Delphi DL-00183 User manual

Category
Oxygen Equipment
Type
User manual

This manual is also suitable for

Portable Oxygen Concentrator
User Manual
12659da delphi POC covers 9/18/08 12:14 PM Page 1
ii
DL-00183 Rev. 3
Manufactured by:
Delphi Medical Systems
5725 Delphi Drive
Troy, Michigan 48098-2815 U.S.A.
CUSTOMER SERVICE
Tel: [1] 888.526.1426
The information in this document is subject to change without notice. This
document contains proprietary information that is protected by copyright. No
part of this document may be reproduced in any manner, in whole or in part
(except for brief excerpts in reviews and scientific papers), without the prior
written consent of Delphi Medical Systems. Before using any Delphi
Medical Systems product, be sure to read carefully and understand all
manuals provided with the product.
For Help
If you have questions about the information in these instructions or about
the safe operation of this device, contact Customer Service at the number
listed above.
0086
European Representative:
MediMark
®
Europe Sarl
11, rue Emile Zola – BP 2332
38033 Grenoble Cedex 2 France
Tel: +33 (0) 4 76 86 43 22
Fax: +33 (0) 4 76 17 19 82
12659da delphi POC covers 9/18/08 12:14 PM Page 2
Classification
This equipment is listed with a nationally recognized testing laboratory and
classified with respect to electric shock, fire, and mechanical hazards in
accordance with the following standards:
•EN60601-1(1990),MedicalElectricalEquipment,Part1:General
RequirementsforSafety+A1(1993)+A2(1995)+A13(1996)
•UL60601-1(1stedition,2006-04-26),MedicalElectricalEquipment,
Part1:GeneralRequirementsforSafety
•IEC60601-1(2ndedition),MedicalElectricalEquipment,Part1:General
RequirementsforSafety,withA1andA2
•CAN/CSAC22.2No.601.1-M90(R2005),MedicalElectricalEquipment,
Part1:GeneralRequirementsforSafety
•IEC60601-1-4(2000-2004),Edition1.1ConsolidatedEdition,Medical
ElectricalEquipment,Parts1-4:GeneralRequirementsforSafety–
CollateralStandard:ProgrammableElectricalMedicalSystems
This equipment is classified as:
•ClassI
•TypeBF
•IPX2
•Continuousoperationattemperaturesof95°F(35°C)downto4F(5°C),
andshort-timeoperationfor25minutesattemperaturesupto104°F
(4C)and93%±2%relativehumidity.
iii
DL-00096 Rev. 7
DL-00096 rev 7.indd 3 6/1/2009 4:48:22 PM
iv
DL-00096 Rev. 6
Explanation of Packaging and
Labeling Symbols
Caution,consultaccompanyingdocuments
Input18VDC,6.7A
Output=90%±3%oxygen
Serial number
Catalognumber
U.S.federallawrestrictsthisdevicetosaleby
or on the order of a physician
Nosmoking
Noopenflamewhendeviceisinuseordonotincinerate
Donotdisassemble
Separate collection for electrical and electronic equipment
TypeBFaccordingtoelectricalsafetyrequirements
Operatingatmosphericpressurelimitation0’to8,000
Storagetemperaturelimitation-4°Fto140°F
(-20°Cto60°C)
Humiditylimitation5%to93%±2%non-condensing
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
Usenooilorgrease
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
iv
DL-00096 Rev. 7
DL-00096 rev 7.indd 4 6/1/2009 4:48:23 PM
v
DL-00096 Rev. 7
Donotuseifpackagingisdamaged
Handle with care
Dateofmanufacture
Manufacturernameandaddress
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
DL-00122
Revision: 3
ECO 10180
Print Colors:
Light Grey- Match GE Plast ics GY7D412 (Background), approx PMS427C
Yellow- PMS 123C (Symbol)
Black- PMS Process Black (Symbol & Text)
Red- PMS 485C (Symbols)
6.7A
CAUTION:
3
Consult
accompanying
documents
Keep
Dry
Do not
Disassemble
No Oil
Do Not Remove Filter
When Unit is Operating
WARNING:
or Grease
No
Smoking
No Open
Flame
Artwork, Product Label, Matchbox
= 90% ± 3% Oxygen
Keepdry(ThissymbolreferstotheIPX2
classificationofthedevice)
Disposeofusedbatteryproperly
Donotremovefilterwhileunitisoperating
Medical Equipment
UL60601-1 CAN/CSA C22.2 No. 601.1
See Accompanying Documents
3RUE
Medicalequipmentwithrespecttoelectricshock,
fire,andmechanicalhazardsinaccordancewithUL
60601-1,CAN/CSAC22.2No.601.1
WARNING!Donotusethisdevicewithoutthe
carryingcase.Useofthisdeviceoutsideitscarrying
casemayresultinanelectricshockhazard.
CautionHotSurface
DL-00096 rev 7.indd 5 6/1/2009 4:48:25 PM
vi
DL-00096 Rev. 7
Introduction
MANUAL VERSION
Pleaserefertothismanualfordetailedinstructionsonwarnings,precautions,
specifications,andadditionalinformation.PrintedinU.S.A.
ElectroniccopiesofthecurrentversionofthePortable Oxygen
Concentrator User Manual canbefoundatwww.delphimedical.com.
GENERAL INFORMATION
ThisusermanualprovidesinformationforusersofthePortableOxygen
Concentrator.Forthesakeofbrevity,theterms“concentrator,”“unit,”
or“device”aresometimesusedinthisdocumenttorefertothePortable
OxygenConcentrator.
WARNING:
STATEMENTS THAT DESCRIBE SERIOUS ADVERSE REACTIONS
AND POTENTIAL SAFETY HAZARDS.
PRECAUTION:
STATEMENTS THAT CALL ATTENTION TO INFORMATION REGARDING
ANY SPECIAL CARE TO BE EXERCISED BY THE PRACTITIONER AND/
OR PATIENT FOR THE SAFE AND EFFECTIVE USE OF THE DEVICE.
IMPORTANT:
Statements calling attention to additional significant information
about the device or a procedure.
IMPORTANT:
Users should read this entire manual before operating the Portable
Oxygen Concentrator. Failure to do so could result in personal injury
and/or death. If you have questions about the information in this
user manual or about the safe operation of this system, contact
Customer Service.
TYPOGRAPHICAL CONVENTIONS
These instructions contain warnings, precautions, and notes to help call
attentiontothemostimportantsafetyandoperationalaspectsofthesystem.
Tohelpidentifytheseitemswhentheyoccurinthetext,theyareshown
usingthefollowingtypographicalconventions:
Note: Statements that provide supplemental information.
DL-00096 rev 7.indd 6 6/1/2009 4:48:25 PM
vii
DL-00096 Rev. 7
INDICATIONS FOR USE
ThePortableOxygenConcentratorisintendedtoprovidesupplemental
oxygeninahome,institutional,ortravelenvironment.
INTENDED USE
ThePortableOxygenConcentratorisintendedtodeliverconcentratedoxygen
for adult patients with chronic pulmonary diseases such as chronic bronchitis,
emphysema, asthma, or lung cancer, those in the terminal stage of cancer, or
anypatientrequiringsupplementaloxygen.Thedeviceisportable,enabling
patientswhoneedanoxygendevicetobetreatedathomeaccordingtoa
clinician’sprescriptionordirection.ThePortableOxygenConcentratorisa
prescription-onlydevice,isnotintendedforuseinlife-supportingor
life-sustainingsituations,andisprovidednon-sterile.Itisdesignedtobe
usedindoorsoroutdoors.
CONTRAINDICATIONS
ThePortableOxygenConcentratorisnotintendedtobeused:
•inlife-supportingorlife-sustainingsituations
•inanoperatingorsurgicalenvironment
•withanon-adultpopulation
•inconjunctionwithflammableanestheticorflammablematerials
DL-00096 rev 7.indd 7 6/1/2009 4:48:25 PM
viii
DL-00096 Rev. 7
Warnings and Precautions
WARNINGS OVERVIEW
PRECAUTION:
1. U.S.FEDERALLAWRESTRICTSTHISDEVICETOSALEBYOR
ONTHEORDEROFAPHYSICIAN.
2. ITISRECOMMENDEDANALTERNATESOURCEOFOXYGEN
BEAVAILABLEINCASEOFPOWEROUTAGEORMECHANICAL
FAILURE.CONSULTPROVIDERORCLINICIANFORAN
APPROPRIATEBACKUPSYSTEM.
3. NON-PRESCRIBEDOXYGENTHERAPYCANBEHAZARDOUS
UNDERCERTAINCIRCUMSTANCES.USETHISDEVICEONLY
WHENPRESCRIBEDBYACLINICIAN.
PRECAUTIONS OVERVIEW
WARNING:
1. CRITICAL! EXPLOSION HAZARD. DO NOT USE IN THE
PRESENCE OF FLAMMABLE ANESTHETICS!
2. DO NOT USE A PORTABLE OXYGEN CONCENTRATOR OR ANY
ACCESSORY THAT SHOWS ANY SIGN OF DAMAGE.
3. DO NOT ALLOW SMOKING OR OPEN FLAMES NEAR THIS
DEVICE. DO NOT USE THIS DEVICE IN THE PRESENCE OF
POLLUTANTS OR FUMES.
4. DO NOT SUBMERGE THIS DEVICE IN LIQUID. DO NOT EXPOSE
TO WATER OR PRECIPITATION. DO NOT EXPOSE TO DUSTY
CONDITIONS.
5. DO NOT USE LUBRICANTS ON THIS DEVICE OR ANY OF ITS
ACCESSORIES.
6. ALWAYS ENSURE BATTERY IS INSERTED BEFORE USING THIS
DEVICE.
7. IF FEELING ILL OR EXPERIENCING DISCOMFORT WHILE USING
THIS DEVICE, CONTACT CLINICIAN IMMEDIATELY.
8. DO NOT DISASSEMBLE THIS DEVICE OR ANY OF ITS
ACCESSORIES. DO NOT ATTEMPT ANY MAINTENANCE OTHER
THAN TASKS DESCRIBED IN “TROUBLESHOOTING” (PAGE 20).
DISASSEMBLY CAN CREATE AN ELECTRIC SHOCK HAZARD
AND WILL VOID THE WARRANTY. CONTACT CUSTOMER
SERVICE FOR SERVICING BY AUTHORIZED PERSONNEL.
9. RISK OF ELECTRIC SHOCK. DO NOT OPERATE THIS DEVICE
WITHOUT THE CARRYING CASE.
10. THIS DEVICE IS FOR CONTINUOUS OPERATION AT
TEMPERATURES OF 95°F (35°C) DOWN TO 41°F (5°C), AND
SHORT-TIME OPERATION FOR 25 MINUTES AT TEMPERATURES
UP TO 104°F (40°C) AND 93% ± 2% RELATIVE HUMIDITY.
DL-00096 rev 7.indd 8 6/1/2009 4:48:25 PM
ix
DL-00096 Rev. 7
PRECAUTION:
4. IFYOUAREUNABLETOHEARORSEEALARMS,DONOTHAVE
NORMALTACTILESENSITIVITY,ORCANNOTCOMMUNICATE
DISCOMFORT,CONSULTCLINICIANBEFOREUSINGTHIS
DEVICE.
5. THISDEVICEISNOTDESIGNEDFORUSEWITHAHUMIDIFIER
ORNEBULIZER.IFAHUMIDIFIERORNEBULIZERISUSEDWITH
THISDEVICE,PERFORMANCEMAYBEDIMINISHEDANDTHE
DEVICEMAYBEDAMAGED.
6. PATIENTSWITHAFASTBREATHINGRATEREQUIRINGA
HIGHEROXYGENSETTINGMAYREQUIREMOREOXYGEN
THANTHISDEVICECANPRODUCE.THISDEVICEMAYNOT
BEAPPROPRIATEINTHATCASE.CONSULTCLINICIANFOR
ALTERNATIVETREATMENT.
7. ONLYUSEAPPROVEDACCESSORIESWITHTHISDEVICE.REFER
TOTHEAPPROVEDACCESSORIESGUIDEFORACOMPLETE
LISTOFACCESSORIESANDCANNULAAPPROVEDFORUSE
WITHTHISDEVICE.USINGUNAPPROVEDACCESSORIESOR
CANNULAMAYIMPAIRTHEPERFORMANCEOFTHISDEVICE.
8. REPLACETHECANNULAONAREGULARBASIS.CHECKWITH
CUSTOMERSERVICEORCLINICIANTODETERMINEHOW
OFTENTHECANNULASHOULDBEREPLACED.
9. ELECTRICALCORDAND/ORTUBINGLYINGONTHEFLOOR
COULDPRESENTATRIPPINGHAZARD.
10.NEVERLEAVETHISDEVICEINAHOTENVIRONMENT.NEVER
LEAVETHISDEVICEINALOW-TEMPERATUREENVIRONMENT.
EXTREMEHIGHORLOWTEMPERATURESCANDAMAGETHIS
DEVICE.
11.IFOXYGENCONCENTRATIONDROPSBELOWTHESPECIFIED
LEVEL,ANALARMWILLINDICATETHISCONDITION.IFALARM
PERSISTS,STOPUSINGTHISDEVICE,SWITCHTOAN
ALTERNATESOURCEOFOXYGEN,ANDCONTACTCUSTOMER
SERVICE.
12.DONOTUSECLEANINGAGENTSOTHERTHANTHOSE
SPECIFIEDINTHISMANUAL.ALLOWTHECLEANINGSOLUTION
TODRYFROMTHECLEANEDSURFACEBEFOREUSE.
13.ALWAYSDISCONNECTPOWERANDTURNOFFTHISDEVICE
BEFORECLEANING.SEE“MAINTENANCEANDCLEANING”
(PAGE12).
14.DONOTOBSTRUCTAIRINTAKEOREXHAUSTWHENOPERATING
THISDEVICE.BLOCKAGECANCAUSEBUILDUPOFINTERNAL
HEATANDSHUTDOWNORDAMAGETHISDEVICE.
15.ALWAYSUSEINAWELL-VENTILATEDLOCATION.
16.DONOTOPERATETHISDEVICEWITHOUTTHEINPUTFILTER
INPLACE.IFFILTERISREMOVED,PARTICLESMAYBEDRAWN
INTOTHESYSTEMANDMAYDAMAGETHISDEVICE.
DL-00096 rev 7.indd 9 6/1/2009 4:48:25 PM
x
DL-00096 Rev. 7
PRECAUTION:
17.ALWAYSOPERATEDEVICEATTHESETTINGPRESCRIBEDBYA
CLINICIAN.DONOTALTERTHESETTINGUNLESSPRESCRIBED
BYACLINICIAN.
18.THISDEVICEISDESIGNEDFORUSEBYONEPATIENTATA
TIME.
19.THISDEVICEMAYNOTREACHSPECIFIEDOXYGEN
CONCENTRATIONPURITYUNTILITHASBEENINUSEFORUP
TO10MINUTES.
20.ALWAYSFOLLOWTHEMAINTENANCESCHEDULEAS
SPECIFIEDIN“ROUTINEMAINTENANCE”(PAGE13).
21.IFTHISDEVICEINDICATESANABNORMALCONDITION,SEE
“TROUBLESHOOTING”(PAGE20).
22.REMOVEBATTERYIFTHISDEVICEISNOTGOINGTOBEUSED
FORMORETHANSEVENDAYS.STOREBATTERYINACOOL,
DRYPLACE.
23.ALWAYSFOLLOWCANNULAMANUFACTURER’SINSTRUCTIONS
FORPROPERUSE.
24.ALWAYSTURNOFFTHISDEVICEWHENNOTINUSE.
25.DONOTUSETHISDEVICEWHILESLEEPINGUNLESSUNDER
THESUPERVISIONOFACLINICIAN.
26.DONOTPLACEOBJECTSONTOPOFTHISDEVICE.
27.THISDEVICEISRATEDIPX2.DONOTUSEINDUSTY
CONDITIONS.
28.USECAUTIONWHENTOUCHINGTHISDEVICEINHIGHAMBIENT
TEMPERATURES.
29.CHECKTHATTHISDEVICEOPERATESONBATTERYAFTER
DISCONNECTINGFROMTHEPOWERSOURCE.
30.ONLYCHARGEBATTERYINTHISDEVICEORINANAPPROVED
CHARGER.(SEEAPPROVEDACCESSORIESGUIDE.)
IMPORTANT:
1. If an extension cord is necessary, use a UL listed 15 amp or higher
cord. Do not connect any other devices on the same extension cord.
2. Inhale through nose for the concentrator to work most effectively.
Inhaling through mouth may result in less effective oxygen therapy.
3. This device utilizes an oxygen-conserving delivery method, and the
setting on this device may differ from your setting on a continuous
flow oxygen concentrator (if you have been prescribed one). Your
clinician will provide you with specific instructions based on your
specific health condition and other variables.
DL-00096 rev 7.indd 10 6/1/2009 4:48:25 PM
xi
DL-00096 Rev. 7
Contents
Classification........................................................................................ iii
ExplanationofPackagingandLabelingSymbols................................. iv
Introduction.......................................................................................... vi
ManualVersion............................................................................. vi
GeneralInformation...................................................................... vi
TypographicalConventions........................................................... vi
IndicationsforUse........................................................................vii
IntendedUse.................................................................................vii
Contraindications...........................................................................vii
WarningsandPrecautions...................................................................viii
WarningsOverview......................................................................viii
PrecautionsOverview..................................................................viii
BeforeYouGetStarted.........................................................................1
PowerSupply.................................................................................... 1
ChargingtheBattery...................................................................... 2
FeaturesandControls........................................................................... 4
Accessories.................................................................................... 6
CarryingCase.......................................................................... 6
PullCart................................................................................... 7
OperatingthePortableOxygenConcentrator...................................... 9
ConnectingNasalCannula................................................................ 9
TurningOn........................................................................................10
AdjustingSetting..............................................................................10
RespondingtoAlarms......................................................................11
TurningOff.......................................................................................11
TravelingbyAirWiththePortableOxygenConcentrator................11
MaintenanceandCleaning..................................................................12
VerifyingtheAlarmSystem.............................................................12
BatteryCare.....................................................................................12
Environment/Storage.......................................................................12
RoutineMaintenance.......................................................................13
CleaningandChangingFilters......................................................13
InputFilter..............................................................................13
PatientFilter...........................................................................14
CompressorFilter...................................................................14
ExteriorCleaning..............................................................................15
AccessoryCleaning..........................................................................15
DeviceRepair...................................................................................15
DL-00096 rev 7.indd 11 6/1/2009 4:48:25 PM
xii
DL-00096 Rev. 7
DeviceDisposal....................................................................................15
AlarmIndicators...................................................................................16
Alarms..............................................................................................16
CriticalHighPriorityAlarms..........................................................16
HighPriorityAlarms......................................................................17
MediumPriorityAlarms................................................................17
LowPriorityAlarms......................................................................18
OtherMessages...........................................................................18
Troubleshooting...................................................................................20
Appendix1:TechnicalDescription...................................................... 21
Appendix2:TechnicalInformation......................................................22
Trademarks..........................................................................................25
Disclaimer............................................................................................25
AirlineTravelGuidelines......................................................................26
DL-00096 rev 7.indd 12 6/1/2009 4:48:25 PM
Before You Get Started
Alwaysinspectthedeviceanditsaccessoriesforanysignofdamagebeforeusing.
Note: While box or packaging may exhibit some damage, e.g., tears or dents, the device may
still be in a usable condition.
Ifthedeviceoranyaccessoryshowsanysignofdamage,contactyour
careprovider.
Beforeyougetstarted,checktomakesureyouhavethefollowing:
•Concentrator
•Battery
•Carryingcase
•AC/DCpowersupply
SeeApprovedAccessoriesGuideformoreinformation.
WARNING:
DO NOT USE A PORTABLE OXYGEN CONCENTRATOR OR ANY
ACCESSORY THAT SHOWS ANY SIGN OF DAMAGE.
POWER SUPPLY
ThePortableOxygenConcentratorcanalwaysbeusedwhendirectly
connectedtoapowersource.However,toenhanceitsportability,the
PortableOxygenConcentratorisequippedwitharechargeablelithium-ion
internalbattery.
WARNING:
ALWAYS ENSURE BATTERY IS INSERTED BEFORE USING THIS
DEVICE.
1
DL-00096 Rev. 7
Note: Optional power supplies are available for various global use and travel. See Approved
Accessories Guide for complete list.
DL-00096 rev 7.indd 1 6/1/2009 4:48:26 PM
2
DL-00096 Rev. 7
Charging the Battery
PriortousingthePortableOxygenConcentratorforthefirsttime,installthe
batteryasshowninFigure1.Thebatterywilllatchwhenfullyseated.
Figure 1.
PRECAUTION:
ONLYCHARGEBATTERYINTHISDEVICEORINANAPPROVED
CHARGER.(SEEAPPROVEDACCESSORIESGUIDE.)
ConnecttheACpowersupply(consultApprovedAccessoriesGuide)by
plugging the round connector into the receptacle on the side of the concen-
trator.SeeFigure2.PlugtheotherendoftheACpowersupplyintoapower
outlet.Alwaysusecautionwheninsertingthepowersupplytoawalloutlet.
Chargerrequirementsareuniversaltosupportawidevarietyofinternational
markets,soitcanbepluggedintoanoutletwith100-240VAC,50-60Hz.
Allowthebatterytochargeforaminimumofthreehoursbeforeuse.Once
thebatteryiscompletelycharged,thedevicecanrunonbatteryforapproxi-
mately2-3hoursatasettingof2.0.
Note: Battery run time may vary based on breathing rate, age of battery, and environmental
conditions. See displayed text on device for battery charge status.
DL-00096 rev 7.indd 2 6/1/2009 4:48:26 PM
DL-00096 Rev. 7
3
Figure 2.
Note: Ensure power status icon (see Figure 6, page 5) indicates power is connected. If not, check
that cord is plugged in completely. (See Troubleshooting” on page 20 for more information.)
Tomaximizebatterylifeandruntime,usewhileconnectedtoapower
sourcewheneverpossible.Theinternalbatterywillautomaticallycharge
whenevertheconcentratorisconnectedtoapowersource.TheLCD
displaywillindicatewhetherthedeviceisoperatingonbatteryorexternal
ACpower.
Thefullychargedbatterywillretainsomelevelofchargeforuptoseven
daysinthisdevicewhennotinuse.
Note: Battery damage may result if the concentrator’s battery is allowed to discharge completely.
PRECAUTION:
REMOVEBATTERYIFTHISDEVICEISNOTGOINGTOBEUSEDFOR MORE
THANSEVENDAYS.STOREBATTERYINACOOL,DRYPLACE.
PRECAUTION:
CHECKTHATTHISDEVICEOPERATESONBATTERYAFTER
DISCONNECTINGFROMTHEPOWERSOURCE.
Note: When not using the battery inside the unit, be sure to store it in the protective sleeve that
was provided with the original package.
DL-00096 rev 7.indd 3 6/1/2009 4:48:27 PM
4
DL-00096 Rev. 7
Display screen
Features and Controls
Figure 3.
Figure 4.
Power key – turns device on or off
Previous/next
key – changes
setting
Mode key – changes
display information
Alarm mute key – temporarily
silences audible alarms or shows
the last highest priority alarm
Cannula port
Green LED
indicates
device is on
Yellow LED
indicates caution or
attention required
Red LED – indicates
a warning of danger
and/or a need for
urgent action
Alarm mute indicator – indicates alarm buzzer
has been muted
Power supply port
Patient filter
Exhaust
DL-00096 rev 7.indd 4 6/1/2009 4:48:28 PM
5
DL-00096 Rev. 7
Figure 5.
Input filter
Battery
Compressor filter
Air intake
Power status (battery)
Power status (external)
Figure 6.
Setting
Setting
DL-00096 rev 7.indd 5 6/1/2009 4:48:30 PM
6
DL-00096 Rev. 7
ACCESSORIES
AvarietyofaccessoriescanenhancetheportabilityanduseofthePortable
OxygenConcentrator.Inadditiontothedevice,thepackagecontains
accessoriestogetstartedandinstructionsforuse.RefertotheApproved
AccessoriesGuideforthecompletelistofavailableaccessories.
Only use a nasal cannula with the following specifications:
•7’(2.1m)long
•Highflow
•Three-fluted
•Crush-resistant
•Largeinternaldiameterbore
•Straightnon-taperedtips
•Suitableforupto10lpm
Note: Recommended model is Salter Labs 1600 HF cannula.
PRECAUTION:
ONLYUSEAPPROVEDACCESSORIESWITHTHISDEVICE.REFERTO
THEAPPROVEDACCESSORIESGUIDEFORACOMPLETELISTOF
ACCESSORIESANDCANNULAAPPROVEDFORUSEWITHTHISDEVICE.
USINGUNAPPROVEDACCESSORIESORCANNULAMAYIMPAIRTHE
PERFORMANCEOFTHISDEVICE.
ContactCustomerServiceforupdatedinformationandaccessoriesorif
additional,optional,orreplacementaccessoriesareneeded.
Carrying Case
Toreducetheriskofelectricshock,alwaysusethePortableOxygen
Concentratorwithacarryingcase(whichalsofurtherreducesthedevicenoise
duringoperation).Insertthedeviceinthecasesuchthatthedisplayscreenis
visiblethroughtheclearplasticwindowandsecurelyfastenallstraps.When
properlyinserted,itshouldlooklikeFigure7.Whenusingthecarryingcase,
donotcarrytheconcentratorbyitshandle;usethecase’scarryingstrap.
WARNING:
RISK OF ELECTRIC SHOCK. DO NOT OPERATE THIS DEVICE WITHOUT
THE CARRYING CASE.
DL-00096 rev 7.indd 6 6/1/2009 4:48:30 PM
7
DL-00096 Rev. 7
Figure 7.
Pull Cart
WhenusingthePortableOxygenConcentratorwithapullcart,attachand
securetheconcentratorwiththestrapsasshowninFigure8.Thehandlecan
bepulledoutandadjustedforcomfort.
Figure 8.
Note: It is recommended that patients use the pull cart to transport the device whenever possible.
DL-00096 rev 7.indd 7 6/1/2009 4:48:31 PM
8
DL-00096 Rev. 7
Intentionally blank
DL-00096 rev 7.indd 8 6/1/2009 4:48:31 PM
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40

Delphi DL-00183 User manual

Category
Oxygen Equipment
Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI