uebe visocor OM50 Instructions For Use Manual

Category
Blood pressure units
Type
Instructions For Use Manual
EN-32 EN-33
english
english
Table of contentsTable of contents
Initial operation of the unit
Operating the unit
Safety instructions
Important instructions for patients
Important technical details
Inserting/replacing batteries
Attaching the cuff
Measuring blood pressure
Unit description
Important instructions for use
Using the memory
Displays
Irregular pulse waves and cardiac arrhythmia
Activating and setting the date/time
Intended use
Original spare parts and accessories
What are normal values for blood pressure?
Maintaining the unit
Warranty
What you should know about blood pressure
General provisions
Technical information
B
G
H
D
Systolic and diastolic blood pressure values
Why you should measure blood pressure regularly
Why you measure different values
F
Explanation of symbols
Disposal
E
Failure and error messages
Customer service
Technical data
Technical inspection / calibration check
C
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
1.
2.
3.
4.
1.
2.
3.
4.
5.
1.
2.
1.
2.
A
34
35
35
36
38
38
39
40
41
41
42
44
46
47
48
50
50
50
50
51
52
52
54
55
57
57
58
58
58
59
60
EN-35
english
B
Safety instructions
A
EN-34
english
Intended use
Thank you for choosing the visocor
®
OM50 upper arm blood pressure monitor (also referred to in the following as the
unit).
It’s determined for measuring blood pressure and pulse at the upper arm, those of people 15 years of age and over. This
unit is recommended for patients with unstable blood pressure for domestic self administration blood pressure control and
treatment support.
Using the pressure variations caused by the pulse at falling cuff pressure, the micro-computer calculates the values systole,
diastole, and pulse. Afterwards, the cuff is deated completely.
A measurement value memory stores the most recent 60 measurement results in each case.
These instructions are intended to help you use the unit reliably and effectively. Keep the instructions together with the unit
at all times. The unit must be used in accordance with the methods detailed in these instructions and may not be used for
other purposes. Read these instructions carefully before using the unit.
1. Important instructions for patients
• Blood pressure measurements on children require specialist knowledge! Please consult your doctor if you wish to
measure the blood pressure of a child. Do not under any circumstances use the unit on a baby/infant.
• The cuff has been specially developed for this unit and must not be used with other units.
• The measurement results of automatically measuring blood pressure monitors may be falsied by pregnancy, irregular
heartbeat or arteriosclerosis. Measure your own blood pressure in consultation with your doctor.
• Do not under any circumstances place the cuff on or over any critical point, e.g. wound, aneurysm, etc. Risk of injuries!
The supply via an intravascular access (infusion) or other medical monitoring devices might be interrupted.
• The unit contains small parts which could be swallowed by children. It should therefore be kept out of the reach of
children at all times. The airhose poses a risk of strangulation. Please follow instructions regarding the handling of cuff
to avoid such risk.
• Make sure the air hose is not kinked. A kinked air hose can prevent the cuff from deating and thereby interrupt the
blood ow in the arm for too long.
Intended use
• Do not use the unit without rst consulting your doctor if you are undergoing dialysis treatment or taking anticoagu-
lants, platelet aggregation inhibitors or steroids. Internal bleeding may be caused under these conditions.
The unit has a build in electronic for collection, preparation, analysis and supervision of pulse waves (oscillometric
measurement).
B
EN-36 EN-37
english
B
english
Safety instructionsSafety instructions
• Please refer to the chapter “Important instructions for use” (Page EN-40) before conducting any measurements
yourself.
2. Important technical details
• A consistently good power supply to the unit is essential for trouble-free blood pressure measurements.
- Please use long-life alkaline batteries only (LR6).
- When replacing batteries, always replace all the batteries at the same time.
- You will need 4 x 1.5 V AA/LR6 batteries. Rechargeable batteries have a voltage of only 1.2 V and are thus not
suitable.
- When using the unit with a mains adapter, please use only the type A1 mains adapter specially tested for medical
devices.
- If the unit is not used for an extended period of time, please remove the batteries. Essentially, all batteries can leak.
• The unit must only be operated with original parts. The warranty will be invalidated if the unit is damaged by non-ap-
proved accessories!
• Using the unit in the immediate vicinity of mobile phones, microwave appliances or other devices with strong electro-
magnetic elds may result in impaired functioning. Maintain a minimum distance of 3 m from such devices when using
this unit.
• The display of the pulse frequency is not suitable for checking the frequency of cardiac pacemakers. Cardiac pacemak-
ers and blood pressure monitors do not inuence each other in their mode of operation.
• Never open or modify the unit or the cuff - this is a medical device and may only be opened by authorised specialists.
(Exception: replacing batteries). If the unit has been opened, it must be subjected to a metrological inspection by an
authorised institution.
• The cuff may be inated on the arm only.
• Please comply with the specied operating conditions for measurement. See Technical data, Page EN-55.
• The inating and measuring procedure can be stopped by pressing the Start/Stop button. In this case, the unit stops
the ination procedure and deates the cuff.
• Measuring your own blood pressure does not constitute treatment. On no account modify of your own accord the
dosage of drugs/medication prescribed by your doctor.
• If you have had a mastectomy, do not measure blood pressure on the side of the body where the breast has been
removed together with the lymph nodes from the armpit.
- Make sure when operating the unit with the mains adapter that the unit can be disconnected from the mains power
supply at all times.
• If the unit has been stored at temperatures below 0°C, let it stand at room temperature for at least 1 hour before use.
C
EN-38 EN-39
english
C
english
Operating the unitOperating the unit
1. Unit description
Memory recall button
Start/Stop button
Displays
Battery compartment
Cuff socket
Connecting socket for mains adapter
Air hose
Measurement arrow for arm
circumference
Cuff
Artery marking
Marking for arm circumference
1
2
3
4
5
6
7
8
9
10
11
2. Displays
Memory location or date/time
SYS = Systole (upper blood pressure value)
DIA = Diastole (lower blood pressure value)
PUL 1/min = Pulse
calculated pulse frequency per minute
Pulse signal display or irregular pulse waves (Page
EN-47)
Battery check display
Deating cuff
Inating cuff
Memory code
Failure and error messages (Page EN-52)
1
2
3
4
5
6
7
8
9
1
2
3
4
56
7
8
9
2
1
3
4
9
8
765
Cuff type VM5 Cuff type VL5
1110 1110
12
Air connector
12
C
EN-40 EN-41
english
C
english
Operating the unitOperating the unit
3. Important instructions for use
• Refrain from drinking alcoholic or caffeinated beverages and smoking at least one hour before measuring.
• Please rest at least 5 minutes before measuring. Depending on the previous degree of stress/exertion, this can even
require up to one hour.
• Expose the upper arm. Clothing must not impair the blood ow in or from the arm, as this inuences the blood pres-
sure at the measurement point.
- For this purpose, sit at a table (if possible the height of a dining table, not a coffee table!).
• Body posture must be relaxed:
- Rest your back against the backrest of the chair.
- Rest your forearm completely at, with the palm of your hand facing upwards.
- Place your feet on the oor and do not cross your legs.
• Unusual measurement results are possible with all automatically measuring blood pressure measurement systems from
time to time. Check yourself: did you comply with the instructions for use above? If necessary, repeat the measurement
after allowing the blood circulation in the arm to recuperate briey; rest for about 3-5 minutes for this purpose.
Tip: Remain seated, your monitor will switch off automatically about 3 minutes after a measurement. Afterwards, we
recommend that you repeat the measurement.
• Blood pressure is not a xed value. It may uctuate upwards or downwards by more than 20 mmHg on patients within
a few minutes.
4. Initial operation of the unit
Insert the supplied batteries into the unit, if these have not been inserted already.
If the unit is to be operated from the mains power supply, the cable plug of the mains adapter (not included in the scope
of delivery) must be inserted into the connection socket of the unit. The batteries are switched off automatically. Please use
only the UEBE type A1 mains adapter. See original spare parts and accessories, Page EN-57.
5. Inserting/replacing batteries
• Opening the battery compartment:
Open the cover of the battery compartment on the underside of the unit.
• It is absolutely essential to sit still and remain quiet. Irregular movements, vibrations (shaking), talking and breathing
heavily will affect and may even falsify the measurement. Look out for irregular pulse signals in the display; if neces-
sary, repeat the measurement under more favourable conditions.
• In order to rule out differences between left and right and to obtain comparable measurement results, it is important
always to take blood pressure measurements on the same arm. Ask your doctor which arm he/she recommends you
take your measurements on.
C
EN-42 EN-43
english
C
english
Operating the unitOperating the unit
When the display shows the “Flat battery “ symbol for the
rst time, the unit will still have enough charge for about 30
measurements. Please change the batteries during this period.
6. Activating and setting the date/time
The unit has a date/time function which you can activate where necessary. This function is deactivated when the unit is
delivered.
Activating the date/time function
To activate the date/time function, press the Start/Stop button and then immediately, while the full display is visible, the
Memory button. This function can be deactivated at any time in the same way.
Setting the date/time
The four-digit number representing the year appears in the display rst. Press the Memory
button to set the year. Press the Start/Stop button to conrm the selected year and proceed to
setting the date.
• Inserting batteries:
Remove the old batteries from the unit and insert the new ones. Please observe proper polarity (marking in battery
compartment).
• Closing the battery compartment:
Close the battery compartment by clipping the battery cover back into the unit.
Set the month with the Memory button and conrm this with the
Start/Stop button. Set the day, hours and minutes in the same way.
Changing the date/time
To change the date/time function, take out a battery and wait
until the display goes blank. Now you can reactivate the date/time
function and reset the date and time again.
C
EN-44 EN-45
english
C
english
Operating the unitOperating the unit
7. Attaching the cuff
• Expose the upper arm.
2-3 cm
• Push the cuff onto the arm (Fig. 1) until the lower edge of the cuff is
2-3 cm above the crook of the arm (Fig. 2).
• When using the unit on the left arm, the air hose runs to the unit
through the middle of the crook of the arm so that the 4 cm long
artery marking is automatically located centrally above the pulse
position (Fig. 3).
• When using the unit on the right arm, the cuff must be rotated to
the left until the artery marking is on the pulse position. The air hose
then runs along the inside of the upper arm.
• The cuff should be tightened to such an extent that 2 ngers can still be inserted
between arm and cuff. Tip:
- Bend your arm slightly.
- The upper arm muscle is tensioned slightly.
- This slightly increases the circumference of the arm.
• Now pull the free cuff end tight and close the Velcro fastener (Fig. 4)
Fig. 1
Fig. 2 Fig. 3
Fig. 4
When the date/time function is activated, the time is also shown in the display when the unit
is switched off.
• Check if the measurement arrow on the cuff is inside the “marking for arm circumference“ at the edge of the cuff.
• Lay your extended and relaxed arm with the cuff on the dining table and do not move it during the measurement; do
not talk. The palm should be facing upwards.
• Connect the cuff connector to the cuff socket on the left-hand side of the unit. Make sure the connector is fully inserted
into the unit.
Deactivating the date/time function
To deactivate the date/time function, press the Start/Stop button and then immediately the
Memory button.
C
EN-46 EN-47
english
C
english
Operating the unitOperating the unit
8. Measuring blood pressure
• Switch on the unit by pressing the Start/Stop button. The following display is visible.
• After the unit has completed its calibration against the ambient air pressure, the cuff is auto-
matically inated to approx. 180 mmHg. If that cuff pressure is not sufcient for measuring
the blood pressure level, the device automatically increases the cuff pressure in steps of 30
mmHg until a sufcient cuff pressure has been reached.
• For an ination pressure exceeding 180 mmHg the ination process can also be controlled
manually: Keep the Start/Stop button pressed immediately after the ination has started
and release it when the cuff pressure has reached 40 mmHg above the expected maximum
systolic blood pressure. The ination stops with releasing the button.
• The measurement being completed is indicated by a long beep. The determined values for
systole, diastole, and pulse are shown on the display.
• The unit switches off automatically after about 3 minutes.
9. Irregular pulse waves and cardiac arrhythmia
If the “symbol for irregular pulse waves ashes after a measurement, the unit has
recorded irregular pulses during the measurement. This can be set off by an irregular heartbeat
(arrhythmia), disturbance caused by movement, talking or even breathing in deeply. The
symbol is stored with the respective measurement.
If the symbol appears more frequently, this must be discussed with your doctor! Measurement
with irregular pulse waves should therefore be viewed as critical and repeated under more
favourable conditions.
• The actual measurement takes place with the start of deation. The “” symbol in the
display ashes until the pulse frequency is displayed.
C
EN-48 EN-49
english
C
english
Operating the unitOperating the unit
10. Using the memory
Measured results are automatically stored in the memory. The memory can store up to 60 results and the average value.
When more than 60 measured values have been stored, the oldest value (No. 60) is deleted to allow the latest value (No.
1) to be recorded.
Recalling data
Press the Memory button to call up data. The average value of the stored results is
displayed with “A”.
When the Memory button is pressed again, the value measured last (No. 1) appears.
Press the Memory button repeatedly to call up further measured values.
When the date/time function is activated, the measured values are displayed alter-
nately with memory number, date and time.
The stored data remain visible in the display for around 30 seconds. The unit then
switches itself off.
Deleting data
To delete individual values, press the Memory button several times to obtain the measured value required.
Then press the Memory button again and hold it down for 8-10 seconds until the value rst ashes and then disappears.
To delete the entiry memory, recall the relevant average value display (A) and press the Memory button again, holding it
down until the mean value rst ashes and then disappears.
D
EN-50 EN-51
english
D
english
What you should know about blood pressureWhat you should know about blood pressure
1. Systolic and diastolic blood pressure values
The cardiovascular system has the important function of supplying all organs and tissues in the body with sufcient
amounts of blood and of transporting metabolites. To do so, the heart contracts and expands at a regular rate of about 60
to 80 times per minute. The pressure of the owing blood on the artery walls caused by the heart contracting is termed
systolic. The pressure in the ensuing relaxation phase, when the heart rells with blood, is termed diastolic. During daily
measurement you determine both values.
2. Why you measure different values
Our blood pressure responds to internal and external inuences like a sensitive measuring instrument. It can be affected by
even slight changes. This explains why values measured by a doctor or pharmacist are higher than those measured at home
in the environment you are used to. Changes in the weather, climate changes, or physical or psychological stress can have
effects as well.
3. Why you should measure blood pressure regularly
Even the time of day has an inuence on your blood pressure. During the day the values are generally higher than during
the periods of rest at night. One-off and irregular measurements therefore say little about your actual blood pressure. A
reliable assessment is possible only when measurements are taken regularly. Discuss the measurement values with your
doctor.
4. What are normal values for blood pressure?
The World Health Organisation (WHO) established the following limits in mmHg (millimetre hydrargyrum) to classify the
blood pressure values.
WHO 2003 Systolic pressure
= Upper value mmHg**
Diastolic pressure
= Lower value mmHg**
High pressure from 140* from 90*
Limit 120 to 139 80 to 89
Normal value below 120 below 80
* It is sufcient for one of the values to be raised. ** millimetres of mercury
E
EN-52 EN-53
english
E
english
Technical informationTechnical information
1. Failure and error messages
Failure encountered Possible cause Corrective action
Display Err - 300 Excessive cuff pressure.
Cuff inated to maximum due to movement
of the arm or body.
The hose may be kinked or the air connec-
tor blocked.
- Repeat measurement
- Do not move arm
- Do not talk
Check that hose with air connector is clear
(unobstructed)
Display Err - 1 Measurement error. Measurement could
not be carried out due to movement of the
arm or body.
- Repeat measurement
- Do not move arm
- Do not talk
Display Err - 2 Ination error, pressure built up too quickly
or too slowly.
Check that air connector is correctly tted,
repeat measurement.
Fault was recognised. Cuff is not correctly
tted.
- Repeat measurement
- Do not move arm
- Do not talk
Display Err - 3 Deation rate too high or too low. Keep still during measurement. Call the Service
Hotline if this fault occurs repeatedly.
Display 0
0
Start/Stop button accidentally pressed while
battery was being changed.
Switch unit off and then on again with Start/
Stop button; if necessary.
Failure encountered Possible cause Corrective action
Irregular heartbeat, disruptive movements,
shaking, wobbling (objects), breathing in
deeply, etc.
Repeat measurement after 3-5 minutes rest.
Possible factors (depending on the severity of
the arrhythmia) affecting the measurement
results must be discussed with your doctor.
Measured values are
too high or too low
Too little rest before the measurement.
Movement or talking during measurement.
Repeat measurement after a break of approx.
3-5 mins. Observe instructions for use P. EN-40.
Has the proper cuff been selected? Cuffs
that are too small result in excessive blood
pressure values.
Measure your arm circumference in the middle
of the upper arm and compare this with the
details given on the cuff.
Incorrect measurement position. Repeat measurement, ensuring correct body
posture. (Page EN-40)
Consecutive measure-
ments produce different
measured values
(readings).
Blood pressure is not a xed quantity.
It may vary in the upward or downward
direction within a few minutes by more
than 20 mmHg.
-------
Pulse values are too
high or too low
Too little rest before the measurement.
Movement or talking during measurement.
Repeat measurement after a break of approx.
3-5 mins. Observe instructions for use P. EN-40.
Measurement was
interrupted and cuff
rst deated and then
inated
Unit detects a fault or that diastolic value
is too low.
Unit shows no fault. Repeat measurement.
Movement during measurement If necessary, interrupt measurement, measure
again after a break of 5 minutes.
E
EN-54 EN-55
english
E
english
Technical informationTechnical information
Failure encountered Possible cause Possible cause
No display after unit is
switched on
Batteries inserted incorrectly. Check position of batteries.
Batteries at. Change batteries.
Battery contacts dirty. Clean battery contacts.
No display after Memo-
ry button is pressed
No measured values stored. Take a new measurement.
Measurement
interrupted
Batteries at Change batteries.
Date/time not shown in
display
- Function accidentally deactivated
- Function was not reactivated after battery
replacement
Activate date and time as described on P. EN-42.
2. Customer service
The unit may only be repaired by the manufacturer or by an expressly authorised organisation. Please contact:
UEBE Medical GmbH
Zum Ottersberg 9
97877 Wertheim, Germany
Phone: +49 (0) 9342/924040
Fax: +49 (0) 9342/924080
E-mail: info@uebe.com
Internet: www.uebe.com
3. Technical data
Unit type: Digital automatic unit with electric pump for measuring blood pressure on upper
arm
Dimensions: L = 100 mm x W = 152 mm x H = 61 mm
Weight: 295 g not including batteries
Display: LCD display (liquid crystal display) for measured values and check displays
Memory: 60 measured values (stored automatically) and average value (A)
Measurement procedure: Oscillometric determination of systole, diastole and pulse
Reference procedure of clinical
testing:
Auscultatory measurement
Inating pressure: Initially up to approx. 180 mmHg, then in stages of 30 mmHg
Pressure display range: 0-300 mmHg
Measurement range Systolic: 50-250 mmHg
Diastolic: 40-150 mmHg
Pulse: 40-160 pulses/min
Error limits: Blood pressure measurement: corresponds to EN 1060 Part 3
Pressure measurement: ± 3 mmHg
Pulse measurement: ± 5%
Serial number: The unit bears a serial number SN which provides
cl
ear identication.
E
EN-56 EN-57
english
E
english
Technical informationTechnical information
Power supply: Battery type: 4 x 1.5 V AA alkali manganese (LR 6) or lithium (FR 6) round cells,
service life: more than 800 measurements in 2 years
Optionally stabilised type A1 mains adapter, output 6 VDC, minimum 600 mA
Cuff: Cuff VM5 (for arm circumference 22-32cm) or optionally
cuff VL5 (for arm circumference 32-42 cm)
Operating conditions: Ambient temperature 10 to 40 °C, rel. air humidity up to 85% not condensing, air
pressure 700 to 1040 hPa
Storage and transport conditions: Ambient temperature -20 to 50 °C, rel. air humidity up to 85% not condensing
Air deation valve: Electronically controlled
Automatic switch-off approx. 3 minutes after end of measurement
IP code: IP 20: Protected against solid foreign particles with a diameter of more than 12.5
mm, no protection against water.
Expected service life: 5 years
Classication: Internal power supply by battery
4. Original spare parts and accessories:
Subject to technical modications.
The following original spare parts and accessories are available from specialist dealers:
• Cuff 22-32 cm type VM5
Part. no. 2505001
PZN 01650881
• Mains adapter type A1
Part. no. 2501020
PZN 00179482
• Cuff 32-42 cm type VL5
Part. no. 2505002
PZN 01650875
Manufacturer: UEBE Medical GmbH
Zum Ottersberg 9, 97877 Wertheim, Germany
5. Technical inspection / calibration check
Basically, a metrological inspection is recommended at intervals of 2 years. However, professional users in Germany are
obligated to comply with the aforementioned according to “Regulation for Operators of Medical Devices“.
This can be implemented either by UEBE Medical GmbH, an authority responsible for metrology, or authorised maintenance
services. For this, please observe your national provisions.
Upon request, responsible authorities or authorised maintenance services receive a “Test instruction for metrological
inspection” from the manufacturer.
Important: No modications, e.g. opening the unit (except to replace the batteries) may be made to this unit without the
manufacturer’s prior permission.
EN-59
english
G
Maintaining the unit
F
EN-58
english
General provisions
1. Explanation of symbols
0123
50 °C
-20 °C
85%
0%
This product complies with the Council Directive 93/42/EC from 5 September 2007 regarding medical devices
and bears the mark CE 0123 (TÜV SÜD Product Service GmbH).
Degree of protection against electric shock:
TYPE BF
Please observe instructions for use
Storage and transportation conditions
Ambient temperature -20 to +50 °C
Relative humidity up to 85 %,
non-condensing
Manufacturer
If using the optionally available mains adapter:
2. Disposal
Batteries and technical appliances must not be disposed of with domestic waste, but should be handed in at the
appropriate collection and disposal points.
• The unit contains sensitive parts and must be protected against strong variations in temperature, air humidity, dust and
direct sunlight.
• The unit is neither impact-resistant nor shock-proof. We recommend that you have the intactness and accuracy of the
display checked after heavier falls or impacts.
• The unit is not waterproof.
• If the unit becomes dirty during use, use a soft, dry cloth to clean the unit. Do not use benzene, thinners or other strong
solvents.
• Do not scrub or machine-wash the cuff. If the cuff gets dirty during use, use a synthetic cleaner and softly rub the
surface.
• If the unit has been opened, it must be subjected to a metrological inspection conducted by an authorised
organisation.
• Make sure that liquid cannot get into the air hose. Dry thoroughly.
Keep dry
Mains adapter specications:
Equipment class II (double-insulated)
For indoor use only
Internal fuse
Internal polarity positive
Maintaining the unit
• Make sure when storing the unit that no heavy objects are left lying on the unit or on the cuff and that the air hose is
not kinked.
H
EN-60 EN-61
english
H
english
WarrantyWarranty
The device has been manufactured and tested with great care. However, in the unlikely event of a defect being detected
after delivery, we provide warranty in accordance with the following terms and conditions:
During the warranty period of 2 years from the date of purchase we reserve the right either to repair any such defect at our
expense or to supply a perfect replacement unit. The cost of returning the unit to our factory shall be borne by the sender.
UEBE shall refuse to accept return deliveries that have not been paid for by the sender.
Excluded from the warranty are parts subject to normal wear and tear as well as damage caused by non-compliance with
the instructions for use, improper handling (e.g. unsuitable power sources, breakages, leaking batteries) and/or disassembly
of the unit by the purchaser. Furthermore, no claims for damages against us are substantiated by the warranty.
Warranty claims can only be advanced in the warranty period and by presenting proof of purchase. In the event of a
warranty claim, the unit must be sent to the following address together with the proof of purchase and a description of the
complaint:
The cost of returning the unit to our factory shall be borne by the sender. UEBE shall refuse to accept return deliveries that
have not been paid for by the sender.
UEBE Medical GmbH
Service-Center
Zum Ottersberg 9
97877 Wertheim
Germany
In the case of defectiveness of the goods, the contractual rights of the purchaser to claim against the seller in accordance
with § 437 German Civil Code are not limited by the warranty.
Please note:
In the event of a warranty claim it is essential to attach the proof of purchase.
Warranty furnished
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33

uebe visocor OM50 Instructions For Use Manual

Category
Blood pressure units
Type
Instructions For Use Manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI

in other languages