Hach SIGMATAX 2 User manual

Type
User manual
DOC023.52.03103
SIGMATAX 2
User manual
05/2018, Edition 3
3
Table of Contents
Section 1 Specifications ........................................................................................................................ 5
Section 2 General Information............................................................................................................... 7
2.1 Safety information............................................................................................................................... 7
2.1.1 Use of hazard information.......................................................................................................... 7
2.1.2 Precautionary labels .................................................................................................................. 7
2.2 Application areas ................................................................................................................................ 8
2.3 Functional description......................................................................................................................... 8
2.4 Instrument design ............................................................................................................................... 9
Section 3 Installation............................................................................................................................ 11
3.1 Mechanical installation...................................................................................................................... 12
3.1.1 Install the sampling probe........................................................................................................ 14
3.1.2 Second mounting point and extension pipe............................................................................. 16
3.1.3 Install the open loop control unit .............................................................................................. 18
3.1.4 Prepare the connection tube.................................................................................................... 20
3.1.5 Connect the connection tube ................................................................................................... 21
3.2 Electrical installation ......................................................................................................................... 22
3.2.1 Connect the pipe trace heater and fault indication contact ...................................................... 22
3.2.2 Connect the fiber optic cable ................................................................................................... 24
3.3 Internal tubing ................................................................................................................................... 24
Section 4 Start Up................................................................................................................................. 27
Section 5 Operation.............................................................................................................................. 29
5.1 Operation .......................................................................................................................................... 29
5.2 Menu overview.................................................................................................................................. 30
5.3 Important instrument settings ........................................................................................................... 31
5.3.1 Switch on the heater ................................................................................................................ 31
5.3.2 Set the processing interval....................................................................................................... 31
5.3.3 Set the display contrast............................................................................................................ 31
5.3.4 Switch on the pressure display ................................................................................................ 31
5.3.5 Access service mode............................................................................................................... 32
Section 6 Maintenance......................................................................................................................... 33
6.1 Maintenance tasks............................................................................................................................ 34
6.1.1 After 1–4 weeks (as necessary)............................................................................................... 34
6.1.2 1×/month (or as necessary) ..................................................................................................... 35
Section 7 Troubleshooting................................................................................................................... 37
Section 8 Replacement parts and accessories ............................................................................... 39
8.1 Product contents............................................................................................................................... 39
8.2 Further accessories .......................................................................................................................... 39
8.3 Replacement and wear parts............................................................................................................ 39
Section 9 Warranty and liability........................................................................................................... 41
4
Table of Contents
5
Section 1 Specifications
From software version 1.6. Subject to change.
Measurement
Principle of operation Samples delivered under pressure, homogenized using ultrasound
Interval 12 to 20 minutes
Sample quantity For up to 2 PHOSPHAX sigma or TOCTAX process photometers
Tube length 10 m, 20 m, 30 m (393.7 in, 787.4 in, 1181.1 in)
Suction height Maximum 7 m (275.6 in) (with tube of length 30 m (1181.1 in): maximum 6 m 236.2 in)
Environmental conditions
Ambient temperature
Control unit: +5 °C to +40 °C (+41 °F to +104 °F)
Connection tube: –20 °C to +40 °C (–4 °F to +104 °F)
(Without additional insulation; in the event of low external temperatures, correct operation can
only be guaranteed with additional insulation. More information available on request)
Medium temperature +5 °C to + 30 °C (+41 °F to +86 °F) (probe)
Equipment properties
Protection type in accordance with IP 54 (Control unit)
Outputs
Fault indication contact, potential-free, 24 V, maximum 1 A
Service interface: RS 232
Power connector 230 V AC / 50 Hz
Power consumption
Device with system components : 600 VA
Recommended fuse: 10 A, time-delay
Line Heater off Heater on (–20 °C (–4 °F))
Continuous
operation
At startup
10 m (393.7 in) 150 VA 250 VA 1300 VA
20 m (787.4 in) 150 VA 350 VA 2300 VA
30 m (1181.1 in) 150 VA 470 VA 3000 VA
Inspection interval: 3 months
Maintenance requirement: Approximately 0.5 h/week, typically
Dimensions:
Open loop control unit: W × H × D 366 × 560 × 212 mm (14.4 × 22 × 8.3 in
Probe: Ø × L 133 × 404 mm (5.2 × 15.9 in
Earth (approximate)
Control unit: 12 kg
Sampling probe 10 m (393.7 in) 7.5 kg
Sampling probe 20 m (787.4 in) 15 kg
Sampling probe 30 m
(1181.1 in)
22 kg
Mounting poles: 10 kg
6
Specifications
7
Section 2 General Information
2.1 Safety information
Please read this entire manual before unpacking, setting up, or operating this equipment.
Pay attention to all danger and caution statements. Failure to do so could result in serious
injury to the operator or damage to the equipment.
To prevent damage to or impairment of the device's protection equipment, the device may
only be used or installed as described in this manual.
2.1.1 Use of hazard information
Note: Information that supplements points in the main text.
2.1.2 Precautionary labels
Observe all labels and notices attached to the instrument and/or delivery packaging.
Personal injury or damage to the instrument could occur if not observed.
DANGER
Indicates a potentially or imminently hazardous situation that, if not avoided, can result in death or
serious injury.
WARNING
Indicates a potentially or imminently dangerous situation that, if it is not avoided, can lead to
death or to serious injuries.
CAUTION
Indicates a possible dangerous situation that can have minor or moderate injuries as the result.
NOTICE
Indicates a situation that, if it is not avoided, can lead to damage to the device. Information that
requires special emphasis.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety
information.
This symbol may be found on an enclosure or barrier within the product and indicates a risk of electric shock
and/or death by electrocution.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal
systems after 12 August 2005. In conformity with European local and national regulations (EU Directive
2002/96/EC), European electrical equipment users must now return old or end-of life equipment to the
manufacturer for disposal at no charge to the user.
Note: You obtain instructions on the correct disposal of all (marked and not marked) electrical products that
were supplied or manufactured by Hach-Lange at your relevant Hach-Lange sales office.
When carrying or transporting the instrument/instrument components and if the total weight is more than 18 kg,
make sure that suitable lifting equipment is used and/or that the instrument/instrument components are carried
by 2
people.
18-32 kg (39.7-70.5 lbs)
8
General Information
2.2 Application areas
Automatic sample removal and homogenization system for supplying the PHOSPHAX S
sigma (total phosphorus), TOCTAX (total organic carbon, TOC) and astro TOC (total
organic carbon, TOC) process photometers with samples from water containing solid
particles with a diameter < 0.5 mm.
2.3 Functional description
A sampling probe is immersed in the water and draws only the sample volume required for
the analysis without coming into contact with pumps. The sample is conveyed into a small
storage vessel in the open loop control unit and is homogenized using an ultrasound
generator.
NOTICE
Any use other than use in accordance with requirements defined in the user manual leads to the
loss of the warranty claims and can lead to personal injury and property damage, for which the
manufacturer assumes no liability.
9
General Information
2.4 Instrument design
Figure 1 Control unit
1 Sample storage vessel 5 Pressure sensor line 9 Overflow drain tube
2 Sample storage vessel lid 6 Display with keypad 10 Sample storage vessel drain tube
3 Sample tube 7 Sample removal for process
instruments
11 Connection tube from the probe
4 Fiber optic cable for
PHOSPHAX
sigma or TOCTAX
8 Mains connection cable 12 Ultrasound transducer
10
General Information
Figure 2 Sampling probe
1 Connection tube 3 Sampling vessel 5 Air outlet
2 Probe cover 4 Valve cover 6 Drain valve
11
Section 3 Installation
DANGER
Only qualified experts may conduct the tasks described in this section.
DANGER
Select a suitable attachment point for the instrument.
Plan out the mechanical mount·before positioning poles or drilling holes. Make sure the mount
has a sufficient bearing capacity. The dowels must be selected and authorized according to the
condition of the wall.
The manufacturer shall accept no liability if the instrument is installed incorrectly.
Plan how to lay cables and tubes and their path in advance. Lay the tubes, data cables and
power cables without any bends and so they do not pose a tripping risk.
Do not connect the electrical supply to the mains·until the instrument is completely wired and
protected against short circuits.
Sufficiently protect the electrical power supply against short circuits.
For the external power supply, always connect a residual-current circuit breaker·(trip current
max.: 30 mA) between the mains and the system.
If the instrument is to be installed outdoors, connect a surge arrester between the mains and the
system.
Products intended by the manufacturer for outdoor use offer a higher level of protection against
the penetration of liquids and dust. If these products are connected to a mains outlet with a cable
and plug·rather than a permanently connected cable, the plug and outlet are much more
susceptible to liquid and dust penetration. The operator must sufficiently protect the plug and
outlet against liquid and dust penetration in accordance with local safety regulations. If the
instrument is to be used outdoors, it must be connected to a suitable outlet with a protection type
of at least IP44 (splash protection).
WARNING
Electrical and fire hazards. Only use the mains cable supplied. Only qualified experts may
perform the tasks described in this section of the manual, while adhering to all locally valid safety
regulations.
NOTICE
Protect the instrument from extreme temperatures from heaters, direct sunlight and other heat
sources.
CAUTION
Beware of the weight (open loop control unit: approximately 12 kg; sampling probe up to
approximately 22 kg) of the instrument and the probe. Do not attempt to carry the instrument and
probe without assistance. Use only suitable lifting devices for the transport.
12
Installation
3.1 Mechanical installation
Figure 3 Dimensions (in mm)
1 Suspension points 3 Wood screw, 5 x 40 (3×)
2 Fixing point 4 Nylon plug, 8 x 40 (3×)
13
Installation
Figure 4 Installation overview
1 SIGMATAX 2 open loop control unit 4 Mounting attachment 7 SIGMATAX 2 sampling probe
2 Connection tube (10–30 m) 5 Base 8 Direction of flow
3 Retaining clamp 6 Mounting pipe, 2 m (78.7 in)
14
Installation
3.1.1 Install the sampling probe
Installation with rim mounting LZX414.00.00000
Figure 5 Base and mounting attachment
1 Express anchor 3 Mounting attachment
2 Base 4 Cylinder-head screw, M8 × 40 (4×)
15
Installation
Installation using pole mount LZX414.00.00000
Figure 6 Sampling probe and mounting pipe
NOTICE
When carrying out an installation variant that deviates from this example, make sure that the
connection tube from the sampling probe is guided vertically upward for at least 0.5
m (19.6 in).
1 Mounting pipe, 2 m (78.7 in) 3 Retaining clamp
2 EPDM flat seal 4 Cylinder-head screw, M5 × 20 (3×)
16
Installation
3.1.2 Second mounting point and extension pipe
Figure 7 Second mounting point LZX456
1 Mounting pipe, 2 m (78.7 in) 3 Retaining clamp
2 EPDM flat seal 4 Cylinder-head screw, M5 × 20 (3×)
17
Installation
Figure 8 Installation with extension pipe.
NOTICE
When carrying out an installation variant that deviates from this example, make sure that the
connection tube from the sampling probe is guided vertically upward for at least 0.5
m (19.6 in).
1 Mounting pipe, 2 m (78.7 in) 4 Retaining clamp
2 EPDM flat seal 5 Cylinder-head screw, M5 × 20 (3×)
3 Extension pipe
18
Installation
3.1.3 Install the open loop control unit
Make sure that an area of 5 cm is kept free around the entire instrument to enable air to
circulate (distance from the wall of the building, cable channel etc.).
NOTICE
Install the open loop control unit in a dry location out of direct sunlight.
NOTICE
The Teflon sample lines must be laid without sharp bends. Shortening or extending the lines will
lead to problems when the samples are transported.
Each process instrument must have a separate drain tube!
The individual device drains may only be combined in a larger drain pipe with regular water
throughput.
19
Installation
Figure 9 Open loop control unit and process instrument
1 SIGMATAX 2 open loop control unit 5 Separate drain pipe for process
instrument
9 Drain tube for process instrument
2 Fiber optic cable to the process
instruments
6 SIGMATAX 2 connection tube 10 SIGMATAX 2 drain tubes
3 PHOSPHAX sigma or TOCTAX 7 SIGMATAX 2 mains connection
cable
4 Shared drain pipe 8 Teflon sample line
20
Installation
3.1.4 Prepare the connection tube
Figure 10 Connection tube
DANGER
The pipe trace heater in the connection tube is operated at 230 V. If any damage occurs to the
tube casing, the entire system must be taken out of operation immediately.
1 Protective tube 3 Connection tube
2 Heat-shrink tube 4 PG 16 screw connection
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44

Hach SIGMATAX 2 User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI