  • Hello! I am an AI chatbot trained to assist you with the Invacare HomeFill HF2ECE9 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Invacar1.0L/1.7LCylinderwith
DualLumenConserver
ForHomeFill®OxygenSystem
enHF2ECE6,HF2ECE9
UserManual
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuturereference.
2014/68/EU
93/42/EEC
©2017InvacareCorporation
Allrightsreserved.Republication,duplicationormodicationinwholeorinpartisprohibited
withoutpriorwrittenpermissionfromInvacare.Trademarksareidentiedbyand®.Alltrademarks
areownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted.
MakingLife’sExperiencesPossibleisaregisteredtrademarkintheU.S.A.
WD-40and3–IN-ONEareregisteredtrademarksoftheWD-40Company.
Contents
1General.........................................4
1.1Symbols.....................................4
1.2IntendedUse.................................5
1.3UserProle...................................5
1.4ServiceLife...................................5
1.5Warranty.....................................5
1.6ComponentIdentication........................6
2Safety..........................................7
2.1LabelLocations................................7
2.2GeneralGuidelines.............................8
3Setup...........................................14
3.1DealerSetup..................................14
3.2UserSetup...................................14
3.3ReceivingInspection............................14
3.4CylinderPrellInspection........................14
3.4.1TheHydrostaticTestingDate...................15
3.4.2ExternalExamination.........................16
4Usage..........................................17
4.1BeforeUse...................................17
4.2Connecting/DisconnectingtheCylinderto/fromthe
Compressor.................................17
4.3CylinderAssemblyOperation......................17
5Maintenance.....................................21
5.1Maintenance..................................21
5.2EmptyingtheCylinderAssembly...................22
5.3Cleaning.....................................23
5.4Disinfecting...................................23
5.5Reuse.......................................23
5.6Storage......................................24
5.7OptionalAccessories............................24
6AfterUse........................................25
6.1Disposal.....................................25
6.2WearandTear................................25
7Troubleshooting...................................26
7.1Troubleshooting...............................26
8Technicaldata....................................27
8.1TechnicalDescription............................27
8.2Specications.................................28
Invacar1.0L/1.7LCylinderwithDualLumenConserver
1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazards
orunsafepracticeswhichcouldresultinpersonalinjuryor
propertydamage.Seetheinformationbelowfordenitions
ofthesignalwords.
DANGER!
Dangerindicatesanimminentlyhazardous
situationwhich,ifnotavoided,willresultin
deathorseriousinjury.
WARNING!
Warningindicatesapotentiallyhazardous
situationwhich,ifnotavoided,couldresultin
deathorseriousinjury.
CAUTION!
Cautionindicatesapotentiallyhazardous
situationwhich,ifnotavoided,mayresultin
propertydamageorminorinjuryorboth.
Indicatesahazardoussituationthatcouldresult
indamagetopropertyifitisnotavoided.
Givesusefultips,recommendationsandinformation
forefcient,trouble-freeuse.
Readthemanual
Noopename
DONOTSmoke
Oxidizer
Recycle
DONOTdisposeofinhouseholdwaste
Manufacturer
EURepresentative
ThisproductcomplieswithDirective93/42/EEC
concerningmedicaldevices.
Thelaunchdateofthisproductisstatedinthe
CEdeclarationofconformity.
4
1176148-D
General
1.2IntendedUse
Thiscylinderisintendedtostore,regulate,anddeliver
supplementaloxygentopatientswhomayhavedifculty
extractingoxygenfromtheairtheybreathe.Itisforpatients
whowouldnormallyreceivetheoxygenviaanasalcannula.
ThedevicedeliversUSP93±3%oxygenatanumberof
differentowsettingsandmayincreasethedryingoutof
airways.Thisproductistobeusedasanoxygensupplement
andisnotintendedtobelifesupportingorlifesustaining.
DANGER!
RiskofInjuryorDeath
Thisproductistobeusedasanoxygen
supplementandisnotintendedtobe
life-supportingorlife-sustaining.ONLYusethis
productifthepatientiscapableofspontaneous
breath,abletoinhaleandexhalewithouttheuse
ofamachine.
WARNING!
RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduse
andspecicationshasnotbeentestedandmay
leadtoproductdamage,lossofproductfunction,
orpersonalinjury.
DONOTusethisproductinanywayotherthan
describedinthespecicationsandintended
usesectionsofthismanual.
1.3UserProle
Nospecicproductknowledgeortrainingisrequired,other
thanwhatiscontainedinthismanual,tooperatethe
product.Itisnotintendedforsupplyingoxygentherapy
duringrelling.Thisproductisintendedtobeusedbyor
underadultsupervisiononlyafterreadingandunderstanding
theinstructionsandwarningsofthisusermanual.
Onlypersonnelinstructedandtrainedintheuseofthe
HomeFill®ConservingCylindershouldoperatethisproduct.
Theuseofthisproductislimitedtotheoxygenpatient.
CylindersMUSTbeusedonlybytheoxygenpatientandare
nottobedistributedtoanyotherindividualforanypurpose.
ThisproductisNOTtobeusedbypatientswhobreathe
throughtheirmouths.
1.4ServiceLife
Theexpectedservicelifeoftheconserverisveyearswhen
usedinaccordancewiththesafetyinstructions,maintenance
intervalsandcorrectusestatedinthismanual.Refertothe
Maintenancesection.
Theexpectedservicelifeofthecylinderisbasedontesting
thecylindereverytenyearsandusingitinaccordancewith
thesafetyinstructions,maintenanceintervalsandcorrect
usestatedinthismanual.Refertothefollowingprocedures:
TheHydrostaticTestingDate,ExternalExamination,
Maintenance
1.5Warranty
Termsandconditionsofthewarrantyarepartofthegeneral
termsandconditionsparticulartotheindividualcountriesin
whichthisproductissold.
1176148-D5
Invacar1.0L/1.7LCylinderwithDualLumenConserver
1.6ComponentIdentication
Item
Description
A
CylinderAssembly
B
Cylinder
C
ContinuousFlowConserver
D
FlowDial
E
IndicatorWindow
F
Switch
G
OutletConnection
H
CylinderFillport
I
ShutoffValve
J
CylinderGauge
K
CylinderGaugeNeedle
61176148-D
Safety
2Safety
2.1LabelLocations
Serialnumberlocatedaboveproductlabelshown
here.
1176148-D
7
Invacar1.0L/1.7LCylinderwithDualLumenConserver
2.2GeneralGuidelines
Inordertoensurethesafeinstallation,assemblyand
operationoftheproducttheseinstructionsMUSTbe
followed.
DANGER!
RiskofDeath,InjuryorDamage
Improperuseofthisproductmaycause
death,injuryordamage.Thissectioncontains
importantinformationforthesafeoperation
anduseofthisproduct.
DONOTusethisproductoranyavailable
optionalequipmentwithoutrstcompletely
readingandunderstandingtheseinstructions
andanyadditionalinstructionalmaterialsuch
asusermanuals,servicemanualsorinstruction
sheetssuppliedwiththisproductoroptional
equipment.
Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare
professional,dealerortechnicalpersonnel
beforeattemptingtousethisequipment.
CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproduct
isnotworkingcorrectly,contactatechnician
orInvacareforrepair .
THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER!
RiskofDeath,InjuryorDamage
Usethiscylinderinaccordancewiththe
HomeFillcompressorusermanual.
DONOThandlethecylinderassemblyor
usethecontentsuntilyouareprofessionally
trained,includingemergencyprocedures.
Trainingrequiredtooperatetheproductis
containedintheusermanual.
Thisproductisnotintendedforsupplying
oxygentherapyduringrelling.
Thisproductisintendedtobeusedbyorunder
adultsupervision.
OxygenCylinderslledbytheInvacareHomeFill
oxygensystemshallbeusedforpersonaluse
only.“Nottobelledforresaleoruseby
professionalusers.
ThiscylinderistobeusedwiththeInvacare
HomeFilloxygensystem.Thiscylinder
contains93±3%oxygenproducedbyan
oxygenconcentratorusingthepressureswing
adsorptionprocess.
ALWAYSopenthevalveslowlyandclosethe
valveaftereachuseandwhenempty.
81176148-D
Safety
DANGER!
RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,
greasysubstancesandothercombustiblesare
easilyignitedandburnwithgreatintensityin
oxygenenrichedairandwhenincontactwith
oxygenunderpressure.Toavoidre,death,
injuryordamage:
DONOTSMOKEwhileusingthisproduct.
DONOTusenearOPENFLAMEorIGNITION
SOURCES.
DONOTuseanylubricantsonthecylinder
unlessrecommendedbyInvacare.
NOSMOKINGsignsshouldbeprominently
displayed.
Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic
electricitycreatedbyanytypeoffriction.
Keepallmatches,lightedcigarettes,electronic
cigarettesorothersourcesofignitionoutof
theroominwhichthiscylinderislocatedand
awayfromwhereoxygenisbeingdelivered.
Keeptheoxygentubing,cord,andcylinder
outfromundersuchitemsasblankets,
bedcoverings,chaircushions,clothing,and
awayfromheatedorhotsurfacesincluding
spaceheaters,stoves,andsimilarelectrical
appliances.
CAUTION!
Federal(statutory)lawrestrictsthisdevicesale
toorontheorderofamedicalpractitioner
licensedbyagovernmentalagencywherehe/she
practices.
ONLYalicensedmedicalpractitionermayorder
thepurchaseoruseofthisdevice.
1176148-D9
Invacar1.0L/1.7LCylinderwithDualLumenConserver
OperatingInformation
DANGER!
RiskofFire
Dirtorcontaminantsonthecylinderassemblyor
conservercancausere,injuryordeath.
DONOTuseifdirtorcontaminantsarepresent
onoraroundcylinderorconserver .
DONOTuseorllcylinderassemblyifthe
cylinderllportisdirtyorcontaminated.
Contactyourproviderforserviceorrepair.
DONOTallowoilfromyourhandsorother
sourcestocomeintocontactwiththe
compressorcouplingorcylinderassembly
ttings.Thesesubstancesmaybecome
ammableinthepresenceofoxygenandcause
injury.Examplesofitemstoavoidarehand
creamorlotion,cookingoil,suntanoilor
sunblockandsimilaroilyproducts.Ifyouuse
thesetypesofproductsInvacaresuggestsyou
washyourhandspriortousingtheHomeFill
system.Ifthecylinderllportshouldcome
incontactwithsuchsubstances,contact
yourproviderforservice.Avoidtouchingthe
cylinderllport.
DANGER!
RiskofFire
Dirtorcontaminantsonthecylinderassemblyor
conservercancausere,injuryordeath.
DONOTcleanwitharomatichydrocarbons,
including,butnotlimitedtobenzeneand
toluene.
NEVERoilorlubricatethecompressorcoupler
orcylinderllport.NEVERuseanypenetrating
oilorlubricantsuchasWD-40®or3-IN-ONE®.
Ifconnectiondifcultyisexperienced,verify
thatthecylinderllportisbeinginserted
straightdownintothecompressorcoupler .
Ifcontinueddifcultyisexperienced,contact
yourproviderforassistance.
101176148-D
Safety
DANGER!
RiskofFire
Dirtorcontaminantsonthecylinderassemblyor
conservercancausere,injuryordeath.
DONOTusecleaning/disinfectingsolutionson
thellporttting.
DONOTimmerseproductinanykindofliquid.
DONOTallowanykindofliquidtogetinside
openingsoftheconserver .
DONOTspraycleaning/disinfectingsolutions
onproduct.
DONOTautoclave.
DONOTgassterilizewithethyleneoxide.
DONOTattempttorepaircylinderassembly.
AllrepairsMUSTbeperformedbyInvacare
CorporationorinanInvacareapprovedrepair
center .
Storecylinderassemblyinaclean,dryandcool
areawhennotinuse.
DANGER!
RiskInjuryorDeath
Toavoidchokingoringestionofchemicalsfrom
airwaycontamination:
DONOTusetheproductinthepresence
ofpollutants,smoke,fumes,ammable
anesthetics,cleaningagents,orchemical
vapors.
WARNING!
RiskofInjuryorDeath
Contaminantsintheoxygenmaycauseinjuryor
death.
DONOTuseiftheoxygenhasanunusualodor.
DONOTuseifparticulatesareinhaled.
ContactInvacareoradealerforservice.
WARNING!
RiskofInjuryorDamage
Toavoidinjuryordamagefromproductmisuse:
Theowdialshouldbeturnedwithnormal
pressurefromone’shand.Iftheowdialis
stifftoturn,donotforce,otherwisedamage
orinjurycouldoccur .
DONOTuseaowoutletfordrivingany
medicalequipment.
1176148-D
11
Invacar1.0L/1.7LCylinderwithDualLumenConserver
CAUTION!
RiskofInjuryorDamage
Toreducetheriskofinjuryordamagefrom
droppedorimpactedcylinderassemblies:
Usetwohandswhenhandling/transporting
oxygencylinders.
OnlyuseInvacareCorporationcarryingbag.
DONOTallowcylinderassembliestotiporfall.
DONOTdropcylinderassemblies.
Securegascylinderssotheycannotfall.
Ensurecylinderisalwaysinuprightposition
whennotconnectedtotheHomeFill
compressor .
DONOTallowcylinderstoimpactinanimate
objects.
CAUTION!
RiskofDamage
Toavoidproductdamage:
DONOTblocktheoutletttingorkinkthe
cannulawhentheproductisinuse,asthis
maydamagetheunit.
DONOTllthecylinderassemblyoutdoorsor
anywherebelow13°C(55°F).
DANGER!
RiskofInjuryorDeath
Thecannulacancausetripping,falling,orother
injuryifimproperlypositionedandsecured.To
avoidinjuryordeath:
ThecannulaMUSTberoutedandsecured
properly.
DONOTpositionthecannulaaroundtheneck.
Ensurethepatientcanmovefreelywhile
wearingthecannula.
Avoidpositioningcannulasacrossareasofhigh
foottrafc(i.e.aisles,doorways,hallways,etc.)
WARNING!
RiskofInjuryorDeath
Toavoidchokingand/orstrangulationfromtubing
entanglement:
Keepchildrenandpetsawayfromnasal
cannulaandtubing.
Closesupervisionisnecessarywhenthenasal
cannulaisusedbyornearchildrenand/or
disabledpersons.
12
1176148-D
Safety
WARNING!
RiskofInjuryorDeath
Toreducetheriskofinjuryordeathfromillness:
Beforetherstpatientuse,theproviderMUST
disinfecttheproduct.Refertothedisinfecting
section.
Replacethenasalcannulaonaregularbasis.
Checkwithyourequipmentprovideror
physiciantodeterminehowoftenthecannula
shouldbereplaced.
DONOTsharecannulasbetweenpatients.
Thecylinderposesnobiocompatibilityhazardsand
islatexandphthalatefree.Thebiocompatibility
ofcylinderaccessoriesmayvarybymanufacturer .
Invacareaccessoriessuchasthecylinderbagand
Invacareoxygentubingarelatexandphthalatefree.
Refertothepackagingonanycylinderaccessoriesto
determinethebiocompatibilityhazard.
ThecylinderassemblyMUSTbeemptiedofgaspriorto
shippingbycourierorpostalservice.RefertoEmptyingthe
CylinderAssembly.
Becausethecylindercontainsmagnetic,ferrousmaterial,it
isnotsuitabletobeinanMRIroom.
Besureallconnectionsaretightandleakfree.DONOTuse
aliquidleakdetectortotestforleaks.Leaksareindicated
bylossofpressurewhenthecylinderassemblyisnotinuse
andbyhissingsounds.Thisproductisequippedwitharelief
valve.Ifyouhearaloudhissingorpoppingsoundcoming
fromtheproduct,discontinueuseofproduct,turntheow
dialtothe“0”setting,andcontactyourdealer/HomeCare
equipmentprovider .
ALWAYSconrmprescribeddosebeforeadministeringto
patientandmonitoronafrequentbasis.
NOOXYGENisdeliveredwhentheconserverowdial
settingis“0”.
DONOTuseahumidierwiththisproduct.
Keepcylinderinawellventilatedplace.
Thisproductisdesignedtooperatewithastandardor
highowduallumen,adultoxygennasalcannulawitha
maximumlengthof2.1m(7ft).
Thecannulaisforsinglepatientuseonly.
CannulaMUSTbeinstalledonoutletconnectionpriortouse.
RefertotheHomeFillUserManualforadditionalwarnings,
cautionsandsafetyinstructions.
1176148-D13
Invacar1.0L/1.7LCylinderwithDualLumenConserver
3Setup
3.1DealerSetup
WARNING!
Beforetherstpatientuse,theproviderMUST
disinfecttheproduct.RefertotheDisinfecting
section.
Beforepatientuse,theproviderMUSTcheck
thehydrostatictestingdate.Refertothe
HydrostaticTestingDatesection.
WARNING!
RiskofInjury
Formedicalapplications,itisveryimportant
toselecttheprescribedlevelofoxygenow.
Uninterrupteduseofhighconcentrationsof
oxygenoveralongduration,withoutmonitoring
itseffectonoxygencontentofarterialblood,
maybeharmful.
UseONLYasdirectedbyalicensedpractitioner.
DONOTincreaseordecreasetheowunlessa
changehasbeenprescribedbyyourphysician
ortherapist.
3.2UserSetup
WARNING!
Whenreceivingthecylinderfromtheprovider,
theuser/patientMUSTperformareceiving
inspection.RefertotheReceivingInspection
section.
3.3ReceivingInspection
Removeproductfrompackageandinspectfordamage.
Ifthereisanydamage,DONOTuse.Contactyour
dealer/HomeCareequipmentprovider .
3.4CylinderPrellInspection
WARNING!
RiskofInjuryorDamage
Oxygencylindersshallbeusedforpersonal
useonly.Nottobelledforresaleoruseby
professionalusers.
FillingofcylindersisperformedbyInvacare
HomeFilloxygensystemonly.
ThiscylinderistobeusedwiththeHomeFill
system.Thiscylindercontains93±3%oxygen
producedbyanoxygenconcentratorusingthe
pressureswingadsorptionprocess.
DONOThandlethecylinderorusecontents
untilyouareprofessionallytrained,including
emergencyprocedures.
Usethiscylinderinaccordancewiththe
HomeFillusermanual.
Openandclosetheshutoffvalveslowly.Close
theshutoffvalveaftereachuseandwhen
empty.
Securethecylinderduringstorageanduse.DO
NOTdrop.
Nosmokingincylinderarea.Keepawayfrom
heat,ameorspark.
Keepoutofthereachofchildren.
14
1176148-D
Setup
DANGER!
RiskofInjuryorDeath
Improperllingorabuseofthiscylinder,orfailure
tofollowthesesafetyinstructionsmaycause
seriousbodilyinjuryordeath.Highpressure
oxidizinggasvigorouslyacceleratescombustion.
DONOTalterormodifycylinderorrelated
components.
DONOTuseoilorgrease.Keepcylinder,
shutoffvalve,allequipmentandconnections
freeofoilandgreasetoavoidviolentignition
resultinginseriousbodilyinjuryordeath.
3.4.1TheHydrostaticTestingDate
AluminumcylindersMUSTundergotestingeverytenyears.
DANGER!
RiskofDeath,InjuryorDamagefromFire
Toavoidre,death,injuryordamage:
Cylindersthathavebeenrenishedand/or
exposedtoelevatedtemperaturesMUSTbe
hydrostaticallytestedbeforelling.Cylinder
temperatureshouldnotexceed52°C(125°F).
Aluminumcylinderssubjecttotheactionof
reorheatedtemperaturesinexcessof177°C
(350°F)MUSTbewithdrawnfromserviceand
condemnedbytrainedpersonnel.DONOT
usecausticpaintstrippers.Valveandsafety
reliefdevicesMUSTberemovedandreplaced
bytrainedandauthorizedpersonnel.DONOT
alterorchangevalveassembly.
AllcylindersMUSTbeinspectedbefore
attemptingtoll;otherwise,injuryordamage
mayoccur.
WARNING!
RiskofInjury
Toreducetheriskofinjury:
DONOTllcylindersthathavenotbeentested
inthepasttenyears.Otherwise,serious
personalinjurymayresult.Contactyour
dealer/HomeCareequipmentproviderfor
replacement.
1176148-D15
Invacar1.0L/1.7LCylinderwithDualLumenConserver
3.4.2ExternalExamination
WARNING!
RiskofInjuryorDamage
Toreducetheriskofinjuryorproductdamage:
DONOTusethecylinderifanyofthefollowing
itemsarefound.Contactyourdealer/Home
Careequipmentproviderforreplacement.
1.Examinetheoutsideofthecylinderforthefollowing
conditions.Iftheyexist,contactyourproviderfor
service.
Dentsordings,arcburns,oilorgrease,anyother
signsofdamagethatmightcauseacylindertobe
unacceptableorunsafeforuse
2.Examinethecylinderforevidenceofreorthermal
damage.Evidenceincludescharringorblisteringof
thepaint,orotherprotectivecoatingorheatsensitive
indicator .Ifreorthermaldamageisfound,contact
yourdealer/HomeCareequipmentproviderfora
replacement.
3.Inspectthecylinderllportforthefollowing.Ifthey
exist,contactyourproviderforareplacement.
Debris,oilorgrease,noticeablesignsofdamage,
signsofcorrosioninsidethevalve,signsofexcessive
heatorredamage,signsofpressureleaks
161176148-D
Usage
4Usage
4.1BeforeUse
WARNING!
RiskofInjuryorDamage
Defectiveorinoperablecylindersmaycause
injuryordamage.
AllcylindersMUSTbeinspectedbefore
attemptingtoll.
Ifthecylinderisfoundtobedefectiveor
doesnotworkproperly,contactanInvacare
dealerorqualiedtechnicianforserviceor
replacement.
Beforellingacylinder,readandfollowtheinformationin
thefollowingsections:
qCylinderPrellInspection
qExternalExamination
4.2Connecting/DisconnectingtheCylinder
to/fromtheCompressor
RefertotheHomeFillCompressorusermanualfor
operatinginstructionstoconnectanddisconnectthe
cylindertotheHomeFillCompressor.
4.3CylinderAssemblyOperation
WARNING!
RiskofFire
NEVERsmokeinanareawhereoxygenisbeing
administered.
NEVERusenearanytypeofameor
ammable/explosivesubstances,vaporsor
atmosphere.
1176148-D
17
Invacar1.0L/1.7LCylinderwithDualLumenConserver
1.Priortoeachuse,inspectproductforvisibledamage.
RefertotheExternalExaminationsection.DONOTuse
ifanydamageisfound.
Ifforanyreason,anylabelbecomesillegibleor
lost,contactyourdealer/HomeCareequipment
provider .
2.InspectthecylindergaugeJtoensurethepressure
isatleast1500psig(10342kPa).Ifbelow1500
psig(10342kPa),thenrellthecylinderBusingthe
HomeFillsystem.
WARNING!
Whencylindergaugeneedleoncylinder
gaugeentersredzone,itisrecommended
torellthecylinderontheInvacare
HomeFillCompressor .
3.Positionthecylindersotheoxygencylindervalve
outletGispointingawayfromtheuserandanyother
person(s).
WARNING!
RiskofInjury
Toreducetheriskofinjury:
DONOTdirectowofoxygenatany
person,orammablematerialwhen
adjustingtheowdial.
4.TurntheowdialDtothe“0”setting.
5.Attachaduallumenadultoxygennasalcannulawith
amaximumlengthof2.1m(7feet)totheproducts
twooutletconnectionsaccordingtothecannula
manufacturersinstructions.
Itdoesnotmatterwhichtubeisinstalledon
whichoutletconnection.
DONOTusepediatric,lowownasalcannulas
oroxygenmaskswiththisproduct.
DANGER!
RiskofDeath
Thecannulacancausetripping,falling,or
otherinjuryifimproperlypositionedand
secured.Toavoidinjuryordeath:
ThecannulaMUSTberoutedandsecured
properly.
DONOTpositionthecannulaaroundthe
neck.Ensurethepatientcanmovefreely
whilewearingthecannula.
Avoidpositioningcannulasacrossareasof
hightrafc(i.e.aisles,doorways,hallways,
etc.)
181176148-D
Usage
6.Placethecannulaoveryourearsandpositionthe
prongsinyournoseasinstructedbyyourHealthCare
providerorcannulamanufacturer.
7.Ensuretherearenokinksortwistsinthecannulatubing.
8.OpentheShutoffvalveIknobslowlytowardstheON
direction(counterclockwise)
Thiswillallowcylinderpressuretotravelthrough
theconserver.
9.UsetheswitchFtoselectoneofthefollowingmodes:
ContinuousFlowL
ConservingPulseM
Theconserverislabeled“CONTINUOUS”and
“PULSE”.
Therstthreeows(0.25,0.50and0.75L/min)
areincontinuousmodeatalltimes,regardless
oftheswitchposition.
10.TurnowdialuntiltheindicatorwindowEisaligned
withtheprescribedsetting.
Whentherearetwonumbersdisplayedinthe
indicatorwindow,thelargernumberisthe
PULSEowsettingandthesmallernumberis
theCONTINUOUSowsetting.
WARNING!
Toavoidinjurytopatient,ALWAYSconrm
prescribedsettingbeforeadministeringto
patientandmonitorowonafrequent
basis.
OnlyusetheInvacareCorporationcarrying
bag.
1176148-D19
Invacar1.0L/1.7LCylinderwithDualLumenConserver
11.Breathethroughthenoseandfeelapulseofoxygen
asyouinhale.
Whenonapulsesetting,thereisow,ora
pulse,onlyatthebeginningofeachbreath.If
youdonotfeelthepulseatthebeginningof
eachbreath,checkthesetting.
Becauseeachpatient’sbreathingpatternis
differentandtheenvironmentvaries,itmaybe
difculttofeelsomelowsettingpulses.
Ifoperatingthecylinderbelow1.7ºC(35ºF)
usethe2L/Mincontinuoussettingonly.
Whentheunitisrstswitchedfrompulseto
continuousow,asingleinhalationisrequired
toinitiatethecontinuousow.
12.Whennishedusingthecylinder,performthefollowing
steps:
a.Turntheowdialtothe“0”setting.
b.Slowlyclosetheshutoffvalve(clockwise)fully.
201176148-D
/