacts Model 831
Standard version Model 831.XX
PTB 99 ATEX 2219 X (enclosure 2) and ZELM 03 ATEX 0128 X (enclosure 4)
Depending on the number of switches and on the case diameter, either
Model SJ2-N ... or Model SJ3.5-...-N... is used.
Safety pattern version Models 831.XX-SN or -S1N
PTB 00 ATEX 2049 X (enclosure 3) and ZELM 03 ATEX 0128 X (enclosure 4)
Depending on the number of switches and on the case diameter, either
Models SJ 2-SN..., SJ 2-S1N..., SJ 3.5-SN.. or SJ 3.5-S1N... are used.
The sensor type fitted is stated on the product label of the pressure gauge.
The permissible limits for U
i
, I
i
and P
i
for intrinsically safe supply circuits
depend on the sensor type. They can be taken from the corresponding
EC-type examination certificates.
(see also data sheet AC 08.01)
Suitable switch amplifiers are e.g.:
GB
8. Explosion protection ... 11. Disposal
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 13
For further information on explosion protection see manufacturer´s declaration
of conformity for the thermometer, EC-type examination certificate and
operating instructions for the electrical switch contacts (enclosed, where
applicable, for the respective instrument version).
9. Maintenance and servicing / cleaning
The instruments require no maintenance or servicing.
The indicator and switching function should be checked once or twice every
12 months. For this the instrument must be disconnected from the process
and checked using a temperature calibrator.
The instruments should be cleaned with a damp cloth, moistened with soap
solution. When cleaning the inside of the terminal box, the mains power must
be disconnected. All parts must be dry before the power is reconnected.
10. Repairs
Repairs are only to be carried out by the manufacturer or appropriately trained
personnel.
For further details see WIKA data sheet AC 08.01 or the data sheet for the
respective basic gauge.
11. Disposal
Disposal of instrument components and packaging materials should be in
accordance with the respective waste treatment and disposal regulations of
the region or country to which the instrument is supplied.
GB
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 14
Enclosure 1:
GB
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 15
Enclosure 2:
GB
Enclosure 4:
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 17
GB
Enclosure 4:
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 19
GB
Enclosure 5:
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 21
GB
Enclosure 5:
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 23
GB
Enclosure 6:
07/2010 P based on 07/2009 GB
WIKA operating instructions gas actuated thermometers 25
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40

WIKA 73 tag:model:74 tag:model:74-8xx tag:model:75 Operating instructions

Type
Operating instructions
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI