Negative Pressure Recording Machine
(ASM- 100)
Users manual
1. Product Feature
1) This machine was designed to measure differential pressure (negative and positive pressure)
[This machine is specialized to measure differential air pressure (Between positive and negative pressure]
2) This machine can measure differential pressure in 0.001 Pa scale with Swiss-made
high-performance static pressure sensor.
3) This machine alarms when average air pressure per 1sec is below -2.5 pa
4) This machine provides user friendly User Interface and display setting
2. Name and function of each part
Component Storage bag for machine, negative pressure recording machine, adaptor, power wires,
measuring hose (silicon)
-1-
-Providing information through touch
screen
-Controlling sound volume through
system management setting
-Controlling on/off
power
Serial port
-Controlling external device
Implementing system Update
Adapter connector
- Battery charging,
connecting to the power
Measuring Sensor
- It measure
Negative pressure,
Differential pressure
and others
Power switch
LCD
Speaker
To charge battery and connect to power(power connection): please connect adapter
connector and power
3. Machine connection(connecting) instruction
Connect to the Measuring hose
1) When measurements are taken with measuring machine is inside of hospital room
(when measuring machine is inside of hospital room)
- After connecting silicon hose to left socket (workplace(external/outside), please position/
place the opposite side of hose outside of the room.
2) When measurements are taken with measuring machine is outside of hospital room
(when measuring machine is outside of hospital room)
- After connecting silicon hose to right socket (workplace(internal/inside), please position
/place the opposite side of hose inside of the room
-2-
4. User manual for display screen
Main menu screen
Switch to the screen for negative
pressure measurement
Switch to the screen for
managing stored data
Please press “save” after
finishing adjustment of time
and date
Select main measurement
scale (primary measurement
scale is Pa)
Please type the name
of hospital
Controlling interval of
data storage setting
-3-
Select the function
(negative pressure, Differential
pressure, temperature)
*Developer Menu
Switch to the screen for
Date, Time, measurement
scale and system setting
menu
Setting current negative
pressure point as “zero point”
Shut down program before
powering off machine
System management screen
Data management screen
Measurement screen
-4-
Showing current data with graph
Deleting current data
Searching previous data with date, address,
name of
workplace
Press “start” to record
measurement
Press “finish” to
finish measurement
Switch to main menu screen
Controlling and connecting to negative
pressure machine
Switch to data management screen
-5-
5. Start to measuring Negative Pressure
Press “measure” button and type workplace (hospital room)
Press “start” button to record and measure negative pressure.
-6-
Press “finish” button to complete measurement
Recording/ Measuring can be paused during measurement
Please be careful for any liquid
This machine is not waterproof therefore any liquid or
moisture may cause damage on components and circuits of
the machine.
If flooded, do not turn on the power and contact the service
center.
If the machine is exposed to fire for a long time, there is a risk
of explosion.
Thank you for purchasing our Asko's product.
The warranty period for products is one year from the date of
purchase according to the Fair Trade Commission Notice consumer
dispute settlement standards.
However, service charges will be levied if the service is subject to
the paid service standard even during the free service period..
Please read and observe the precautions to protect your safety and
prevent property damage.
Please do not repair and disassemble the machine
arbitrarily, except technical representative.
A/S does not be provided when user repair and disassemble
the machine arbitrarily
Please be careful for artificial shock
Do not shock the machine intentionally
Do not push the LCD too hard
Since screen is a touchscreen, excessive force can cause screen
damage.
Safety Precautions
Be careful of the fire
Product Warranty Information
Service Center
Contact us
Unit 401, Seunglim Bidg.. 137 Achasan-ro, Seongdong-gu,
Seoul, Republic of Korea (Post Code : 04797)
Tel. 1855-1191 Home page. www.asmedi.net
E-mail. asmedi@asko119.com
www.asmedi.net
All equipment and consumables produced by Asman Co., Ltd. meet the
relevant statutory standards.