LeCroy WavePro HD User manual

Category
Measuring, testing & control
Type
User manual
Operator's Manual
WavePro HD
Oscilloscopes
WavePro HD Oscilloscopes Operator's Manual
© 2022 Teledyne LeCroy, Inc. All rights reserved.
Users are permitted to duplicate and distribute Teledyne LeCroy, Inc. documentation for internal educational purposes only.
Resale or unauthorized duplication of Teledyne LeCroy publications is strictly prohibited.
Teledyne LeCroy is a trademark of Teledyne LeCroy, Inc., Inc. Other product or brand names are trademarks or requested
trademarks of their respective holders. Information in this publication supersedes all earlier versions. Specifications are
subject to change without notice.
November, 2022
waveprohd-om-eng_21nov22.pdf
Contents
About This Manual iv
Safety 1
Symbols 1
Precautions 1
Operating Environment 2
Measuring Terminal Ratings (C1-C4 and Ext) 2
Cooling 2
Cleaning 2
Power 3
Oscilloscope Overview and Set Up 4
Front of Oscilloscope 4
Back of Oscilloscope 5
Side of Oscilloscope 6
Front Panel 7
Signal Interfaces 10
Probes 11
Powering On/Off 12
Connecting to Other Devices/Systems 13
Configuring the User Button 14
MAUI User Interface 15
Automation 15
Volatile vs. Non-Volatile Settings 15
Touch Screen 16
Display 22
Controlling Traces 26
Zooming 33
MAUI with OneTouch 38
Windows 10 Gestures 43
Print/Screen Capture 44
Acquisition 45
Auto Setup 45
Vertical 46
Digital (Mixed Signal) 51
Timebase 55
Trigger 66
Cursors 79
Cursor Types 79
Apply and Position Cursors 82
Standard Cursors Dialog 83
XY Cursors Dialog 83
Measure 85
Measure Table 86
Measure Dialog 87
Parameter Set Up 91
i
WavePro HD Oscilloscopes Operator's Manual
List of Standard Measurements 99
Parameter Math 109
Using Web Editor 111
Fast MultiWave Port 114
Using Histograms 118
Using Trends 121
Using Tracks 122
Track vs. Trend 123
EMC Pulse Parameter Software Package 123
Math 125
Math Function Set Up 125
Math Dialog 127
List of Standard Math Operators 127
Average Function 132
Copy Function 133
ERes Function 134
Fast Wave Port Function 136
FFT Function 141
Interpolate Function 143
Sparse Function 143
Rescale Function 144
Phistogram Function 149
Ptrace Functions 150
Memories 151
Saving Memories 151
Restoring Memories 152
Analysis Tools 153
WaveScan 153
PASS/FAIL Testing 159
Save/Recall Data (File Menu) 163
Save 163
Recall 169
LabNotebook 171
Report Generator 176
File Sharing Dialog (send email) 177
Print 178
Email & Report Settings 179
Using the File Browser 180
Trace File Format 182
Utilities 187
Utilities Dialogs 187
Disk Utilities 191
Preferences Dialogs 192
Maintenance 197
Restart/Reboot Instrument 197
Restore Default Setup 197
ii
Changing Screen Settings 197
Touch Screen Calibration 198
Windows 10 External Display Setup 199
Software and File Management 200
Switching Windows Users 201
MAUI Firmware Update 201
Technical Support 203
Returning a Product for Service 204
Certifications 205
EMC Compliance 205
Safety Compliance 206
Environmental Compliance 207
ISO Certification 207
Intellectual Property 207
Index 209
iii
WavePro HD Oscilloscopes Operator's Manual
About This Manual
Thank you for purchasing a Teledyne LeCroy oscilloscope. We're certain you'll be pleased with the detailed features
unique to our instruments.
This manual covers the operation and maintenance of all instruments in the WavePro HD series.
Documentation for implementing COM Automation and remote control, and for using optional software packages
sold for Teledyne LeCroy instruments can be downloaded from our free technical library at:
teledynelecroy.com/support/techlib
Our website maintains the most current product specifications and should be checked for updates. Detailed
specifications are listed on the product datasheet.
Note: Specifications are subject to change without notice.
iv
Safety
Safety
Symbols
These symbols appear on the instrument or in documentation to alert you to important safety concerns:
Caution of potential damage to instrument or Warning of potential bodily injury. Refer to manual. Do not
proceed until the information is fully understood and conditions are met.
Caution, high voltage; risk of electric shock or burn.
Caution, contains parts/assemblies susceptible to damage by Electrostatic Discharge (ESD).
Frame or chassis terminal (ground connection).
Alternating current.
Standby power (front of instrument).
Precautions
Observe generally accepted safety procedures in addition to the precautions listed here. The overall safety of any
system incorporating this product is the responsibility of the assembler of the system.
Use indoors only .
Use only within the operational environment listed. Do not use in wet or explosive atmospheres.
Maintain ground. The AC inlet ground is connected directly to the chassis of the instrument. To avoid electric shock,
connect only to a mating outlet with a safety ground contact.
Caution: Interrupting the protective conductor inside or outside the oscilloscope, or disconnecting the safety
ground terminal, creates a hazardous situation. Intentional interruption is prohibited.
Connect and disconnect properly. Do not connect/disconnect probes, test leads, or cables while they are connected
to a live voltage source.
Observe all terminal ratings. Do not apply a voltage to any input that exceeds the maximum rating of that input. Refer
to the body of the instrument for maximum input ratings.
Use only the power cord shipped with the instrument and certified for the country of use.
Keep product surfaces clean and dry. See Cleaning.
Do not remove the covers or inside parts. Refer all maintenance to qualified service personnel.
Exercise care when lifting.
Do not operate with suspected failures. Do not use the product if any part is damaged. Cease operation immediately
and secure the instrumentfrom inadvertent use.
1
WavePro HD Oscilloscopes Operator's Manual
Operating Environment
Temperature:5 °C to 40 °C
Humidity: Maximum relative humidity 90% up to 31 °C,
decreasing linearly to 50% relative humidity at 40 °C
Altitude: Up to 10,000 ft (3,048 m) at or below 30 °C
Measuring Terminal Ratings (C1-C4 and Ext)
Maximum Input Voltage:50 Ω coupling ≤ 5 Vrms
1 MΩ coupling ≤ 400 Vpk max. (Peak AC ≤ 10 kHz + DC)
derating at 15 dB/decade from 10 kHz to 1.6 MHz,
10 Vpk max.above 1.6 MHz
Caution: Measuring terminals have no rated measurement category per IEC/EN 61010-1:2010. Measuring
terminals are not intended to be connected directly to supply mains.
Cooling
The instrument relies on forced air cooling with internal fans and vents. The internal fan control circuitry regulates
the fan speed based on the ambient temperature. This is performed automatically after start-up.
Caution: Do not block the cooling vents.
Take care to avoid restricting the airflow to any part. In a benchtop configuration, leave a minimum of 15 cm (6
inches) around the sides between the instrument and the nearest object. The feet provide adequate bottom
clearance. Follow rackmount instructions for proper rack spacing.
Cleaning
Clean only the exterior of the instrumentusing a soft cloth moistened with water or an isopropyl alcohol solution. Do
not use harsh chemicals or abrasive elements. Under no circumstances submerge the instrument or allow moisture
to penetrate it. Dry thoroughly before connecting a live voltage source.
Caution: Unplug the power cord before cleaning. Do not attempt to clean internal parts.
2
Safety
Power
The instrument automatically adapts to the line voltage; manual voltage selection is not required.
AC Power Source:100-240 VAC (±10%) at 50/60 Hz (±10%) or
100-120 VAC (±10%) at 400 Hz (±5%)
Automatic AC voltage selection
Maximum Consumption:* 525 W (525 VA)
Nominal Consumption:400 W (400 VA)
Standby Consumption:10 W
* All PC peripherals and active probes installed on four channels.
The provided power cords mate to a compatible power inlet on the instrument for making line voltage and safety
ground connections. The AC inlet ground is connected directly to the chassis of the instrument. For adequate
protection again electric shock, connect to a mating outlet with a safety ground contact.
3
WavePro HD Oscilloscopes Operator's Manual
Oscilloscope Overview and Set Up
Front of Oscilloscope
A. Touch Screen display
B. Front Panel controls
C. Standby Power button
D. Calibration and Ground terminals
E. USB 3.1 Gen1 Ports (3)
F. ProBus2 analog inputs (4)
G. External trigger input
H. Auxiliary output
I. Mixed Signal interface
J. Tilting feet
4
Oscilloscope Overview and Set Up
Back of Oscilloscope
A. Built-in carrying handle
B. Removable solid state drive
C. AC power Inlet
D. Kensington lock
E. 10 MHz Ref In and Ref Out for
reference clock
F. USBTMC Port
5
WavePro HD Oscilloscopes Operator's Manual
Side of Oscilloscope
G. Speaker Out and Mic In
H. USB 3.1 Gen1 Ports (4)
I. Ethernet Ports
J. HDMI and DisplayPort Outputs
to external monitors
6
Oscilloscope Overview and Set Up
Front Panel
Front panel controls duplicate functionality available through the touch screen and
are described here only briefly.
Knobs on the front panel function one way if turned and another if pushed like a
button. The first label describes the knob’s “turn” action, the second label its “push”
action. Actions performed from the front panel always apply to the active trace.
Many buttons light to show the active traces and functions.
Trigger Controls
The front panel Level knob changes the trigger threshold level (V). The number is
shown on the Trigger descriptor box. Pushing the knob sets the trigger level to the
50% point of the input signal.
The READY indicator is lit when the trigger is armed. TRIG'D is lit momentarily when
a trigger occurs.
Setup opens the Trigger Setup dialog. Press it again to close the dialog.
Auto sets Auto mode, which triggers the oscilloscope after a set time, even if the
trigger conditions are not met.
Normal sets Normal mode, which triggers the oscilloscope each time a signal is
present that meets the trigger conditions.
Single sets Single trigger mode. The first press readies the oscilloscope to trigger.
The second press arms and triggers the oscilloscope once (single-shot acquisition)
when the input signal meets the trigger conditions.
Stop prevents the scope from triggering on a signal. If you boot up the instrument
with the trigger in Stop mode, a "No trace available" message is shown. Press the
Auto button to display a trace.
Horizontal Controls
Turn the Delay knob to change the Trigger Delay value (S). Push the knob to reset
Delay to zero.
If the trace source is an input channel, turn the Horizontal Adjust knob to set the
Time/division (S) of the oscilloscope acquisition system. The value is shown on the
Timebase descriptor box. When using this control, the oscilloscope allocates
memory as needed to maintain the highest sample rate possible for the timebase
setting. If the active trace is a math, zoom or memory, use this knob to change the
Horizontal Scale, effectively "zooming" in or out. The value is shown on the corresponding descriptor box. Push the
knob to adjust scale in fine increments; push it again to return to stepped increments.
Vertical Controls
The Offset knob adjusts the zero level of the trace (this makes it appear to move up or down relative to the center
axis of the grid). The value appears on the trace descriptor box. Push it to reset Offset to zero.
7
WavePro HD Oscilloscopes Operator's Manual
The Vertical Adjust knob sets Vertical Gain (V/div) for channel traces or Vertical Scale for math, zoom and memory
traces. The value appears on the trace descriptor box. Push the knob to toggle between fine and stepped
increments.
The Channel buttons (1-4) turn on a channel that is off, or activate a channel that is already on. When the channel is
active, pushing its channel button turns it off. A lit button shows the active channel.
The Zoom (Z) button creates a Quick Zoom of each open trace.The resulting zoom traces are 1/10 of the source
channel scale and centered on the display. Use the Horizontal knobs to adjust this, or touch the zoom descriptor box
twice to display the zoom dialog controls. Touch the Zoom button again to turn off the zooms.
The Digital (D) button enables digital input on -MS models.
Math, Zoom, and Mem Indicators
These LEDs light to indicate when a type of trace other than an input channel is currently active and controllable
using the Front Panel. Touch a trace descriptor box to make it active and controllable.
Cursor Controls
Cursors identify specific voltage and time values on the waveform. The white cursor markers help make these points
more visible, as well as provide a simple way to reposition them. A readout of the Vertical values appears on the
trace descriptor box, while the Horizontal values appear below the Timebase descriptor box.
There are five preset cursor types, each with a unique appearance on the display. These are described in more detail
in the Cursors section.
Press the Cursor Type button to apply or remove cursors. Continue pressing to cycle through all cursor types until
the desired type is found ("Off" occurs when no cursors are visible).
Turn the Coarse knob to adjust the position of absolute cursors (dashed-dotted lines). Push it to return the cursor to
the default 2.5 division setting. If you're using the Both Rel cursor type, each push will toggle between the Horizontal
and Vertical absolute cursors. With the correct cursor selected (highlighted), turn the knob to adjust the position.
Turn the Fine knob to adjust the position of relative cursors (dashed lines). Push it to return the cursor to the default
7.5 division setting, or to select the other relative cursor when using the Both Rel cursor type.
When a numeric value field is selected, these knobs serve as the Adjust knobs. Turn the Coarse knob to make large-
increment adjustments; turn the Fine knob to make small-increment adjustments. Push either knob to return to the
default setting.
When a waveform trace is selected (by touching the descriptor box or the trace itself), these knobs
increase/decrease trace intensity.
Miscellaneous Controls
These miscellaneous controls also appear along the top and bottom rows of the front panel.
Auto Setup performs an Auto Setup. After the first press, you will be prompted for a confirmation. Press the button
again or use the touch screen to confirm.
Default Setup resets the oscilloscope to the factory default configuration. After the first press, you will be prompted
for a confirmation. Press the button again or use the touch screen to confirm.
Clear Sweeps resets the acquisition counter and any cumulative measurements.
Touch Screen toggles on/off touch screen functionality.
8
Oscilloscope Overview and Set Up
Save performs whatever action was last selected on the Save dialog: save a LabNotebook, screen image, setup or
waveform file.
User can be configured as you choose, for example:
lTo save a screen capture using your current Screen Image Preferences.
lTo save other file types: LabNotebooks, setup or waveforms.
See Configuring the User Button for instructions on making the selection.
9
WavePro HD Oscilloscopes Operator's Manual
Signal Interfaces
Teledyne LeCroy instruments offer a variety of interfaces to input analog or digital signals. See the oscilloscope
product page at teledynelecroy.com for a list of compatible input devices.
ProBus2 Interface
WavePro HD analog channel inputs are equipped with the ProBus2 interface.
Like the ProBus interface that it replaced, the ProBus2 interface contains a 6-pin power and communication
connection and a BNC signal connection to the probe, with sense rings for detecting passive probes. A special airline
BNC connector enables the interface to support bandwidths up to 8 GHz. It offers both 50 Ω/1 MΩ input impedance,
and provides probe power and control for a wide range of probes such as high impedance passive probes, high
impedance active probes, current probes, high voltage probes and differential probes.
Note: ProBus2 interfaces are compatible with legacy ProBus probes, as well as newer Probus2 probes. SMA
cables can be connected with the use of a precision adapter available from Teledyne LeCroy Service.
The ProBus2 interface completely integrates the probe with the channel. Upon connecting a Teledyne LeCroy probe,
the probe type is recognized and some setup information, such as input coupling and attenuation, is performed
automatically. This information is displayed on the Probe Dialog, behind the Channel (Cn) dialog. System (probe plus
instrument) gain settings are automatically calculated and displayed based on the probe attenuation.
ProBus2 may also have a BNC-terminated cable connected directly to it. Depending on the BNC connector used on
the cable, the interface is rated for up to 8 GHz with 50 Ω coupling or 1 GHz with 1 MΩ coupling.
Note: Operational bandwidth is equal to the maximum input frequency of your oscilloscope model. See the
product datasheet.
Other Analog Inputs
EXT In can be used to input an external trigger pulse.
REF In can be used to input an external reference clock signal.
These inputs have a simple BNC interface with no power supply. See your product datasheet for voltage and
frequency ratings.
Mixed Signal Inputs
The digital leadset shipped with the -MS model oscilloscopes connects to the Mixed Signal Input on the front of the
oscilloscope to input of up-to-16 lines of digital data. Physical lines can be preconfigured into different logical
groups, Digitaln, corresponding to a bus and renamed appropriately depending on the group. The transitions for each
line may be viewed through different displays.
See Digital Setup Using theDigital Leadset for detailed instructions.
10
Oscilloscope Overview and Set Up
Probes
The oscilloscope is compatible with the included passive probes and most Teledyne LeCroy active probes that are
rated for the instrument’s bandwidth. Probe specifications and documentation are available at
teledynelecroy.com/probes.
Passive Probes
The passive probes supplied are matched to the input impedance of the instrument but may need further
compensation. Follow the directions in the probe instruction manual to compensate the frequency response of the
probes.
If using other passive probes than those supplied, be sure to perform a low frequency calibration before using them
to measure signal. You can use the signal from the Cal Out hook on the front of the oscilloscope.
Active Probes
Teledyne LeCroy offers a variety of active probes for use with your oscilloscope. Most active probes match probe to
oscilloscope response automatically using probe response data stored in an on-board EEPROM. This ensures the
best possible combined probe plus oscilloscope channel frequency response without the need to perform any de-
embedding procedure.
Be aware that many active probes require a minimum oscilloscope firmware version to be fully operational. See the
probe documentation.
11
WavePro HD Oscilloscopes Operator's Manual
Powering On/Off
Press the Power button to turn on the instrument. If the button is not easily accessible (e.g., in a rackmount),
you can go to Utilities > Preferences and enable Power on AC. Then, the instrument will start whenever the AC power
supply is turned on.
Caution: Do not power on or calibrate with a signal attached.
Powering on will automatically start all services and load the oscilloscope application software. After a 20 min. warm
up and calibration period, the oscilloscope will be fully operational.
To power down, you can quickly press the Power button again, but the safest way to power down is to use the File >
Shutdown menu option, which will always execute a proper shut down process and preserve settings. Holding the
Power button will execute a “hard” shut down (as on a computer), which we do not recommend doing because it
does not allow the operating system to close properly, and setup data may be lost.
Caution: Never power off by pulling the power cord or turning off the AC supply without first shutting down
the oscilloscope application properly.
The Power button does not disconnect the instrument from the AC power supply. The only way to fully power down
the instrument is to unplug the AC power cord or otherwise turn off the AC power supply.
We recommend unplugging the instrument if it will remain unused for a long period of time.
Caution: Do not change the instrument’s Windows®Power Options setting from the default Never to
System Standby or System Hibernate. Doing so can cause the system to fail.
12
Oscilloscope Overview and Set Up
Connecting to Other Devices/Systems
Use the menu options listed below to configure connections to other devices.
LAN
The instrument is preset to accept a DHCP network address over a TCP/IP connection. Connect a cable from an
Ethernet port on the side panel to a network access device. Go to Utilities > Utilities Setup > Remote to find the IP
address.
To assign a static IP address, choose Net Connections from the Remote dialog. Use the standard Windows
networking dialogs to configure the device address.
Choose File > File Sharing and open the Email & Report Settings dialog to configure email settings.
Audio/USB Peripherals
Connect the device to the appropriate port on the front or side of the instrument. These connections are "plug-and-
play" and do not require any additional configuration.
Printer
MAUI oscilloscopes support USB printers compatible with the instrument's Windows OS. Go to File > Print Setup to
configure printer settings. Select Properties to open the Windows Print dialog.
External Monitor
You may operate the instrument using the built-in touch screen or attach an external monitor for extended desktop
operation. See your product datasheet for the supported monitor resolution.
Connect the monitor cable to a video output on the instrument. You can use an adaptor if the monitor cable has a
different interface. Go to Display > Display Setup > Open Monitor Control Panel to configure the display. Be sure to
select the instrument as the primary display.
To use the Extend Grids feature, configure the second monitor to extend, not duplicate, the oscilloscope display. If
the external monitor is touch screen enabled, the MAUI user interface can be controlled through touch on the
external monitor as well as the oscilloscope. See Windows 10 External Display Setup for additional instructions on
setting up external monitors with Windows 10 oscilloscopes.
Remote Control
Go to Utilities > Utilities Setup > Remote to configure remote control. Connect the oscilloscope to the
network/controller using the cable type required by your selection.
lVICP( TCP/IP) and VXI-11(LXI) over Ethernet are supported standard, as is USBTMC.
lGPIB is supported with the use of the optional USB-GPIB adapter.
Note: Choose TCP/IP for remote control using MAUI Studio Pro. You can make the Ethernet connection
over a LAN or connect directly to the MAUI Studio host PC.
13
WavePro HD Oscilloscopes Operator's Manual
Reference Clock
To input/output a reference clock signal, connect a BNC cable from the Ref In/Out connector to the other
instrument. Go to Timebase > Horizontal Setup > Clock Source to configure the clock.
Auxiliary Output
To output signal to another instrument, connect a BNC cable from Aux Out to the other device. Go to Utilities >
Utilities Setup > Aux Output to configure the output.
Configuring the User Button
The front panel User button can be configured to perform your choice of actions, such as saving a screen image or
other file type.
1. Go to Utilities > Front Panel Setup, then touch Select User Action and make your User button selection.
2. Optionally, touch Configure Action to make any other setting related to that action, such as Screen Image or
Print preferences.
14
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221

LeCroy WavePro HD User manual

Category
Measuring, testing & control
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI