BIO-TEK FL600
FLUORESCENCE
PLATE READER
HARDWARE GUIDE
29?D5;
9>CDBE=5>DC, 9>3
29?D5;
9>CDBE=5>DC, 9>3
HIGHLAND PARK, BOX 998, WINOOSKI, VT 05404-0998 USA.
(800) 451-5172 (802) 655-4040 FAX: (802) 655-7941
BIO-TEK FL600
FLUORESCENCE
PLATE READER
For Research & Development Use Only
HARDWARE GUIDE
Part No. 6001000
June, 1998
Revision D
Bio-Tek FL600 iii
Notices
BIO-TEK INSTRUMENTS, INC.
Highland Park, Box 998
Winooski, Vermont, USA
05404-0998
Customer Service & Sales: 888-451-5171
Outside U.S.: 802-655-4040
Service: 800-24-BIOTK
Sales FAX: 802-655-7941
Service FAX: 802-655-3399
Internet: http://www.biotek.com
Lionheart Technologies, Europe N.V.
Sneeuwbeslaan 33
B-2610 Wilrijk
Belgium
TEL: (32) 3-8290455
FAX: (32) 3-8254749
Copyright
Copyright 1997, Bio-Tek Instruments, Incorporated. No part of this
publication may be reproduced, transmitted, transcribed, stored in a retrieval
system, or translated into any language without the written permission of Bio-Tek
Instruments, Incorporated.
Trademarks
Bio-Tek
®
is a registered trademark, and FL600™ and KC4™ are trademarks of
Bio-Tek Instruments, Inc.. Windows
®
and Windows 95 are trademarks of the
Microsoft Corporation.
Restrictions and Liabilities
Information in this document is subject to change, and does not represent a
commitment by Bio-Tek Instruments, Inc. Changes made to the information in
this document will be incorporated in new editions of the publication. No
responsibility is assumed by Bio-Tek for the use or reliability of software or
equipment that is not supplied by Bio-Tek, or its affiliated dealers.
Bio-Tek Part No. 6001000
iv Bio-Tek FL600
Revision History
Revision Date Changes
A 4/97 Release to production
B 10/97 Added Stay Clear safety symbol. Corrected Electrical, Optical,
and Performance Specifications sections. Added CE Mark
information. Deleted Appendix A.
C3/98
Preface: Added warranty. Page 2-1: Modified instructions for
Setting Up the FL600 Reader. Page 5-2: Changed electrical
specification for fuses. Page 5-11: Modified text and data on
serial communications to reflect parity differences between KC4
reader software and FL600 diagnostic software. Page 5-13: On
filter table, modified use of excitation filters and added two new
filter wavelengths.
D6/98
Preface: Changed phone numbers and e-mail addresses.
Chapter 4: Changed air filter maintenance procedure. Added
lamp replacement procedure.
Bio-Tek FL600 v
Warranty
This Warranty is limited and applies only to new products, except for computer-based
software which is covered under a separate Warranty Policy, manufactured by
Bio-Tek
®
Instruments, Inc. (“Bio-Tek”). Bio-Tek makes no warranty whatsoever
regarding the condition of used products.
Bio-Tek warrants the instrument (hereinafter collectively referred to as “Products” or
“Product”) for a period of one (1) year from the original purchase date against defective
materials or workmanship. This Warranty is limited to the original purchaser (the
“Purchaser”) and cannot be assigned or transferred. All claims under this Limited
Warranty must be made in writing to Bio-Tek, Attention: Service Department.
Purchaser must ship the Product to Bio-Tek, postage pre-paid. Bio-Tek shall either
repair or replace with new or like-new, at its option and without cost to the Purchaser,
any Product which in Bio-Tek’s sole judgment is defective by reason of defects in the
materials or workmanship.
This Warranty is VOID if the Product has been damaged by accident or misuse, or has
been damaged by abuse or negligence in the operation or maintenance of the Product,
including without limitation unsafe operation, operation by untrained personnel, and
failure to perform routine maintenance. This Warranty is VOID if the Product has been
repaired or altered by persons not authorized by Bio-Tek, or if the Product has had the
serial number altered, effaced, or removed. This Warranty is VOID if any of the
Products has not been connected, installed or adjusted strictly in accordance with
written directions furnished by Bio-Tek. Batteries, fuses, lightbulbs, and other
“consumable” items used in any of the Products are not covered by this Warranty.
Software utilized in conjunction with any of the Products is not covered by the terms of
this Warranty but may be covered under a separate Bio-Tek software warranty.
We will continue to stock parts for a maximum period of five (5) years after the
manufacturer of any equipment has been discontinued. Parts shall include all materials,
charts, instructions, diagrams, and accessories that were furnished with the standard
models.
THIS WARRANTY CONTAINS THE ENTIRE OBLIGATION OF BIO-TEK
INSTRUMENTS, INC., AND NO OTHER WARRANTIES, EXPRESSED, IMPLIED,
OR STATUTORY ARE GIVEN. PURCHASER AGREES TO ASSUME ALL
LIABILITY FOR ANY DAMAGES AND/OR BODILY INJURY OR DEATH
WHICH MAY RESULT FROM THE USE OR MISUSE OF ANY EQUIPMENT OR
INSTRUMENT BY THE PURCHASER, HIS EMPLOYEES, AGENTS, OR
CUSTOMERS, OTHER THAN THE EXPRESS WARRANTY CONTAINED
HEREIN. WE SHALL NOT BE RESPONSIBLE FOR ANY DIRECT OR
CONSEQUENTIAL DAMAGES OF ANY KIND. THIS WARRANTY SHALL NOT
BE CHANGED OR MODIFIED IN ANY WAY WITHOUT THE EXPRESS
WRITTEN PERMISSION OF AN OFFICER OF BIO-TEK INSTRUMENTS, INC.
vi Bio-Tek FL600
Warnings and Precautions
When operated in a safe environment, according to the instructions in this manual,
there are no known hazards associated with the instrument. However, the
operator should be aware of certain situations which could result in serious injury.
Hazard Warnings
Warning! Power Rating. The instrument must be connected to a power
receptacle that provides voltage and current within the specified rating for the
system. Use of an incompatible power receptacle may produce electrical shock
and fire hazards.
Warning! Electrical Grounding. Never use a two-prong plug adapter to
connect primary power to the instrument. Use of a two-prong adapter disconnects
the utility ground, creating a severe shock hazard. Always connect the system
power cord directly to a three-prong receptacle with a functional ground.
Warning! Internal Voltage. Always turn off the power switch and unplug the
power cord before cleaning the outer surfaces of the instrument or changing
filters.
Warning! Liquids. Avoid spilling liquids on the instrument; fluid seepage into
internal components creates a potential shock hazard. Wipe up all spills
immediately. Do not operate the instrument if internal components have been
exposed to fluid. If the microplate appears to have been exposed to fluid, the
reading chamber may also have been exposed. To protect the incubator, remove
the top cover and clean the interior.
Precautions
The following precautions are provided to help you avoid damaging the system:
Caution: Service. The system should be serviced by Bio-Tek-authorized service
personnel. Only qualified technical personnel should perform troubleshooting and
service procedures on internal components.
Caution: Replacement Fuses. Use replacement fuses with the required current
rating and specification. Improper fuses or short-circuiting the fuse holders may
cause fire or damage the instrument.
Bio-Tek FL600 vii
Caution: Environmental Conditions. Do not expose the system to temperature
extremes. System performance may be adversely affected.
Caution: Filter Settings. The software filter settings must match the actual
filters used in the reader, in order to prevent possible damage to the
photomultiplier and/or degradation of data.
Safety Symbols
The following warning and informational symbols may be found in various
locations on the instrument. Only qualified personnel who recognize shock
hazards and are familiar with the safety precautions should use this instrument.
Read the manual carefully before operating this instrument.
Direct current
Courant continue
Gleichstrom
Corriente continua
Corrente continua
Alternating current
Courant alternatif
Dreiphasen-Wechselstrom
Corriente Atterna
Corrente alternata
Both direct and alternating current
Courant continu et courant alternatif
Allstrom (Gleich - und Wechselstrom
Corriente continua y corriente alterna
Corrente continua e corrente alternata
Earth ground terminal
Borne de terre
Erde (Bettriebserde)
Borne de Tierra
Terra (di funzionamento)
Protective conductor terminal
Borne de terre de protection
Schutzleiteranscluss
Borne de Tierra de Protección
Terra di protezione
On (Supply)
Marche (alimentation)
Ein (Verbindung mit dem Netz)
Connectado
Chiuso
viii Bio-Tek FL600
Off (Supply)
Arrest (alimentation)
Aus (Trennung vom Netz)
Desconectado
Aperto (sconnessione dalla rete di alimentazione)
Caution (Refer to accompanying documents)
Attention (voir documents d'accompanement)
Achtung siehe Begleitpapiere
Atención (vease los documentos incluidos)
Attenzione, consultare la doc annessa
Caution, risk of electric shock
Attention, risque de choc electrique
Gefährliche elektrische Spannung
Atención, riesgo de sacudida eléctrica
Alta tensione (in questo documento
Alta tensione non significa "tensione pericolosa"
come definito in IEC 417)
Stay clear
Ne pas toucher svp.
Bitte nicht berühren
No tocario por favor
Si prega di non toccare
Bio-Tek FL600 ix
Contents
1. Introducing the Bio-Tek FL600
Hardware Requirements 1-3
Software Requirements 1-4
2. Getting Started
Setting Up the FL600 Reader 2-1
Installing KC4™ and FL600 Diagnostic Software 2-2
Installing KC4 2-2
Installing the Hardware Key 2-2
Installing the FL600 Diagnostic Software (Diagnostics Use)2-2
Checking the FL600 System 2-4
Using the FL600 Diagnostic Software 2-4
Filter Wavelength Locations 2-5
Checking the COM Port 2-5
Running System Diagnostics 2-7
What to Do if the Tests Fail 2-8
Exiting the FL600 Diagnostic Software 2-9
3. Replacing Filters and Adjusting Probe Height
Adding and Replacing Filters 3-1
Changing the Filters Inside the Reader 3-2
Setting the Filter Positions in KC4 Software 3-6
Utilizing the Filter Adapter System 3-7
Removing the Filter from the Adapter 3-7
Inserting the Replacement Filter into the Adapter 3-8
Adjusting Top Optical Probe Height 3-8
FL600 Interior Configuration Diagrams 3-10
4. Bio-Tek FL600 Maintenance
Maintaining the Reader 4-1
How to Maintain the Exterior 4-1
How to Replace the Air Filter 4-1
How to Replace the Lamp 4-3
x Bio-Tek FL600
5. Technical Information
Bio-Tek FL600 Fluorescence/Absorbance Plate Reader 5-1
Physical Dimensions 5-1
Operating Conditions 5-1
Electrical Specifications 5-2
Optical Specifications 5-2
Scanning Mechanism 5-3
Temperature Control 5-4
Performance Specifications 5-4
CE Mark Approval 5-8
Computer Specifications for FL600 Diagnostic Software 5-10
FL600 Serial Communications 5-11
Ordering Filters 5-12
Fluorescence Filters 5-12
Absorbance Filters 5-14
&KDSWHU
,QWURGXFWLRQ
WR
WKH
%LR7HN
)/
This chapter introduces the Bio-Tek FL600 Fluorescence
Plate Reader. You'll find information on hardware and
software requirements and optional equipment
suggestions.
,QWURGXFLQJ7KH)/
Thank you for selecting the Bio-Tek FL600. The FL600 is a powerful
fluorescence and absorbance measurement system which measures
through the use of a photomultiplier tube housed inside of a light-
impermeable detection compartment. The FL600 is capable of quantifying
both soluble and cell-associated fluorescence. The system consists of:
The Bio-Tek FL600 Fluorescence Plate Reader
KC4™ reader control and data reduction software
FL600 System Software for diagnostics
Accessories Kit
Serial Communications Cable (P/N 75034)
Both software packages are for Windows-based PCs and provides the user
with full command of the operation of the X-Y scanner. FL600’s precision
X-Y mechanism scans a microplate with a sensitive optical detection
system to quantify the fluorescence or absorbance in each well.
Introduction to the Bio-Tek FL600
1-2
Virtually any microplate type from 6 to 384 wells can be used with the
FL600. Users can select any microplate from the FL600’s built-in list of
industry-standard plates, or they can define their own unique plate
format(s).
The FL600 is designed to meet international radio frequency emission
standards, EMC immunity standards, and user safety standards.
The Bio-Tek FL600 should be installed in a well-ventilated area, free from
dust and fumes. Keep the system away from direct sunlight and local heat
sources.
Introduction to the Bio-Tek FL600
1-3
+DUGZDUH5HTXLUHPHQWV
0LQLPXP&RPSXWHU5HTXLUHPHQWVIRU)/6\VWHP6RIWZDUH
DQG.&6RIWZDUHIRU:LQGRZVDQG
IBM
486 PC or clone 33 MHz clock
640 KB base memory, extended to 8 MB
Color graphics monitor (VGA)
Color graphics adapter (VGA)
120 MB hard drive
High-density floppy disk drive 3½"
1 serial (COM) and 1 parallel (LPT) port
Mouse and keyboard ports
Microsoft or compatible mouse
5HFRPPHQGHG&RPSXWHU5HTXLUHPHQWVIRU)/6\VWHP
6RIWZDUHDQG.&6RIWZDUHIRU:LQGRZVDQG
IBM 486 PC or clone 100 MHz clock
640 KB base memory, extended to 8 MB
Color graphics monitor (VGA)
Color graphics adapter (VGA)
120 MB hard drive
High-density floppy disk drive 3½"
1 serial (COM) and 1 parallel (LPT) port
Mouse and keyboard ports
Microsoft or compatible mouse
Surge/spike protector
Introduction to the Bio-Tek FL600
1-4
5HFRPPHQGHG&RPSXWHU5HTXLUHPHQWV)RU:LQGRZV

IBM 486 / 100 MHz PC or better compatible
Memory: 640 KB base, extended to 16 MB minimum
250 MB hard drive
3ULQWHU
Microsoft Windows compatible printer
6RIWZDUH5HTXLUHPHQWV
FL600 System Software, version 2 or higher
KC4™ reader control and data reduction software
MS-DOS version 5.0 or later
Microsoft Windows version 3.1 or later, or Microsoft
Windows 95 (with 16 MB memory)
&KDSWHU
*HWWLQJ6WDUWHG
This chapter discusses how to set up the Bio-Tek FL600
hardware, check the communication port and run the
diagnostic tests, install and configure the FL600 software,
and exit the FL600 system software.
6HWWLQJ8S7KH)/5HDGHU
CAUTION: It is recommended that two people lift the FL600, which weighs
about 23
kg (50.6 lb.). Care must be taken when moving the device or removing it from the
packaging. Hold the device firmly from the bottom of the case when lifting it.
Setting up the Bio-Tek FL600 Fluorescence Plate Reader is simply a matter of
connecting the system to the computer via the computer’s serial port, installing
the software, and setting up communications parameters. One of the serial ports
most likely already has a mouse connected to it (the FL600 software, since it
operates under Windows, requires a mouse).
Step 1:
Set the incubator power switch to the appropriate voltage as
follows:
110V for 100-120V˜ operation,
220V for 200-240V˜ operation
Step 2:
Turn off the PC.
Step 3:
To connect the FL600, first attach the 9-pin D-type cable into the
computer’s serial port (COM1, COM2, COM3, or COM4).
Step 4:
Plug the other end of the cable into the Serial Data Port located at
the rear of the FL600 Fluorescence Plate Reader.
Step 5:
Connect the power cable to the back of the Plate Reader.
Step 6:
Plug all power cords into grounded, electrical outlets (use of a surge
protector is strongly recommended).
Getting Started
2-2
The FL600 and all computer hardware must be electrically grounded and
connected to a viable surge protector to avoid dangerous shocks and to
avoid destroying the electronic components through electrical spikes and
surges.
,QVWDOOLQJ.&DQG)/'LDJQRVWLF6RIWZDUH
The FL600 is supplied with two Windows™ software packages, KC4™
and FL600 Diagnostic Software.
KC4 controls the reader, reduces the fluorescence and absorbance data,
and furnishes the reports.
The FL600 Diagnostic software provides diagnostics to test the reader.
Although it contains other functions, these are undocumented and
unsupported.
,QVWDOOLQJ.&
Please consult your KC4™ User’s Guide. Place disk 1 into the floppy
drive. Run Setup.exe by choosing the Run option from the File Menu of
your Windows program. This routine will guide you through the
installation of KC4 onto your hard disk.
,QVWDOOLQJWKH+DUGZDUH.H\
To use KC4 to control the reader and print data, the hardware key needs to
be installed on the parallel port of your computer. Please consult your
KC4 Operations manual for further instruction.
,QVWDOOLQJWKH)/'LDJQRVWLF6RIWZDUH
'LDJQRVWLFV8VH
To install the FL600 software, follow these steps:
Step 1:
Turn on your computer, start Windows, and activate Windows’
Program Manager, if necessary.
Step 2:
Insert the FL600 Setup diskette into the floppy drive.
Step 3:
Select
Run
from the Windows’
File
menu.
Step 4:
Type drive: setup, (i.e., b: setup).
Getting Started
2-3
Setup will indicate to you that all FL600 diagnostic software files will be
installed into the drive and directory C:\FL600. You will be given the
opportunity to change either the drive or the directory (or both). Setup
creates a new group called Bio-Tek FL600. Note that this is different from
the KC4™ group created by the KC4™ software.
NOTE: Be sure that TrueType Fonts are enabled in the Fonts Section of the
Windows Control Panel. FL600 requires certain TrueType fonts when
printing reports. For more information on activating TrueType Fonts,
please check your Windows User’s Guide.
Getting Started
2-4
&KHFNLQJWKH)/6\VWHP
Before you begin to use the FL600 software, you should first check that
the FL600 settings are correct.
8VLQJWKH)/'LDJQRVWLF6RIWZDUH
Step 1:
To start the FL600 diagnostic software:
From within Windows, double-click on the FL600 icon.
Depending on your computer speed, the FL600 diagnostic
software should boot up from within 15 to 60 seconds.
Longer than that might indicate that your computer system
needs more RAM or clock speed.
The FL600 Plate Layout screen will appear:
Figure 2-1: FL600 Plate Layout Screen
Getting Started
2-5
)LOWHU:DYHOHQJWK/RFDWLRQV
Refer to Chapter 3 for instructions on determining filter locations. Record
the filter locations to be entered into KC4.
&KHFNLQJWKH&203RUW
Next, turn on the FL600 Reader using the switch located on the right side
of the front panel and check the COM Port Setting.
NOTE: When the Reader has been switched on, a green light next the switch
lights up and the plate transport moves out, ready to load a plate.
Step 1:
Pull down the
Instrument
menu and select
Communications
. A
dialog box appears showing you the currently selected
communications port, through which the software is attempting to
communicate with the FL600 Reader:
Figure 2-3: Communications Port Settings Screen
/