  • Hello! I'm a chat assistant, and I've reviewed the user manual for the MEAN WELL MPM-15 series medical power supplies. This document details the specifications, features, and safety standards for these compact, high-reliability power supplies designed for medical applications. I'm ready to answer your questions regarding their operation, usage, and technical details.
  • What is the input voltage range for this power supply?
    What are the safety standards this power supply complies with?
    Is this power supply suitable for BF applications?
    What kind of protections does this power supply have?
File Name:MPM-15-SPEC 2022-12-19
Model Encoding
Description
Features
MPM-15 series
Output voltage
Rated wattage
Series name
Portable medical device
Mobile clinical workstation
Medical computer monitor
Medical examination instrument
Applications
-5
MPM - 15
15W High Reliable Green Medical Encapsulated Type
MPM-15 is a 15W high density and small size (52 4 27 2 24mm) AC/DC module type medical power supply .* .*
series offered in pin type. It features the operation for 80~264VAC, a low no load power consumption less
than 0.1W, a high efficiency up to 87%, Class (no FG) double insulation, outstanding dissipation and high
lifespan thanks to the interior potting, 2G anti-vibration, high EMC performance, 4KVAC isolation, etc. The
design observes IEC BS EN/EN60601 1 and ANSI/AAMI ES60601 1version three with 2xMOPP level and ultra-low /- -
leakage current (<80 A). It is very suitable for BF (patient contact) type medical device or relevant equipment. μ
++
2.06" .07" x1 compact size
Medical safety approved (2 x MOPP) according to
ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
Suitable for BF application with appropriate system consideration
No load power consumption<0.1W
Extremely low leakage current
Wide operating temp. range -40 ~ +85
EMI class B for class configuration
Protections:
Short circuit / Overload / Over voltage / Over temperature
No minimum load required
3 years warranty
BS EN/EN60335-1
EN 60601-1
IEC 60601-1
8
IEC60601-1BS EN/EN60601-1ANSI/AAMI ES60601-1
TPTC004
User’s Manual
https://www.meanwell.com/serviceGTIN.aspx
MW Search:
GTIN CODE
Downloaded from Arrow.com.
File Name:MPM-15-SPEC 2022-12-19
OPERATING ALTITUDE Note.8 5000 meters
SPECIFICATIONSPECIFICATION
MODEL
DC VOLTAGE
RATED CURRENT
CURRENT RANGE Note.2
RATED POWER
OUTPUT
LINE REGULATION
LOAD REGULATION
SETUP, RISE TIME
HOLD UP TIME (Typ.)
VOLTAGE RANGE Note.6
FREQUENCY RANGE
EFFICIENCY (Typ.)
INPUT
INRUSH CURRENT (Typ.)
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
WORKING HUMIDITY
WORKING TEMP.
OVER TEMPERATURE
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
MTBF
DIMENSIONOTHERS
NOTE
PACKING
OVERLOAD
OVER VOLTAGE
AC CURRENT (Typ.)
20 ~ 90% RH non-condensing
-40 ~ +85 (Refer to "Derating Curve")
-40 ~ +85 , 10 ~ 95% RH non-condensing
±℃ ℃0.03%/ (0 ~ 60 )
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
SAFETY STANDARDS
RIPPLE & NOISE (max.) Note.4
VOLTAGE TOLERANCE Note.5
ENVIRONMENT
SAFETY &
EMC
(Note.9)
PROTECTION
Protection type : Shut off o/p voltage, clamping by zener diode
Protection type : Shut down o/p voltage, recovers automatically after temperature goes down
I/P-O/P:4KVAC
I/P-O/P:100M Ohms / 500VDC / 25 / 70% RH
LEAKAGE CURRENT (max.) Note.7
ISOLATION LEVEL Primary-Secondary: 2xMOPP
Parameter
Parameter
ESD
Standard
Standard
BS EN/EN55011 (CISPR11)
BS EN/EN55011 (CISPR11)
BS EN/EN61000-3-2
BS EN/EN61000-3-3
BS EN/EN61000-4-2
BS EN/EN61000-4-3
BS EN/EN61000-4-4
BS EN/EN61000-4-5
BS EN/EN61000-4-6
BS EN/EN61000-4-8
BS EN/EN61000-4-11
Test Level / Note
Test Level / Note
Class B
Class B
Class A
-----
EMC IMMUNITY
EMC EMISSION
100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods
SOLDERING TEMPERATURE
1500ms, 30ms/230VAC 1500ms, 30ms/115VAC at full load
40ms/230VAC 10ms/115VAC at full load
8 ~ 264VAC0 113V~370VDC
47 ~ 440Hz
0.6A/115VAC 0.3A/230VAC
COLD START 20A/115VAC 45A/230VAC
110% ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
MPM-15-3.3
Touch current <80 A/264VAC
μ
MPM-15-5 MPM-15-12 MPM-15-15 MPM-15-24
±0.3%
±0.5%
±0.3%
±0.5%
±0.3%
±0.5%
±0.5%
±0.5%
±0.5%
±2.0%
5V3.3V 12V 15V 24V
3A3.5A 1.25A 1A 0.63A
0 ~ 3A0 ~ 3.5A 0 ~ 1.25A 0 ~ 1A 0 ~ 0.63A
15W11.6W 15W 15W 15.1W
13.8 ~ 16.2V 17.3 ~ 20.3V 27.6 ~ 32.4V
5.8~ 6.8V3.8 ~ 5V
3.85A 3.3A 1.38A 1.1A 0.69A
MPM-15 series
±2.0% ±2.0% ±2.0% ±2.0%
±0.5%
0.056Kg; 240pcs/14.4Kg/0.94CUFT
PEAK CURRENT
15W High Reliable Green Medical Encapsulated Type
PEAK LOAD(10sec.)
Note.3
12.7W 16.5W 16.6W 16.5W 16.6W
IEC60601-1, TUV BS EN/EN60601-1, IEC60335-1, Dekra BS EN/EN60335-1 , ULANSI/AAMI ES60601-1, EAC TP TC 004
(3.1 version), CAN/CSA-C22 3 Edition approved
rd
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. No minimum load required.
3. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.
4. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1 f & 47 f parallel capacitor.μμ
5. Tolerance : includes set up tolerance, line regulation and load regulation.
6. Derating may be needed under low input voltages. Please check the derating curve for more details.
7. Touch current was measured from primary input to DC output.
8. The ambient temperature derating of 3.5 /1000m with fanless models and of 5 /1000m with fan models for operating altitude higher than 2000m(6500ft).℃℃
9. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still
meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies.
(as available on http://www.meanwell.com)
Product Liability Disclaimer For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx
Level 4, 15KV air ; Level 4, 8KV contact
Level 3, 1KV/Line-Line
52.4*27.2*24mm (L*W*H) 2.06 *1.07 *0.94 inch
or ″″″
BS EN/ BS EN/EN60601-1-2EN55035,
EFT bursts
Surge susceptibility
Conducted susceptibility
Magnetic field immunity
Conducted emission
Radiated emission
Harmonic current
Voltage flicker
Level 3, 10V/m( 80MHz~2.7GHz )
Level 4, 30A/m
Level 3, 10V
Level 3, 2KV
Voltage dip, interruption
150mVp-p150mVp-p 150mVp-p 180mVp-p 180mVp-p
85.5%83.5% 86.5% 87% 86.5%
Table 9, 9~28V/m( 385MHz~5.78GHz )
RF field susceptibility
7319.8K hrs min. Telcordia SR-332 (Bellcore) ; 1210.0K hrs min. MIL-HDBK-217F (25 )
Wave soldering: 265 ,5s (max.); Manual soldering: 390 ,3s (max.)℃℃
Downloaded from Arrow.com.Downloaded from Arrow.com.
File Name:MPM-15-SPEC 2022-12-19
Block Diagram
fosc : 100KHz
Output Derating VS Input Voltage
INPUT VOLTAGE (VAC) 60Hz
80 10095 120115 160140 200180 240220 264
LOAD (%)
90
100
80
70
60
50
40
Derating Curve
MPM-15 series
AMBIENT TEMPERATURE ( )
LOAD (%)
3.3V,5V 12V~24V
15W High Reliable Green Medical Encapsulated Type
FILTER
I/P EMI
FILTER
RECTIFIERS
&
RECTIFIERS
CONTROL
PWM DETECTION
CIRCUIT
&
FILTER
+V
-V
SWITCHING
POWER
O.L.P.
50
100
30
70
(HORIZONTAL)
-30 -20 -10 0 10 20 30 40 50 60 70 80 85-40 55
80
Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.
File Name:MPM-15-SPEC 2022-12-19
MPM-15 series
Mechanical Specification Case No.219A Unit:(mm)
P/N diameter:1.04
24
1.04
3.5 1mm±
45
52.4
BOTTOM VIEW
AC/N -V
+V
AC/L
3.4
3.2
3.2 8
20.8
27.2
SIDE VIEW