Tank Drain Assembly
Gate Valve
Facility Drain
OR
Open
Drain Pipe
Closed
Vented Drain
Drain/Sewer Connection
MT10 Tank
From 1/4-inch
Vent Condensation
Drain Port
Note:Note: Install 10MM OD Urethane Tubing
(P/N 54509) between the drip leg and
facility sewer drain.
Heat Exhaust Vent Connection
Heat Exchanger
Clamps
P/N 89324
Vent Hose
P/N 56057
Exhaust Vent
Connection
RJ-45
Connector
Electrical Box
Detail
Power Switch with
Circuit Breaker
Pump Direct Power & RJ-45 Control/Status Connections
Supplied 6-Foot BX Cable
from Pump Electrical Box
Black
(208 - 230V)
White
(208 - 230V)
Green
Yellow
(Ground) Use supplied 10 Foot CAT5e
Network Cable (P/N 419342)
for each Pump to MMC RJ-45
Control/Status Connection
MOJAVE Monitor Master Controller
(MMC-M)
IEC AC Power
Receptacle
MMC Power & Vacuum Port Connections
Sample
Vacuum Port
(1/4-inch Elbow)
Tank Outlet
Sample
Vacuum Port
(1/4-inch Elbow)
1/4 inch OD
Urethane Tubing
(P/N 51453)
To
Hospital Grade
NEMA 5-15 or 6-15
Wall Receptacle
Line Cord
© 2022 Air Techniques, Inc., • P/N H5272M, Rev. A • March 2022
Corporate Headquarters
1295 Walt Whitman Road | Melville, New York 11747- 3062, USA
Phone: 1-800-247-8324 | Fax: 1-888-247-8481
www.airtechniques.com
CT20-M Tank
(Used with three or four pumps.)
Washout Port
Tank Outlet to
Pump
Tank
Inlet from
Operatory
Master
Controller
Tank Drain
Check
Valve Gate Valve
Multiple Pump System Connection
MT10-M Tank
(Used with one or two pumps.)
Dual Pump Tank Outlet
Assembly P/N H5441 with two
check valves H5234 supplied
by MOJAVE Dual System
Installation Kit, P/N MIK2
3 or 4 pumps are connected using a tank outlet manifold
fabricated during site construction with piping supplied by user
and check valves supplied by MOJAVE Triple and Quad System
Installation Kit, P/N MIK4.
From Tank
Outlet
To Pump Air
Inlet Filter
4
6
5
7
Exhaust Vent
Assembly
P/N H5302
Flat Side Installed
Facing Up
Clear Hose to
Drain
Check Valve Assembly Side View
Clear Hose
P/N 54521
Note: Not included in
-NHE models
Each pump is wired directly to a dedicated 230V, 20 AMP
(30 AMP for V7) single phase 50/60 Hz circuit using a
disconnect box or circuit breaker panel with approved
ground.
Check
Valve
Optional Washout Solenoid Installation
Using the two supplied thread forming screws, install the solenoid onto the
rear of the MT10-M Tank.
Install the supplied 3/8 inch tubing from the facility water source to the IN fitting.
Connect the short 3/8 inch tubing between the OUT fitting and the Washout
Fitting on top of the tank.
Washout Solenoid
is Pre-Wired to
the MMC.
OUT to Washout
Fitting
IN from Facility
Water
Flow Direction
3/8-inch
Washout Port
to Washout
Solenoid