OPERATION
2-4
2.3 Power On/Off
To turn on the MiniRAE 2000 portable VOC monitor, press
[MODE] key for one second and release. The audio buzzer will
beep once and the air pump will turn on. The display will show
“ON!..” and then “Ver n.nn” to indicate the unit’s current
firmware version number. Next displayed are the serial number,
the model number, Operating mode, current date and time, unit
internal temperature, gas selected, high low, STEL, TWA/AVG
alarm limits, battery voltage, and shut off voltage. Also
displayed are internal mode settings such as User mode, Alarm
mode, datalog time remaining and log periods in the respective
order.
To turn off the MiniRAE 2000 portable VOC monitor, press
and hold the [MODE] key for 5 seconds. The monitor will beep
once per second during the power-down sequence with a count
down timer showing the number of remaining seconds. The
message “Off!..” flashes on the LCD display and the display
will go blank indicating that the monitor is turned off.
Data protection during power off
When the monitor is turned off, all the current real time data
including last measured value are erased. However, the datalog
data is preserved in non-volatile memory. Even if the battery is
disconnected, the datalog data will not be lost. While the power
is off, the real time clock will continue to operate until the
battery is completely drained (usually in 4-5 days without any
charging). If the battery is completely drained or is
disconnected from the monitor for more than 30 minutes, the
real time clock will be lost. In this case, the user needs to enter
the real time clock information again, as described in Section 4,
or send the PC clock during configuration through the PC
communication.