Midmark M3 Steam Sterilizer Parts Manual

Type
Parts Manual
FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY
Model Numbers:
Service and
Parts Manual
M3 Steam Sterilizer
M3
-001 thru -004
-040 thru -042
004-0520-00 (6/27/18)
SA104100-1i
ii© Midmark Corporation 2010 SF-1889 [Revised: 7/12/17]
Table of ContentsTable of Contents
GENERAL INFORMATION
Symbols .......................................................... iv
Ordering Parts ................................................ iv
Serial Number Location .................................. iv
Speciî‚¿cations .................................................. v
Scheduled Maintenance ................................. vi
Cycle Parameters ...........................................vii
TROUBLESHOOTING
Error Codes ................................................. A-2
Power Up / Select Cycle Mode .................. A-15
Close Mode ............................................... A-16
Heat Up Mode ........................................... A-17
Sterilization Mode ...................................... A-18
Vent Mode ................................................. A-19
Open Mode ................................................ A-20
Dry Mode ................................................... A-21
TESTING & REPAIR
Activating Service Diagnostics .................... B-2
Test Selection Screen .................................. B-3
Vent Valve Test ............................................ B-4
Door Motor Test ........................................... B-6
Water Pump Test / Priming Instructions ...... B-9
Boiler Test .................................................. B-12
Dry Heater Test.......................................... B-14
Condensation Heater Test ......................... B-16
Door / Tray Position Switch
Display Screen .......................................... B-18
Chamber Pressure Display Screen ........... B-19
Temperature Display Screen ..................... B-20
Water Level / External Tank
Sensor Display Screen .............................. B-21
Error Code Display Screen ........................ B-22
Key Test Display Screen............................ B-23
Five Point Leak Test .................................. B-24
Temperature Sensor Resistance Test ....... B-25
Inlet Switch Module Test ............................ B-27
General InfoSection A
Section E
Section B
Section C
Section D
(*) Indicates multiple pages due to model / serial number break(s).
ACCESS PROCEDURES
Top Cover Removal / Installation ................. C-2
WIRING DIAGRAMS
M3 ................................................................ D-2
EXPLODED VIEWS / PARTS LIST
Components Overview ................................ E-2
Main Enclosure ............................................ E-3
Main Components ............................................ E-4
Door Tray Latch Mechanism ....................... E-5*
Boiler Assembly ........................................... E-6
Pump Solenoid Assembly ............................ E-7
Reservoir Assembly ..................................... E-8
PC Board Assembly..................................... E-9
Switch Harness Assembly ............................. E-10
Plumbing and Tubing Components ............E-11
Door / Tray Accessory ............................... E-12
External Condensing Tank......................... E-13
Shipping Packaging ................................... E-14
Label Location ........................................... E-15
Section B
TESTING & REPAIR (CONTINUED...)
Photo Interrupter ........................................ B-28
iii© Midmark Corporation 2010 SF-1889 [Revised: mo/dd/yr]
General InformationGeneral Information
WARNING
Indicates a potentially hazardous situation which
could result in serious injury if not avoided.
Equipment Alert
Indicates a potentially hazardous situation which
could result in equipment damage if not avoided.
Caution
Indicates a potentially hazardous situation which may
result in minor or moderate injury if not avoided. It
may also be used to alert against unsafe practices
Symbols
Indicates the function / component is working properly.
No action required.
Indicates the function / component is working,
but a problem exists.
Indicates the function is not working at all, or that the
component is faulty.
Ordering Parts
The following information is required when ordering parts:
 • Serialnumber&modelnumber
 • Partnumberfordesiredpart
[Refer to Section E: Exploded Views & Parts Lists]
Non-warranty parts orders may be faxed to Midmark using
the Fax Order Form in the back of this manual.
For warranty parts orders, call Midmark’s Technical Service
Department with the required information.
Hours:8:00amto5:00p.m.EST(MondaythruFriday)
Phone:1-800-Midmark(1-800-643-6275)
Serial Number Location
Note
Used for special instructions or additional information.
SA103100i
Serial Number VXXXXXX
VXXXXXX
VXXXXXX
Thesymbolsbelowmaybeusedinthismanualtorepresent
theoperationalstatusofsterilizerfunctionsandcomponents.
iv© Midmark Corporation 2010 SF-1889 [Revised: 7/2/18
General Information
Weights, Dimensions, Electrical Speciî‚¿cations
Note
To ensure unit is properly grounded, it must be connected to a
matching grounded, dedicated, correctly polarized receptacle.
Electrical Ratings:
115 VAC Unit
CircuitRequirements*......................120VAC,50/60Hz,12amp
Max.PowerConsumption................1400watts
230 VAC Unit
CircuitRequirements*.....................230VAC,50/60Hz,6amp
Max.PowerConsumption...............1400watts
*Power source must have over voltage limits less than 1500 watts from
mains to ground. (Installation Category ll in accordance to IEC 664)
Fuse Ratings:
115 VAC Unit
F1....................................................15Amp,250V,FastActing,1/4"x11/4"
F2....................................................0.25Amp,250V,Slo-blo,1/4"x11/4"
230 VAC Unit
F1....................................................8Amp,250V,FastActing,5x20mm
F2....................................................0.125Amp,250V,Slo-blo,5x20mm
Certiî‚¿cations:
ASMEBoiler&PressureVesselCode,SectionVIII,Division1
CanadianRegistrationNumberAvailable
UL61010-12ndEdition
IEC61010-2-0401stEdition
CAN/CSA-C22.2No.61010-12ndEdition
FCCPart15,Sub-partB
Physical Dimensions:
OverallLength(A):................................21in.(53.3cm)
OverallWidth(B):..................................17.8in.(45.2cm)
OverallHeight(C):.................................6.9in.(17.5cm)
ShippingCartonLength:.......................25in.(63.5cm)
ShippingCartonWidth:.........................22in.(55.9cm)
ShippingCartonHeight:........................16.6in(42.2cm)
CounterArea:.......................................24in.(61cm)deepx22in.(55.9cm)wide
ChamberVolume:.................................0.49gal(1.8liter)
Weight:
EmptyReservoir:.........................................71lbs.(32.2kg)
FullReservoir:.............................................80lbs(36.3kg)
WithShippingCarton:..................................80lbs(36.3kg)
Water Reservoir Capacity:........................1.20gal(4.5liter)
Pressure Relief Valve Setting:....................40PSI(275.8kPA)
Chamber Pressure:
@270°F(132°C).................................27.1PSI(186.2kPA)
SA105900i
C
A
B
v© Midmark Corporation 2010 SF-1889 [Revised: mo/dd/yr]
Scheduled Maintenance / Cleaning Chart
Interval Inspection / Service Description
Periodically
Cleaning
Cleanexternalsurfaces,tray,&chamberwithmildsoapanddistilledwater.Wipedry.
Cleandoorgasket&matingsurfacewithadampcloth.
Inspectgasketfordamage,replacegasketifnecessary.
Emptywaterfromcondensingtank.Cleantankwithdilutedbleachsolution(1/4 cup bleach : 1 gallon water)&abrush.
Rinsetank,thenlltotheminimumwaterlevelindicatorline.
ObviousDamage Visuallyinspectcomponentsfordamagethatcouldcauseproblemsduringoperationorunsafeoperation.
Drain/RellReservoir Usingthedrainhose,drainallwaterfromthereservoir.
Rellreservoirwithdistilledwater.
Hardware Allfastenersmustbepresentandfastenedsecurely.
ElectricalSystem
Inspectpowercordandallwiringfordamage.
Besureallelectricalconnectionsaretight.
Monthly Cleaning
Remove&cleanlterusingamildsoapsolution&abrush,Placeinultrasoniccleanerifnecessary.
Rinsewithdistilledwater.
Cleancondensingtanklevelsensorswithmildsoapsolution.
Wipedry.
DateofService:
Location:
ServiceTechnician:
ModelNumber:
SerialNumber:
Notes:
vi© Midmark Corporation 2010 SF-1889 [Revised: mo/dd/yr]
CYCLE
STERILIZATION
PARAMETERS
DRY TIME ITEMS TO BE STERILIZED
Unwrapped
Temperature:
270°F(132° C)
Pressure:
27.1psi(186kPa)
Time:
3:30Minutes
Time:
25Minutes
•DentalInstruments/handpiecesloose
onatray.
•Otheritemsmanufacturersrecommendfor
exposureat270°F(132°C),looseonatray.
Pouches
Temperature:
270°F(132° C)
Pressure:
27.1psi(186kPa)
Time:
5:30Minutes
Time:
30Minutes
•DentalInstruments/handpiecesinpouches,
wrapped,orinawrappedcassette.
•Otheritemsmanufacturersrecommendforexposure
at 270°F(132°C),inpouches,wrapped,orina
wrappedcassette.
Low Temp
Temperature:
250°F(121° C)
Pressure:
15.0psi(104kPa)
Time:
20:00Minutes
Time:
50Minutes
•Rubberorplasticdentalitems,dentalinstruments
/handpieceslooseonatray,inpouches,wrapped,
orinawrappedorunwrappedcassette.
•Otheritemsmanufacturersrecommendforexposureat
250°F(121°C),looseonatray,inpouches,wrapped,
orinawrappedorunwrappedcassette.
Cycle Parameters
Thetablebelowshowsthetemperature/pressure/timeparametersforthepre-setcycles.
(ATTENTION: Before sterilizing any item in the M3, refer to Loading the Tray in the user's guide).
A-1© Midmark Corporation 2010 SF-1889 [Revised: 11/18/14]
Section ASection A
Troubleshooting
Mode / System
Error Codes Troubleshooting .............A-2
Power Up / Select Cycle Mode ........ A-15
Close Mode .....................................A-16
Heat-Up Mode .................................A-17
Sterilization Mode ............................A-18
Vent Mode .......................................A-19
Open Mode ...................................... A-20
Dry Mode .........................................A-21
Mode / System
Error Codes Troubleshooting .............A-2
Power Up / Select Cycle Mode ........ A-15
Close Mode .....................................A-16
Heat-Up Mode .................................A-17
Sterilization Mode ............................A-18
Vent Mode .......................................A-19
Open Mode ...................................... A-20
Dry Mode .........................................A-21
A-2© Midmark Corporation 2010 SF-1889 [Revised: 7/2/18]
Error Codes
Ifasystemmalfunctionisdetectedduringacycle,anumericerrorcodewillappearonthe
displaypanel.Eachdigitintheerrorcodeprovidesinformationabouttheproblemthatoccurred.
Error Codes
C010: (System Power Loss) .....................A-3
C060: (System Hardware) .......................A-4
C100 Series: (Stop Key) .............................A-4
C230 Series: (Water Low) .....................A-4
C320 Series: (Door Closed) ...................A-5
C350 Series: (Tray Switch) ....................A-6
C380 Series: (Door Open) ..................... A-6
C440 Series: (Ext. Condensing Tank Full) .A-7
C530 Series: (Steam Temp. Low) ............ A-7
C540 Series: (Steam Temp. High) ...........A-8
C560 Series: (Steam Temp. Hardware).....A-8
C570 Series: (Steam Temp. Over limit) .....A-8
C630 Series: (Pressure Low) .................A-9
C640 Series: (Pressure High) .................A-9
C660 Series: (Pressure Hardware) .......... A-10
C670 Series: (Pressure Over limit) ........A-10
C760 Series: (Boiler Temp. Hardware) ...A-10
C770 Series: (Boiler Temp. Over limit) .... A-11
C860 Series: (Dryer Temp. Hardware) .... A-11
C870 Series: (Dryer Temp. Over limit) ....A-12
C880 Series: (Dryer Hi-Limit Open) .......A-12
C980 Series: (Boiler Hi-Limit Open) .......A-12
Instruments Not Drying .................... A-12
Biological Strip Error ........................ A-13
Maintenance Message ....................A-14
Does Not Print (Optional Printer) ..... A-14
Chamber Fills With Water ................A-14
Error Codes
C010: (System Power Loss) .....................A-3
C060: (System Hardware) .......................A-4
C100 Series: (Stop Key) .............................A-4
C230 Series: (Water Low) .....................A-4
C320 Series: (Door Closed) ...................A-5
C350 Series: (Tray Switch) ....................A-6
C380 Series: (Door Open) ..................... A-6
C440 Series: (Ext. Condensing Tank Full) .A-7
C530 Series: (Steam Temp. Low) ............ A-7
C540 Series: (Steam Temp. High) ...........A-8
C560 Series: (Steam Temp. Hardware).....A-8
C570 Series: (Steam Temp. Over limit) .....A-8
C630 Series: (Pressure Low) .................A-9
C640 Series: (Pressure High) .................A-9
C660 Series: (Pressure Hardware) .......... A-10
C670 Series: (Pressure Over limit) ........A-10
C760 Series: (Boiler Temp. Hardware) ...A-10
C770 Series: (Boiler Temp. Over limit) .... A-11
C860 Series: (Dryer Temp. Hardware) .... A-11
C870 Series: (Dryer Temp. Over limit) ....A-12
C880 Series: (Dryer Hi-Limit Open) .......A-12
C980 Series: (Boiler Hi-Limit Open) .......A-12
Instruments Not Drying .................... A-12
Biological Strip Error ........................ A-13
Maintenance Message ....................A-14
Does Not Print (Optional Printer) ..... A-14
Chamber Fills With Water ................A-14
First Digit = Where?
Therstdigitindicatesthecomponent
orsystemwheretheproblemoccurs.
(example: 3 = door)
Example:
Second Digit = What?
Theseconddigitindicateswhat
problemorsymptomwasdetected.
(example: 8 = Open)
Third Digit = When?
Thethirddigitindicateswhen
theproblemwasdetected.
(example: 2 = Close Mode)
The table below cross-references the numeric error code with the Component, Problem, and Mode.
0=GeneralSystem 0(Blank)* 0=PowerUpMode
1=StopButton 1=PowerLoss 1=SelectCycleMode
2=WaterLevelSensor 2=Closed 2=CloseMode
3=DoorSwitch 3=Low 3=HeatUpMode
4=Ext.CondensingTank 4=High 4=SterilizeMode
5=TemperatureSensor 5=TraySwitchOpen 5=VentMode
6=PressureSensor 6=Hardware 6=OpenMode
7=BoilerTemperature 7=OverLimit 7=DryMode
8=DryerTemperature 8=Open 8=CompleteMode
9=HighLimitThermostat 9(notused) 9=Misc.
First Digit
(Component)
Second Digit
(Problem)
Third Digit
(Mode)
C382:CLOSE MODE
DOOR OPEN
*0isusedinthe100serieserrorcodes
toindicatetheSTOPkeywaspressed.
A-3© Midmark Corporation 2010 SF-1889 [Revised: 4/23/13]
Troubleshooting Chart
Problem Display / Symptom Probable Cause Check Correction
Sterilizer has no power.
Touchpad&displaydonot
work.(Switchfailureunder
probablecausewouldonlybe
onnewerstylePCboards)
Sterilizernotpluggedin. Checkpowercordconnectionsatwall
outletandonbackofsterilizer.
Securepowercordconnections.
Powerswitchturnedo. Checkpowerswitch. Turnpowerswitchon.
Nopoweratwalloutlet. Checkvoltageatwalloutlet. Resetfacilitycircuitbreaker.The
sterilizershouldbepluggedintoa
dedicated20ampcircuit.
F1fuseonPCboardis
blownoropen.
CheckF1fuseforcontinuity. Replacefuse.
Inletswitchmodulefailure.
Checkinletswitchmodule.
Referto:InletSwitchModuleTest
Performcontinuitytestoninletswitch
module.
Sterilizer has power, but
touchpad & display do
not work.
Touchpad&displaydonot
work.
F2fuseonPCboardis
blownoropen.
CheckF2fuseforcontinuity. Replacefuse.
RibboncablefromJ15on
mainPCboardtoJ3on
displayPCboardislooseor
disconnected.
CheckribboncableconnectionsatJ15
onmainPCboard&J3ondisplayPC
board.
Secureribboncableconnections.
J15/J3ribboncablehas
openlead(s).
CheckJ15/J3ribboncablefor
continuity.
Replaceribboncable.
MainPCboardis
malfunctioning.
Checkfor5VDCbetweenTP1&TP4,
andonJ15acrosspins1&2onmainPC
board.
If5VDCisnotpresentatthese
points,replacemainPCboard.
Displayisblankorundined
charcters&internittentbeeps.
DisplayPCboardis
malfunctioning
Checkfor5VDCbetweenR1&J1on
displayPCboard.
If5VDCisnotpresent,replace
displayPCboard.
Displayworksbuttouch
paddoesn't.
RibboncablefromJ2to
touchpadisdisconnected.
CheckJ2Plugconnection. AssureJ2Plugpinsarealignedand
connectedsecurely.
J2ribboncabletotouchpad
hasopenlead(s).
CheckcontinuityofJ2ribboncable
andtouchpad.
Replacetouchpadandribboncable
assembly.
Unit has power, but the
door is not open in Select
Cycle Mode, and there is
no error code associated
with this condition.
Displayisnormal. Doormotorconnectingrodis
broken.
Removetopcoverandverifythatthe
doormotorconnectingrodisintact.
Replacedoormotorconnectingrod.
C010
C010:POWERUPMODE
SYSTEMPOWERLOSS
…
ITEMSNOTSTERILE
PUSHSTOPTORESTART
Powerinterruptionduringany
modeinwhichsterilization
maybecompromised.
Userturnedpoweroduring
ChamberClose,HeatUp,or
SterilizationMode.
Askquestionsaboutusagepatterns Trainuseronproperoperation.
Unitpluggedintoanon
dedicated,20ampcircuit.
Checkifunitispluggedintoadedicated
20ampcircuit.
Connectunittoadedicated20amp
circuit.
Someotherevent(suchasa
storm,apoweroutage,ect.)
interruptedpower.
Askquestionsaboutrecentelectrical
events(storms,outages,ect.).
Ifthisisacommonoccurrence,iso-
latefromsourceofelectricalinterfer-
ence.
A-4© Midmark Corporation 2010 SF-1889 [Revised: 8/16/13]
Troubleshooting Chart
Problem Display / Symptom Probable Cause Check Correction
C060
C060:POWERUPMODE
SYSTEMHARDWARE
…
TURNPOWEROFF/ON
Atpower-up,memoryor
otherhardwarecannotbe
properlyreadorinitialized.
(Ifoccational)powerhas
beenrapidlycycledothen
onagainbyuser(possiblyto
clearanothererror).
Askquestionsaboutusagepatterns. Trainuseronproperoperation.
(Ifoccational)powerhas
beenrapidlycycledothen
onagainbyanelectrical
interruption(suchasastorm,
poweroutage,ect.).
Checkifsterilizerispluggedintoadedi-
cated20ampcircuit.
Connectunittoadedicated20ampcircuit.
Askquestionsaboutrecentelectrical
events.
Ifthisisacommonoccurrence,isolatefrom
sourceofelectricalinterference.
(Ifpersistent)damaged
componentonPCboard.
Turnpowerswitcho.Wait20seconds,
andturnpowerbackon.Repeatseveral
timesinarow.
ReplacePCboardifnecessary.
C099
C099:MISCMODE
??? ???
…
UNPLUG/RE-PLUGUNIT
Anerrorthatcanoccur
whileconnectedtoaPC,
ect.Thiserrorwillneverbe
causedbynormalusage.
Codewasgeneratedatthe
factoryduringtestingwhen
anexternaldevicesuchasa
PCorlaptopwasconnected
tothesterilizermainPC
board.
Nochecknecessary. ContactMidmark'sTechnicalService
Department.
C102: Close Mode
C103: Heatup Mode
C104: Sterilize Mode
C105: Vent Mode
C106: Open Mode
C10X:XXXXMODE
STOPKEY
…
ITEMSNOTSTERILE
PUSHSTOPTORESTART
TheSTOPkeywassensed
duringacycle.
UserpressedtheSTOPkey
duringthecycle.
Askquestionsaboutusagepatterns. Trainuseronproperoperation.
PressSTOPkeytoclearerror.
Note:Itmaybenecessarytowaitbriey
(upto1minute)whilechamberpressure&
temperaturedissipate.
Membraneswitch(touchpad)
isdamaged.
Performkeytest.Referto:
Service Diagnostics Key Test.
Replacemembraneswitch.
C232: Close Mode
C232:CLOSEMODE
WATERLOW
…
ITEMSNOTSTERILE
PUSHSTOPTORESTART
Reservoirwaterlevel
sensordetectsthesupply
waterislow.
Reservoirwaterlevellow. Checkwaterlevelinreservoir. Fillreservoirwithdistilledwater.
Faultywaterlevelsensor.
Referto:I/O Test Water Level / External
Tank Display Screen.
Replacewaterlevelsensor.
Levelsensorconnection
looseordisconnected,or
brokenwires.
ChecklevelsensorconnectionatJ13on
PCboard,andrelatedleads
AssureJ13plugconnectionissecurely
connectedandrelatedwiresarenot
damaged.
MalfunctioningPCboard. Runservicediagnostics,andcheckfor
4.5to5.0VDCatJ13pins1&2onPC
boardwhenlevelsensorstateisbeing
displayedonLCD.
Ifnovoltage,replacePCboard.
A-5© Midmark Corporation 2010 SF-1889 [Revised: mo/dd/yr]
Troubleshooting Chart
Problem Display / Symptom Probable Cause Check Correction
C321: Select Mode
C326: Open Mode
C32X:XXXXMODE
DOORCLOSED
…
PUSHSTOPTORESTART
DoorMotorOpenSwitchis
not
sensedasbeingopenwhen
shouldbe.
Somethingoutsideoftheunit
isphysicallyinterferingwith
themovementofthedoor.
Verifythatthereisnothingoutsideofthe
unitstoppingmovementofthedoor.
Removeanyexternalimpedimenttodoor
movement.
Thedoormotorcamisnot
makingcontactwiththeDoor
OpenSwitch.
Removetopcover,putunitintoService
Diagnostics,andverifythatthedoormotor
isnotbrokenandphysicallyattemptsto
operate.
Replacedoormotor.
Doorwillnotopenbecause
thereistoomuchpressurein
chamber.
Checkforobstructioninplastictube
connectingunittoexternalcondensing
tank.
Unkinkplastictube,removeobstruction,or
replace.
Checkcoppercoilinexternalcondensing
tankforblockageordamage.
Clearobstructionorreplacecoppercoil.
Checkventportinchamberforblockage.
Checklterforblockage.
Clearobstructioninventport.
Cleanorreplacelter.
Checkventvalveforblockage.
PutunitintoServiceDiagnosticsand
checkventvalveoperation.
Cleanorreplaceventvalve.
PutunitintoServiceDiagnostics,and
verifythatthePCboardisenergizingvent
valvewhenappropriate.
ReplacePCboard,ifnecessary.
PressureInterlockAssy.is
stuckopen.
Verifythatchamberpressureiszeroand
thencheckthestateofthepressure
interlockassy.
Replacepressureinterlockassy.
DoorOpenSwitchisstuckin
theclosedstate.
PutunitintoServiceDiagnostics,and
verifythattheDoorOpenSwitchis
operatingnormally.
Replacedoorswitch/harnessassy.
Somethinginsidethe
chamberisphysically
interferingwiththemovement
ofthedoor.
Verifythatthereisnothingjammedinside
thechamberstoppingmovementofthe
doorbymanuallyopeningthedoor.
Turnpowero,removetopcover,and
purgechamberofpressure.Removepivot
pinfromdoormotorconnectingrodand
pivotbracket,thenmanuallyopenthe
door.Removejammedinstruments.
PCBoardisnotenergizing
doormotor.
PutunitintoServiceDiagnosticsandverify
thatthePCboardisenergizingdoormotor
whenappropriate.
ReplacePCboard,ifnecessary.
A-6© Midmark Corporation 2010 SF-1889 [Revised: 8/16/13]
Troubleshooting Chart
Problem Display / Symptom Probable Cause Check Correction
C352: Close Mode
C353: Heatup Mode
C354: Sterilize Mode
C355: Vent Mode
C35X:XXXXMODE
TRAYSWITCHOPEN
…
ITEMSNOTSTERILE
PUSHSTOPTORESTART
TraySwitchnotsensedas
fullyclosedwhenitshouldbe.
Traynotproperlystaged
wheninsertedintochamber.
Askquestionsaboutthewaytheuserloads
thetray.
Trainusertofullystagethetraybefore
attemptingtostartacycle.
Trayhandle'splastickeyis
brokeno.
Ensurethatleftlowersideofplastictray
handlehasa"key"thatsticksoutaboutan
inch.
Replacebottomdoortrayhandle.
Trayswitchisnotpositioned
correctly,andisnotsensing
presenceoftray.
UseServiceDiagnosticstoverifythatthe
trayswitchactuateswhenthetrayinitially
crestsbumpinthetusk.
Adjustthetrayswitchbracket.
Trayswitchisfaulty. UseServiceDiagnosticstoensurethatthe
switchoperatescorrectly.
Replacedoorswitch/harnessassy.
C382: Close Mode
C383: Heatup Mode
C384: Sterilize Mode
C385: Vent Mode
Continue on Next
Page...
C38X:XXXXMODE
DOOROPEN
…
ITEMSNOTSTERILE
PUSHSTOPTORESTART
DoorMotorClosedSwitch
hasbeenactivated,stopping
thedoormotor,buttheDoor
ClawSwitchisnotsensed
asbeingclosed.
Somethingoutsideofthe
unitisphysicallyinterfering
withthemovementofthe
door.
Verifythatthereisnothingoutsideofthe
unitstoppingmovementofthedoor.
Removeanyexternalimpedimenttodoor
movement.
Thedoormotorcamisnot
makingcontactwiththe
DoorClosedSwitch.
Removetopcover,putunitintoService
Diagnostics,andverifythatthedoormotor
isnotbrokenandphysicallyattemptsto
operate.
Replacedoormotor.
Doorwillnotclosebecause
thereistoomuchpressure
inchamber.
Checkforobstructioninplastictube
connectingunittoexternalcondensingtank.
Unkinkplastictube,removeobstruction,or
replace.
Checkcoppercoilinexternalcondensing
tankforblockageordamage.
Clearobstructionorreplacecoppercoil.
Checkventportinchamberforblockage.
Checklterforblockage.
Clearobstructioninventport.
Cleanorreplacelter.
Checkventvalveforblockage.
PutunitintoServiceDiagnosticsandcheck
ventvalveoperation.
Cleanorreplaceventvalve.
PutunitintoServiceDiagnostics,andverify
thatthePCboardisenergizingventvalve
whenappropriate.
ReplacePCboard,ifnecessary.
Doormotorbracketnot
adjustedproperly.
Verifythat,whenthedoormotorisinthe
closedposition,thereisenoughthrowto
physicallyclosethedooralltheway.
Ifthereisnotenoughthrow,adjustthe
doormotormoretowardtherearofthe
unit.
DoorClosedSwitchisstuck
intheclosedstate.
PutunitinServiceDiagnostics,andverify
thatthedoorclosedswitchisoperating
normally.
Replacedoorswitch/harnessassy.
A-7© Midmark Corporation 2010 SF-1889 [Revised: 8/16/13]
Troubleshooting Chart
Problem Display / Symptom Probable Cause Check Correction
C382: Close Mode
C383: Heatup Mode
C384: Sterilize Mode
C385: Vent Mode
(Continued)
DoorMotorClosedSwitch
hasbeenactivated,stopping
thedoormotor,buttheDoor
ClawSwitchisnotsensed
asbeingclosed.
Somethinginsidethe
chamberisphysically
interferingwiththemove-
mentofthedoor.
Verifythatthereisnothingjammedinside
thechamberstoppingmovementofthe
door.
Removeanyjammedinstruments.
PCBoardisnotenergizing
doormotor.
PutunitintoServiceDiagnosticsandverify
thatthePCboardisenergizingdoormotor
whenappropriate.
ReplacePCboard,ifnecessary.
C442: Close Mode
C442:CLOSEMODE
EXTERNALTANKFULL
…
PUSHSTOPTORESTART
Theexternaltankwaterlevel
sensordetectedwaterlevel
intheexternaltankistoo
hightoperformnextcycle.
Externalcondensingtankis
full.
Visuallychecktanktoseeiffull. Emptyexternalcondensingtank..
Corrosiononcondensing
tanklevelsensors.
Visuallyinspectexternalcondensingtank
waterlevelsensors.
Cleansensors,coppercoil,andinsideof
condensingtankwithacleancloth.
Levelsensorsorattached
wiringdamaged.
Checklevelsensorsandattachedwiring
fordamage.
Replacelevelsensorsand/orwiring.
Systemisnotprocessing
dataproperly.
GointoServiceDiagnosticstoverifysensor
operation.
Referto:I/O Test Water Level / External
Tank Display Screen.
C533: Heatup Mode
C534: Sterilize Mode
C53X:XXXXMODE
STEAMTEMPLOW
…
ITEMSNOTSTERILE
TURNPOWEROFF/ON
Duringheatupmode,
sterilizationmodeisnot
reachedwithin15minutes.
DuringSterilizationMode,
steamtemperaturedropsto
apointlessthanorequalto
thesetpointchosenduring
SelectCycleMode.
Lowwaterinreservoir Checkwaterlevelinreservoir. Fillreservoirwithdistilledwater.
Lowwaterinreservoirnot
caughtbyaC23xerror.
Referto:I/O Test Water Level / External
Tank Display Screen.
Replacewaterlevelsensor.
Waterpumpnotprimed.
Referto:
Water pump priming instructions.
Primewaterpump.
Steamleakage.
Referto:Five point leak check. Repairsteamleak.
Chambertemperaturesensor
notfunctioningproperly.
Referto:
Temperature Sensor Resistance Test.
Cleanorreplacechambertemperature
sensor.
Boilertemperaturesensor
notfunctioningproperly.
Referto:
Temperature Sensor Resistance Test.
Replaceboilertemperaturesensor.
Boilernotfunctioning
properly.
Referto:
Service Diagnostics Boiler Test.
Replaceboilerifnecessary.
PCBoardisnotprocessing
temperatureinput(s)properly.
Verifyboilerandchambertemperature
sensorsareworkingproperly.
ReplacePCboardifnecessary.
Defectivecheckvalvein
waterpump.
Checkforlargeairpocketformingintubing,
thatmovesfromtheboilertowaterpump.
Replacewaterpump.
A-8© Midmark Corporation 2010 SF-1889 [Revised: 11/18/14]
Troubleshooting Chart
Problem Display / Symptom Probable Cause Check Correction
C540: Power Up Mode
C544: Sterilize Mode
C546: Open Mode
C54X:XXXXMODE
STEAMTEMPHIGH
…
ITEMSNOTSTERILE
TURNPOWEROFF/ON
DuringSterilizationMode,
steamtemperaturerisestoa
pointgreaterthanorequalto
thesetpoint+8°F.
Steamnotbeingreleased
fromchamber.
Checkforobstructioninplastictube
connectingunittoexternalcondensingtank.
Unkinkplastictube,removeobstruction,
orreplace.
Checkcoppercoilinexternalcondensing
tankforblockageordamage.
Clearobstructionorreplacecoppercoil.
Checkventportinchamberforblockage.
Checklterforblockage.
Clearobstructioninventport.
Cleanorreplacelter.
Checkventvalveforblockage.
PutunitintoServiceDiagnosticsandcheck
ventvalveoperation.
Cleanorreplaceventvalve.
Chambertemperaturesensor
notfunctioningproperly.
Referto:
Temperature Sensor Resistance Test.
Cleanorreplacechambertemperature
sensor.
Boilertemperaturesensor
notfunctioningproperly.
Referto:
Temperature Sensor Resistance Test.
Replaceboilertemperaturesensor.
PCBoardisnotprocessing
temperatureinput(s)properly.
Verifyboilerandchambertemperature
sensorsareworkingproperly.
ReplacePCboard,ifnecessary.
C560: Power Up Mode
C561: Select Mode
C562: Close Mode
C563: Heatup Mode
C564: Sterilize Mode
C565: Vent Mode
C566: Open Mode
C567: Dry Mode
C568: Complete Mode
C56X:XXXXMODE
STEAMTEMPHARDWARE
…
ITEMSNOTSTERILE
TURNPOWEROFF/ON
A/DconverteronPCBoard
reportsasteamtemperature
inputvalueoutsidethe
normallimits.
Chambertemperaturesensor
disconnectedfromPCBoard.
Checkchambertemperaturesensorwire
connectionatJ11onPCboard.
Securewireconnection.
Chambertemperaturesensor
notfunctioningproperly.
Referto:
Temperature Sensor Resistance Test.
Cleanorreplacechambertemperature
sensor.
PCBoardisnotprocessing
temperatureinput(s)properly
Verifychambertemperaturesensoris
workingproperly.
ReplacePCboard,ifnecessary.
C570: Power Up Mode
C571: Select Mode
C572: Close Mode
C573: Heatup Mode
C574: Sterilize Mode
C575: Vent Mode
C576: Open Mode
C577: Dry Mode
C578: Complete Mode
Continue on Next
Page...
C57X:XXXXMODE
STEAMTEMPOVERLIMIT
…
ITEMSNOTSTERILE
TURNPOWEROFF/ON
Steamtemperature
greaterthan285°F
(140.6° C)
Steamnotbeingreleased
fromchamber.
Checkforobstructioninplastictube
connectingunittoexternalcondensingtank
Unkinkplastictube,removeobstruction,
orreplace.
Checkcoppercoilinexternalcondensing
tankforblockageordamage.
Clearobstructionorreplacecoppercoil.
Checkventportinchamberforblockage.
Checklterforblockage.
Clearobstructioninventport.
Cleanorreplacelter.
Checkventvalveforblockage.
PutunitintoServiceDiagnosticsandcheck
ventvalveoperation.
Cleanorreplaceventvalve.
A-9© Midmark Corporation 2010 SF-1889 [Revised: 11/18/14]
Troubleshooting Chart
Problem Display / Symptom Probable Cause Check Correction
C570: Power Up Mode
C571: Select Mode
C572: Close Mode
C573: Heatup Mode
C574: Sterilize Mode
C575: Vent Mode
C576: Open Mode
C577: Dry Mode
C578: Complete Mode
(Continued)
C57X:XXXXMODE
STEAMTEMPOVERLIMIT
…
ITEMSNOTSTERILE
TURNPOWEROFF/ON
Steamtemperature
isgreaterthan285°F
(140.6° C)
Chambertemperaturesensor
notfunctioningproperly.
Referto:
Temperature Sensor Resistance Test.
Cleanorreplacechambertemperature
sensor.
Boilertemperaturesensor
notfunctioningproperly.
Referto:
Temperature Sensor Resistance Test.
Replaceboilertemperaturesensor.
PCBoardisnotprocessing
temperatureinput(s)properly.
Verifyboilerandchambertemperature
sensorsareworkingproperly.
ReplacePCboard,ifnecessary.
C630: Power Up Mode
C631: Select Cycle
Mode
C633: Heatup Mode
C636: Open Mode
C637: Dry mode
C638: Complete Mode
C63X:XXXXMODE
PRESSURELOW
…
ITEMSNOTSTERILE
TURNPOWEROFF/ON
DuringHeatupMode,the
rstventvalveopen
operationhastooccur
within10minutes.
Lowwaterinreservoir Checkwaterlevelinreservoir. Fillreservoirwithdistilledwater.
Lowwaterinreservoirnot
caughtbyaC23xerror.
Referto:I/O Test Water Level / External
Tank Display Screen.
Replacewaterlevelsensor.
Waterpumpnotprimed.
Referto:
Water pump priming instructions.
Primewaterpump.
Steamleakage.
Referto:Five point leak check. Repairsteamleak.
Chambertemperaturesensor
notfunctioningproperly.
Referto:Temperature Sensor Resistance
Test.
Cleanorreplacechambertemperature
sensor.
Boilertemperaturesensor
notfunctioningproperly.
Referto:
Temperature Sensor Resistance Test.
Replaceboilertemperaturesensor.
Boilernotfunctioning
properly.
Referto:
Service Diagnostics Boiler Test.
Replaceboiler,ifnecessary.
Pressuretransducertubing
damagedordisconnected.
Inspectcopper&neoprenetubingfrom
chambertopressuretransducer.
Repairorreplacetubing,ifnecessary.
PCBoardmalfuctioning. Varifythattherearenootherwaysthat
pressurecouldnotbetransmittedfrom
chambertopressuretransducer.
ReplacePCboard,ifnecessary.
Defectivecheckvalvein
waterpump.
Checkforlargeairpocketformingintubing,
thatmovesfromtheboilertowaterpump.
Replacewaterpump.
Pressuretransducerportin
chamberblocked
Checkforobstruction. Removeobstruction.
Photointerrupterobstructed.
Referto: Photo Interrupter Clearobstructionandclean.Verifyhard-
wareisworkingbacktoprocessor.
A-10© Midmark Corporation 2010 SF-1889 [Revised: 11/18/14]
Troubleshooting Chart
Problem Display / Symptom Probable Cause Check Correction
C641: Select Mode
C642: Close Mode
C645: Vent Mode
C646: Open Mode
C647: Dry Mode
C648: Complete Mode
C64X:XXXXMODE
PRESSUREHIGH
…
ITEMSNOTSTERILE
PUSHSTOPTORESTART
Thegaugepressureinthe
chamberisgreaterthan
2.0PSIg(13.8kPag)when
itshouldbenearzero.
Steamnotbeingreleased
fromchamber.
Checkforobstructioninplastictube
connectingunittoexternalcondensingtank.
Unkinkplastictube,removeobstruction,
orreplace.
Checkcoppercoilinexternalcondensing
tankforblockageordamage.
Clearobstructionorreplacecoppercoil.
Checkventportinchamberforblockage.
Checklterforblockage.
Clearobstructioninventport.
Cleanorreplacelter.
Checkventvalveforblockage.
PutunitintoServiceDiagnosticsandcheck
ventvalveoperation.
Cleanorreplaceventvalve.
Pressureisbeingread
abnormallyhighby
pressuretransducer.
Referto:Service Diagnostics
Chamber Pressure Display Screen.
ReplacePCboard,ifnecessary.
C660: Power Up Mode
C661: Select Mode
C662: Close Mode
C663: Heatup Mode
C664: Sterilize Mode
C665: Vent Mode
C666: Open Mode
C667: Dry Mode
C668: Complete Mode
C66X:XXXXMODE
PRESSUREHARDWARE
…
ITEMSNOTSTERILE
TURNPOWEROFF/ON
A/Dconverterreportsa
gaugepressureinputvalue
outsidethenormallimits.
Electricalinterruption. Iferrorisintermittant. None.Ignorecode.
PCBoardmalfunctioning. Iferrorispersistant,thenaproblemexist. ReplacePCboard,ifnecessary.
C670: Power Up Mode
C671: Select Mode
C673: Heatup Mode
C674: Sterilize Mode
C675: Vent Mode
C67XXXXXMODE
PRESSUREOVERLIMIT
…
ITEMSNOTSTERILE
TURNPOWEROFF/ON
Gaugechamberpressure
greaterthan40PSI
(275.8kPa)
Steamnotbeingreleased
fromchamber.
Checkforobstructioninplastictube
connectingunittoexternalcondensingtank
Unkinkplastictube,removeobstruction,
orreplace.
Checkcoppercoilinexternalcondensing
tankforblockageordamage.
Clearobstructionorreplacecoppercoil.
Checkventportinchamberforblockage.
Checklterforblockage.
Clearobstructioninventport.
Cleanorreplacelter.
Checkventvalveforblockage.
PutunitintoServiceDiagnosticsandcheck
ventvalveoperation.
Cleanorreplaceventvalve.
Chambertemperaturesensor
notfunctioningproperly.
Referto:
Temperature Sensor Resistance Test.
Cleanorreplacechambertemperature
sensor.
Boilertemperaturesensor
notfunctioningproperly.
Referto:
Temperature Sensor Resistance Test.
Replaceboilertemperaturesensor.
PCBoardisnotprocessing
temperatureinput(s)properly.
Verifytemperaturesensorsareworking
properly.
ReplacePCboard,ifnecessary.
A-11© Midmark Corporation 2010 SF-1889 [Revised: 11/18/14]
Troubleshooting Chart
Problem Display / Symptom Probable Cause Check Correction
C760: Power Up Mode
C761: Select Mode
C762: Close Mode
C763: Heatup Mode
C764: Sterilize Mode
C765: Vent Mode
C766: Open Mode
C767: Dry Mode
C768: Complete Mode
C76X:XXXXMODE
BOILERTEMPHARDWARE
…
ITEMSNOTSTERILE
TURNPOWEROFF/ON
A/Dconverterreportsboiler
temperatureinputvalue
outsidethenormallimits.
Boilertemperaturesensor
disconnectedatPCboard.
Checkforlooseorbrokenwireconnections
atJ10onPCboard.
Securewireconnections.
Boilertemperaturesensor
notfunctioningproperly.
Referto:
Temperature Sensor Resistance Test.
Replaceboilertemperaturesensor.
Boilernotfunctioning
properly.
Referto:
Service Diagnostics Boiler Test.
Replaceboilerifnecessary.
PCBoardisnotprocessing
temperatureinput(s)properly.
Verifyboilertemperaturesensorisworking
properly.
ReplacePCboard,ifnecessary.
C770: Power Up Mode
C771: Select Mode
C772: Close Mode
C773: Heatup Mode
C774: Sterilize Mode
C775: Vent Mode
C776: Open Mode
C777: Dry Mode
C778: Complete Mode
C77X:XXXXMODE
BOILERTEMPOVERLIMIT
…
ITEMSNOTSTERILE
TURNPOWEROFF/ON
Boilertemperatureisgreater
than340°F(171.1° C)
Steamnotbeingreleased
fromchamber.
Checkforobstructioninplastictube
connectingunittoexternalcondensingtank
Unkinkplastictube,removeobstruction,
orreplace.
Checkcoppercoilinexternalcondensing
tankforblockageordamage.
Clearobstructionorreplacecoppercoil.
Checkventportinchamberforblockage.
Checklterforblockage.
Clearobstructioninventport.
Cleanorreplacelter.
Checkventvalveforblockage.
PutunitintoServiceDiagnosticsandcheck
ventvalveoperation.
Cleanorreplaceventvalve.
Boilertemperaturesensor
disconnectedatPCboard.
Checkforlooseorbrokenwireconnections
atJ10onPCboard.
Securewireconnections.
Boilertemperaturesensor
notfunctioningproperly.
Referto:
Temperature Sensor Resistance Test.
Replaceboilertemperaturesensor.
PCBoardisnotprocessing
temperatureinput(s)properly
Verifyboilertemperaturesensorsis
workingproperly.
ReplacePCboard,ifnecessary.
C860: Power Up Mode
C861: Select Mode
C862: Close Mode
C863: Heatup Mode
C864: Sterilize Mode
C865: Vent Mode
C866: Open Mode
C868: Complete Mode
C86X:XXXXMODE
DRYERTEMPHARDWARE
…
ITEMSNOTSTERILE
TURNPOWEROFF/ON
A/Dconverterreportsadry
heatertemperatureinput
valueoutsidethenormal
limits.
Dryheater(s)disconnected
fromPCboard.
CheckdryheaterconnectionsatJ7&J8
onPCboard.
Secureheaterconnections.
Dryheatersmalfunctioning. PutunitintoServiceDiagnosticsandverify
heateroperation.
Iftheconditionpersist,contactMidmark
TechnicalServiceDepartment.
Dryheater(s)sensingcircuit
onPCBoardmalfunctioning.
Iftheconditionpersist,contactMidmark
TechnicalServiceDepartment.
A-12© Midmark Corporation 2010 SF-1889 [Revised: 11/18/14]
Troubleshooting Chart
Problem Display / Symptom Probable Cause Check Correction
C870: Power Up Mode
C871: Select Mode
C872: Close Mode
C873: Heatup Mode
C874: Sterilize Mode
C875: Vent Mode
C876: Open Mode
C877: Dry Mode
C878: Complete Mode
C87X:XXXXMODE
DRYERTEMPOVERLIMIT
…
ITEMSNOTSTERILE
TURNPOWEROFF/ON
Dryheatertemperature
greaterthan325°F
(162.8° C)
Dryheatersmalfunctioning. PutunitintoServiceDiagnosticsandverify
heateroperation.
Iftheconditionpersist,contactMidmark
TechnicalServiceDepartment.
Dryheater(s)sensingcircuit
onPCBoardmalfunctioning.
Iftheconditionpersist,contactMidmark
TechnicalServiceDepartment.
C881: Select Mode
C887: Dry Mode
C88X:XXXXMODE
DRYERHI-LIMITOPEN
…
ITEMSNOTSTERILE
TURNPOWEROFF/ON
Dryheaterhigh-limit
thermostathasopenedforat
least0.25seconds.
Dryheater(s)disconnected
fromPCboard.
CheckdryheaterconnectionsatJ7&J8
onPCboard.
Secureheaterconnections.
PCboardmalfunctioning.
Referto:Dry Heater Voltage Test. ReplacePCboard,ifnecessary.
Dryheater(s)malfunctioning.
Referto:Dry Heater Resistance Test. Iftheconditionpersist,contactMidmark
TechnicalServiceDepartment.
Dryheater(s)sensingcircuit
onPCBoardmalfunctioning.
Iftheconditionpersist,contactMidmark
TechnicalServiceDepartment.
C981: Select Mode
C982: Close Mode
C983: Heatup Mode
C984: Sterilize Mode
C986: Open Mode
C987: Dry Mode
C98X:XXXXMODE
BOILERHI-LIMITOPEN
…
ITEMSNOTSTERILE
TURNPOWEROFF/ON
Boileroverheatthermostat
hasopenedforatleast0.25
seconds&needstobe
manually reset.
Waterpumpnotprimed. Referto:
Water pump priming instructions.
Primewaterpump.
Boilertemperaturesensor
notfunctioningproperly.
Referto:
Temperature Sensor Resistance Test.
Replaceboilertemperaturesensor.
Boileroverheatthermostat
disconnectedatPCboard.
CheckwireconnectionsatJ3onPCboard. Securewireconnections.
Looseorbrokenwire
connectionsatboiler
overheatthermostat.
Checkwireconnectionsatboiler
overheatthermostat.
Securewireconnections.
Boileroverheatthermostat
trippedopenprematurely.
Checkifoverheatthermostatcanbe
manuallyresetwithacooledboiler.
Replaceoverheatthermostat.
Sterilizer is not drying
instruments properly.
Continue on Next
Page...
Instrumentsarewetafter
drycycleiscomplete.
Sterilizernotlevel. Checkifsterilizerislevel. Placesterilizeronalevelsurface.
Sterilizerisoverloaded. Checkifchamberisoverloaded. Reduceloadsize.
Poucheswereplacedin
chamberwithpapersideup.
Checkifpouchesarepapersideup. Placepouchesplasticsideup.
Pouchesarestackedontop
ofeachother.
Checkifpouchesareoverlappingor
stackedontopofeachother.
Pouchesmayoverlapslightly,butitems
mustnotbelayered.
A-13© Midmark Corporation 2010 SF-1889 [Revised: mo/dd/yr]
Troubleshooting Chart
Problem Display / Symptom Probable Cause Check Correction
Sterilizer is not drying
instruments properly.
(continued)
Instrumentsarewetafter
drycycleiscomplete.
Doortraywasremovedfrom
chamberandreinserted.
Checkifdoortraywasremovedfrom
chamberbeforecompletionofthedry
cycle.
Instructoperatorthatoncethedoortray
hasbeenremovedfromchamber,thedry
cyclewillbeterminated.
Steamnotbeingreleased
fromchamber.
Checkforobstructioninplastictube
connectingunittoexternalcondensingtank
Unkinkplastictube,removeobstruction,
orreplace.
Checkcoppercoilinexternalcondensing
tankforblockageordamage
Clearobstructionorreplacecoppercoil.
Checkventportinchamberforblockage.
Checklterforblockage
Clearobstructioninventport.
Cleanorreplacelter.
Checkventvalveforblockage.
PutunitintoServiceDiagnosticsandcheck
ventvalveoperation.
Cleanorreplaceventvalve.
Doortrayisnotopeningtoits
fulleststagedextent.
Checkifdoortrayisopeningenoughto
allowsteamtoescapefromchamber.
Adjustdoormotorthrow.
Surfaceheater(s)
malfunctioning.
RunServiceDiagnosticsandcheck
operationofsurfaceheater(s).
ContactMidmark'sTechnicalService
Department.
Biological indicator or
chemical indicator
strip designates unit
not sterilizing.
Positivebiological
indicator.
Nocolorchange,or
incompletechangehas
occurredonchemical
indicatorstrip.
Sterilizeroverloaded. Checkifchamberisoverloaded. Reduceloadsize.
Improperusage. Checkuseroperatingprocedure. Demonstrateproperoperationtouser.
Wrongtypeofbiological
indicatororchemical
indicatorbeingusedforthis
typeofsterilizer.
Checkbiologicalorchemicalindicatorto
assureitisforsteamsterilizationandis
ratedforthecycletimesandtemperature
beingused.
RefertotheUsersGuideforproper
sterilizationmonitoringguidelines.
Biologicalindicatoror
chemicalindicatorbeing
usedispasttheexpiration
date.
Checktheindicatorand/orpackagingto
determinewhattheexpirationdateis.
Useonlybiologicalandchemical
indicatorsthathaven'texpired.
Chemicalindicatorhascome
intocontactwithwaterin
sterilizer.
Checkwithoperatoronthepositioningof
indicatorinthechamber.
Demonstratecorrectpositioningof
indicatortouser.
Biologicalorchemical
indicatorsarebeingstoredin
adampand/orhot
environment.
Checkconditionoftheareawhere
indicatorsarebeingstored.
Informoperatortofollowstorage
recommendationsofindicator
manufacturer.
Sterilizermalfunction PutunitintoServiceDiagnosticsandcheck
forstorederrorcodes,correcttemperature,
andpressurereadings.
Repairasneededbasedonresultsof
ServiceDiagnosticstestresults.
A-14© Midmark Corporation 2010 SF-1889 [Revised: mo/dd/yr]
Troubleshooting Chart
Problem Display / Symptom Probable Cause Check Correction
Maintenance
message is displayed.
PERFORMPERIODICor
MONTHLYMAINTENANCE
Sterilizersoftwaresenses
thatthecyclecounthas
reached7,14,or21days
frominitialstart-upor
monthlymaintenance.
Refertotheuserguidefor
periodicmaintenance
procedures.
Performperiodicmaintenance.
Sterilizersoftwaresenses
thatthecyclecounthas
reached28daysfrominitial
start-upormonthly
maintenance
Refertotheuserguidefor
monthlymaintenance
procedures.
Performmonthlymaintenance.
Printer (Optional)
does not print.
Printerdoesnotgenerate
data.
Printerisoutofpaper. Liftprintercoverandcheck
conditionofpaper.
Installnewrollofpaper(refertouserguide).
Printercartridgeisdry. Checkconditionofprinter
cartridge.
Replaceprintercartridge(refertouserguide).
Printerwireharnessis
disconnected.
Checkprinterwireharness
connectionsonprinterandon
backofsterilizer.
Secureprinterwireharnessconnections.
Printerwireharnesshas
brokenoropenleads.
Checkprinterwireharness
leadsforcontinuity.
Replaceprinterwireharness.
Printerismalfunctioning. Checkfor5VDCatTP3&TP4
onPCboard.
Ifvoltageispresent,replaceprinter.
PCboardismalfunctioning. Checkfor5VDCatTP3&TP4
onPCboard.
Ifvoltageisnotpresent,replacePCboard.
Chamber î‚¿lls with
water.
Chamberllswithwater
whencycleisabortedwith
thedoorclosed.
Unitlostpowerinheatupor
sterilizemodesduetoa
powerfailureoruserturned
powero.Chamberlled
withwaterfromthereservoir
asunitcooled.
Askquestionsaboutrecent
usagepatternsandelectrical
events(storms,outages,ect.).
Replacecheckvalveoncoppertubingbetweenthe
boilerandthepressureinterlockassembly.
English / Metric
display units can
not be changed from
display panel.
WheninUserDiagnosic
mode,English/Metric
displayunitscannotbe
changed.
Version1.1.4mainPCB. IfmainPCBversionis1.1.4. Onversion1.1.4mainPCB,ip#7dipswitchON.
ITEMS NOT STERILE messagewillbedisplayedifanerroroccursinanyofthefollowingmodes:
Closemode,HeatUpMode,orSterilizeMode.Otherwise,messagewillnotappear.
Thismessageisintendedtoindicatetotheuserthattheitemsinthechambermayormaynot
besterile,andmustbereprocessed.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94

Midmark M3 Steam Sterilizer Parts Manual

Type
Parts Manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI