R&J LASER Physiolaser olympic 500 Instructions Manual

Category
Skin care appliances
Type
Instructions Manual
Physiolaser olympic 500
Version 1.4 June 2007
1
Physiolaser olympic 500
Instructions for Veterinary Use
Physiolaser olympic 500
Version 1.4 June 2007
2
Contents
General Notes .............................................................................................................. 3
Physiolaser principle of function .................................................................................. 3
Safety standard ............................................................................................................ 4
Safety Precautions ....................................................................................................... 5
Device out of use.......................................................................................................... 5
Description of the Device / Proper Usage .................................................................... 6
Technical data .............................................................................................................. 6
Content of delivery ....................................................................................................... 6
Proper usage and field of application ........................................................................... 7
Indications .................................................................................................................... 7
Annual inspection of the device.................................................................................... 8
Explanation of the pictograms ...................................................................................... 9
Location of the labels ................................................................................................. 10
Laser probes .............................................................................................................. 11
Sockets and connections on the front cover and probe.............................................. 12
Buttons on the single laser probe ............................................................................... 12
Buttons on the cluster laser probe.............................................................................. 12
Cluster probes ............................................................................................................ 13
LED indicator, display................................................................................................. 14
Display after starting the device with the Key switch .................................................. 15
Button functions on the display (yellow color shows active function).......................... 16
Pre-settings, default setting........................................................................................ 21
Starting operation ....................................................................................................... 23
Operational modes ..................................................................................................... 25
Before starting the therapy mode ............................................................................... 25
Selecting the settings for the laser probe: Power, Energy, Time ................................ 26
Pilot beam, laser control LED on the probes .............................................................. 28
Acoustic control .......................................................................................................... 28
Laser Test .................................................................................................................. 29
NOGIER/BAHR program............................................................................................ 29
REININGER program ................................................................................................. 30
Free programmable frequencies ................................................................................ 31
Therapy programmes (based on Nogier / Bahr frequencies)...................................... 32
Diagnostic mode......................................................................................................... 33
Charging the battery................................................................................................... 35
International plugs ...................................................................................................... 36
New installation of the software, update..................................................................... 36
Error messages .......................................................................................................... 36
Applicators, optional ................................................................................................... 37
Laser-protection goggles............................................................................................ 38
Cleaning / Disinfecting................................................................................................ 38
Maintenance/Calibration............................................................................................. 38
Disposal...................................................................................................................... 38
Warranty..................................................................................................................... 38
Electromagnetic compatibility - Warning instructions ................................................ 39
Manufacturer .............................................................................................................. 41
Physiolaser olympic 500
Version 1.4 June 2007
3
General Notes
These operating instructions must be read carefully before starting the device.
These operating instructions describe the correct usage of the Physiolaser olympic
and also point out any potential risks. These operating instructions are to be read and
observed by all persons who operate, use, service and inspect the device.
The laser device should only be operated by authorized, trained personnel that have
been properly instructed in using the device and have been made aware of the dangers
of laser radiation as well as the accident prevention regulation BGV 11 ”Laser radiation”
and the DIN VDE 0837 directive ”Radiation Safety for Laser Devices”.
Please contact your supplier, representative or REIMERS & JANSSEN GmbH in case
you have any related questions.
REIMERS & JANSSEN GmbH
Medical Laser Technology
Frohnacker 8 79297 Winden, Germany
Tel. +49-7682-6558 Fax +49-7682-6640v
[email protected] www.rj-laser.com
Certified according to EN ISO 13485:2003
Exclusion of liability
The Physiolaser olympic should only be used for the purpose described in these
operating instructions.
The manufacturer will not be liable for any damage or injuries resulting from usage not
contained in these operating instructions.
The manufacturer reserves the right to make modifications in keeping with technological
advances.
Physiolaser principle of function
The Physiolaser olympic will emit monochromatic, coherent laser light to the cells and
tissue. The photon energy will be absorbed by the cells (e.g. mitochondrials), which
leads e.g. to a higher oxygen level and further to an improvement of the metabolism.
Additionally special impulse frequencies are applied (acc. to Dr. NOGIER, France,
BAHR, Germany, REININGER, Austria) in order to stimulate the body, cell and tissue.
The special features of the Physiolaser olympic are proven through several years of
intensive research and offer an exact supply of photon energy to the area being treated.
Further information on the wide range of uses for laser therapy can be found in technical
literature and with scheduled seminars RJ-LASER GmbH, www.rj-laser.com.
Physiolaser olympic 500
Version 1.4 June 2007
4
Safety standard
The Physiolaser olympic is a high-quality, microprocessor-controlled, multi-function
device. It combines a high-output, universal therapy laser with a diagnostic device for
measuring skin conductance values.
The Physiolaser olympic software offers users a wide range of application
possibilities. The special features of the Physiolaser olympic are proven through
several years of intensive research and offer an exact supply of photon energy to the
body.
The Physiolaser olympic meets the requirements of the following International
standards:
EN 60601-1: 1996 Safety of Medical Electrical Equipment
EN 60601-2-22: 1996 Safety of Diagnostic / Therapeutic Lasers
EN 60825-1: 2003 Safety of Laser Devices
DIN EN 60601-1-2 Electro magnetic compatibility
DIN EN 60601-1-4 Programable Electronic Device
Further information on the wide range of uses for laser therapy can be found in technical
literature and in seminars of REIMERS & JANSSEN GmbH.
Physiolaser olympic 500
Version 1.4 June 2007
5
Safety Precautions
Caution - Using the operating equipment or settings in a way not described here
can result in dangerous exposure to radiation.
Shutdown in an emergency
The laser can be shut down in the event of an emergency by pressing the Emergency
button or by pulling out the safety plug.
Personnel
The Physiolaser olympic should only be operated by trained medical personnel. The
personnel must be instructed in safe handling procedures and also made aware of the
potential risks of laser radiation.
Radiation injuries
Do not look directly into the laser beam outlet. Laser radiation can cause eye
damage. Protective gear must be worn inside the treatment room during treatment at all
times, caution during therapy in the head area.
Registering (valid for Germany, please refer to your national regulations)
The Physiolaser olympic operator must be registered at the professional association
as an operator of a class 3 B laser device.
Laser safety representative: The operator must be able to name a laser safety
representative.
Treatment room
The laser may only be operated in closed rooms. The treatment room in which laser
therapy is performed must meet the requirements of the Accident Prevention Regulation
BGV 11 ”Laser radiation”. (Attach a warning sign to all inputs in accordance with DIN
IEC 76 (CO) 6. Reflecting objects, mirrors and chrome parts must be removed.)
Safeguarding against unauthorized usage
To safeguard against unauthorized usage, the Physiolaser olympic is equipped with a
removable key, safety plug and safety code. The laser can only be operated when the
key and safety plug is inserted or the code is activated. If the Physiolaser olympic
laser is not being used, always remove the key and store it separately from the device
to prevent unauthorized usage.
Device out of use
If you assume that the device can not be used properly any more and could endanger
user or patient, it should be set out of use and has to be transferred to the manufacturer
or the authorized service station for repair. Such case could be e.g.:
Visible damage of the housing of charger or main unit
The device does not work properly
The device was stored or transported under harmful conditions
Physiolaser olympic 500
Version 1.4 June 2007
6
Description of the Device / Proper Usage
Technical data
Laser therapy device for mobile and local use Physiolaser olympic 500
Laser class 3B
Protection class IP 21 (drop proof)
Absolute failure of the measurement of the laser
power
5%
Modulation frequency 0,1 Hz - 18 688 Hz (± 0,1%)
Precision of the electrical skin resistance
measurement
± 10 %
Risk class according to RL 93/42/EWG IIa
Current supply Charger Model FRIWO FW7555M/08
Weight 2 kg
Dimensions (L x B x H) 25 x 10 x 15 cm
Operating conditions
Room temperature + 10°C to +40°C
Relative humidity 30% to 75%
Atmospheric pressure 700 hPa to 1060 hPa
Transport and storage
Room temperature -20° C to +40 C
Relative humidity 30% to 75%
Atmospheric pressure 700 hPa to 1060 hPa
Content of delivery
incl. Current supply charger, safety goggles, manual, therapy guide, carrying case,
”Laser Radiation” warning sign (DIN IEC 76 (CO) 6).
For the usage an optional laser probe is required :
- Laser probe
- Multi-cluster probe
- Diagnostic probe incl. Hand electrode
Physiolaser olympic 500
Version 1.4 June 2007
7
Proper usage and field of application
The Physiolaser is a therapeutic medical device which is to be used exclusively in
medical offices, clinics and rehab centers, and should only be operated under constant
supervision.
The Physiolaser is designed for a veterinary therapy, treatment of wounds, joints
and pain. In order to perform the therapy for pain and healing, the probe tip has to be
placed with light contact to the skin.
In case of infected skin, eczema, open wounds, keep 1-2 cm distance or cover the area
with plastic wrap and keep the probe tip in light contact to the area
Use the cluster probe for the treatment of large areas and joints and the single probe on
small areas, joints and acupuncture points.
Therapy recommendation: Start with a setting of 4-6 Joule/cm2 and proceed on the
following treatment day at the same energy if improved or increase until healing is
performed.
You can modify the treatment parameters according to scientific research (e.g.
recommended average dosage of 4 Joule/cm2) or according to the reactions at any time.
Please refer to the training manual to obtain further information regarding the
applications.
Indications
Acupuncture: Stimulation of acupuncture points
Tissue: Ulceration, wound healing
Bones/joints: Traumata, carpal tunnel syndrome, joint pain, fracture healing
Nerves: Nerve pain, neuritis
Pain: Reduction of pain
Contra indications
-ophthalmologic therapy
-over cancerous or malignant areas
When the Physiolaser is used for symptomatic treatment, and is contraindicated
where analgesia may mask progressive pathology, and should not be used where the
clinician would normally avoid the use of any other analgesia in order to retain the
beneficial aspects of pain.
It is conceivable that laser-mediated vasodilatation may worsen a haemorrhage.
Physiolaser olympic 500
Version 1.4 June 2007
8
Precautions
Patients may suffer an exacerbation of symptoms after laser therapy in conjunction with
a recent steroid injection. For this reason laser should not be used within 2-3 weeks of a
recent steroid injection on or near the same site.
Annual inspection of the device
Due to the advanced technology and precision of the Physiolaser must under go a
safety inspection and calibration at least one year from the purchase date (various test
are performed in order to insure that the device is operating within the tolerances that
the device was originally designed). This annual inspection/calibration must be
submitted to an authorized facility. Repairs or maintenance performed by an
unauthorized facility or personnel will void your warranty.
Inclusive with your device will be shipping labels and instructions for deliver to the repair
and maintenance facility. For repair and maintenance, devices must be sent directly to:
REIMERS & JANSSEN GmbH
Medical Laser Technology
Frohnacker 8 79297 Winden, Germany
Tel. +49-7682-6558 Fax +49-7682-6640
[email protected] www.rj-laser.com
Certified according to EN ISO 13485:2003
Physiolaser olympic 500
Version 1.4 June 2007
9
Explanation of the pictograms
IP21
Equipment rating against harmful ingress of water as
defined in IEC529. (see std. 60601-1 clause 6.1(L) )
Protection against electric shock regarding allowable
leakage current limitations are specified in standard
60601-1.
0123
The device meets the requirements of 93/42/EEC for
medical products.
“Attention” Observe the instructions in the manual for
safe and effective usage of this device.
Symbol for the laser beam: ”Warning for laser beam”.
Explanatory Label Symbol (International):
" Visible and invisible laser radiation" including the
standard which was used for the classification of the
laser.
Singel probe
Multi cluster probe
Labels according to EN 60825-1:
- Laser aperture and laser beam direction
Type plate with technical data and address of the
manufacturer
Physiolaser olympic 500
Version 1.4 June 2007
10
Location of the labels
Front panel: Symbol for the laser beam
”Warning for laser beam”.
Left: Explanatory Label Symbol (International):
"Laser radiation" including the standard which was
used for the classification of the laser.
Right: Label indicating technical data, serial number
and address of the manufacturer.
Aperture labels according to
EN 60825-1:
Front (side on the cluster probe): Laser
aperture/laser beam direction and explanatory label
Rear: Label indicating technical data as probe type,
serial number, power, pulse.
Physiolaser olympic 500
Version 1.4 June 2007
11
Laser probes
Single laser probes
Nr.: Wave
length Output power Laser type
Impulse duration NOHD *) m
a <= 1,5 Beam
divergence
510 785 nm max. 50 mW Continuous (CW) 0,10 0,140 rad
511 810 nm max. 500 mW Continuous (CW) 0,29 0,157 rad
512 670 nm max. 200 mW Continuous (CW) 0,14 0,209 rad
513 655 nm Max. 50 mW Continuous (CW) 0,09 0,157 rad
515 904 nm max.90 W Impulse 200 ns 0,05 0,190 rad
Multi-cluster probes
Nr.: Wave
length Output power Laser type
Impulse duration NOHD *) m
a <= 1,5 Beam
divergence
516 904 nm 5 x max. 30 W Impulse 200 ns 0,06 0,190 rad
517 785 nm
655 nm
655 nm
4 x max. 55 mW
4 x max. 40 mW
4 x max. 5 mW
Continuous (CW)
Continuous (CW)
Continuous (CW)
0,10
0,09
0,04
0,140 rad
0,157 rad
0,155 rad
518 904 nm 12 x max. 10 W Impulse 200 ns 0,05 0,190 rad
*) Safety distance (NOHD). Distance at which the radiation intensity or the radiation is equivalent to the limiting value
of the maximum permitted radiation of the cornea (MZB).
For more information, please refer to the device plate.
The controller of the Physiolaser olympic will recognize automatically which probe is
connected and will adjust all relevant settings. Please observe that when the output is
set, the laser diodes’ maximum laser output may differ from one another. The probes
can be identified through their name plates, the first two digits in each number specify
the maximum laser output in mW (CW) or W (imp). You will find the recognized probe
shown in the display as:
Physiolaser olympic 500
Version 1.4 June 2007
12
Sockets and connections on the front cover and probe
Buttons on the single laser probe
Laser Start/Stop
(Press: Short= full power, long= reduced power)
Selection of single frequencies
Selection of frequency main groups. Nogier/BAHR,
REININGER, free frequencies, therapy programmes
(Press: Short= rows, long= main groups)
Buttons on the cluster laser probe
Laser Start/Stop
(Press: Short= full power, long= reduced power)
Display
Slot for the memor
y
card
Lase
rTest
Sockets for the laser
p
robe
Socket for the
diagnostic probe,
hand electrode
Remote laser
stop
Ke
y
switch
Emergency stop
Dis
p
la
y
Current connected
Scroll
up/down
Frequency
group
Laser start/stop
Physiolaser olympic 500
Version 1.4 June 2007
13
Selection of frequencies
Cluster probes
Universalsonde
4x5 mW/655 nm + 4x40 mW/655 nm + 4x55
mW/785 nm
12x10 W/904 nm
Brush mit Fiberoptiken
4x5 mW/655 nm + 4x40 mW/655 nm + 4x55
mW/785 nm
12x10 W/904 nm
Derma
4x5 mW/635 nm + 6x45 mW/655 nm + 2x55
mW/785 nm
Kompakt
5x 30 W/904 nm, Impuls 200 nsec.
Physiolaser olympic 500
Version 1.4 June 2007
14
LED indicator, display
Main unit Display and LED
Color Status Meaning
Buttons blue off Buttons are ready for use
Buttons yellow on Button is active
Stand-by
Ready
yellow on Show Stand-by or Ready mode
Laser data blue off Connected probe is recognized but not active
Laser data yellow on Connected probe is recognized and active
LED red on
Main unit is connected with the power supply to
the mains current
30% red flash accumulator will soon be empty, please charge
Probe
LED red Flashes
on
2 seconds during start procedure
Laser is active
Pilot-LED red on Illuminates the area to be radiated (distance: 30
mm) 2 seconds before emitting the beam
Display white on Shows data: Frequency, power, time, Joule
Physiolaser olympic 500
Version 1.4 June 2007
15
Display after starting the device with the Key switch
Touching the buttons on the display will activate the functions.
Ready (therapy mode), frequency
selection, laser setting, diagnosis
Pre settings, manufacturer´s data, version no.,
Language setting.
Indicates the
Stand by mode
Frequency group
Single frequency
Battery indicator
Stand by button
Frequency row
Frequency mix
Laser settings Single and multi
cluster
p
robe
Physiolaser olympic 500
Version 1.4 June 2007
16
Button functions on the display (yellow color shows active function)
Display after pressing the Ready button. Active
buttons are yellow, inactive blue.
Upper buttons are corresponding with the frequency
groups , thereafter frequency rows and single
frequencies.
The green line above Stand-by indicates the condition
of the battery.
Lower buttons are corresponding with Stand by mode,
setting of Power/Energy/Time and probe selection.
Two probes connected, data valid for the single probe
(yellow).
Frequency selection: NOGIER
Frequency B´ for the single probe.
Frequency mix
Frequency selection NOGIER B´+ C´
Here frequency B´+ C´ for the single probe
In order to mix follow the steps:
- Select the first frequency and press the mix button
once (>B´) and again to go to safe the first frequency
(B´ >Freq. 2),
- thereafter select the second frequency (>2).
Physiolaser olympic 500
Version 1.4 June 2007
17
Frequency selection: BAHR1
Frequency 5´ for the multi-cluster probe.
Frequency selection: BAHR2
Hysteria frequency for the multi-cluster probe
Frequency selection: Chakra 1-8 acc. to BAHR
Sexual Chakra
Physiolaser olympic 500
Version 1.4 June 2007
18
Frequency selection: Chakra 9-15 acc. to BAHR
Protection Chakra
Frequency selection: Meridian 1 acc. REININGER
Small Intestine.
Frequency selection: Meridian 2 acc. REININGER
Liver.
Frequency selection: Reininger
Reininger 1
RI 2
Reininger I to II have a similar display design.
Physiolaser olympic 500
Version 1.4 June 2007
19
Frequency selection: Reininger
Level 2
K-HWP
Frequency selection: Reininger
Anti-Freq. 2
Anti inflammation
Frequency selection: Reininger
Viral/Bacterial
Anti viral
Physiolaser olympic 500
Version 1.4 June 2007
20
Frequency selection: Therapy program
General 1
Nerves
Available rows:
General 1 and 2
Orthopaedics
Derma
Dental
Vet
The mix button is not available for the therapy
programs.
Frequency selection: Free Frequency
Free Freq. 6
F 6
Available are 10 memories each with 11 spaces.
Individual programming of frequency and name is
possible, please press the button Edit.
Selection of power, energy, time, duty cycle and
therapy beep.
Press one of the buttons to change the value.
Enter will store the value.
Therapy leads back to the main therapy menu.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41