  • Hello! I am an AI chatbot trained to assist you with the Invacare IVC Bariatric Bed User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
IVC™BariatricBed
enModelBAR600IVC
UserManual
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuturereference.
©2015Invacare®Corporation
Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithout
priorwrittenpermissionfromInvacare.Trademarksareidentifiedby™and®.Alltrademarksare
ownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted.
Contents
1General........................................4
1.1Symbols.....................................4
2Overview.......................................5
2.1LabelLocations................................5
2.2IVCBariatricBedSpecifications....................7
2.3BedComponents...............................7
3Safety..........................................8
3.1GeneralGuidelines.............................8
3.2MattressCompressionTest.......................15
3.3Disposal.....................................16
4Set-up.........................................17
4.1Receiving.....................................17
4.2Unpacking....................................17
4.3Inspecting....................................19
4.4Assembling/DisassemblingHeadandFootSections.......19
4.5InstallingCasters...............................21
4.6Installing/RemovingBedEnds......................22
4.7ConnectingCrossoverCable......................23
4.8Assembling/Installing/RemovingDriveShaft............25
4.9InspectingtheBed..............................27
4.10StoringtheBed...............................28
5Operation......................................29
5.1OperatingBariatricBed..........................29
5.2EmergencyCrank..............................30
5.3OptionalBedRails..............................31
6Maintenance....................................33
6.1ServiceLife...................................33
6.2WearandTearItems............................33
6.3WashingBedEnds..............................34
6.4CleaningHeadandFootSections...................35
6.5MaintenanceandCleaningChecklist.................36
7Troubleshooting.................................37
7.1Troubleshooting:Electrical........................37
7.2Troubleshooting:Mechanical......................39
8Warranty.......................................40
8.1LimitedWarranty..............................40
IVC™BariatricBed
1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazardsorunsafe
practiceswhichcouldresultinpersonalinjuryorpropertydamage.
Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious
injury.
WARNING!
Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious
injury.
CAUTION!
Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage
orminorinjuryorboth.
Givesusefultips,recommendationsandinformationfor
efficient,trouble-freeuse.
4
1123842-I-05
Overview
2Overview
2.1LabelLocations
1123842-I-055
IVC™BariatricBed
61123842-I-05
Overview
2.2IVCBariatricBedSpecifications
BEDENDHEIGHT
MINIMUM:
26inches
MAXIMUM:34inches
DECKHEIGHT
MINIMUM:
(DOWNPOSITION):
14inches
MAXIMUM(UPPOSITION):22inches
BED
OVERALLLENGTH:88inches
WIDTH:42inches
SPLITSPRINGSTOENDS
HEAD:43inches
FOOT:
45inches
SHIPPINGWEIGHT
275lbs
WEIGHTCAPACITY
OCCUPANT:600lbs
RATEDMAXIMUM:
(SAFEWORKINGLOAD)
700lbs(Includingaccessories,
mattress,occupantandany
otherobjectspositionedon
thebed)
Heightfromfloortobeddeck.
Measurementsmayvary±2inches.
Forhoushold/residentialorpublicoccupancyuseperUL
962.
BedisNOTforhospitaluseandwasnotdesignedtomeet
hospitalstandards.
Onlyusebedindoors.
WARNING!
RiskofInjuryorDamage
TheweightlimitoftheoccupantofthebedCANNOT
exceed600pounds.Toavoidinjuryand/ordamagefrom
bedcollapse:
TheTOTALweightlimitofthebed(including
accessories,mattress,occupantandanyotherobjects
positionedonthebed)cannotexceed700pounds
(SafeWorkingLoad).
WARNING!
RiskofInjuryorDamage
Patientsthatweighlessthan351poundshavean
increasedriskofentrapment.Toavoidinjuryordamage
fromentrapment:
Patientsthatweighlessthan351poundsshouldnot
usetheIVCBariatricBed.
2.3BedComponents
Headspringwithonemotor(BAR5000IVC)
Footspringwithtwomotors(BAR5490IVC)
BedEnds(BAR5301IVC)
1123842-I-057
IVC™BariatricBed
3Safety
3.1GeneralGuidelines
DANGER!
RiskOfDeath,Injury,OrDamage
Improperuseofthisproductmaycauseinjuryor
damage.
Ifyouareunabletounderstandthewarnings,cautions
orinstructions,contactahealthcareprofessional,
dealerortechnicalpersonnelbeforeattemptingto
usethisequipment.
Donotusethisproductoranyavailableoptional
equipmentwithoutfirstcompletelyreadingand
understandingtheseinstructionsandanyadditional
instructionalmaterialsuchasusermanuals,service
manualsorinstructionsheetssuppliedwiththis
productoroptionalequipment.
Continueduseoftheproductwithdamagedpartscould
leadtotheproductmalfunctioning,causinginjurytothe
userand/orcaregiver.
CheckALLproductcomponentsandcartonfor
damage,andtestcomponentsbeforeuse.DONOT
useproductifcomponentsaredamagedorifproduct
isnotworkingproperly.Contactaqualifiedtechnician
orInvacareforrepair.
Toavoidinjuryorproductdamage:
Set-upandAssemblyinstructionsMUSTbeperformed
byqualifiedtechniciansonly.“QualifiedTechnician
Only”sectionsaremarkedaccordingly.
DANGER!
RiskOfDeath,Injury,OrDamage
Lightedcigarettescancauseafireorotherrelated
hazardsresultingindeath,injury,ordamage.
Donotsmokewhileusingthisproduct.
Donotallowotherstosmokenearthisproduct.
DANGER!
RiskOfDeath,InjuryOrDamage
Bedaccessoriesdesignedbyothermanufacturershave
NOTbeentestedbyInvacare.UseofNON-Invacare
bedaccessoriesmayresultininjuryordeath.
UseONLYInvacarerails,mattresses,bedextenders
andotheraccessorieswithInvacarebedproducts.
THEINFORMATIONCONTAINEDINTHISDOCUMENT
ISSUBJECTTOCHANGEWITHOUTNOTICE.
DANGER!
RiskOfInjuryOrDeath
Toreducetheriskofelectricalshock:
Alwaysunplugtheproductfromtheelectricaloutlet
beforecleaning.
81123842-I-05
Safety
DANGER!
RiskOfDeath,Injury,OrDamage
Toreducetheriskofshock:
Unplugfromtheoutletbeforeservicing.
Closesupervisionisnecessarywhentheproductis
usedby,ornearchildrenand/ordisabledpersons.
Usethisproductonlyforitsintendeduseasdescribed
intheseinstructions.DONOTuseattachmentsnot
recommendedbythemanufacturer.
Ifliquidisspilledinoraroundtheproduct,unplug
beforecleaning.Cleanupthespillandallowthe
productortheareaaroundtheproducttodry
thoroughlybeforeusingthecontrolsagain.
DANGER!
RiskOfDeath,Injury,OrDamage
Toavoidshock,productdamageandpersonalinjury:
DONOTuseoutdoors.
Connectthisproducttoaproperlygroundedoutlet
only.Referto5.1Operation,page29
.
DONOTplacevideoequipmentsuchastelevisions
orcomputermonitorsonproduct.
OperatingInformation
DANGER!
RiskOfDeath,Injury,OrDamage
Toavoidinjuryordamagewhenoperatingproduct:
Afteranyadjustments,repairorserviceandbefore
use,makesureallattachinghardwareistightened
securely.
Closesupervisionisnecessarywhentheproductis
usednearchildrenorpetsORwhenthisproduct
isoperatedbyornearpeoplewithphysical/mental
disabilities.Physicallychallengedindividualswho
cannotpreventthemselvesfromrolling/climbingout
oftheproductmayrequirealternativesafemeansof
restraint.
Theproductisnotdesignedtobeusedasapatient
transport.Whentransportingapatient,usean
approvedpatienttransport.Otherwise,injuryor
damagemayresult.
1123842-I-059
IVC™BariatricBed
WARNING!
RiskOfInjuryOrDamage
Toavoidinjuryordamagewhenoperating/movingbed:
ALWAYSensurethattheindividualutilizingthebed
ispositionedproperlywithintheconfinesofthebed.
DONOTletanyextremitiesprotrudeovertheside
orbetweenthebedrailswhenperformingthese
functions.
Beawareofpotentialpinchpointswhenoperatingbed
railsorotheraccessories.
Makesureheadandfootsprings/sectionsare
connectedsecurelytothebedendsbeforeuse.
Thebedisequippedwithlockingcasters.Before
usingthebed,testthewheellockstoensurecasters
lockcorrectly.Whentransferringintooroutofthe
Bariatricbed,ALWAYSlockthecasters.Evenwith
castersproperlylocked,someflooringsurfaces(such
astileorwood,)willallowthebedtomoveunder
someconditions.Usingthebedonthesetypesof
surfacesMUSTbeevaluatedbythecareprovider.
DONOTtransferoccupantifbedisnotstable.
Otherwise,injuryand/ordamagemayoccur.
DONOTmoveproductwithlockingcastersin
LOCKEDposition.
DANGER!
RiskOfDeath,Injury,OrDamage
Toavoidpinching,crushandcollapsehazardsfrombed
deckcontactwithobjectsunderthebed:
DONOTplaceanyobjectsorletanyindividual
underneaththebedorinbetweentheraisedbed
framecomponentsatanytime.
Toavoidchokingfromsmalldetachablepartscontained
underneaththebed:
Closelysupervisechildren,pets,orpeoplewith
physical/mentaldisabilities.
RefertotheBedRailEntrapmentRiskNotification
Guideatwww.invacare.comforadditionalsafety
information.
DANGER!
RiskOfDeath,Injury,OrDamage
Patiententrapmentfromtheuseofbedsiderailsmay
causeinjuryordeath.Toavoidpatiententrapment:
TheInvacaremattressMUSTfitfirmlyagainstthe
bedframeANDbedsiderailstopreventpatient
entrapment.Followthemanufacturer’sinstructions.
Monitorpatientfrequently.Readandunderstandthe
UserManualpriortousingtheInvacarebedandbed
rails.Usermanualsareavailableatwww.invacare.com
oryourdealer.
Afteranyadjustments,repairorserviceandbefore
use,makesureallattachinghardwareistightened
securely.
Mattressplatform,bedends,pendant,andaccessories
areappliedparts.
101123842-I-05
Safety
DANGER!
RiskOfDeath,Injury,OrDamage
Conditionssuchasrestlessness,mentaldeterioration
anddementiaorseizuredisorders(uncontrolledbody
movement),sleepingproblems,andincontinencecan
significantlyimpactapatient'sriskofentrapment.
Pediatricpatientsorpatientswithsmallbodysizemay
alsohaveanincreasedriskofentrapment.
Monitorpatientswiththeseconditionsfrequently.
Placethemattressdeckintheflatpositionwhenleft
unattended.
WARNING!
RiskOfInjuryOrAdverseHealthConsequences
Fallhazardexistsduetobedheight.
Alwaysleavebedinlowestheightwhenunattended.
ReplacementParts/Accessories
DANGER!
RiskOfDeath,Injury,OrDamage
Toavoidfirehazardsandtubingentanglement:
Usemaskedornasaltypeoxygenadministering
equipmentonlyinconjunctionwiththemanual/electric
bed.Theuseofanyothertypeofoxygenadministering
equipmentcanresultinafirehazard.
Whenusingnasalormaskedtypeoxygenadministering
equipment,theoxygenorairtubingMUSTberouted
andsecuredproperlytoensurethatthetubingdoes
notbecomeentangledand/orseveredduringnormal
operationofmanual/electricbed.
DonotusethisbedinActiveTreatment“Ether
Environments.”
DANGER!
RiskOfDeath,Injury,OrDamage
Toavoidfirehazards,injuryordamage:
DONOTuseproductwithflammablegases.
Explosionmayoccur.
WARNING!
RiskOfInjuryOrDamage
Toavoidinjuryordamagefrommisuseorentrapment:
Readandunderstandtheinstructionsorlabelprovided
withmattress.
Ifyoudonotunderstandtheinstructionsorif
instructionsaremissing,contactthemattress
manufacturer.
DANGER!
RiskOfDeath,Injury,OrDamage
Toavoidentrapment,productdamage,and/orpersonal
injury:
Invacarerecommendsthatthemattressbecentered
onthebedframe.Otherwise,individualsmaybecome
trappedbetweenthebedrailandthebedframe.
ALWAYStesttomakesurethatthesiderails
areproperlyandsecurelyinplacebeforeusing.
Otherwise,injuryordamagemayoccur.
1123842-I-05
11
IVC™BariatricBed
DANGER!
RiskOfInjuryOrAdverseHealthConsequences
FallhazardexistsduetouseofNON-Invacare
mattresses.Non-Invacaremattressesarepotentially
incompatiblewithInvacarebeds.
Toavoidinjury,ensurethatonlyInvacaremattresses
areusedwithInvacarebedsatalltimes.
DANGER!
RiskOfDeathorInjury
Anincreasedriskofpatiententrapmentmayoccurover
timeduetomattresscompression.
Periodicallymonitorgapsbetweenthebed,mattress,
and/orbedrail.Wheregapsoccur,patiententrapment
ispossibleandthemattressshouldbereplaced.
DANGER!
RiskOfDeath,Injury,OrDamage
Toavoidentrapment,productdamage,and/orpersonal
injury:
Whenusedwithamanual/electricbed,thebedrails
DONOTfallunderanyweightlimitations.Bedrails
canbedeformedorbrokenifexcessivesidepressure
isexertedonthebedrails.Thesebedrailsareused
forthepurposeofpreventinganindividualfrom
inadvertentlyrollingoutofbed.Thebedrailsare
notintendednormaybeusedforrestraintpurposes.
Ifanindividualiscapableofinjuringhimself/herself,
aphysicianorahealthcareprofessionalshouldbe
consultedforalternativemeansofsaferestraint.
Afterraising/loweringthehead/footendoftheelectric
bed,checkthedistancebetweenthebottomofthe
bedrailandthemattress.Ifthereisexcessivedistance
betweenthebottomofthebedrailandthemattress
inwhichindividualsmaybecomeentangled,adjust
theheightofthebedrail(ifapplicable),orprovide
alternativemeansofpatientprotection.
DONOTusethesiderailsaspushhandlesformoving
theelectricbed.
Replacementmattressesandbedsiderailswith
dimensionsdifferentfromtheoriginalequipment
suppliedorspecifiedbythebedframemanufacturer
arenotinterchangeable.Variationsinbedsiderail
design,widthandthicknessorfirmnessofthemattress
couldcause/contributetoentrapment.
12
1123842-I-05
Safety
Electrical
WARNING!
RiskOfDeath,Injury,OrDamage
Shockhazardsandriskoffireexistduetouseof
improperextensioncordanduseofthreeprong
adapters.
Toavoidinjuryorproductdamage,whenusingan
extensioncord,useonlyaULapprovedthreeprong
extensioncordhavingatleast16AWG(American
WireGauge)wireandthesameorhigherelectrical
ratingastheproductbeingconnected.
DONOTusethreeprongtotwoprongadapters
WARNING!
RiskOfInjuryOrDamage
Toavoidproductdamageandpersonalinjury:
Disconnectpendantwhennotinuseandstoreout
ofreachofchildren.
RepairorServiceInformation
GroundingInformation
WARNING!
RiskOfShockInjury
Electricbedsareequippedwiththreeprong(grounding)
plugsforprotectionagainstpossibleshockhazards.
Whereatwoprongwallreceptacleisencountered,
itisthepersonalresponsibilityandobligationofthe
customertocontactaqualifiedelectricianandhavethe
twoprongreceptaclereplacedwithaproperlygrounded
threeprongwallreceptacleinaccordancewiththe
NationalElectricalCode.
Ifyoumustuseanextensioncord,useonlyathree
prongextensioncordhavingthesameorhigher
electricalratingastheproductbeingconnected.DO
NOT,underanycircumstances,cutorremovethe
roundgroundingprongfromanyplugusedwithor
forInvacareproducts.
JunctionboxMUSTbesecurelyconnectedtothebed
withmetalscrewsandthespacerprovidedforproper
grounding.Otherwise,electricalshockmayoccur.
WARNING!
RiskOfInjuryorDamage
Toavoidinjuryordamagefromimpropergroundingor
usingdamagedparts:
DONOToperatetheproductifthecordoroutlet
isdamaged.
DONOToperatetheproductifanymotorhas
malfunctionedorisdamagedinanymanner.
1123842-I-0513
IVC™BariatricBed
RadioFrequencyInterference
UseofportableandmobileRFcommunicationsequipment
canaffectmedicalelectricalequipment
WARNING!
RiskOfProductDamage
ElectronicequipmentmaybeinfluencedbyRadio
FrequencyInterference(RFI).
Cautionshouldbeexercisedwithregardtotheuse
ofportablecommunicationsequipmentinthearea
aroundsuchequipment.
IfRFIcauseserraticbehavior,unplugtheproduct
immediately.Leaveunpluggedwhiletransmissionis
inprogress.
WARNING!
RiskOfProductDamage
Theuseofaccessories,tranducers,cordsandcables
otherthanthosespecifiedbythemanufacturer,may
resultinincreasedemissionsordecreasedimmunityof
thisbed
Thisproductshouldnotbeusedadjacenttoorstacked
withotherequipmentandthatifadjacentorstacked
useisnecessary,thisproductshouldbeobservedto
verifynormaloperationintheconfigurationinwhich
itwillbeused.
WeightLimitations
DANGER!
RiskOfDeath,Injury,OrDamage
Toavoidbedcollapseandresultinginjuryordamage:
TheweightlimitoftheoccupantofthebedCANNOT
exceed600pounds.TheTOTALweightlimitofthe
bed(includingaccessories,mattress,occupantandany
otherobjectspositionedonthebed)cannotexceed
700pounds(SafeWorkingLoad).
NEVERpermitmorethanonepersononthebedat
anytime.
Bodyweightshouldbeevenlydistributedoverthe
surfaceofthebed.DONOTlay,sitorleaninsuch
awaythatyourentirebodyweightisplacedonlyon
raisedheadorfootsectionsofthebed.Thisincludes
whenrepositioningortransferringinoroutofbed.
14
1123842-I-05
Safety
3.2MattressCompressionTest
DANGER!
RiskofDeath,Injury,orDamage
Toavoidinjuryordamagefrommisuseorentrapment:
Properpatientassessmentandmonitoring,andproper
maintenanceanduseofequipmentisrequiredto
reducetheriskofentrapment.Variationsinbedrail
dimensions,andmattressthickness,sizeordensity
couldincreasetheriskofentrapment.VisittheFDA
websiteathttp://www.fda.govtolearnabouttherisks
ofentrapment.Review“AGuidetoBedSafety”,
publishedbytheHospitalBedSafetyWorkgroup,
locatedatwww.invacare.com.Usethelinklocated
undereachbedrailproductentrytoaccessthisbed
safetyguide.
Refertotheusermanualsforbedsandrailsfor
additionalproductandsafetyinformation.
Patiententrapmentfromuseofbedsiderails
maycauseinjuryordeath.TheInvacaremattress
MUSTfitfirmlyagainstthebedframeANDbed
siderailstopreventpatiententrapment.Follow
themanufacturer’sinstructions.Monitorpatient
frequently.ReadandunderstandtheUserManual
priortousingthisequipment.Invacaremanualsare
availableatwww.invacare.comoryourdealer.
Anincreasedriskofpatiententrapmentmayoccur
overtimeduetomattresscompression.Periodically
monitorgapsbetweenthebed,mattress,and/orbed
rail.Wheregapsoccur,patiententrapmentispossible
andthemattressshouldbereplaced.
1.Removeallbeddingfromthemattress.
2.Measurethethicknessofthemattress(mattressdepth)inan
uncompressedarea.
3.Layastraightedgeoverthecompressedareaofthemattressso
eachendisroughlyatthesameheight.
4.Measurefromthebottomofthestraightedgetothelowestpoint
ofthecompressedareatofindthedepthofthecompressed
area(compressiondepth).
5.Putthemeasurementsyouhaveobtainedintofollowingformula
(compressiondepthdividedbymattressdepthmultipledby100)
todeterminetheamountofcompression(compresspercentage):
1123842-I-0515
IVC™BariatricBed
3.3Disposal
Followlocalgoverningordinancesandrecyclingplansregarding
disposaloftheproductorcomponentsnormallyusedinoperation.
Theproductdoesnotgeneratewasteorresidueinoperation.Any
accessoriesnotpartoftheproductMUSTbehandledinaccordance
withtheindividualproductmarkingfordisposal.
Thisproducthasbeensuppliedfromanenvironmentallyaware
manufacturerthatcomplieswiththeWasteElectricalandElectronic
Equipment(WEEE)Directive2012/19/EC.
Thisproductmaycontainsubstancesthatcouldbeharmfultothe
environmentifdisposedofinplaces(landfills)thatarenotappropriate
accordingtolegislation.
Pleasebeenvironmentallyresponsibleandrecyclethisproduct
throughyourrecyclingfacilityatitsendoflife.
161123842-I-05
Set-up
4Set-up
4.1Receiving
WARNING!
RiskofInjuryorDamage
“QualifiedTechnicianOnly”sectionsaremarked
accordingly.
Set-upandAssemblyinstructionsMUSTbeperformed
byqualifiedtechniciansonly.
1.Countcartonsbeforesigningshippingpapers.
2.Designatesomeoneatthefacilitytoreceiveandinspectall
merchandise.Usethemasterorderacknowledgmenttocheck
offitemsastheyarereceived.
3.Comparethenumberofcartonsshippedwiththenumberof
cartonsactuallyreceivedandnotesuchontheconsigneecopy
ofthefreightbill.Makesuretohavethetruckdriversignthe
copyacknowledgingthecount.
4.Ifshortageoccurs,IMMEDIATELYcontactInvacareCustomer
Servicedepartmentat(800)333-6900.
4.2Unpacking
Thebediscontainedinthreecartons.
DANGER!
RiskofInjuryorDamage
Continueduseoftheproductwithdamagedpartscould
leadtothebedmalfunctioning,causinginjurytotheuser
and/orcaregiver.
CheckALLproductcomponentsandcartonfor
damage,andtestcomponentsbeforeuse.Incaseof
damageoriftheproductisnotworkingproperly,
contactaqualifiedtechnicianorInvacareforrepair.
1.Removeanyloosepackingfromthecartons.
2.Carefullyremoveallcomponentsfromthecartons.
WARNING!
RiskofInjuryorDamage
Toavoiddamageorpersonalinjuryfrommisuse:
DONOTplugthepowercordintoapower
sourceuntilassemblyiscomplete.
DONOTattempttooperateproductcontrols
priortocompletionofassembly.
1123842-I-0517
IVC™BariatricBed
UnpackingBedEnds
1.ThebedendcartonAincludes:
TwobedendsB
FourcastersC(twolockingandtwonon-lockingcasters)
OnedriveshaftassemblyintwopiecesD
UnpackingFootandHeadSections
1.ThefootsectioncartonEincludes:
OnefootsectionF
Oneemergencycrankhandle
2.TheheadsectioncartonGincludes:
OneheadsectionH
NotshownPackinginserts,junctionboxwithcord,
motorswithcablesattached,pendant,emergencycrank
handle,andusermanual.
181123842-I-05
Set-up
4.3Inspecting
DANGER!
RiskOfInjuryOrDamage
Toavoidinjuryordamage,inspectthefollowingand
contactaqualifiedtechnicianorInvacareifanyofthe
followingissuesarepresent:
Pendantcoveringiscrackedordamaged.
Powercordorplughavecutsordamage.
Motorcableshavecutsordamage.
Junctionboxorconnectorsaredamaged.
DANGER!
RiskOfInjuryOrDamage
Toavoidinjuryordamage:
Ensurethattheconnectorsarepluggedintothe
junctionboxandfitproperly.
4.4Assembling/DisassemblingHeadandFoot
Sections
WARNING!
RiskofInjuryorDamage
Toavoidinjuryordamage:
DONOTattempttooperateproductcontrolsprior
tocompletionofassembly.
Assembling
Notoolsarerequiredtoassemblethisproduct.
1.Removetheheadsectionofthebedfromthepackingcarton.
Placetheheadsectiononaflatsurfacesothatthemotor
guardsonthebottomoftheframearesupportingthe
headsection.
2.Removethefootsectionofthebedfromthepackingcarton.
3.Removetie-wrapsAsecuringthecablesforshipping.
Placethefootsectiononaflatsurfacesothatthemotor
guardsonthebottomoftheframearesupportingthe
footsection.
4.Aligntheinsideoftheheadsectionwiththeinsideofthefoot
section.
Lockingpinsarelocatedonthefootsectionframeand
willmakeanaudibleclickwhenproperlyengagedwith
theheadsectionframe.
5.Pushtheheadsectiontowardsthefootsectionsothatthehead
sectionframeslidesintothefootsectionframeandthelocking
pinsengage.
6.Ensurethelockingpinsareproperlyengagedbyattemptingto
pulltheheadsectionfromthefootsection.Ifthesections
separate,repeatSTEP5-6untilthelockingpinsaresecure.
1123842-I-0519
IVC™BariatricBed
7.Installthebedends.Referto4.6Installing/RemovingBedEnds,
page22
.
8.Connectthecrossovercable.Referto4.7Connecting
CrossoverCable,page23
.
9.Assembleandinstallthedriveshaft.Referto4.8
Assembling/Installing/RemovingDriveShaft,page25
.
10.Activatethebedfunctions.Referto4.9InspectingtheBed,page
27
.
Disassembling
1.LowertheheadAandfootBsectionscompletely.Referto5.1
OperatingBariatricBed,page29.
2.Lowerthebedtothelowestheight.
3.Removethedriveshaft.Referto4.8Assembling/Installing/Re-
movingDriveShaft,page25.
4.Disconnectthecrossovercable.Referto4.7Connecting
CrossoverCable,page23
.
5.Removebothbedends.Referto4.6Installing/RemovingBed
Ends,page22.
Placethebedonaflatsurfacesothatthemotorguards
onthebottomoftheframearesupportingthebed.
6.Startingwithonesideofbed,pullthereleasepinCandpullthe
headsectionsothatitstartstoseparateslightlyfromthefoot
section.
7.Ontheoppositesideofbed,pullthereleasepin.
8.Separatetheheadsectionfromthefootsection.
Motorguardslocatedonundersideofframesections.
201123842-I-05
/