Prepare the reagents
• The following reagents are ready for use: 1‑Coated microplate MVV/CAEV, 2‑Negative
Control MVV/CAEV, 3‑Positive Control MVV/CAEV , B1‑Sample DB MVV/CAEV,
B2‑Conjugate diluent MVV/CAEV, C‑Substrate solution MVV/CAEV, and D‑Stop
solution MVV/CAEV.
• A‑Wash (10x) MVV/CAEV must be diluted to 1:10 with distilled or deionized water. The
diluted wash solution can be kept for 8 days at 2–8°C.
Example: For one strip, add 3mL of the A‑Wash (10x) MVV/CAEV to 27 mL of water.
For one plate, add 35mL of the A - Wash (10x) MVV/CAEV to 315 mL of water. Mix after
dilution.
Note: Due to the high concentration of salts, A‑Wash (10x) MVV/CAEV can contain
crystals. Before dilution, shake the bottle to dissolve any crystals.
• 4‑Conjugate (100x) MVV/CAEV must be diluted to 1:100 with the B2‑Conjugate diluent
MVV/CAEV. Mix after dilution. Use the diluted Conjugate MVV/CAEV immediately after
preparation.
Perform the ELISA procedure
IMPORTANT! Specific requirements apply for the use of the kit in diagnostic procedures in
France, see Appendix A, “Recommended protocol in France”.
Note: All reagents should be equilibrated to room temperature before use. It is recommended
to use disposable containers for distribution (diluted Wash, diluted Conjugate MVV/CAEV,
C‑Substrate solution MVV/CAEV, and D‑Stop solution MVV/CAEV).
Use undiluted samples and controls.
Note: After opening a new plastic bag, it is
recommended to immediately put the unused plate or
strips in a plastic bag with the provided desiccant
sachet. Store protected from light.
1. Add 50μL of the 3‑Positive Control MVV/CAEV
reagent to the appropriate wells of the assay plate
(for example, add to A1 and B1 wells).
2. Add 50μL of the 2‑Negative Control MVV/CAEV
reagent to the appropriate wells of the assay plate
(for example, add to C1 and D1 wells).
3. Add 50μL of the test samples to the remaining
wells.
4. Tap or gently agitate each of the long sides of the
loaded assay plate to make sure the samples coat
the bottom of the wells.
5. Cover the plate with an adhesive plate film.
1Incubate the samples
and controls
4 PrioCHECK™ MAEDI-VISNA & CAEV Ab Serum Kit Instructions for Use