Cannon miniAV-X 1 Bath 115V 50/60 HZ Installation guide

Type
Installation guide

This manual is also suitable for

THANK YOU for your purchase of the miniAV automatic viscometer from CANNON Instrument
Company. This guide is designed to provide information about the installation of the miniAV-HT.
All page numbers referenced in these instructions are from the miniAV Instruction & Operation
Manual.
1. Unpack the miniAV-HT viscometer
Remove all components from the shipping container(s).
CAUTION
Do NOT lift the miniAV instrument by the modular side panels (“ears”). Always
support the weight of the unit from the BASE.
Remove all packing materials from the components.
Verify reception of shipped materials by comparing equipment items
with packing/parts list(s). Report missing items to Cannon Instrument
Company immediately.
Inspect each component for signs of damage. Report damages to the
shipper and Cannon Instrument Company immediately.
Retain all packing materials until the instrument is connected and
functioning properly.
If any component(s) must be returned to Cannon Instrument Company,
the damaged item(s) should be packaged in the original shipping
container. Refer to the final page of this guide for instructions on
returning defective equipment. Customers outside the United States
should contact the local Cannon dealer for procedures on returning
products to Cannon.
Review operator safety information on pages 2-4 of the miniAV
I
nstruction & Operation Manual
Verify electrical specifications for Mains on page 4 of the miniAV
I
nstruction & Operation Manual.
- 4 -
CANNON® miniAV-HT Installation
Guide 2139 High Tech Road
State College, PA, 16803 | USA
800-676-6232 | 814-353-8000
- 1 -
miniAV-HT viscometer Installation
Guide
2139 High Tech Drive
State College, PA 16803 | USA
814-353-8000 | 800-676-6232
service@cannoninstrument.com
cannoninstrument.com
T
B
B
A
B
Power Supply
Waste Receiver
A
Tubing Detail
A = 6 mm transparent tubing
B = 1/8” transparent tubing
2.
Connect tubing and cables as indicated in the image. Cut tubing
to length as necessary.
3. Fill bath vessel to the fill line with silicone fluid.
4. Install VISCPRO Software on the controlling PC using
the installation CD or thumb drive.
miniAV-HT
Waste Vessel
Power Supply
Solvent
Vessel
5.
Supply power to the miniAV instrume
nt.
6. Load VISCPRO software.
7. Set the desired control
temperature.
8. Calibrate the instrument per
manual instructions or confirm
calibration data included with your
installation CD.
9. Enter sample ID information into
VISCPRO and begin running
samples.
- 2 -
miniAV-HT Installation Guide
|
Cannon Instrument Company®
|
2139 High Tech Rd
|
State College, PA
|
16803
|
USA
|
814-353-8007
|
fax 814-353-8007
|
cannoninstrument.com
- 3 -
Above:
Filling the miniAV-HT
bath vessel with bath fluid using
the provided glass funnel.
  • Page 1 1
  • Page 2 2

Cannon miniAV-X 1 Bath 115V 50/60 HZ Installation guide

Type
Installation guide
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI