Thermo Fisher Scientific 84i Owner's manual

Type
Owner's manual
Model 84i
Hg Permeation Source
Instruction Manual
Part Number 114051-00
24Mar2016
© 2013 Thermo Fisher Scientific Inc. All rights reserved.
Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please
consult your local sales representative for details.
Thermo Fisher Scientific
Air Quality Instruments
27 Forge Parkway
Franklin, MA 02038
1-508-520-0430
www.thermo.com/aqi
Thermo Fisher Scientific WEEE Compliance
WEEE Compliance
This product is required to comply with the European Union’s Waste
Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is
marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more
recycling/disposal companies in each EU Member State, and this product
should be disposed of or recycled through them. Further information on
Thermo Fisher Scientific’s compliance with these Directives, the recyclers
in your country, and information on Thermo Fisher Scientific products
which may assist the detection of substances subject to the RoHS Directive
are available at: www.thermo.com/WEEERoHS.
Thermo Fisher Scientific Model 84i Instruction Manual i
About This Manual
This manual provides information about installing, maintaining, and
servicing the Model 84i Hg Permeation Source. It also contains important
alerts to ensure safe operation and prevent equipment damage. The manual
is organized into the following chapters and appendixes to provide direct
access to specific operation and service information.
Chapter 1 “Introduction provides an overview of the product,
describes the principle of operation, and lists the specifications.
Chapter 2 “Installation” describes how to install and connect the Hg
Permeation Source into the Thermo Scientific Mercury System.
Chapter 3 “Operation” describes the associated menu-driven software
and permeation/generator calculation.
Chapter 4 “Preventive Maintenance, Troubleshooting and Servicing
presents safety alerts for technicians working on the instrument,
preventive maintenance information, troubleshooting tips, and
component replacement information. It also includes contact
information for product support and technical information.
Appendix A “Warranty” is a copy of the warranty statement.
Appendix B “C-Link Protocol Commands” provides a list of the 84i
C-Link protocol commands that can be used to remotely control an
analyzer using a host device such as a PC or datalogger.
Appendix C “MODBUS Protocol” provides details of the MODBUS
Protocol for the 84i.
Review the following information carefully before using the Model 84i.
This manual provides specific information on how to operate the
instrument, however if the instrument is used in a manner not specified by
the manufacturer, the protection provided by the equipment may be
impaired.
Safety
About This Manual
Safety and Equipment Damage Alerts
ii Model 84i Instruction Manual Thermo Fisher Scientific
This manual contains important information to alert you to potential safety
hazards and risks of equipment damage. Refer to the following types of
alerts you may see in this manual.
Safety and Equipment Damage Alert Descriptions
Alert Description
DANGER A hazard is present that will result in death or serious
personal injury if the warning is ignored.
WARNING A hazard is present or an unsafe practice can result in
serious personal injury if the warning is ignored.
CAUTION The hazard or unsafe practice could result in minor to
moderate personal injury if the warning is ignored.
Equipment Damage The hazard or unsafe practice could result in property
damage if the warning is ignored.
Safety and Equipment Damage Alerts in this Manual
Alert Description
WARNING The service procedures in this manual are restricted to
qualified service personnel only.
If the equipment is operated in a manner not specified by
the manufacturer, the protection provided by the
equipment may be impaired.
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly ground antistatic
wrist strap must be worn while handling any internal
component. If an antistatic wrist strap is not available, be
sure to touch the instrument chassis before touching any
internal components. When the instrument is unplugged,
the chassis is not at earth ground.
Safety and Equipment
Damage Alerts
About This Manual
WEEE Symbol
Thermo Fisher Scientific Model 84i Instruction Manual iii
The following symbol and description identify the WEEE marking used on
the instrument and in the associated documentation.
Symbol Description
Marking of electrical and electronic equipment which applies to electrical and
electronic equipment falling under the Directive 2002/96/EC (WEEE) and the
equipment that has been put on the market after 13 August 2005.
Service is available from exclusive distributors worldwide. Contact one of
the phone numbers below for product support and technical information
or visit us on the web at www.thermoscientific.com/aqi.
1-866-282-0430 Toll Free
1-508-520-0430 International
We continue to support our customers with advanced online resources.
Our Air Quality Instruments Online Library allows our customers access to
product documents and information on a constant basis.
Available 24-hours a day and seven-days a week, the online library provides
quick access to information regardless of time zone or office hours.
To register for an account or log in, please visit
www.thermoscientific.com/aqilibrary.
W
EEE Symbol
W
here to Get Help
About This Manual
Where to Get Help
iv Model 84i Instruction Manual Thermo Fisher Scientific
Thermo Fisher Scientific Model 84i Instruction Manual v
Contents
Introduction ........................................................................................................ 1-1
Principle of Operation ........................................................................ 1-1
Component Description ..................................................................... 1-3
Permeation Oven ............................................................................. 1-4
Capillary Block Assembly ................................................................. 1-4
Pneumatic Assembly ........................................................................ 1-5
Manifold .......................................................................................... 1-6
Valves ............................................................................................... 1-6
Scrubber ........................................................................................... 1-7
Flow Switch ..................................................................................... 1-7
Pressure Sensor ................................................................................. 1-7
Regulators ........................................................................................ 1-7
Ambient Temperature Thermistor ................................................... 1-7
84i Control Board ............................................................................ 1-7
Power Receptacle ............................................................................. 1-7
LEDs ................................................................................................ 1-7
RS 485 Link ..................................................................................... 1-8
Specifications ...................................................................................... 1-8
Installation ......................................................................................................... 2-1
Equipment Required ........................................................................... 2-1
System Preparation .............................................................................. 2-2
Sequence of Work, Mechanical ........................................................... 2-2
Sequence of Work, Software ............................................................... 2-9
Operation ............................................................................................................ 3-1
Turning On the Permeation Option ................................................... 3-2
Gas Modes Menu ................................................................................ 3-3
84i Permeation Oven Settings Menu .................................................. 3-4
Permeation Oven Enable .................................................................. 3-4
Permeation Gas Setpoint .................................................................. 3-4
Capillary Setpoint ............................................................................ 3-5
84i Connection B ............................................................................. 3-5
Temperature Diagnostics .................................................................... 3-6
84i Perm Gas Temperature .............................................................. 3-6
84i Perm Heater Temperature ......................................................... 3-6
84i Capillary Temperature ............................................................... 3-6
84i Ambient Temperature ................................................................ 3-6
Voltage Diagnostics ............................................................................. 3-6
Chapter 1
Chapter 2
Chapter 3
Contents
vi Model 84i Instruction Manual Thermo Fisher Scientific
Alarms ................................................................................................. 3-7
Alarms Detected ............................................................................... 3-7
Permeation Gas Temperature ........................................................... 3-7
Permeation Heater Temperature ...................................................... 3-7
Capillary Temperature ..................................................................... 3-7
Pressure ............................................................................................ 3-7
Flow ................................................................................................. 3-8
Board Status ..................................................................................... 3-8
Performing an Audit using the Auto Permeation Schedule .................. 3-8
Schedule ........................................................................................... 3-9
Next Time ....................................................................................... 3-9
Period Days ...................................................................................... 3-9
Total Duration ................................................................................. 3-9
INST Zero Duration Minutes .......................................................... 3-9
INST Span Duration Minutes ......................................................... 3-9
PERM Span Duration Minutes ........................................................ 3-9
INST Hg Span Level ........................................................................ 3-9
Zero/Span Averaging Seconds ........................................................ 3-10
System Configurations ...................................................................... 3-10
Configuration for Auditing One Mercury Freedom System ........... 3-10
Configuration for Auditing Two Mercury Freedom Systems using
Connections A and B ..................................................................... 3-12
General Procedure for Running an Audit .......................................... 3-13
Running an Audit on the Primary Mercury Freedom System ............ 3-14
Running an Audit on a Secondary Mercury Freedom System ............ 3-17
Permeation/Generator Ratio ............................................................. 3-19
Obtaining the Base Perm/Gen Ratio ................................................. 3-19
Data Logging .................................................................................... 3-21
Shutting Off the 84i .......................................................................... 3-21
Stand-alone Configuration ................................................................ 3-21
Preventive Maintenance, Troubleshooting and Servicing ....................... 4-1
Safety Precautions ............................................................................... 4-1
Replacement Parts List ........................................................................ 4-2
Hg Scrubber Replacement ................................................................... 4-3
Fan and Fan Filter Guard Replacement ............................................... 4-4
DC Power Supply Replacement .......................................................... 4-5
Solenoid Valve Replacement ............................................................... 4-6
Permeation Oven Replacement ........................................................... 4-7
Pressure Transducer Assembly Replacement ...................................... 4-10
Plumbing Flow Splitter Assembly Replacement ................................. 4-11
Precision Regulator Assembly Replacement ....................................... 4-14
Measurement Interface Board Replacement ...................................... 4-16
LED PCB Replacement .................................................................... 4-18
Capillary Block Heater Assembly Replacement ................................. 4-19
Service Locations ............................................................................... 4-20
Chapter 4
Contents
Thermo Fisher Scientific Model 84i Instruction Manual vii
Warranty ............................................................................................................ A-1
C-Link Protocol Commands ............................................................................ B-1
MODBUS Protocol ............................................................................................ C-1
Appendix A
Appendix B
Appendix C
Contents
viii Model 84i Instruction Manual Thermo Fisher Scientific
Thermo Fisher Scientific Model 84i Instruction Manual ix
Figures
Figure 1–1. Model 84i Hg Permeation Source Flow Diagram ........................... 1-2
Figure 1–2. Model 84i Component Layout ......................................................... 1-3
Figure 1–3. Permeation Oven ............................................................................. 1-4
Figure 1–4. Capillary Block Assembly ................................................................ 1-5
Figure 1–5. Pneumatic Assembly ....................................................................... 1-6
Figure 2–1. 84i Mounting Location .................................................................... 2-3
Figure 2–2. Slide Rail Mounting Holes .............................................................. 2-3
Figure 2–3. 84i Permeation Source Mounted in Rack ....................................... 2-4
Figure 2–4. 81i CDA Line Removed .................................................................... 2-5
Figure 2–5. 81i CDA Fitting Installation ............................................................. 2-6
Figure 2–6. Retrofitted 81i CDA Plumbing to Support 84i ................................ 2-6
Figure 3–1. Tubing and Electrical Configuration .............................................. 3-11
Figure 3–2. RS-485 Connections Diagram ....................................................... 3-12
Figure 3–3. Physical and Electrical Connections for auditing two Mercury
Freedom Systems using one 84i Permeation Source ........................................ 3-13
Figure 3–4. Typical Quality Assurance Audit Graph ........................................ 3-20
Figure 3–5. Typical Correlation between the averaged Base Perm/Gen Ratio and
Subsequent Perm/Gen Ratios ............................................................................ 3-20
Figure 3–6. Digital In Connection Diagram ...................................................... 3-22
Figure 4–1. Replacing the Fan ............................................................................ 4-4
Figure 4–2. Replacing the DC Power Supply ...................................................... 4-5
Figure 4–3. Permeation Oven Removal .............................................................. 4-8
Figure 4–4. Removing and Replacing the Glass Chamber Assembly ................ 4-9
Figure 4–5. Replacing the Pressure Transducer Assembly .............................. 4-11
Figure 4–6. Flow Splitter Assembly ................................................................. 4-12
Figure 4–7. Capillary Block Assembly .............................................................. 4-13
Figure 4–8. Pneumatic Assembly ..................................................................... 4-15
Figure 4–9. Model 84i Component Layout ....................................................... 4-17
Figure B–1. Flag Status ...................................................................................... B-6
x Model 84i Instruction Manual Thermo Fisher Scientific
Thermo Fisher Scientific Model 84i Instruction Manual xi
Tables
Table 1–1. 84i Valve Configurations .................................................................. 1-6
Table 1–2. 84i LEDs ............................................................................................ 1-7
Table 1–3. Model 84i Specifications ................................................................. 1-8
Table 2–1. 84i Hg Permeation Source Parts List ................................................ 2-1
Table 4–1. Model 84i Hg Permeation Source Replacement Parts .................... 4-2
Table C–1. Read Coils for 84i ............................................................................. C-1
Table C–2. Read Registers for 84i ..................................................................... C-1
Table C–3. Write Coils for 84i ............................................................................ C-2
xii Model 84i Instruction Manual Thermo Fisher Scientific
Thermo Fisher Scientific Model 84i Instruction Manual 1-1
Chapter 1
Introduction
The Thermo Scientific™ 84i Hg Permeation Source compliments the
Mercury Freedom System by periodically assessing and qualifying the
constancy of the 81i’s mercury concentration output. The 80i Mercury
Analyzer is used to control the functions of the 84i. An option in the model
80i software has been added to allow the user to access and manage the 84i
Hg Permeation Source. For details see the following topics:
Principle of Operation” on page 1-1
Component Description” on page 1-3
Specifications” on page 1-8
The Model 84i Hg Permeation Source uses a process by which a specific
and consistent concentration of Mercury is generated from a permeation
assembly (Figure 1–1). The Hg permeation tube generates a known and
reliable concentration of mercury at a constant temperature and flow. The
84i is located in the Mercury System rack. The generated 84i mercury
concentration, as measured by the 80i Mercury Analyzer, will be used to
confirm the reliability of the 81i Mercury Calibrator output in accord with
requirements.
In compliance with the “Interim Traceability Protocol for Qualification
and Certification of Elemental Mercury Gas Generators”, periodic quality
assessments of Mercury Gas Generators are required following initial
certification of an elemental Hg Generator.
Principle of
Operation
Introduction
Principle of Operation
1-2 Model 84i Instruction Manual Thermo Fisher Scientific
Figure 1–1. Model 84i Hg Permeation Source Flow Diagram
Introduction
Component Description
Thermo Fisher Scientific Model 84i Instruction Manual 1-3
This section describes the function and location of the 84i components.
Refer to Figure 1–2 to locate the instrument components.
Figure 1–2. Model 84i Component Layout
Component
Description
Pneumatic
Assembly
LED Display
Permeation Oven
Capillary Bloc
k
Power
Supply
PCB Board
Introduction
Component Description
1-4 Model 84i Instruction Manual Thermo Fisher Scientific
The Permeation Oven supplies a reliable and consistent mercury
concentration. It contains a permeation tube, permeation oven heater, oven
thermistor and a gas thermistor. When held at a constant temperature the
permeation tube outgases a steady amount of mercury that is transported to
the 80i Mercury Analyzer. A fixed and continuous flow rate and tight
control of the permeation oven temperature assures a repeatable and stable
mercury concentration.
Figure 1–3. Permeation Oven
The Heated Capillary Block maintains a consistent and reliable flow
through the permeation oven. The heated Capillary Block Assembly (84i
Flow Splitter) contains two capillaries, a heater, thermistor and an inline
screen. Both capillaries receive Zero Air supplied from the rear panel. Air
flow going through the 0.004-inch sized capillary leads into the permeation
oven. The air flow from the 0.006-inch capillary acts as permeation
dilution air and recombines with the Hg permeation air flow post
permeation oven. The Capillary Block Assembly is kept at a constant
temperature. Also, the precision pressure regulator, located upstream in the
Pneumatic Assembly, supplies the Capillary Block with air at a constant
pressure. With constant temperature and pressure, the flows of the two air
streams remain constant giving a stable and continuous mercury
concentration.
Permeation Oven
Capillary Block
Assembly
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86

Thermo Fisher Scientific 84i Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI