The physical and toxicological properties of the products in this kit have not been
thoroughly investigated. Follow prudent laboratory practices and use general
laboratory safety equipment (eye protection, labcoat, and lab gloves) and good
personal hygiene when working with these or any laboratory reagents. Refer to the
Safety Data Sheet for more information.
Proper laboratory practice is necessary as previously amplied PCR product is the
most likely potential source of contamination. We strongly recommend two separate
work areas be used to minimize the risk of cross-contamination during the assay
procedure. It is essential to adhere to workow recommendations. PCR reactions
should be set up in the Pre-PCR Area only. Personnel should not reenter the Pre-PCR
Clean Area once exposed to PCR products without rst showering and changing into
clean clothes.
Carefully reading and following the protocol as wrien is essential. The KaryoStat™
Assay has been validated using the reagents and suppliers listed. Substitution of
reagents and not following detailed procedures are not recommended as your results
could be suboptimal.
• Follow standard procedures and single-direction workow for the Pre-PCR
laboratory area.
• Use dedicated equipment for each area (e.g., thermal cyclers, microfuges, pipees
and tips, ice buckets, etc.)
• Place all reagents and master stocks in use area. Do not move equipment between
Pre- and Post-PCR Rooms, e.g., ice buckets, pipees, etc.
• Use separate copy of assay procedure in Pre- and Post-PCR areas.
• Follow procedures for re-entry of Pre-PCR Clean Room from post-PCR.
• If pre-PCR work is done in a laminar ow hood or PCR cabinet, then additionally
ensure:
– laminar ow hood is always on
– UV lamp is on when not in use
• Use kit after its expiration date.
• Use reagents after more than four freeze-thaw cycles.
• Use any water other than Nuclease-free Water supplied with the CytoScan™
Optima Kit.
• Reuse a plate seal.
Precautions
Pre/post-PCR
Do not