5.2.1Cell
Navigation: Home > Setup > Hardware > Cell
The measuring cell is connected via an internal interface in the compact device. The instrument can be extend
ed to create a 2–cell instrument (see also "Setup: Maintenance & Service > Add external cell (page33)").
– Touch the button Cell in the Hardware dialog.
With single cell instruments you go directly to the list with the cell parameters
In the case of two cell instruments the connected cells are listed. You have to touch one of the displayed
cells so that the Cell parameters dialog opens.
The Cell parameters dialog is opened.
Parameters Description Displayed if
Cell Name of the measuring cell -
Type Measuring cell type -
Status Shows whether the cell is connected. -
Serial number Serial number of the cell -
If an adjustment has been performed, an entry in the adjustment set appears in the setup of the corresponding
cell. If you touch the entry, the data of the most recent adjustment that was executed with the set will be dis
played.
Parameters Description Displayed if
Adjustment set By touching the set entry, you obtain the values for the most
recent adjustment of this set.
-
Monitoring
adjustment set
Monitoring of the service life of the adjustment can be activated.
The process, if the service life of a set has expired, is defined in
the method function "Measure".
-
Adjustment inter
val
Validity of adjustment "Monitoring adjustment
set" activated
Reminder Before the set expires, a message appears indicating that the
adjustment or test has expired.
"Monitoring adjustment
set / test set" activated
Days before expi
ration
Specifies the number of days after which a warning is triggered. "Reminder" activated
If a test has been performed, an entry in the test set appears in the setup of the corresponding cell. If you touch
the entry, the data of the most recent adjustment that were executed with the set will be displayed.
Parameters Description Displayed if
Test set By touching the set you go to the values for the most recent test
for this set.
-
Monitoring test
set
Monitoring of the service life of the test can be activated. The
process, if the service life of a set has expired, is defined in the
method function "Measure".
-
Test interval Service life of test "Monitoring test set"
activated
Reminder Before the set expires, a message appears indicating that the
adjustment or test has expired.
"Monitoring adjustment
set / test set" activated
Days before expi
ration
Specifies the number of days after which a warning is triggered. "Reminder" activated
Via the button History you go to a list that contains a maximum of ten adjustment or test entries for the selected
set. In the process the measured refractive index values are displayed (in the adjustment mode air and water
only the value for water). The history can be displayed in graphic form via the button Graph. You can view the
history data by touching an entry.
5.2.2Automation
Navigation: Home > Setup > Hardware > Automation
The automation units that can be installed are listed below:
17Setup