Enraf-Nonius StatUS Pack 400 User manual

Type
User manual
StatUSTM Pack 400
EN109-1629761-43 IFU
November 8, 2022
Instructions for use
Page 2 of 36 EN109-1629761-43 IFU
Table of contents
1 Introduction ...................................................................................................................................................... 3
2 Symbols .............................................................................................................................................................. 4
3 Device components ....................................................................................................................................... 5
4 Package contents ............................................................................................................................................ 6
5 Installation ......................................................................................................................................................... 7
6 Intended use and intended user ............................................................................................................... 8
7 Indications ......................................................................................................................................................... 9
8 Contraindications ......................................................................................................................................... 10
9 Precautionary instructions ........................................................................................................................ 11
10 Operation ........................................................................................................................................................ 13
11 Application information............................................................................................................................. 21
12 Maintenance and troubleshooting ........................................................................................................ 23
13 Specifications ................................................................................................................................................. 26
14 Contact ............................................................................................................................................................. 29
15 Product liability ............................................................................................................................................. 29
16 Appendix 1 Library ...................................................................................................................................... 30
Page 3 of 36 EN109-1629761-43 IFU
1 Introduction
Foreword
This manual has been written for the owners and operators of the StatUS™ Pack 400.
It contains general instructions on operation, precautionary practices, maintenance and parts
information. In order to maximize the use, efficiency and lifespan of your unit, please read this manual
thoroughly and become familiar with the controls as well as the accessories before operating the unit.
Specifications put forth in this manual were in effect at the time of publication. However, owing to
Enraf-Nonius BV’s policy of continual improvement, changes to these specifications may be made at
any time without obligation on the part of Enraf-Nonius BV.
It is important to read the following instructions carefully before using this device
properly and safely. The manufacturer cannot be held responsible for the results of using
this device for any purposes other than described in these operating instructions.
Device Description
The StatUS™ Pack 400 is a feature (accessory) for the application of StatUS™ therapy and works only
when connected to a Sonopuls 490 / 492 device. StatUS stands for Stationary Ultra Sound: the
application of ultrasound without moving the ultrasound treatment head (= stationary).
Comment: the StatUS™ Pack 400 is not designed for use in combination with the Vacotron 460.
Page 4 of 36 EN109-1629761-43 IFU
2 Symbols
Symbol used
Page 5 of 36 EN109-1629761-43 IFU
3 Device components
Page 6 of 36 EN109-1629761-43 IFU
4 Package contents
1629902
StatUS™ Pack 400
Standard accessories
0169990
Holder for StatUS™ applicator (pre-assembled)
0629902
StatUS™-applicator 400
3442941
Gelpad StatUS™ (240x)
3442942
Fixation ring gelpad StatUS™ (3x)
1629767
Instructions for Use - StatUS™ Pack 400 (CD-ROM)
1629769
Information booklet StatUS™ Pack 400
Optional accessories
3442941
Gelpad StatUS™ (240x)
3442942
Fixation ring gelpad StatUS™ (3x)
Ordering information
For the ordering data of the 4-series, StatUS™ Pack 400, standard accessories and additional
accessories we refer to the website www.enraf-nonius.com.
Page 7 of 36 EN109-1629761-43 IFU
5 Installation
The StatUS™ Pack 400 is a feature (accessory) for the application of StatUS™ therapy and only works
when connected to a Sonopuls 490 device (art. no. 1498901, 1498902, 1498903) or a Sonopuls 492
(art.no. 1498911, 1498912, 1498913).*
To connect the StatUS™ Pack 400 with the Sonopuls 490 / 492 place the StatUS™ Pack 400
underneath the Sonopulse 490 / 492.
Connect the StatUS™ Pack 400 connector to the Sonopuls 490 / 492 to connector port B.
The StatUS™ Pack 400 does not have its own power supply and is powered by the Sonopuls 490 / 492.
Operating the device is also done (see Ch. 10) via the control panel (touchscreen) of the Sonopuls 490
/ 492.
Comment: the StatUS™ Pack 400 is not designed for use in combination with the Vacotron 460.
* = with software V009 or higher.
[A] Connection for (traditional) ultrasound treatment head (left side)
[B] Connection for StatUS applicator (right side)
The StatUS applicator / StatUS™ Pack 400) should be connected at the front (connection B).
The StatUS™ Pack 400 must be placed below the Sonopuls 490 / 492 device.
The StatUS applicator/ultrasound treatment head is a precision instrument. Effective pooling
of properties was an important focus during development and production. Careless
handling (jarring or dropping) may have a negative effect on these properties and should
therefore be avoided.
A maximum of one channel can be active. You can not work simultaneously with applicator
A and B.
Page 8 of 36 EN109-1629761-43 IFU
6 Intended use and intended user
Ultrasound is a mechanical energy consisting of high-frequency vibrations applied by means of an
ultrasound applicator. These vibrations pass through the tissue of the body and are gradually
absorbed and transformed into heat. The resulting temperature increase triggers biological changes
to occur in the tissue for the relief of pain, relaxation of muscle spasms and reduction of joint
contractures.
The Sonopuls 490 / 492 with StatUSTM Pack 400 device is intended to be used, and shall only be used,
by or under the supervision of professional users in the field of physical therapy and rehabilitation,
who understand the benefits and limitations of ultrasound therapy.
Page 9 of 36 EN109-1629761-43 IFU
7 Indications
Ultrasound is indicated for conditions that benefit from the application of deep heat: relief of pain,
muscle spasms and joint contractures. The objective of therapeutic ultrasound in the treatment of
selected medical conditions associated with the chronic and sub chronic conditions of
bursitis/capsulitis, epicondylitis, ligament sprains, tendinitis, scar tissue healing and muscle strain,
is to reduce pain.
Page 10 of 36 EN109-1629761-43 IFU
8 Contraindications
Contraindications
The established contraindications to heat therapy itself
In an area of the body where a malignancy is known to be present
Over or near bone growth centers until bone growth is complete
Over the thoracic area if the patient is using a cardiac pacemaker
Over a healing fracture*
Over ischemic tissues in individuals with vascular disease where the blood supply would be
unable to follow the increase in metabolic demand and tissue necrosis might result
In the presence of metal implants of any type*
Patients with sensory loss on the area to be treated
The gonads or to the developing foetus
The heart
The brains
The testicles
The eyes
Ultrasound should not be used on unconscious patients
* = Does not apply to LIPUS (Low Intensity Pulsed Ultrasound)
Precautions and Warnings
Precaution should be taken when using therapeutic ultrasound on patients with hemorrhagic
diatheses.
Ultrasound treatment presents a potential safety hazard in patients whose pain response has been
decreased because of disease, previous surgery, ionizing radiation therapy, chemotherapy, or
general or regional anaesthesia. It could cause burns. Do not use on insensitive areas or in the
presence of poor circulation.
Large thermal doses could result in regions of thermal aseptic necrosis which could not be
apparent on inspection of the skin.
Always ensure proper hygiene. (see chapter 12 Maintenance and troubleshooting)
Only apply the applicator on intact skin. When treating damaged skin (for example ulcers), only
place the applicator on the edges of the wound, never on the wound itself.
See also chapter 9 of the user manual, Precautionary Instructions, for general Warnings and
Precautions.
Relevant Hazards
Use of ultrasound in treating areas above the shoulders may pose relevant hazards. While it is
recognized that certain specific conditions involving the eyes can and have been treated by
qualified specialists, with knowledge and experience to administer such treatments, such
application carries with it recognized hazards of applying heat to the eyes.
Treatment of the thyroid, as well as lymph nodes in the neck, may expose the patient to as yet
undetermined effects inasmuch as the safety of such treatments has not yet been established.
Potential adverse effects
Cataracts
Male sterility
Enhanced drug activity
Thermal stress
Additionally some information on how the ultrasound should be applied is presented in chapter 11
“Application information” of the user manual.
Page 11 of 36 EN109-1629761-43 IFU
9 Precautionary instructions
Before administering any treatment to a patient you should become acquainted with the
operating procedures for each mode of treatment available, as well as the indications,
contraindications, warnings and precautions. Consult other resources for additional
information regarding the application of ultrasound.
Federal law (USA only) restricts this device to sale by, or on the order of, a physician or
licensed practitioner. This device should be used only under the continued supervision of a
physician or licensed practitioner.
This device should be kept out of the reach of children.
Do not operate the unit in an environment of short-wave or micro-wave diathermy use as this
can result in burns beneath the electrodes.
This device is not suitable for use in the presence of flammable anesthetics mixture with air,
oxygen, or nitrous oxide.
Do not expose the unit to direct sunlight, heat radiated from a heat radiator, excessive
amounts of dust, moisture, vibrations and mechanical shocks.
In the case of ingress of liquids, unplug the unit from the mains supply and have it checked
by an authorized person (see the paragraph on technical maintenance).
Do not operate the StatUS™ Pack 400 when connected to any unit other than Enraf-Nonius
BV devices.
Inspect the StatUS applicator cables and associated connectors before each use.
The StatUS applicator / StatUS™ Pack 400) should be connected at the front (connection B).
The StatUS™ Pack 400 must be placed below the Sonopuls 490 / 492 device.
The StatUS applicator/ultrasound treatment head is a precision instrument. Effective pooling
of properties was an important focus during development and production. Careless handling
(jarring or dropping) may have a negative effect on these properties and should therefore be
avoided.
A maximum of one channel can be active. You can not work simultaneously with applicator A
and B.
The ultrasound applicator is a precision instrument. Great care has been taken during the
development and in production to obtain the best possible beam characteristics. Rough
treatment (jarring or dropping) can adversely affect these characteristics, and must therefore
be avoided.
Use only the original Enraf-Nonius gelpads! Do not use liquid gel as this will be sucked into
the vacuum tubes and pump, causing malfunction! Any warranty claim will be rejected if
liquid gel residue if found in any of these components.
The gelpads are intended for once-only use.
Since the StatUS™ applicator / ultrasound treatment head transfers ultrasound energy into
the tissue at a most efficient and effective way, we advise you to use lower intensity settings
than the usual intensity settings for 'traditional' (semi-stationary or dynamic) ultrasound.
The StatUS applicator and the cable should be regularly inspected for damage, such as cracks,
which could cause fluid to penetrate.
Do not rinse the StatUS applicator under running water!
Never immerse the StatUS applicator in water or other (cleaning) liquids!
For proper cleaning of the StatUS applicator’s contact surface, the rubber cuff may be folded
back on one movement.
If necessary, you can push up the StatUS applicator a bit from beneath.
Never fold the entire cuff!
In order to guarantee good transmission of the ultrasound energy, you must use original Enraf-
Nonius gelpads.
Page 12 of 36 EN109-1629761-43 IFU
Do not use regular (liquid) Sonogel because this will be sucked up into the vacuum pump!
Warranty claims in which the remnants of regular ultrasound gel are found in the pump, hoses
or connections will be rejected by Enraf-Nonius.
To perform well the StatUS™ Pack 400 should be checked this once a year by a service
technician.
No attempt should be made to disassemble the unit. Maintenance and repair should be carried
out by authorized personnel only. Enraf-Nonius will not be held responsible for the results of
maintenance or repairs by unauthorized persons.
Page 13 of 36 EN109-1629761-43 IFU
10 Operation
Basic operation
The parameters for the application of static ultrasound (StatUS therapy) conform for the most part to
the parameters for conventional ultrasound therapy as provided by the Sonopuls 490 / 492. The
parameters for StatUS therapy are entered via the (touschscreen) control panel of the Sonopuls 490 /
492.
The following is an overview of the specific operating instructions for the application of static
ultrasound therapy with the StatUS applicator. For a general description of the settings and operation
of the Sonopuls 490 / 492 please see the manual (art.no.1498751) for the 4-series. This manual (CD-
ROM) is included with the Sonopuls 490 / 492.
Home menu
The Sonopuls 490 / 492 devices are equipped with touch-
screen operation. Select the desired function by pushing an
icon.
Image: main menu Sonopuls 492
Note: The available therapy choices may vary by model and are depending on the accessories
connected to terminal A and / or B. If a StatUS™ Pack 400 is connected (connection B), then the
therapy choice “StatUS Therapy” is accessible. When disconnecting the StatUS Pack 400, the function
“StatUS therapy” remains visible in the main menu, until the moment that you stop the power supply.
After switching on the device, the function “StatUS therapy” is no longer visible. Once the StatUS
applicator has been reconnected, the function will appear again in the main menu.
StatUS™ Therapy
Select StatUS therapy from the main menu
Then you can choose from “Clinical Protocols”, “Favorites” and “Manual”.
o Under „Clinical Protocols“ you can retrieve the 52 pre-programmed treatment protocols.
o Under “Favorites” you can retrieve your stored protocols.
o Under 'Manual Operation', you can set all the parameters.
Page 14 of 36 EN109-1629761-43 IFU
Clinical Protocols
The Sonopuls 490 / 492 has 52 treatment suggestions for the most common treatments for
ultrasound.
Push the icon “Clinical
Protocols” to activate the
list.
Select in the list the treatment
suggestion or:
Push icon for more
therapy information about
the protocol
Use the central controller
to scroll through the
pages.
The first page is text followed
by one or more illustrations
Note: the numbers under
Literature are referring to
relevant publications like
shown in the literature list
(Appendix 1).
Push the icon (“accept”).
The screen shows all
parameters in once glance
The parameters can be
changed at all times.
Press button [1] to set
vacuum pressure.
1
Page 15 of 36 EN109-1629761-43 IFU
Place the StatUS applicator
on the treatment area and
increase the vacuum pressure
using the central regulator
Once there is sufficient
contact the pause icon
disappears and treatment
time starts counting
down.
If contact is insufficient, it
will pause the treatment
time.
In this case, increase the
vacuum pressure until the
contact is restored.
BE AWARE:
Once the treatment time has elapsed, the vacuum pressure will still be maintained for 1
minute max. This prevents that the StatUS applicator immediately disengages at the end of
treatment (and possibly drops). The switch-off time for the vacuum pressure can be set in
increments of 1 minute / can be changed via the “sleep setting after…” button in the menu
“System settings”.
In general, the Amplitude Modulation and Duty Cycle Modulation are always switched on
(=standard setting) *. However, both modulations can be switched on and off manually (see
images below). Caution is advised though. The owner or user of a StatUS Pack 400 will remain
responsible and liable at all times for the use of StatUS, when one or both modulations are
switched off temporarily (also refer to chapter 6, chapter 10 paragraph Manual Operation -
Selecting the parameters: Duty Cycle modulation and Amplitude modulation, chapter 13
paragraph StatUS therapy - Amplitude modulation).
* applies when using the function "Manual" . In some protocols , the modulations are not active.
Page 16 of 36 EN109-1629761-43 IFU
Display if both modulations are
enabled ( active).
Display if Amplitude Modulation
is disabled (inactive).
Display if Duty Cycle
Modulation is disabled
(inactive).
Manual operation
Setting parameters: frequencies
Press the "Manual
Operation" button
Select the parameter by
pressing the relevant
button.
Example: Press button [2]
to change the pulse
frequency
Change the frequency by
using the central controller.
The ultrasound frequency can be set/adjusted in the same way (1 or 3 MHz).
2
Page 17 of 36 EN109-1629761-43 IFU
Selecting the parameters: duty cycle modulation and amplitude modulation
When using StatUS therapy, the Duty Cycle Modulation (as well as the Amplitude Modulation) is
always active (see chapter 13). This means that in general the device is always in pulsed modus and
that the Duty Cycle (and the amplitude) vary continuously. Because of this, the button for selecting the
pulse mode [3] is not accessible.
Press button [4] to change
the duty cycle.
There are 7 options: 100,
80, 50, 33, 20, 10 and 5%.
Select the duty cycle with
the central regulator.
In this example, the duty
cycle drops from 20% to
10, then goes back up to
20%.
The icon indicates that the
Duty Cycle modulation has
been switched on (is active).
The Duty Cycle modulation and Amplitude modulation can be switched on and off manually (using
the central regulator). Caution is advised though. The owner or user of a StatUS Pack 400 will remain
responsible and liable at all times for the use of StatUS Pack 400, when one or both modulations are
switched off temporarily (also refer to chapter 6, chapter 10 paragraph Clinical protocols, chapter 13
paragraph StatUS therapy - Amplitude modulation).
4
45
3
Page 18 of 36 EN109-1629761-43 IFU
Selecting the parameters: treatment time
Press button [5] to change
the treatment time.
Change treatment time
with the central controller.
Selecting the parameters: intensity
Press button [6] to change
the intensity.
Change the intensity with
the central regulator.
Note: a pop-up appears if the
selected intensity exceeds the
mean value.
Press the button to
continue or to adjust the
intensity.
5
6
Page 19 of 36 EN109-1629761-43 IFU
Favorites
Storing protocols
Prior to the implementation of an appropriate protocol or a manually set treatment settings can be
saved as 'favorites'.
First set the desired
parameters.
Press button [7] to store the
parameters.
Enter a name for the
program when you save it.
Press the button
(“confirm”).
7
Page 20 of 36 EN109-1629761-43 IFU
Loading favorites
Press the button
“Favorites”.
Select the desired
programme from the list of
favorites.
The parameter screen with
the stored parameters
appears.
Next, enter the vacuum
pressure and position the
StatUS applicator.
The treatment starts
automatically when there is
sufficient contact.
Delete favorites
Press the button
“Favorites”.
First press the ‘Waste bin’
button [8].
Press the name of the
favorite you wish to delete.
(e.g. my protocol)
The favorite will be deleted
and removes from the list.
8
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36

Enraf-Nonius StatUS Pack 400 User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI