11. Function mode
F00: Simply pass through.
F01: UV lamp lighting time set (initial: 000)
(UV lamp is optionally provided)
F02: UV lamp 24hours sterilization mode set
(UV lamp is optionally provided)
F03: Service code input (code: 384)
F04: UV lamp life span check (UV lamp is optionally provided)
F05: Displays the CO
2
/O
2
sensor voltage
F06: Changes the bottom heater output ratio (initial:5)
F07: Changes the door heater output ratio (initial:6)
F08: Auto changer, Auto Zero mode set (initial : 000)
F10: Displays a coefficient for CO
2
zero point adjustment
F11: Displays a coefficient for CO
2
span point adjustment
F12: Displays a ambient temperature sensor temperature
F13: Initializes non volatile memory
F14: LEDs and drivers performance check
F15: Displays a temperature zero point adjustment value (initial: 008)
F21: Communication device address (ID) set (initial: 000)
F22: Communication parameter set (initial: 000)
F24: Interlocks the remote alarm with buzzer (initial: 000)
F25: Ring Back time set (initial: 030)
F26: Displays a coefficient for O
2
zero point adjustment (low density side)
F27: Displays a coefficient for O
2
span point adjustment (low density side)
F28: Displays a coefficient for O
2
zero point adjustment (high density side)
F29: Displays a coefficient for O
2
span point adjustment (high density side)
F30: Displays a ROM version
Direction for use: In PV display, press CAL key for more than 5seconds to enter the device
into temperature calibration mode. Input a desired function code with
key and key then press ENT key.
F01: UV lamp lighting time set (initial: 0min, settable range:0~30min)
Note) UV lamp is optionally provided.
<Direction
for use>
Input “F01” on the temperature display then press ENT key to show a present SV
on the CO
2
display. Input the desired value with key and key then press
ENT key to store the value and reverts to PV.
If a value out of the range is input then ENT key is pressed, the buzzer sounds
for a second to keep the previous setting.
<Action> After the door open-and-shut, UV lamp is lit in appointed time then automatically
goes off. There is approx. 3~5seconds delay to active the mode, due the device
would check water level in the humidifying pan just after the door shut.
If the door is opened during UV lamp is lit, the lamp goes off.
F02: UV lamp 24hours sterilization mode set
Note) UV lamp is optionally provided.
<Direction
for use>
Input “F02” on temperature display to show “000” on CO
2
display.
Input “001” with key and key to enter the device into 24hours sterilization
mode.
Press ENT key when “000” is displayed, simply reverts to PV display.
<Action> When 24hours sterilization mode is set in F02, UV lamp is lit and automatically
goes off after 24hours has passed. UV lamp goes off and 24hours sterilization
mode would be cancelled if the door is opened during the sterilization.
If 24hours sterilization mode is reset during the sterilization, it would restart
24hours sterilization mode at the time.
Note) If this mode is set, UV indicator lamp should be lit even if UV lamp is not fitted.