CT Injection System
CT Injection System
Operation Manual
Personalized Patient Protocol Technology (P3T)
P3T Pulmonary Angiography (PA) Module
for Certegra® Workstation
i
Personalized Patient Protocol Technology (P3T)
P3T Pulmonary Angiography (PA) Module
Operation Manual
Report any serious incident that has occurred in relation to this device to
Bayer (radiology.bayer.com/contact) and to your local European competent
authority (or, where applicable, to the appropriate regulatory authority of the
country in which the incident has occurred).
Hardware operating specifications and feature availability may vary by
country. Check with your local product representative and country-specific
operating instructions.
A glossary of the symbols can be found in Section 2 of this manual.
Personalized Patient Protocol Technology (P3T): P3T Pulmonary Angiography (PA) Module
ii
iii
Table of Contents
1 Introduction ......................................................................................................................1 - 1
1.1 Purpose ............................................................................................................................................................ 1 - 1
1.2 Essential Claims................................................................................................................................................ 1 - 1
1.3 Intended Use..................................................................................................................................................... 1 - 1
1.4 Indications for Use............................................................................................................................................. 1 - 1
1.5 Restricted Sale.................................................................................................................................................. 1 - 1
1.6 Disclaimers ....................................................................................................................................................... 1 - 1
2 Symbols ............................................................................................................................2 - 3
2.1 Warning Symbols .............................................................................................................................................. 2 - 3
2.2 General Symbols ............................................................................................................................................... 2 - 3
3 Product Overview ............................................................................................................3 - 5
3.1 Product Overview .............................................................................................................................................. 3 - 5
3.2 Theory of Operation........................................................................................................................................... 3 - 5
3.2.1 Introduction to Presets............................................................................................................................ 3 - 5
4 P3T Preset Setup Options...............................................................................................4 - 7
4.1 Setup Options ................................................................................................................................................... 4 - 7
4.2 Must Review, Must Enter................................................................................................................................. 4 - 11
4.3 DualFlow......................................................................................................................................................... 4 - 12
5 Managing Presets..........................................................................................................5 - 13
5.1 Create New Preset .......................................................................................................................................... 5 - 13
5.2 Modify a Preset ............................................................................................................................................... 5 - 15
5.3 Delete a Preset................................................................................................................................................ 5 - 17
5.4 Rearrange Preset Order................................................................................................................................... 5 - 18
6 Using Presets .................................................................................................................6 - 19
6.1 Using a Preset to Generate a Protocol.............................................................................................................. 6 - 19
6.2 Modify P3T Input Parameters .......................................................................................................................... 6 - 21
6.3 Modify a P3T Protocol After a Transit Bolus ..................................................................................................... 6 - 22
7 Common P3T Work Flows ............................................................................................7 - 23
7.1 Minimum Injection Duration ............................................................................................................................ 7 - 23
7.2 Transit Bolus................................................................................................................................................... 7 - 23
7.3 Bolus Tracking ................................................................................................................................................ 7 - 24
7.4 DualFlow Configuration ................................................................................................................................... 7 - 25
Personalized Patient Protocol Technology (P3T): P3T Pulmonary Angiography (PA) Module
iv
1 - 1
1 Introduction
1.1 Purpose
This document provides instructions for using the Personalized Patient Protocol Technology (P3T)
Pulmonary Angiography (PA) (P3T PA) on the MEDRAD
®
Stellant CT Injection System with Certegra
®
Workstation (Stellant CWS) and on the MEDRAD
®
Stellant FLEX CT Injection System with Certegra
®
Workstation (Stellant FLEX).
Further instructions on loading and injecting, as well as other important information can be found in the
operations manuals for the specific injection system.
Please read all the information included in this manual. It is important to understand this information to
help ensure safe and effective usage of this product.
1.2 Essential Claims
P3T PA computes individual contrast injection protocols and scan timing for individualized
dosing.
P3T PA increases the consistency of individualized injection protocols amongst clinicians.
1.3 Intended Use
P3T PA is intended for use with CT Angiography. P3T PA computes individual contrast injection
protocols and scan timing, based on patient characteristics, scanner parameters and contrast
concentration.
1.4 Indications for Use
P3T PA is indicated for use with CT Angiography of the cardiac structures, coronary arteries, chambers
of the heart, pulmonary vasculature, thoracic and abdominal aorta.
NOTE: P3T PA has not been validated for anatomical regions outside of the stated Indica-
tions For Use.
1.5 Restricted Sale
Federal (USA) law restricts this device to sale by or on the order of a physician.
1.6 Disclaimers
Screen images in this manual are for illustration purposes only. Actual screens may vary.
Personalized Patient Protocol Technology (P3T): P3T Pulmonary Angiography (PA) Module
1 - 2
2 - 3
2 Symbols
This section includes the symbols and icons for the following:
"Warning Symbols"
"General Symbols"
2.1 Warning Symbols
2.2 General Symbols
Indicates that the information is a warning. Warnings advise the user of
circumstances that could result in injury or death to the patient or operator.
Read and understand the warnings before operating the injection system.
Indicates that the information is a caution. Cautions advise the user of
circumstances that could result in minor or moderate injury to the patient or
operator. Read and understand the cautions before operating the injection
system.
Indicates that the information is a notice. Notices advise the user of
circumstances that could result in damage to the device. Read and understand
the notices before operating the injection system.
Note
Indicates that the information that follows is additional important information or
a tip that will help the user recover from an error or reference to related
information within the manual.
WARNING
CAUTION
NOTICE
Manufacturer (ISO 15223-1, 5.1.1)
Authorized Representative in the European Community (ISO 15223-1, 5.1.2)
Personalized Patient Protocol Technology (P3T): P3T Pulmonary Angiography (PA) Module
2 - 4
3 - 5
3 Product Overview
"Product Overview"
"Theory of Operation"
3.1 Product Overview
P3T allows users to create individualized contrast injection protocols based on patient characteristics,
contrast media properties and other procedure parameters. Contrast dosage is determined based on
the iodine to be delivered to the patient (for example, a heavier patient will receive more grams of
iodine while a lighter patient will receive fewer grams of iodine).
P3T PA is a module available as part of the P3T platform that enables optimizing of contrast injection
protocols for CT angiography of pulmonary vasculature. This module automates the calculation of
individualized contrast injection protocols that would otherwise have to be performed manually by the
user.
The user enters the patient's weight, the iodine concentration of the contrast media, and other key
parameters for the contrast injection protocol. Based on this input, P3T generates a protocol
specifically tailored to the individual patient by delivering a measured amount of contrast at a flow rate
that will allow all of the iodine to be delivered within the duration of the scan.
NOTE: P3T does not determine the appropriate contrast timing (i.e., flow rate or injection
duration). Physicians must determine the contrast timing.
The overall injection protocol consists of two sub-protocols: an optional Pre-Diagnostic Protocol, and a
Diagnostic Protocol. The Pre-Diagnostic Protocol includes an optional Test Injection and/or Transit
Bolus phase.
The Transit Bolus phase injects a small amount of contrast or saline and is used to help determine the
time to peak enhancement in a specific area of interest and the peak enhancement level achieved
(information that is then entered into P3T to help configure the final diagnostic protocol).
NOTE: A Test Injection is available. A Test Inject phase injects a small amount of saline and
is used to check vein patency. Refer to the specific injection system operation
manual for more information.
3.2 Theory of Operation
3.2.1 Introduction to Presets
P3T Presets are user-defined templates that specify values for the input parameters required to
generate a P3T protocol. Presets enable the user to reuse specific input parameters without needing to
enter or modify input parameter values.
To create a personalized P3T protocol for each patient, the user recalls the defined preset from
memory and adds patient and procedure-specific information such as patient weight and contrast
concentration.
NOTE: Refer to “Using Presets” on page 6 - 19 for information on using presets.
Personalized Patient Protocol Technology (P3T): P3T Pulmonary Angiography (PA) Module
3 - 6
4 - 7
4 P3T Preset Setup Options
4.1 Setup Options
The Preset setup options allow users to configure the module according to work flow, policies, and
preferences of each facility.
Table 4 - 1: P3T Setup Options
UI Display
Order
Name Description Selectable Value
Factory
Default
1 Contrast Type
Defines the contrast concentration value used for
the protocol.
A pick-list of user-
defined contrast
types. For more
information on
defining contrast
types, refer to the
specific injection
system operation
manual. *
350 mg/mL
2 Test Inject Includes a test injection as part of the protocol. Yes, No * No
3 Test Inject Fluid
Defines the fluid to be used during the test
injection.
Contrast, Saline Saline
4
Test Inject
Delivery Method
Selects the way the test injection is calculated
and delivery - by volume or duration.
Volume, Duration Volume
5
Test Inject
Volume
Sets the volume for the test injection when the
Test Inject Delivery Method is set to Volume.
10 mL- 50 mL in
increments of 1 mL *
20 mL
6
Test Inject
Duration
Sets the duration of the test injection when the
Test Inject Delivery Method is set to Duration
3 seconds -15
seconds in
increments of 1
second *
5 seconds
7 Transit Bolus
Configures the preset to include a transit bolus in
the pre-diagnostic sub-protocol. A transit bolus
consists of a contrast bolus follow by a saline
bolus.
Yes, No * No
8
Transit Bolus
Contrast
Volume
Sets the default volume for a contrast bolus
portion of the Transit Bolus if the Transit Bolus
parameter is set to Yes.
10mL - 25mL in
increments of 1mL *
20 mL
9
Transit Bolus
Saline Volume
Sets the default volume for a saline bolus portion
of the Transit Bolus if the Transit Bolus parameter
is set to Yes.
0mL - 50mL in
increments of 1mL *
40 mL
*A user can modify this parameter during protocol usage.
** A user can modify this parameter during protocol usage and after the Transit Bolus.
Personalized Patient Protocol Technology (P3T): P3T Pulmonary Angiography (PA) Module
4 - 8
10 Weight Bins
Enables a user to create seven custom weight
bins. Units can be in pounds or kilograms
depending on system settings.
Upper limit of each
Weight Bin can be
defined from 20 –
320 kg (44 – 705 lb).
Weight Bins
(kg)
< 40 kg,
40–59 kg,
60–74 kg,
75–94 kg,
95–109 kg,
110–125
kg,
> 125 kg
Weight Bins
(lb)
< 86 lb,
86-130 lb,
131-163 lb,
164-207 lb,
208-240 lb,
241-276 lb,
>276 lb,
11
Maximum Flow
Rate
Sets the maximum flow rate allowed for the
protocol. If the P3T protocol is generated such
that it exceeds the max flow rate value, the
protocol will automatically be adjust so that the
max flow rate value is not exceeded and the user
is notified of the protocol adjustment.
5 mL/s - 10 mL/s in
increments of 0.1
mL/s **
6.0 mL/s
12
Duration
Adjustment
Used to calculate the duration for the first phase
of the diagnostic protocol. The Duration
Adjustment is added to the Scan Duration to
determine the injection duration of the first
phase. Contrast injection durations should be
sufficiently long to account for the dispersive
effects of the cardio pulmonary system. It is
therefore advisable to include additional injection
duration to the bolus to avoid suboptimal
enhancement towards end of the scan. The
Duration Adjustment setup option enables the
user to configure this additional injection
duration. Since P3T computes the flow rate of the
contrast phase in the Diagnostic Protocol based
on contrast volume and injection duration, a
larger value of Duration Adjustment results in a
longer injection duration and a lower flow rate.
0 seconds -10
seconds in
increments of 1
second
4 seconds
Table 4 - 1: P3T Setup Options
UI Display
Order
Name Description Selectable Value
Factory
Default
*A user can modify this parameter during protocol usage.
** A user can modify this parameter during protocol usage and after the Transit Bolus.
P3T Preset Setup Options
4 - 9
13 Weight Factors
Enables the user to edit the default weight
factors, which determine the gI/kg used in the
first contrast phase of the diagnostic protocol.
0.05 to 1.0 gI/kg, in
increments of 0.001
0.5, 0.4,
0.375, 0.35,
0.31, 0.30
gI/kg
14
Minimum
Injection
Duration
Specifies the minimum injection duration of the
diagnostic contrast phase. In order to reduce
contrast volume, short minimum injection
duration value may be desired for very fast scans
(3-5 seconds) when the user is comfortable with
higher flow rates for a patient. The minimum
injection duration requirement is enforced after
the Duration Adjustment is added to the Scan
Duration. For example, in the case of a 5 second
Scan Duration, a 4 second Duration Adjustment,
and a Min Injection Duration of 10 seconds: the
calculated injection duration of the first phase of
the P3T protocol is 9 seconds, but would be
increased to 10 seconds based on the Min
Injection Duration specified.
6 seconds -16
seconds in
increments of 1
second
13 seconds
15
Saline Flush
Volume
Sets the volume for the saline flush phase. Set
the volume to zero to eliminate the saline flush
phase
0 mL - 50 mL in
increments of 1 mL *
30 mL
16
Protocol
Truncation
Adjustment
Sets the minimum time between when a phase
containing contrast ends and when the scan
ends. Because of the time it takes for the contrast
to travel from the injection site to the region of
interest, contrast injected too close to the end of
the scan does not contribute to the enhancement
in that region
0 seconds - 10
seconds in
increments of 1
second
3 seconds
17
Maximum
Iodine Load
Sets the maximum iodine load for the first
contrast phase of the diagnostic injection only.
20 gI - 80 gI in
increments of 1 gI *
38 gI
18
Minimum Iodine
Load
Sets the minimum iodine load for the first
contrast phase of the diagnostic injection only.
0 gI - 50 gI in
increments of 1 gI
20 gI
19
Maximum
Diagnostic
Contrast
Volume
Sets the maximum contrast volume allowed for
the protocol. If the P3T protocol is generated such
that it exceeds the max contrast volume value,
the protocol will automatically be adjusted so that
the max volume is not exceeded and the user is
notified of the protocol adjustment.
79 mL - 194 mL in
increments of 1mL
**
194 mL
Table 4 - 1: P3T Setup Options
UI Display
Order
Name Description Selectable Value
Factory
Default
*A user can modify this parameter during protocol usage.
** A user can modify this parameter during protocol usage and after the Transit Bolus.
Personalized Patient Protocol Technology (P3T): P3T Pulmonary Angiography (PA) Module
4 - 10
20
Maximum
Diagnostic
Saline Volume
Sets the maximum saline volume allowed for the
protocol. If the P3T protocol is generated such
that it exceeds the maximum saline volume
value, the protocol will automatically be adjusted
so that the max volume is not exceeded and the
user is notified of the protocol adjustment.
0 mL - 194 mL in
increments of 1 mL
**
194 mL
21 Pressure Limit
Defines the pressure limit to be stored with the
Preset
A pick-list of
available pressure
limits (50, 100, 150,
200, 250, 300, 325
psi)
325 psi
22 Scan Duration
Identifies the length of time for the Diagnostic
portion of the scan acquisition.
1 second - 40
seconds
10 seconds
23
Weight Entry
Method
Defines the way in which the user enters patient
weight. For Weight Bins, the system displays five
defined weight ranges for the patient. For Key
Pad, the system display a numeric keypad for
patient weight entry.
NOTE: If the user configures the Weight Entry
Method to use Weight Bins, a representative
weight within the selected range will be used to
calculate the contrast dosage. As a result, the
Key Pad weight entry method is more accurate
compared to the Weight Bin entry method.
Weight Bin, Key Pad Key Pad
24 DualFlow
Enables the use of DualFlow for the P3T Preset.
When Yes is selected, the underlying algorithm
will consider a contrast dilution phase for the
protocol. In some cases, even through the Preset
may include DualFlow, the algorithm may not
generate a protocol with a DualFlow phase (for
example, when scan durations are short and the
scan is completed before the time of the injection
of the DualFlow phase). Since DualFlow is
primarily beneficial for maintaining opacification
of the right heart structures, it may not be
required for pulmonary angiography studies.
Also, by configuring the Preset for P3T PA to not
include DualFlow, initial contrast volume
calculation could be lower than if it were
included.
Yes, No No
Table 4 - 1: P3T Setup Options
UI Display
Order
Name Description Selectable Value
Factory
Default
*A user can modify this parameter during protocol usage.
** A user can modify this parameter during protocol usage and after the Transit Bolus.
P3T Preset Setup Options
4 - 11
4.2 Must Review, Must Enter
As part of the preset, some input parameters can be defined as Must Review or Must Enter.
NOTE: During use of the preset, some input parameters can be modified. The Must
Review designation provides a faster method to highlight specific parameters. Test
Inject is always a Must Review value.
NOTE: Patient weight is always a Must Enter value.
Table 4 - 2: Must Review, Must Enter Options
Check Box Options Example Action and Display During P3T Preset Use
“X” in Must Review box:
Action: User should review the value of the
parameter each time the P3T Preset is used.
But, there is no Required action.
Display: The parameter will move towards
the top of the list and be highlighted in beige
“X” in Must Enter box:
Action: User is Required to enter a value for
the parameter each time the Preset is used.
Display: The parameter will move towards
the top of the list and be highlighted in beige
No “X” in either box
Action: User may enter a value for the
parameter each time the Preset is used. But,
there is no Required action.
Display: The parameter will appear in the
list, but in no particular order
No check boxes and gray
shading
Action: User will not be able to change the
value for the parameter when the Preset is
used. The value can only be changed during
the Preset creation.
Display: The parameter will NOT be
displayed in the list.
Personalized Patient Protocol Technology (P3T): P3T Pulmonary Angiography (PA) Module
4 - 12
4.3 DualFlow
When Yes is selected, the System considers a DualFlow phase for the protocol. In some cases, even
though the module is configured to include DualFlow, the system may not generate a protocol with a
DualFlow phase (for example, when scan durations are short and the scan is completed before the
time of injection of the DualFlow phase).
NOTE: Since DualFlow is primarily beneficial for maintaining opacification of the right
heart structures, it may not be required for pulmonary angiography studies.
NOTE: By configuring P3T to not include DualFlow, initial contrast volume calculations
could be lower than if it were included.
NOTE: Shorter scan times for pulmonary angiography exams can obviate the need for
DualFlow. DualFlow may be appropriate to use for slower scanners and gated PE
exams.
For more information on DualFlow refer to the operation manual for the specific injection system.
5 - 13
5 Managing Presets
This section describes how to:
"Create New Preset"
"Modify a Preset"
"Delete a Preset"
"Rearrange Preset Order"
Features to organize presets within the protocol manager are also discussed including:
How to change the display order of presets
How to move presets from one folder to another
How to create a copy of a preset
Configuration of presets is typically done during initial set up and during injection protocol reviews. It is
recommended that the setup options be reviewed by clinicians responsible for specifying injection
protocols in a facility.
5.1 Create New Preset
1. From the launch menu, select Setup > P3T Preset Setup.
Figure 5 - 1: P3T Preset Setup
2. Select Create New Preset.
3. Under Presets, select P3T PA.
Personalized Patient Protocol Technology (P3T): P3T Pulmonary Angiography (PA) Module
5 - 14
Figure 5 - 2: Create New Preset
The preview pane displays the default values for the preset parameters.
4. Select OK. A preset entry screen displays.
5. Select the Region to Save. Press SAVE.
Figure 5 - 3: Preset Entry Screen
6. Modify a preset.
a. Use the UP or DOWN arrows to scroll through the parameters.
b. Under CURRENT, select the displayed Preset parameter to modify the value.
c. Enter or choose the values from the keypad or pick list.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34