these spacing recommendations could result in oversensing. Clinical risks for not following
these spacing recommendations include, but are not limited to, loss of pacing output,
asynchronous ventricular pacing, and pacemaker mediated tachycardia.
Unipolar lead systems – Unipolar lead systems are not recommended because they are
more susceptible to electromagnetic interference, which may result in inappropriate pacing.
Unipolar lead systems should not be used in the dual-chambered pacing modes because
the current path of one lead system may interfere with the current path of the other.
Dual chamber pacing modes – Do not configure the temporary pacemaker in any of the
dual chamber pacing modes unless both channels are connected to the heart. If the
temporary pacemaker is configured in a dual chamber pacing mode and one of the channels
is not connected, the open channel could pick up unintended noise. The noise can be
interpreted as a sensed event and can lead to events such as, but not limited to,
asynchronous ventricular pacing, unintended pacing inhibition, or ventricular tachycardia.
Avoid use of dual chamber pacing modes without an atrial connection as this could, in rare
circumstances, result in inappropriate delivery of a ventricular pace on an intrinsic T-wave.
Atrial sensing – When programming to a mode that requires atrial sensing, the sensing
threshold should be evaluated for sufficient safety margin.
Place the temporary pacing lead on the right atrial free wall, oriented along the direction of
the myocardial fibers, approximately 1 cm apart. It is important to achieve a sensing
threshold of at least 1.0 mV. Set atrial sensitivity to a minimum of one-half the measured
threshold. The setting ensures a minimum safety margin of two times the sensing threshold.
Failure to follow this procedure can lead to delivery of asynchronous pulses.
Sensing thresholds – Do not use the temporary pacemaker to determine sensing
thresholds for permanently implanted lead systems. When implanting a permanent
pacemaker, Medtronic recommends the use of a pacing system analyzer (PSA).
Sensitivity settings – Since the sensitivity setting determines the smallest signal that can
be sensed by the pacemaker, set the sensitivity dial to one-half the mV value of the patient’s
sensitivity threshold. This setting will provide a 2x safety margin to ensure proper sensing.
A more sensitive setting may be chosen to provide a greater safety margin. However, be
aware that setting the sensitivity value too low (too sensitive) could result in inappropriate
sensing of far field signals (for example, sensing of R-waves or T-waves on the atrial channel
or P-waves on the ventricular channel), leading to inappropriate inhibition of pacing pulses.
Sensitivity threshold testing – Complete the sensitivity threshold testing to determine the
appropriate settings for sensitivity. Clinical risks for failure to perform this step include, but
are not limited to, asynchronous ventricular pacing and pacemaker mediated tachycardia.
Output threshold testing – Complete the output threshold testing to determine the
appropriate settings for output. Clinical risks for failure to perform this step include, but are
Medtronic 5392
16 Technical Manual