CARDELL® MAX-12HD User’s Manual
M1.0 5-15-09
9
current is within safety limits.
• When a defibrillator is used, make sure patient will not touch the ground,
metal or other conductor or device. During defibrillation, never touch the
patient, table or the device.
• When electrosurgery unit is used, make sure the patient lead and cable are
far away from the operating table to reduce the risk of burns caused by
poor connection.
• Before using on another patient, make sure previous monitoring data is
cleared.
• Before using the monitor, make sure it is in normal working condition.
• Use properly grounded power sockets and ensure adequate grounding. If
there is any doubt about the grounding, please use battery operation.
• Check accessories on regular basis and discard damaged accessories
properly.
• To ensure patient’s safety and performance of the product, use only the
manufacturer recommended accessories.
• Service parts must be in conformity with IEC 60601 standards. The
system configuration of monitor must be in conformity with IEC
60601-1-1 medical electric standard, otherwise, it will reduce the safety
of monitor.
• Even while not being used, the battery may also discharge. So check
battery level every month.
• ECG cable socket is for connecting ECG lead wires only. Please do not
connect it to any other signal source. Pay attention to the color label and
marks of ECG lead wires.
• Please clean the monitor and accessories according to instructions.
Always unplug the power cord before cleaning.
• EMC — The device is in conformity with the requirements of IEC
60601-1-2 and related EMC standards. But when electromagnetic power
is extremely high, it may cause interference. Please ensure any device
close to the monitor meets the related EMC requirements. Do not use cell
CAUTION