Troubleshooting Chart
Problem Display / Symptom Probable Cause Check Correction
Sterilizer has no power.
Touchpad&displaydonot
work.(Switchfailureunder
probablecausewouldonlybe
onnewerstylePCboards)
Sterilizernotpluggedin. Checkpowercordconnectionsatwall
outletandonbackofsterilizer.
Securepowercordconnections.
Powerswitchturnedo. Checkpowerswitch. Turnpowerswitchon.
Nopoweratwalloutlet. Checkvoltageatwalloutlet. Resetfacilitycircuitbreaker.The
sterilizershouldbepluggedintoa
dedicated20ampcircuit.
F1fuseonPCboardis
blownoropen.
CheckF1fuseforcontinuity. Replacefuse.
Inletswitchmodulefailure.
Checkinletswitchmodule.
Referto:InletSwitchModuleTest
Performcontinuitytestoninletswitch
module.
Sterilizer has power, but
touchpad & display do
not work.
Touchpad&displaydonot
work.
F2fuseonPCboardis
blownoropen.
CheckF2fuseforcontinuity. Replacefuse.
RibboncablefromJ15on
mainPCboardtoJ3on
displayPCboardislooseor
disconnected.
CheckribboncableconnectionsatJ15
onmainPCboard&J3ondisplayPC
board.
Secureribboncableconnections.
J15/J3ribboncablehas
openlead(s).
CheckJ15/J3ribboncablefor
continuity.
Replaceribboncable.
MainPCboardis
malfunctioning.
Checkfor5VDCbetweenTP1&TP4,
andonJ15acrosspins1&2onmainPC
board.
If5VDCisnotpresentatthese
points,replacemainPCboard.
Displayisblankorundined
charcters&internittentbeeps.
DisplayPCboardis
malfunctioning
Checkfor5VDCbetweenR1&J1on
displayPCboard.
If5VDCisnotpresent,replace
displayPCboard.
Displayworksbuttouch
paddoesn't.
RibboncablefromJ2to
touchpadisdisconnected.
CheckJ2Plugconnection. AssureJ2Plugpinsarealignedand
connectedsecurely.
J2ribboncabletotouchpad
hasopenlead(s).
CheckcontinuityofJ2ribboncable
andtouchpad.
Replacetouchpadandribboncable
assembly.
Unit has power, but the
door is not open in Select
Cycle Mode, and there is
no error code associated
with this condition.
Displayisnormal. Doormotorconnectingrodis
broken.
Removetopcoverandverifythatthe
doormotorconnectingrodisintact.
Replacedoormotorconnectingrod.
C010
C010:POWERUPMODE
SYSTEMPOWERLOSS
…
ITEMSNOTSTERILE
PUSHSTOPTORESTART
Powerinterruptionduringany
modeinwhichsterilization
maybecompromised.
Userturnedpoweroduring
ChamberClose,HeatUp,or
SterilizationMode.
Askquestionsaboutusagepatterns Trainuseronproperoperation.
Unitpluggedintoanon
dedicated,20ampcircuit.
Checkifunitispluggedintoadedicated
20ampcircuit.
Connectunittoadedicated20amp
circuit.
Someotherevent(suchasa
storm,apoweroutage,ect.)
interruptedpower.
Askquestionsaboutrecentelectrical
events(storms,outages,ect.).
Ifthisisacommonoccurrence,iso-
latefromsourceofelectricalinterfer-
ence.