Invacare Platinum 10 User manual

Category
Oxygen Equipment
Type
User manual

This manual is also suitable for

Service Manual
DEALER: Keep this manual. The procedures in this manual
MUST be performed by a qualified technician.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
Platinum™Series
Platinum 5
Platinum 10
Platinum XL - 5 Liter
Oxygen Concentrator
Model IRC5LX, IRC10LX (HomeFill™II Compatible)
Model IRC5LXO
2
, IRC10LXO
2
(With SensO
2
®and HomeFill II Compatible)
ACCESSORIES
Platinum™Series 2 Part No 1110538
WARNING
DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND
UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND
THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT INVACARE
TECHNICAL SERVICES BEFORE ATTEMPTING TO USE THIS EQUIPMENT -
OTHERWISE SERIOUS INJURY OR PROPERTY DAMAGE MAY RESULT.
ACCESSORIES
There are many different types of humidifiers, oxygen tubing, cannulas and masks that
can be used with this device. You should contact your local home care provider for
recommendations on which of these devices will be best for you. They should also give
you advice on the proper usage, maintenance, and cleaning.
WARNING
Invacare products are specifically designed and manufactured for use in conjunction
with Invacare accessories. Accessories designed by other manufacturers have not
been tested by Invacare and are not recommended for use with Invacare products.
TABLE OF CONTENTS
Part No 1110538 3 Platinum™Series
TABLE OF CONTENTS
ACCESSORIES ................................................................................... 2
SPECIAL NOTES ................................................................................ 7
FEATURES ........................................................................................ 8
TYPICAL PRODUCT PARAMETERS .................................................... 9
SECTION 1—IMPORTANT SAFEGUARDS .......................................... 11
Important Safeguards...............................................................................................................................11
Maintenance...............................................................................................................................................12
Radio Frequency Interference ...............................................................................................................12
To Reduce the Risk of Burns, Electrocution, Fire or Injury to Persons......................................12
SECTION 2—INSTALLATION/SEQUENCE OF OPERATION .................. 13
Installation/Verification of Battery Free Power Loss Alarm...........................................................13
Sequence of Operation ...........................................................................................................................13
SensO
2
Oxygen Sensor Technology - Ceramic Zirconia Sensor .................................................14
Technical Description .........................................................................................................................14
Operating Sequence ............................................................................................................................14
SECTION 3—PNEUMATIC DIAGRAM ............................................... 16
SECTION 4—TROUBLESHOOTING .................................................... 17
SECTION 5—CABINET .................................................................... 25
Removing Cabinet ....................................................................................................................................25
SECTION 6—PREVENTIVE MAINTENANCE ...................................... 26
Cleaning the Cabinet Filter ....................................................................................................................26
Replacing the Outlet HEPA Filter.........................................................................................................27
Check the Outlet HEPA Filter..........................................................................................................27
Replace the Outlet HEPA Filter before 06\09\06 ........................................................................27
Replace the Outlet HEPA Filter after 06\09\06............................................................................28
Replacing the Compressor Inlet Filter.................................................................................................28
Replacing the Exhaust Canister/Exhaust Muffler (Platinum 5 Models Manufactured before 7/
24/03) .................................................................................................................................................................29
Replacing the Throttling Muffler ...........................................................................................................31
Cleaning the Heat Exchanger.................................................................................................................31
TABLE OF CONTENTS
Platinum™Series 4 Part No 1110538
TABLE OF CONTENTS
SECTION 7—COMPRESSOR .............................................................. 33
Replacing Compressor Assembly .........................................................................................................33
Replacing Capacitor .................................................................................................................................35
5-Liter Models (Platinum 5 and Platinum XL) ...............................................................................35
10-Liter Model (Platinum 10) ............................................................................................................36
Rebuilding the Thomas Model 2650 and 2660 Compressor - Platinum 5 and 10 Models......37
Removing Compressor and Compressor Heads .........................................................................38
Rebuilding the Compressor...............................................................................................................41
Platinum XL Model - Rebuilding the Gast Compressor..................................................................45
Removing the Head Cover and Valve Plate Assembly.....................................................................46
Removing Head Cover Gaskets, O-Ring Gaskets, Exhaust Valves and Limiter ........................46
Installing the New Intake Valves, Exhaust Valves With Limiters...................................................48
Removing the Existing Retainer Plate, Cups and Cylinders............................................................49
Installing the New Head Cover Gaskets and O-Ring Gaskets ......................................................51
Installing the Head Cover.......................................................................................................................52
SECTION 8—P.E. VALVE .................................................................. 53
Replacing P.E. Valve..................................................................................................................................53
SECTION 9—SIEVE BEDS ................................................................. 54
Replacing Sieve Beds................................................................................................................................54
SECTION 10—CHECK VALVES ......................................................... 56
Replacing Check Valves...........................................................................................................................56
SECTION 11—REGULATOR .............................................................. 58
Replacing Regulator..................................................................................................................................58
Adjusting Regulator..................................................................................................................................60
SECTION 12—HEAT EXCHANGER ................................................... 61
Replacing Heat Exchanger ......................................................................................................................61
SECTION 13—CONTROL PANEL ...................................................... 62
Removing Control Panel.........................................................................................................................62
SECTION 14—COOLING FAN ........................................................... 64
Replacing Cooling Fan .............................................................................................................................64
SECTION 15—P.C. BOARD .............................................................. 66
Replacing P.C. Board................................................................................................................................66
SECTION 16—TRANSFORMER ......................................................... 68
Replacing Transformer Assembly .........................................................................................................68
TABLE OF CONTENTS
Part No 1110538 5 Platinum™Series
TABLE OF CONTENTS
SECTION 17—ON/OFF SWITCH ......................................................... 69
Replacing On/Off Switch.........................................................................................................................69
SECTION 18—FLOWMETER .............................................................. 70
Replacing Flowmeter ...............................................................................................................................70
SECTION 19—HOUR METER ............................................................ 72
Replacing Hour Meter.............................................................................................................................72
SECTION 20— 4-WAY VALVE .......................................................... 73
Replacing 4-Way Valve............................................................................................................................73
Replacing Manifold Assembly (Units Manufactured before 7/24/03)............................................74
Replacing Manifold Assembly (Units Manufactured after 7/24/03) ...............................................77
Replacing Pilot Valve Poppets and O-Rings........................................................................................80
SECTION 21—PURITY INDICATORS ................................................ 83
SensO
2
Alarm Indicators ........................................................................................................................83
Operation ..............................................................................................................................................83
Indicator Lights .....................................................................................................................................83
Indicators - If your unit does not feature the O
2
Sensor ...............................................................84
LX Models Indicator Light Explanation...........................................................................................84
SECTION 22—TIMING ..................................................................... 85
Timing the Platinum 5, XL, and 10 Concentrators...........................................................................85
Autotuning.............................................................................................................................................85
Manual Tuning.......................................................................................................................................87
SECTION 23—LEAK TEST ................................................................ 89
Leak Test ....................................................................................................................................................89
SECTION 24— ALARM TEST ............................................................ 91
Alarm and Sensor Tests..........................................................................................................................91
Power Loss Alarm ...............................................................................................................................91
Low Pressure Test...............................................................................................................................91
Time-Out Test......................................................................................................................................92
High Pressure Test ..............................................................................................................................92
Oxygen Sensor .....................................................................................................................................93
P.E. Valve Coil Test .............................................................................................................................94
Platinum™Series 6 Part No 1110538
SECTION 25— SIEVE BED PRESSURE CHECK ..................................... 96
Installing the Sieve Bed Pressure Tap-In Kit Installation and Checking Bed Pressure .............96
Disassembly ...........................................................................................................................................96
Installation..............................................................................................................................................97
Pressure Testing...................................................................................................................................97
After Testing .........................................................................................................................................98
LIMITED WARRANTY ..................................................................... 99
SPECIAL NOTES
Part No 1110538 7 Platinum™Series
SPECIAL NOTES
Signal words are used in this manual and apply to hazards or unsafe practices which
could result in personal injury or property damage. Refer to the table below for
definitions of the signal words.
NOTICE
THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO
CHANGE WITHOUT NOTICE.
DANGER
Users MUST not smoke while using this device. Keep all matches, lighted cigarettes
or other sources of ignition out of the room in which this product is located. NO
SMOKING signs should be prominently displayed. Textiles and other materials that
normally would not burn are easily ignited and burn with great intensity in oxygen
enriched air. Failure to observe this warning can result in severe fire, property
damage, and cause physical injury or death.
SIGNAL WORD MEANING
DANGER
Danger indicates an imminently hazardous situation which,
if not avoided, will result in death or serious injury.
WARNING
Warning indicates a potentially hazardous situation which, if
not avoided, could result in death or serious injury.
CAUTION
Caution indicates a potentially hazardous situation which, if
not avoided, may result in property damage.
FEATURES
Platinum™Series 8 Part No 1110538
FEATURES
Oxygen Purity and Fault
Indicator Lights (SensO
2
model)/Fault and Power
Indicator Lights
(LX model)
Circuit Breaker
Power Switch
Hour Meter
Cabinet Filter
(one of two
shown)
Flowmeter
Oxygen Outlet
FRONT VIEW
TOP VIEW
ACCESSORIES (NOT SHOWN):
Remote Humidifier/Flowmeter Accessory -
IRCRF5
PreciseR
X
™ Pediatric Humidifier/ Flowmeter
Accessory - IRCPF16
Homefill II home oxygen compressor - IOH200
*NOTE: This outlet fitting is to be used only for filling oxygen cylinders with the
HomeFill II home oxygen compressor. The outlet fitting does not affect concentrator
performance. Refer to the HomeFill II owner's manual, part number 1100873, for
connection and operating instructions. When not in use, the plug provided with the
concentrator should be inserted into the outlet fitting.
Power Cord
Outlet Fitting*
TYPICAL PRODUCT PARAMETERS
Part No 1110538 9 Platinum™Series
TYPICAL PRODUCT PARAMETERS
ELECTRICAL REQUIREMENTS: 120 VAC ± 10% (132 VAC/108 VAC), 60 Hz
RATED CURRENT INPUT: Platinum 5 and XL: 4.3 A
Platinum 10: 4.7 A
SOUND LEVEL: Platinum 5 and XL: 50 dB Average
Platinum 10: 58 dB Average
ALTITUDE: Platinum 5 and XL - Up to 6,000 ft. (1828 meters) above sea
level without degradation of concentration levels. From 6,000
ft. (1828 meters) to 13,129 ft. (4000 meters) below 90%
efficiency
Platinum 10 - Up to 4,000 ft. (1230 meters) above sea level
without degradation of concentration levels. From 4,000 ft.
(1230 meters) to 13,129 ft. (4000 meters) below 90% efficiency.
*OXYGEN OUTPUT
CONCENTRATION LEVELS:
All 5LXO2/5LX models 95.6% to 87% at 0.5 to 5 L/min. and all
10LXO2/10LX models 95.6% to 87% at 2 L to 10 L/min.
* Stated concentration levels achieved after initial warm-up
period (approximately 30 minutes).
MAXIMUM OUTLET PRESSURE: Platinum 5 and XL: 5 ± 0.5 psi. (34.5 ± 3.45 kPa)
Platinum 10: 9 psi ± 0.5 psi. (62.0 kPa ± 3.45 kPa)
FLOW RANGE: Platinum 5 and XL - 0.5 to 5 L/min. For flowrates less than 1
L/min., we recommend the use of the Invacare Pediatric
Flowmeter IRCPF16 Accessory.
Platinum 10 - 2 L/min. to 10 L/min.
LOW FLOW ALARM: Platinum 5 - 0 L/min. to 0.5 L/min.
Platinum 10 - 0 L/min. to 1 L/min.
Rapid Audible Alarm Beeping (No Accessories Connected).
AVERAGE POWER CONSUMPTION: Platinum 5 and XL: 400W
Platinum 10: 585W
PRESSURE RELIEF MECHANISM
OPERATIONAL AT:
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa).
CHANGE IN MAXIMUM
RECOMMENDED FLOW WHEN BACK
PRESSURE OF 7K PA IS APPLIED:
0.7 L/min.
FILTERS: Cabinet (2), Outlet HEPA and Compressor Inlet.
SAFETY SYSTEM: Current overload or line surge shutdown. High temperature
compressor shutdown. High Pressure Alarm w/compressor
shutdown. Low Pressure Alarm w/compressor shutdown.
Battery Free Power Loss Alarm. SensO
2
Oxygen System
(SensO
2
Model) Low Flow Alarm.
WIDTH: 18 3/8 in. ± 3/8 in. (46.7 cm ± 1 cm).
HEIGHT: 26 3/8 in. ± 3/8 in. (67.0 cm ± 1 cm).
DEPTH: 14 3/8 in. ± 3/8 in. (36.5 cm ± 1 cm).
TYPICAL PRODUCT PARAMETERS
Platinum™Series 10 Part No 1110538
WEIGHT: Platinum 5 and XL: 52 lbs ± 2 lbs (23.6 kg ± 1 kg).
Platinum 10: 53 lbs ± 2 lbs (24.0 kg ± 1 kg).
SHIPPING WEIGHT: Platinum 5 and XL: 58 lbs ± 2 lbs (26.3 kg ± 1 kg).
Platinum 10: 60 lbs ± 2 lbs (27.2 kg ± 1 kg).
OPERATING AMBIENT
TEMPERATURE:
50°F - 95°F (10°C - 35°C) at 20-60% relative humidity.
EXHAUST: Platinum 5 and XL: Less than Ambient + 35°F (+19°C).
Platinum 10: Less than Ambient + 45°F (+25°C).
OXYGEN OUTPUT:
Platinum 5 and XL: Less than Ambient + 4°F (+2°C).
Platinum 10: Less than Ambient + 8°F (+4.4°C).
CABINET: Impact Resistant flame-retardant plastic cabinet that conforms
to UL 94-V0.
REGULATORY LISTING: Double Insulated Product ETL certified complying with UL 1097
ETL certified to CSA C22.2 No. 68.
ELECTRICAL: No extension cords.
PLACEMENT: No closer than three inches from any wall, furniture, draperies,
or similar surfaces.
TUBING: Platinum 5 and XL: 7 ft cannula with a maximum 50 ft of
Crush-Proof Tubing (DO NOT pinch).
Platinum 10: Recommended use up to 50 ft high flow tubing
with high flow cannula at all flow rates.
RELATIVE HUMIDITY: 20 to 60%.
TIME OF OPERATION: Up to 24 hours per day.
RECOMMENDED STORAGE AND
SHIPPING TEMPERATURE:
-20°F to 150°F (-29°C to 65°C), at 15-95% relative humidity.
LABEL
SYMBOL
STATUS SENSO
2
MODELS INDICATOR LIGHTS (LED)
SYSTEM OKAY
O
2
over 85% (± 2%)
GREEN Indicator Light
O
2
Between 73% (± 3%) to 85%
(±2%)
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor Failure
Call a qualified technician.
SYSTEM FAILURE
O
2
Below 73% (±3%)
RED Indicator light
Continuous Audible Alarm Sieve-GARD™Compressor Shutdown
Call a qualified technician
LABEL
SYMBOL
STATUS LX MODEL INDICATOR LIGHTS (LED)
SYSTEM OKAY GREEN Indicator Light
SYSTEM FAILURE RED Indicator light
Continuous Audible Alarm Sieve-GARD™ Compressor Shutdown
Call a qualified technician
O
2
I/O
SECTION 1—IMPORTANT SAFEGUARDS
Part No 1110538 11 Platinum™Series
SECTION 1—IMPORTANT
SAFEGUARDS
WARNING
IMPORTANT SAFEGUARDS contains important information for the safe
operation and use of this product. DO NOT use this product or any available
optional equipment without first completely reading and understanding these
instructions and any additional instructional material such as Owner’s Manuals,
Service Manuals or Instruction Sheets supplied with this product or optional
equipment. If you are unable to understand the Warnings, Cautions or Instructions,
contact a healthcare professional, dealer or technical personnel before attempting
to use this equipment - otherwise, injury or damage may occur.
Important Safeguards
The use of oxygen therapy requires that special care be taken to reduce the risk of fire. Any
materials that will burn in air, and some that will not, are easily ignited and burn rapidly
in high concentrations of oxygen. For safety concerns, keep all sources of ignition out of
the room in which this product is located and away from areas where oxygen is being
delivered. Textiles, oil and other combustibles are easily ignited and will burn with great
intensity in oxygen-enriched air. NO SMOKING signs should be prominently displayed.
A spontaneous and violent ignition may occur if oil, grease or greasy substances come in
contact with oxygen under pressure. These substances MUST be kept away from the
oxygen concentrator, tubing and connections, and all other oxygen equipment. DO NOT
use any lubricants unless recommended by Invacare.
For optimum performance, Invacare recommends that each concentrator be on and
running for a minimum of 30 minutes at a time. Shorter periods of operation may be
harmful for maximum product life.
Keep the oxygen tubing, cord, and unit away from heated or hot surfaces, including space
heaters, electric blankets, stoves and similar electrical appliances.
DO NOT move or relocate concentrator by pulling on the power cord.
NEVER drop or insert any object into any opening.
NEVER block the air openings of the product or place it on a soft surface, such as a bed
or couch, where the air opening may be blocked. Keep the openings free from lint, hair
and the like.
Fill humidifier with water to the level shown by the manufacturer. DO NOT overfill.
Invacare recommends that Crush-Proof oxygen tubing be used with this product and
NOT exceed 50 ft. (15.2m) in length. Recommended use for Platinum 10 is up to 50 ft. in
length of high flow tubing.
SECTION 1—IMPORTANT SAFEGUARDS
Platinum™Series 12 Part No 1110538
Maintenance
The Platinum Oxygen Concentrator was specifically designed to minimize routine
preventive maintenance. Only professionals of the healthcare field or persons fully
conversant with this process, such as authorized or factory trained personnel, should
perform preventive maintenance or performance adjustments on the oxygen
concentrator.
Radio Frequency Interference
Most electronic equipment is influenced by Radio Frequency Interference (RFI).
CAUTION should be exercised with regard to the use of portable communications
equipment in the area around such equipment.
To Reduce the Risk of Burns, Electrocution, Fire or
Injury to Persons
DO NOT come in contact with the concentrator while you are wet.
DO NOT place or store product where it can drop into water or other liquids.
DO NOT reach for product that has fallen into water. Unplug IMMEDIATELY.
No adjustments should be made to the flowrate unless prescribed by a physician
therapist.
The product should NEVER be left unattended when it is plugged in.
Close supervision is necessary when this product is used by, on or near children or
physically challenged individuals.
The Platinum concentrator is intended to be used only as an oxygen supplement as
described in the Owner's Manual.
Invacare products are specifically designed and manufactured for use in conjunction with
Invacare accessories. Accessories designed by other manufacturers have not been tested
by Invacare and are not recommended for use with Invacare products.
DO NOT connect the concentrator in parallel or series with other oxygen concentrators or
oxygen therapy devices.
Use of some administration accessories or certain humidifiers not specified for use with
oxygen concentrator may impact the performance.
Avoid creation of any spark near medical oxygen equipment. This includes sparks from
static electricity created by any type of friction.
SECTION 2—INSTALLATION/SEQUENCE OF OPERATION
Part No 1110538 13 Platinum™Series
SECTION 2—INSTALLATION/
SEQUENCE OF OPERATION
Installation/Verification of Battery Free Power Loss
Alarm
When your new Platinum concentrator arrives, it should be checked for proper operating
conditions.
1. If the unit has been in below-freezing temperatures, allow it to warm up to room
temperature before operating.
2. The concentrator may need to be turned on for four to five seconds to charge the
Battery Free Power Loss Alarm. Connect power cord to outlet and turn the
concentrator on. Turn flow control knob counterclockwise and flow will begin
IMMEDIATELY. Set flow rate to five L/min. Let the unit run for 30 minutes, then turn
unit off.
3. Unplug the power cord and press on/off ( ) switch to the On ( ) position. An
intermittent audible alarm will sound. This confirms proper operation of the Battery
Free Power Loss Alarm. Turn On/Off switch Off ( ).
4. Connect power cord to outlet and turn on concentrator. Unit will beep on start-up.
5. Check the oxygen concentration per specifications after 30-40 minutes running time.
Sequence of Operation
Turning on (I) the power switch applies 120 VAC to the compressor motor, hour meter,
transformer, cooling fan and the printed circuit (P.C.) board.
Room air enters the compressor via the cabinet filter and the compressor inlet filter. The
air is compressed by the wobble pistons in the compressor to a pressure of 21 p.s.i. (144.79
kPa) for Platinum 5, XL and 25 p.s.i. (172.36 kPa) for Platinum 10.
As increased pressure creates increased temperature, a heat exchanger is used to lower
the temperature before the air enters the 4-way valve. It is then channeled to a sieve bed
containing the adsorption material. Restriction downstream of the sieve bed causes
pressure to build up inside the sieve bed which is necessary for the adsorption process. A
small amount of relatively pure oxygen enters the top of the second bed through a
restrictive leak in the pressure equalization (P.E.) valve with the balance entering a storage
tank. The nitrogen removed is exhausted back from the bed through the 4-way valve into
room air. A muffler is located at the exhaust end of the valve to muffle the sound of the
exhaust as it exits the concentrator.
The oxygen not being used to exhaust is channeled into the storage tank. The pressurized
oxygen is regulated down to five p.s.i. (34.4 kPa), enters an accurate flow-measuring
device, flows through Outlet HEPA filter and check valve, then out to the patient.
I/O
I
/
O
SECTION 2—INSTALLATION/SEQUENCE OF OPERATION
Platinum™Series 14 Part No 1110538
The electrical activation of the 4-way Valve is accomplished every 8 to 15 seconds by the
pressure sensor and P.C. Board electronics when the pressure reaches a set point of 21
p.s.i. (144.79 kPa) for Platinum 5 and XL; 25 p.s.i. (172.36) for Platinum 10. The time
between cycles is dependent on altitude, flow rate and internal environmental factors.
A P.E. valve opens just prior to the shift of the 4-way valve. This allows highly
concentrated oxygen to enter the just exhausted bed from the top. This additional
pressure allows the bed to start its cycle at a higher pressure. The P.E. valve will close just
after the shift of the 4-way valve.
If main power is lost, the Battery Free Power Loss Alarm will sound a short "BEEP", with
a long pause after. All units are equipped with a diagnostic alarm system that signals if
the pneumatic pressure or electrical systems malfunction. The Troubleshooting Guide in
this manual explains the alarm system signals and reasons, in detail, for your
convenience.
SensO
2
Oxygen Sensor Technology - Ceramic Zirconia
Sensor
Technical Description
The oxygen being produced by the concentrator flows out of the product tank and into
the flowmeter. A small flow of oxygen produced by the unit is sent through a precision
orifice to the oxygen sensor mounted on the printed circuit board.
As the oxygen enters the sensor, it passes through a screen and contacts the sensing disk.
Electric current flowing through a metal film resistor heats the disk in excess of 300°
C.
Oxygen molecules contact the electrode of the disk and pick-up extra electrons to become
oxygen ions. These oxygen ions are attracted to the electrode on the bottom of the zirconia
sensing disk. Because of the crystal structure of the zirconia, only oxygen ions can pass
through. When the oxygen ions reach the bottom electrode, the extra electrons are
released from the oxygen ions and oxygen molecules return to the air. The number of
electrons is directly related to the oxygen concentration. The electrons travel to the P.C.
board where they are counted and the oxygen concentration reading is calculated.
A microprocessor on the P.C. board contains software that interprets the signal being
received from the sensor. It compares the signal to clinically acceptable limits. Signals
outside of the clinically acceptable limits generate responses in the form of lights, audible
indicators, and/or system shut-down.
Operating Sequence
Once the power switch has been turned on ( ), the SensO
2
circuit will wait five minutes
for the concentrator to begin producing clinically acceptable oxygen and the oxygen
sensor to stabilize. The GREEN light will illuminate (indicating normal system operation)
while the oxygen sensor is warming up.
I
/
SECTION 2—INSTALLATION/SEQUENCE OF OPERATION
Part No 1110538 15 Platinum™Series
After five minutes, if the oxygen purity exceeds 85% ± 2%, the GREEN light will continue
to illuminate.
If the oxygen level is not above 85% ± 2% after the first five minutes, the system will
continue to monitor the O
2
and wait for a maximum of 30 minutes from start-up to reach
85% ± 2% before activating an alarm. Environmental factors such as low voltage, high
altitude, or age of the machine will affect the time required to reach 85% ± 2%.
If the oxygen level is not above 85% ± 2% within the first 30 minutes, the oxygen
concentration alarm sequence will activate and the unit will shut down.
When oxygen concentration is above 85% ± 2%, the sensor measures oxygen purity every
10 minutes. If a reading falls below 85% ± 2%, a YELLOW light will illuminate. If the
oxygen purity falls below 73% ± 3% the RED light/Alarm/Shut-Down mode will activate.
SECTION 3—PNEUMATIC DIAGRAM
Platinum™Series 16 Part No 1110538
SECTION 3—PNEUMATIC
DIAGRAM
Plantinum Oxygen Concentrator
Flowmeter
Ouput
HEPA Filter
Output
Check
Valve
Patient
Output
Oxygen Sensor
(SensO
2
models Only)
Pressure Sensor
P.S.I.
O
2
P.C. Board
Pressure
Regulator
Product
Tank
Check
Valves
Air Inlet
Cabinet
Filter(s)
Compressor
Inlet Filter
Sieve
Bed
Sieve
Bed
P.E. Valve
4-way
Valve
Compressor
Relief
Valve
Heat
Solenoid
Pilot
Exhaust
Muffler
LEGEND Pneumatic Connections Electrical Connections
SECTION 4—TROUBLESHOOTING
Part No 1110538 17 Platinum™Series
SECTION 4—TROUBLESHOOTING
SYMPTOM PROBABLE CAUSE SOLUTION
NORMAL OPERATION:
Internal Status Indicators:
RED: Off
GREEN: Off
Unit plugged in, power
switch on. Single beep on
start up.
No Problems System Okay.
POWER LOSS:
Internal Status Indicators:
RED: Off
GREEN: Off
Unit unplugged, power
switch on, alarm off.
No Problems. Battery Free Circuit drained. Plug in cord
and turn power switch ON (I) to recharge.
POWER LOSS:
Internal Status Indicators:
RED: Off
GREEN: Off
Unit plugged in, power
switch, alarm off, battery
Free circuit drained.
No power at outlet. Check electrical outlet with a table lamp or
voltmeter set on 50-200 VAC scale. If
outlet isn't working, check protective
device in home's electrical panel or consult
an electrician. Also ensure that unit is
properly plugged in. DO NOT use
extension cords. Move to another outlet
or circuit.
Power cord:
a. Frayed.
b. Broken or damaged spade.
c. Spade connector from power cord
loose or disconnected (inside back
of unit).
a. Reattach cord.
b. Replace power cord connectors on
plug.
c. Reattach connector.
SECTION 4—TROUBLESHOOTING
Platinum™Series 18 Part No 1110538
POWER LOSS:
(Continued)
Internal Status Indicators:
RED: Off
GREEN: Off
Unit plugged in, power
switch, alarm off. Battery
Free circuit drained
On/Off
( ) switch.
a. Disconnected wire.
b. Faulty switch.
Check all electrical connections to the
ON/OFF
( ) switch for any
disconnected wires.
If the concentrator does not come on at all
and wiring is intact, color code and remove
wires one at a time. Remove and replace
with new one. Transfer wires from old
switch to new switch one at a time to the
matching contact.
Circuit breaker tripped. Reset breaker.
NOTE: Breaker may trip to safeguard
concentrator during a power surge.
If breaker trips IMMEDIATELY, there is a
probable short in the unit. Check for
pinched or charred wires. If the breaker
does not trip, run unit for approximately
two hours. The circuit breaker should be
less than 10 ohms. If breaker trips again,
there is an internal problem. Contact
INVACARE TECHNICAL SERVICE at 1-
800-832-4707.
P.C. board.
a. P.C. board damaged.
b. Loose or damaged connector.
a. Replace P.C. Board. Refer to
Replacing P.C. Board
on page 66.
b. Repair or replace connector.
INTERNAL POWER
LOSS SENSO
2
:
Internal Status Indicators:
RED: Off
GREEN: Off
Alarm may or may not be
on.
Control Panel Indicators:
RED: Off
YELLOW: Off
GREEN: Off
Fan operates,
Compressor not
operating.
Transformer assembly.
a. Faulty.
b. Connector loose or disconnected.
c. Faulty wiring.
a. Replace. Refer to Replacing
Transformer Assembly on page 68.
b. Reattach connector.
c. Replace transformer assembly. Refer
to Replacing Transformer Assembly
on page 68.
P.C. board faulty. Replace P.C. board. Unit requires retiming
after P.C. board replacement. Refer to
Replacing P.C. Board
on page 66.
SYMPTOM PROBABLE CAUSE SOLUTION
I/O
I/O
SECTION 4—TROUBLESHOOTING
Part No 1110538 19 Platinum™Series
HIGH PRESSURE:
Internal Status Indicators:
RED: One Flash
GREEN: Two Flashes
Unit plugged in, power
switch on, continuous
alarm. Compressor shut
down.
P.C. board:
a. Malfunction
b. Disconnected wire.
c. Shifting valve at pressures greater
than 25 p.s.i. for Platinum 5 and XL;
29 PSI for Platinum 10.
a. Set flow to max L/min. for
concentrator. Check voltage across
Pilot Valve 1 on 200 volt scale. If
meter reads 0 volts when unit is
turned on, replace P.C. board
b. Check spade connectors on pilot
valves 1 or 2 and connectors on P.C.
board.
c. Replace P.C. board. Unit requires
retiming after P.C. board
replacement. Refer to Replacing P.C.
Board on page 66.
4-way Valve:
a. Not shifting.
b. Bad coil.
c. Coil resistance.
a. Check voltage. If voltage rises to
approximately 24 volts D.C. when
unit shuts down, P.C. board is
functioning properly. Replace the
valve if shorted or open. Refer to
Replacing 4-Way Valve
on page 73.
b. Replace 4-way valve. Refer to
Replacing 4-Way Valve
on page 73.
c. Coil should read 80 ohm ± 10.
SYMPTOM PROBABLE CAUSE SOLUTION
SECTION 4—TROUBLESHOOTING
Platinum™Series 20 Part No 1110538
LOW PRESSURE:
Internal Status Indicators:
RED: One Flash
GREEN: One Flash
Or
RED: One Flash
GREEN: Three Flashes
Control Panel Indicators:
RED: On
YELLOW: Off
GREEN: Off
Unit plugged in, power
switch on, continuous
audible alarm.
Compressor shut down
(Failure to cycle due to
low pressure).
Compressor inlet filter dirty or plugged. Replace compressor inlet filter. Refer to
Replacing the Compressor Inlet Filter
on
page 28.
Compressor:
a. Leaks at fittings or tubing
b. Leaking or defective relief valve.
c. Insufficient voltage at outlet.
d. Worn cup seals or gaskets.
a. Replace or repair.
b. Repair leak or replace.
c. DO NOT use extension cords. Use
another outlet.
d. Rebuild top end of compressor. Refer
to Rebuilding the Thomas Model
2650 and 2660 Compressor -
Platinum 5 and 10 Models on page 37
or Platinum XL Model - Rebuilding
the Gast Compressor on page 45.
Heat exchanger:
a. Leak at tubing or body chamber.
b. Inspect tubing and heat exchanger.
a. Replace or retighten.
b. Replace or retighten tubing. Replace
heat exchanger. Refer to Replacing
Heat Exchanger on page 61.
Regulator cracked or leaking. Replace regulator. Refer to Replacing
Regulator on page 58.
P.E. valve leaking. Replace P.E. Valve. Check voltage at P.E.
valve connector on 24 volt D.C. scale. The
P.E. Valve activates or energizes
approximately one second prior to the
activation of the 4 way valve with
approximately 24 volts. If voltage is in
excess of 24 volts consistently, replace the
P.C. board. Refer to Replacing P.C. Board
on page 66. If the P.C. board voltage acts
normally, replace the P.E. valve. Refer to
Replacing P.E. Valve
on page 53.
NOTE: Check for leaks starting at the
compressor output through all the
pneumatic connections. Major leaks will
cause system pressures to remain below
adequate shift (exhaust) pressures and will
cause compressor shutdown.
Defective Parts:
Internal Status Indicators:
RED: One Flash GREEN:
Five Flashes
Or
RED: Two Flashes
GREEN: Three Flashes
CONTROL PANEL
INDICATORS:
RED: On
YELLOW: Off
GREEN: Off
Unit plugged in, power
switch on, continuous
audible alarm.
Compressor shut down.
Defective main valve coil. a. Replace main valve.
b. Replace main valve coil.
Connector loose. Reattach connector.
Defective P.E. valve coil. (Resistance 80
ohm’s ± 10.)
Replace P.E. coil if shorted or open on
resistance check. Refer to Replacing P.E.
Valve on page 53.
Defective P.E. valve. Replace P.E. valve. Refer to Replacing P.E.
Valve on page 53.
Connector loose. Reattach connector.
Defective P.C. board. Replace P.C. board. Refer to Replacing P.C.
Board on page 66.
SYMPTOM PROBABLE CAUSE SOLUTION
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100

Invacare Platinum 10 User manual

Category
Oxygen Equipment
Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI