Dwyer Series PLS2 User manual

Type
User manual

Dwyer Series PLS2 Below you will find brief product information for the Dwyer Series PLS2, a versatile electromechanical level switch designed for level monitoring of bulk materials. The PLS2's capabilities include:

  • 3 sensitivity settings for customized performance
  • Both top or side mountable, for flexible installation
  • Brushless synchronous motor, ensuring long-term reliability and efficiency
  • Automatic motor shutoff when the paddle stalls, preventing damage
  • Screw cover for easy access without the hassle of misplaced bolts or screws, simplifying maintenance
  • Wide compatibility with dry powder or bulk materials, making it suitable for various industrial applications

Dwyer Series PLS2 Below you will find brief product information for the Dwyer Series PLS2, a versatile electromechanical level switch designed for level monitoring of bulk materials. The PLS2's capabilities include:

  • 3 sensitivity settings for customized performance
  • Both top or side mountable, for flexible installation
  • Brushless synchronous motor, ensuring long-term reliability and efficiency
  • Automatic motor shutoff when the paddle stalls, preventing damage
  • Screw cover for easy access without the hassle of misplaced bolts or screws, simplifying maintenance
  • Wide compatibility with dry powder or bulk materials, making it suitable for various industrial applications
The Series PLS2 is an electromechanical level switch designed for level
monitoring of bulk materials. The rotating measuring vane is driven by a
brushless synchronous motor at one revolution per minute. As product
builds up the paddle rotation is impeded and the resulting motor torque
activates the output switch and stops the motor. A spring mechanism
reactivates the motor and returns the switch to normal state when the
material no longer impedes the paddle rotation. A torque adjusting
mechanism eliminates the need for different sized paddles. The PLS2 is
designed with the industry standard 1-1/4˝ male NPT connection and unit
can be side or top mounted.
FEATURES
• 3 sensitivity settings
• Top or side mountable
• Brushless synchronous motor: Assures long term reliability and
efficiency
• Motor shuts off when paddle stalls
• Screw cover: For easy access with no worries about losing bolts or
screws
TECHNICAL DATA
Model PLS2 Paddle Level Switch
Specifications - Installation and Operating Instructions
Bulletin L-PLS2
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000 www.dwyer-inst.com
P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: [email protected]
SPECIFICATIONS
Service: Dry powder or bulk materials compatible with wetted
materials.
Wetted Materials:
Paddle: 304 SS;
Exposed shaft: 303 SS;
Shaft seal: NBR;
Process connection: Aluminum.
Temperature Limits:
Process: -13 to 176°F (-25 to 80°C);
Ambient: -4 to 140°F (-20 to 60°C).
Pressure Limit: 11.6 psi (0.8 bar).
Power Requirement: Select by model number: 115 VAC, 230 VAC or
24 VDC.
Power Consumption: AC version: 4 VA; DC version: 2.5 watt.
Enclosure: Aluminum, powder coated.
Enclosure Rating: NEMA 4 (IP66) and rated for Class II & III, Div. 1,
Group E, F, G.
Switch Type: SPDT micro switch.
Electrical Rating: 5 A @ 250 VAC, 3 A @ 30 VDC.
Electric Connections: Screw terminals.
Conduit Connection: 3/4˝ female NPT.
Process Connection: 1-1/4˝ male NPT.
Mounting Orientation: Side or top mounting.
Sensitivity: Minimum material density of 6 lb/ft
3
(96 kg/m
3
).
Permitted Mechanical Loading:
Standard shaft: 67 lbf (300 N) max;
Optional extension: 22 lbf (100 N) max.
Weight: 2.6 lb (1.2 kg).
Agency Approvals: CE, FM.
5-1/8
[130.00]
5
-29/32
[
150.00]
3-5/8
[92.00]
2
5/64
[
10.00]
1-1/4˝
NPT
PLS2-E-1-X
2X 3/4˝
NPT
4-23/32
[120.00]
3-15/61
[100.00]
1-3/8
[35.00]
7-7/8
[200.06]
P
LS2-PL2
PLS2-PL1
3
-15/16
[
100.00]
3-15/16
[100.00]
1
-29/64
[
37.00]
4
-59/64
[
125.00]
INSTALLATION
The PLS2 should be mounted on the side or top of the tank or bin. The
conduit opening should be placed in a downward position, to protect from
moisture entering the unit through the conduit. When selecting a mounting
location, be sure the material can freely flow to and away from the shaft and
paddle. The shaft and paddle should also be out of the direct flow of material
as it fills the bin or tank. If it is necessary to mount in the flow stream, a
protective baffle should be installed on the inner wall of the bin.
Wiring
Make electrical connections per the wiring diagram for output switch options
selected. Wire in accordance with local codes. For hazardous locations, a
sealing fitting must be located within 18˝ of the control.
ASSEMBLY
MOUNTING PROCEDURE
ADJUSTMENT OF THE SPRING FORCE
MAINTENANCE/REPAIR
Upon final installation of the Series PLS2, no routine maintenance is
required. The Series PLS2 is not field serviceable and should be returned
if repair is needed. Field repair should not be attempted and may void
warranty.
WARRANTY/RETURN
Refer to “Terms and Conditions of Sales” in our catalog and on our website.
Contact customer service to receive a Return Goods Authorization number
before shipping the product back for repair. Be sure to include a brief
description of the problem plus any additional application notes.
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000 www.dwyer-inst.com
P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: [email protected]
©Copyright 2014 Dwyer Instruments, Inc. Printed in U.S.A. 7/14 FR# R6-443903-00 Rev. 2
  • Page 1 1
  • Page 2 2

Dwyer Series PLS2 User manual

Type
User manual

Dwyer Series PLS2 Below you will find brief product information for the Dwyer Series PLS2, a versatile electromechanical level switch designed for level monitoring of bulk materials. The PLS2's capabilities include:

  • 3 sensitivity settings for customized performance
  • Both top or side mountable, for flexible installation
  • Brushless synchronous motor, ensuring long-term reliability and efficiency
  • Automatic motor shutoff when the paddle stalls, preventing damage
  • Screw cover for easy access without the hassle of misplaced bolts or screws, simplifying maintenance
  • Wide compatibility with dry powder or bulk materials, making it suitable for various industrial applications

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI