Testo 171-3 Short Operation Manual

Type
Short Operation Manual

This manual is also suitable for

Whilst every effort has been made to ensure the accuracy of the information given in this short manual, no liability is given or implied for any error or
omission contained herein. Due to our policy of improvement, Instrument Solutions reserve the right to change products and/or services without notice.
Date :- 09 January 2009
Testo 175-H2 177-T2 171-3 short user manual.doc
SHORT OPERATIONS MANUAL
BSRIA Instrument Solutions
Old Bracknell Lane West,
Bracknell,
Berkshire,
RG12 7AH
Tel: +44 (0) 1344 459314
Fax: +44 (0) 1344 465556
E mail : info@bis.fm
Manufacturer:- TESTO
Model : 171, 175 & 177 Series
1. INTRODUCTION
The Testo 171, 175 & 177 series of data loggers are
precision instruments designed to measure and record
various parameters including temperature / relative humidity.
Parameters measured are dependant upon model. The
following short form instruction manual covers the basic
operational aspects of the above instruments, for a full instruction manual please contact us.
2. CHOOSING THE LOCATION FOR YOUR INSTRUMENT (IMPORTANT PLEASE READ)
For ambient temperature / relative humidity models the instrument should be sited at a location that is
typical of the ambient conditions and is not affected by any external temperature / humidity sources.
Avoid placing the instrument in or near the following locations:- . Bright sunlight, Air conditioning
inlets & outlets, radiators, computers / monitors or other office equipment (that may give off heat), damp
corners, by fans or sinks and any draughts in general. Also avoid handling the instrument as this may
also affect its readings.
THE INSTRUMENT SHOULD BE KEPT DRY AT ALL TIMES.
For ambient condition monitor logger models, the instrument will take approximately 30min to 1 hour to
acclimatise.
3. LOGGER IDENTIFICATION (IMPORTANT PLEASE READ)
When returning your rental logger please ensure you keep record of where the loggers have been placed.
BSRIA Instrument Solutions rental instruments are identified by two sets of numbers, a group code and a
unique six digit ID e.g. 200123 it is recommended that you use this 6 digit ID code to identify the
instruments location.
Please do not affix any other labels to our instruments.
4. CONFIGURING THE INSTRUMENT
Sections 5 & 6 deal with the various set up options i.e. starting modes, Stopping modes, logging intervals
etc available. These configurations can only be changed using the optional software. BSRIA Instrument
Solutions can pre configure the instrument to the customers requirements prior to the start of the hire at no
extra cost.
Unless otherwise requested by the customer the DEFAULT setup for the instrument is as shown below:-
SAMPLE RATE =
START MODE =
STOP MODE =
5mins,
“KEYPAD” for 177 & 175s and “MAGNETIC” start for 171’s .
Until memory is full
171-3 175-H2 175-S1
BSRIA Instrument Solutions
Quick Start Guide Page 2 of 2
Whilst every effort has been made to ensure the accuracy of the information given in this quick start guide, no liability is given or implied for any
error or omission contained herein. Due to our policy of progressive improvement, BSRIA Instrument Solutions reserve the right to change
products and/or services without notice.
Date :- 09 January 2009
K:\Hire Product Manuals\TESTO\Testo 171 175 177\Testo 175-H2 177-T2 171-3 short user manual.doc
Where the customer has specified that the logger be pre configured for a specific start method and/or
logging interval a note stating this will be enclosed with the instrument/s. It is the responsibility of the hirer
to ensure that the correct starting procedure is followed. BSRIA Instrument Solutions accept no liability for
loggers that have been incorrectly started.
5. STARTING THE LOGGER
The instrument can be configured, using SOFTWARE, to start recording by either of the following ways,
DATE/TIME, KEYPAD (175 / 177 models only), MAGNET (171 models only)
For 175/177 logger models ONCE STARTED REC” or RECORD” is shown on its LC display.
For 171 logger models IMMEDIATELY AFTER STARTING the instrument will beep 3 times. Once
started the logger will only beep once every logging interval. No other indications of the
loggers recording state are evident unless connected to a PC. This beep is very quiet therefore
it is recommended that the user starts the logger in a quiet environment.
DATE/TIME START MODE:- Using software, a start time & Date can be set. Once programmed with this
start mode the logger will start automatically, no further action is required. REC or RECORD is shown on
175/177 models ONCE recording.
KEYPAD START MODE*:- For 177 & 175 loggers (ONLY) the instrument may be configured through
software to start recording using its keypad. Where the instrument has been configured this way TO
START the logger PRESS & HOLD the “GO” button for approx’ 3 - 5 sec’s until REC or RECORD
is displayed.
MAGNETIC START MODE*:- For 171 loggers (ONLY) the instrument may be configured through
software to start recording using a magnet. Magnets can be supplied free of charge on request. Once
configured this way the user simply has to hold the magnet over the TESTO logo located on the
instruments front. If successful the instrument will beep 3 times. No other indications of the loggers
recording state are evident unless connected to a PC.
6. STOPPING THE LOGGER
The logger can be configured (using software) to stop by either of the following ways, UNTIL MEMORY IS
FULL, NUMBER OF LOGS, WRAP AROUND MEMORY, DATE / TIME.
Once stopped the logger can only be restarted using software or the optional Testo 580 Data Stick or a
Test575 printer (only applicable for 175/177 models only).
IMPORTANT:- When restarted the data already in the logger is permanently erased and can not
be recovered.
UNTIL MEMORY IS FULL STOP MODE*:- The logger stops when its memory is full.
NO. OF LOGS STOP MODE:- The logger stops after a predefined number of records.
WRAP-AROUND MEMORY STOP MODE:- The logger never stops, it will record over the oldest data
once it has filled its memory. Once data has been recorded over it can’t be recovered.
DATE / TIME STOP MODE:- The logger stops at a pre defined date / time.
------ End of Manual -----