Eurotech ReliaSENS 18-12 Owner's manual

Type
Owner's manual
ReliaSENS 18
-12-00
Cloud
-connected Environment Monitoring System
USER MANUAL
Rev
1.0 – 16 July 2014 – RESENS-18-12-00_UsrMan_EN_1.0
© 2014 Eurotech SpA - Via Fratelli Solari 3/A - 33020 AMARO (UD) - Italy
Trademarks
All trademarks or registered trademarks, logos, designs and products appearing in this document are property of
their respective owners
Revision history
REVISION
DESCRIPTION
DATE
1.0
First release
16 July 2014
Table of contents ReliaSENS 18-12-00 User manual
3
RESENS
-18-12-00_UsrMan_EN_1.0
Table of contents
Trademarks .................................................................................................................................................................. 2
Revision history ........................................................................................................................................................... 2
TABLE OF CONTENTS ......................................................................................................................................................... 3
1 IMPORTANT USER INFORMATION ................................................................................................................................. 5
1.1 Alerts that can be found throughout this manual .............................................................................................. 5
1.2 Safety notices and warnings ............................................................................................................................. 6
1.2.1 Do not operate in an explosive atmosphere .............................................................................................. 6
1.2.2 Antistatic precautions ................................................................................................................................ 6
1.2.3 Connection to power supply or other devices ........................................................................................... 6
1.2.4 Ventilation .................................................................................................................................................. 7
1.2.5 Cleaning .................................................................................................................................................... 7
1.3 Life support policy ............................................................................................................................................. 7
1.4 Warranty ........................................................................................................................................................... 7
1.5 CE compliance .................................................................................................................................................. 7
1.6 WEEE ............................................................................................................................................................... 8
1.7 RoHS ................................................................................................................................................................ 8
1.8 Technical assistance ........................................................................................................................................ 8
1.8.1 Transportation ........................................................................................................................................... 8
2 WHAT YOU FIND INSIDE THE PACKAGING ..................................................................................................................... 9
3 WHAT YOU HAVE TO SUPPLY .................................................................................................................................... 10
4 RELIASENS DESCRIPTION ....................................................................................................................................... 11
5 RELIASENS SPECIFICATIONS................................................................................................................................... 12
6 INSTALL THE RELIASENS ........................................................................................................................................ 13
6.1 Install procedure ............................................................................................................................................. 14
7 POWER SUPPLY THE RELIASENS ............................................................................................................................ 16
8 ESTABLISH A NETWORK CONNECTION ....................................................................................................................... 17
8.1 Using a wireless 3G/GPRS connection .......................................................................................................... 17
8.2 Using a Wi-Fi connection ................................................................................................................................ 19
8.3 Using a wired LAN connection with static IP .................................................................................................. 21
8.4 Using a wired LAN connection in DHCP ........................................................................................................ 22
9 VIEW THE DATA ....................................................................................................................................................... 23
9.1 View the data from ReliaSENS web application ............................................................................................. 23
9.1.1 Sign in ...................................................................................................................................................... 23
9.1.2 View data in real time .............................................................................................................................. 24
9.1.3 Obtain a data report................................................................................................................................. 25
9.1.4 Have access to raw real-time data .......................................................................................................... 28
9.1.5 Have access to the graphic view of raw real-time data ........................................................................... 28
9.1.6 Enable alert emails .................................................................................................................................. 29
9.1.7 Set the thresholds.................................................................................................................................... 30
9.1.8 Set the units of measure ......................................................................................................................... 30
10 RELIASENS MAINTENANCE OPERATIONS ................................................................................................................. 31
10.1 Never open the device! ............................................................................................................................... 31
10.2 Recalibrate the ReliaSENS ......................................................................................................................... 31
Table of contents ReliaSENS 18-12-00 User manual
4
RESENS-18-12-
00_UsrMan_EN_1.0
10.3 Inspect the ReliaSENS installation periodically .......................................................................................... 31
NOTES ............................................................................................................................................................................ 33
Important user information ReliaSENS 18-12-00 User manual
5
RESENS
-18-12-00_UsrMan_EN_1.0
1 Important user information
Carefully read and understand the instructions in this manual before using this device.
Whenever you have any doubt regarding the operation of this device, consult this manual or contact your local
Eurotech Technical Support Team (see the last page of this manual for details).
Keep this manual for future reference.
To lower the risk of personal injury, electric shock, fire or damage to equipment, observe the following
precautions, as well as using good technical judgment, whenever installing or using the device.
Eurotech has made every effort to ensure the accuracy of this document; however, Eurotech assumes no liability
resulting from any error/omission in this document, or from the use of the information contained herein.
Eurotech reserves the right to revise this document or to make changes to its content at any time without any
obligation to notify any person of such revision or changes.
1.1 Alerts that can be found throughout this manual
SYMBOL
DANGER!
Information highlighting potential electrical shock hazards:
Personal injury or death could occur.
Damage to the system, connected peripheral devices, or software could occur.
Always use appropriate safety precautions. Also ensure that the installation meets all the requirements as
set out for the environment that the equipment will be deployed in.
WARNING!
Information highlighting potential hazards:
Personal injury or death could occur.
Damage to the system, connected peripheral devices, or software could occur.
Always use appropriate safety precautions. Also ensure that the installation meets all the requirements as
set out for the environment that the equipment will be deployed in.
NOTE
These will highlight important features or instructions.
Important user information ReliaSENS 18-12-00 User manual
6
RESENS-18-12-
00_UsrMan_EN_1.0
1.2 Safety notices and warnings
Observe the following safety precautions during all phases of operation, service, and repair of the device. Failure
to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of
design, manufacture, and intended use of the device.
Eurotech assumes no liability for the customer’s failure to comply with these requirements.
The safety precautions listed below represent warnings of certain dangers of which Eurotech is aware. You, as
the user of the device, should follow these warnings and all other safety precautions necessary for the safe
operation of the device in your operating environment.
1.2.1 Do not operate in an explosive atmosphere
WARNING!
Do not operate the equipment in the presence of flammable gases or fumes. Operation of any
electrical equipment in such an environment constitutes a definite safety hazard.
1.2.2 Antistatic precautions
WARNING!
To avoid ESD (Electro Static Discharge) damage, always use appropriate antistatic precautions
when handing any electronic equipment.
1.2.3 Connection to power supply or other devices
DANGER!
Before applying power to the system, thoroughly review all installation, operation, and safety
instructions.
Failure to install the system power supply correctly or to follow all operating instructions correctly
may create an electrical shock hazard, which can result in personal injury or loss of life, and/or
damage to equipment or other property
To avoid injuries, always disconnect power and discharge circuits before touching them.
Only start the device with a power supply that meets the requirements stated on the voltage label. In case of uncertainties about
the required power supply, please contact the Eurotech Technical Support Team or the electricity authority
Before connecting other equipment carefully read any supplied instructions
Always disconnect the power before connecting or disconnecting cables
Do not perform connections with wet hands
Check any power cords for damage before use
Use certified power cables. The power cable must meet the requirements (voltage and current) of the device.
Position cables with care. Avoid positioning cables in places where they may be trampled on or compressed by objects placed on
them. Take particular care of the plug, power-point and outlet of power cable
Avoid overcharging any power outlets
Only apply power to the device or connected equipment after checking that all the above conditions have been met
Important user information ReliaSENS 18-12-00 User manual
7
RESENS
-18-12-00_UsrMan_EN_1.0
1.2.4 Ventilation
WARNING!
Ensure adequate ventilation to avoid overheating
Do not block any ventilation openings
1.2.5 Cleaning
WARNING!
When cleaning the device, remember to:
Ensure sufficient ESD protection during the cleaning process
Remove any power from the device
When cleaning an enclosed system or peripheral use a dry cloth on the external casing
Do not use detergents, aerosol sprays, solvents or abrasive sponges
1.3 Life support policy
WARNING!
Do not use Eurotech products as critical components of life support devices or systems without the
express written approval of Eurotech Spa.
1.4 Warranty
Please contact your local Eurotech Sales Office for detailed warranty terms and conditions.
See the last page of this manual for further details.
1.5 CE compliance
This product is CE marked.
The CE mark on the product indicates that it conforms to the requirements of the EC
directive.
Eurotech is not responsible for the use of its product s together with equipment (e.g.
power supplies, personal computers, etc.) that are not CE marked and compliant with
technical requirements specified in this manual..
Important user information ReliaSENS 18-12-00 User manual
8
RESENS-18-12-
00_UsrMan_EN_1.0
1.6 WEEE
This marking shown on the product or its literature, indicates that it should not be disposed
with other household wastes at the end of its working life. To prevent possible harm to the
environment or human health from uncontrolled waste disposal, please separate this from
other types of wastes and recycle it responsibly to promote the sustainable reuse of
material resources.
Household users should contact either the retailer where they purchased this product, or
their local government office, for details of where and how they can take this item for
environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions of the
purchase contract.
This product should not be mixed with other commercial wastes for disposal.
1.7 RoHS
This device, including all the components, subassemblies and the consumable materials that are an integral part
of the product, have been manufactured in compliance with the European directive 2002/95/EC known as the
RoHS directive (Restrictions of the use of certain Hazardous Substances). This directive targets the reduction of
certain hazardous substances previously used in electrical and electronic equipment (EEE).
1.8 Technical assistance
For any technical questions, or if you cannot isolate a problem with your device, or for any enquiry about repair
and returns policies, feel free to contact your local Eurotech Technical Support Team (see the last page of this
manual for details).
1.8.1 Transportation
When transporting the system, for any reason, it should be packed adequately using its original packaging.
Pay attention to not damage the cables.
Warning:
Any product returned to Eurotech that is damaged due to inappropriate packaging will not be
covered by the warranty!
What you find inside the packaging ReliaSENS 18-12-00 User manual
9
RESENS
-18-12-00_UsrMan_EN_1.0
2 What you find inside the packaging
The packaging includes:
One ReliaSENS unit (front and rear views)
One Ethernet cable
One power supply cable
What you have to supply ReliaSENS 18-12-00 User manual
10
RESENS-18-12-
00_UsrMan_EN_1.0
3 What you have to supply
You have to supply all the material necessary to install the ReliaSENS unit in place.
For example, if you have to install the ReliaSENS on a pole, you will need two suitable stainless steel band
clamps.
The band clamps must conserve their integrity throughout the lifecycle of the ReliaSENS, even if exposed to
extreme weather conditions. Each band clamp must be able to withstand wind forces, the weight of the
ReliaSENS and the weight of the cables for a total amount of around 37 Kg. If an accidental breakage of one of
the band clamp occurs, the other band clamp must be able to safely hold the ReliaSENS in place.
ReliaSENS description ReliaSENS 18-12-00 User manual
11
RESENS
-18-12-00_UsrMan_EN_1.0
4 ReliaSENS description
The ReliaSENS is a compact and easy to install environment monitoring system provided with Ethernet, Wi-Fi, or
Cellular (GPRS/3G) connectivity.
It analyses air quality parameters, electromagnetic fields, and ionizing radiation levels. Then, using the integrated
Eurotech Cloud technology, sends the acquired data to the Cloud.
The ReliaSENS offers you a user-friendly web interface and a mobile app (ReliaSENS, available on Google Play
Store and App Store) compatible with Android and iOS, that enables immediate data access through PCs, tablets
and smartphones.
You can set thresholds and configure the ReliaSENS to send alerts and generate data logs automatically.
You can easily export data and integrate them with existing IT systems to perform statistical analyses and to
identify spatial and time patterns related to pollutant generation and propagation.
ReliaSENS specifications ReliaSENS 18-12-00 User manual
12
RESENS-18-12-
00_UsrMan_EN_1.0
5 ReliaSENS specifications
GASEOUS POLLUTANTS
RANGE
DETECTION LIMITS
NOTES
CO
0 - 1000 ppm 1.5 ppm
Precision: < 3 ppm
CO
2
2000 - 950000 ppm
100 ppm
Precision: 100 ppm
NO
2
0 - 20 ppm 100 ppb
Precision: < 350 ppb
O
3
0 - 2 ppm
5 ppb
Precision: < 3 ppb
VOC
(1)
0 - 50 ppm
5 ppb
Precision: 3%
NON-GASEOUS POLLUTANTS
RANGE
DETECTION TECHNOLOGY
NOTES
PARTICULATE MATTER PM10
0 to 5440 μg/m
3
Light scattering
Accuracy: 10%
IONIZING RADIATIONS:
α, β, γ
counts (cumulative)
10 - 10
6
Bq
(2)
Geiger tube
NON-IONIZING RADIATIONS:
50 Hz
1 400 μT
Integration
Resolution: 1 μT
900 MHz
1.8 GHz
2.1 GHz
2.4 GHZ
0.01 - 10 V/m
0.05 - 10 V/m
1 - 10 V/m
0.1 - 10 V/m
Rf demodulation
Resolution:0.01 V/m
Resolution:0.05 V/m
Resolution:1 V/m
Resolution:0.1 V/m
PHYSICAL CHARACTERISTICS
DESCRIPTION
POWER REQUIREMENTS
12 Vdc, 5 W
DIMENSIONS (W x H x D)
180 x 360 x 150 mm
EXTERNAL AIR TEMPERATURE
-20 to +50 °C
WEIGHT
6 kg
DATA ACCESS & CONSOLE
DESCRIPTION
CONNECTIVITY OPTIONS
GPRS/3G
Wi-Fi (802.11b/g/n)
10/100/1000 Ethernet
MANAGEMENT INTERFACE
Web
APPLICATION GUI
Web
Mobile App (Android, iOS)
(1)
VOC (Volatile Organic Compounds) includes: Benzene, Toluene, Xylene
(2)
10 Bq = natural radiation background - sea level
0.5 MHz - 8 GHz
Install the ReliaSENS ReliaSENS 18-12-00 User manual
13
RESENS
-18-12-00_UsrMan_EN_1.0
6 Install the ReliaSENS
This operation must be carried out by authorized personnel, observing all applicable safety rules. When
applicable, the operator must have the appropriate license for working at the installation height.
The ReliaSENS must be installed at a height of 3 meters or more, for example on a pole, and would not normally
be accessible to people.
The ReliaSENS must be installed by using two suitable stainless steel band clamps (not provided). These band
clamps must conserve their integrity throughout the lifecycle of the ReliaSENS, even if exposed to extreme
weather conditions. Each band clamp must be able to withstand wind forces, the weight of the ReliaSENS and
the weight of the cables for a total amount of around 37 Kg. If an accidental breakage of one of the band clamp
occurs, the other band clamp must be able to safely hold the ReliaSENS in place.
Install the ReliaSENS in such a way that it is protected from lightning strikes. If required, a lightning rod should be
installed in close proximity to the ReliaSENS.
The minimum distance from the human body to which the device complies with the European requirements on
human exposure to electromagnetic fields (Council Recommendation 1999/519/EC) is 8.5 cm.
Install the ReliaSENS ReliaSENS 18-12-00 User manual
14
RESENS-18-12-
00_UsrMan_EN_1.0
6.1 Install procedure
1. Set up two suitable stainless steel band clamps and unfasten them completely
2. Insert the two stainless steel band clamps in the slots
Install the ReliaSENS ReliaSENS 18-12-00 User manual
15
RESENS
-18-12-00_UsrMan_EN_1.0
3. Install the device on the pole making sure to observe the “UP” and “DOWN” directions
4. Fasten the two stainless steel band clamps
5. Connect the power cable (the red connector is for VIN+; the black connector is for VIN–)
6. Connect also the Ethernet cable if your applcation requires a wired LAN connection
TOP
Power supply the ReliaSENS ReliaSENS 18-12-00 User manual
16
RESENS-18-12-
00_UsrMan_EN_1.0
7 Power supply the ReliaSENS
Set up a proper PSU unit to give the ReliaSENS the following power supply parameters:
Specification
Description
Power supply
12 VDC
Power consumption
5 W
Establish a network connection ReliaSENS 18-12-00 User manual
17
RESENS
-18-12-00_UsrMan_EN_1.0
8 Establish a network connection
Your ReliaSENS is already configured to automatically connect to the Eurotech Cloud platform - Everyware
Cloud (EC).
On ReliaSENS three network interfaces are available by default:
1. 3G/GPRS interface (ppp0). This is the preferred option for connectivity since your unit comes with a
configured cellular module and a microSIM card already included. This interface is configured in WAN
mode using DHCP in order to automatically get an IP address, netmask and DNS from your provider
2. Wi-Fi interface (wlan0). This interface is configured in LAN mode in order to assign IP addresses of the
class 172.16.1.X in DHCP mode.
3. Ethernet (eth0). This interface is configured in LAN mode in order to assign IP addresses of the class
172.16.0.X in DHCP mode.
8.1 Using a wireless 3G/GPRS connection
On ReliaSENS the preferred option for connectivity is 3G/GPRS since your unit comes with a configured cellular
module and a microSIM card already included.
So, after the unit is powered on, it automatically connects to the internet and starts sending data that can be
viewed as described in Section 7.
You may want however to reconfigure you cellular interface.
Please, note that the PIN has to be disabled on the microSIM you use.
You can reconfigure the cellular interface from the Eurotech Gateway Admin through the eth0 wired interface.
Follow these steps:
1. Configure the Ethernet interface of your PC in DHCP
2. Connect the Ethernet cable of the ReliaSENS unit to your PC. Use the cable included in the cable kit for this
purpose.
3. After a few minutes your PC will receive an IP address from the ReliaSENS unit in the class 172.16.0.X
4. Browse the Eurotech Gateway Admin interface to: http://172.16.0.1/esf
5. Insert the username and password provided to you by your Eurotech sales representative
6. Select Network in the System menu
Establish a network connection ReliaSENS 18-12-00 User manual
18
RESENS-18-12-
00_UsrMan_EN_1.0
7. Select ppp0 in the Interface menu and enable it for WAN in the Status. Click “Apply”.
8. Select the “Cellular” tab:
9. Insert the parameters specific for your carrier and press “Apply”.
Establish a network connection ReliaSENS 18-12-00 User manual
19
RESENS
-18-12-00_UsrMan_EN_1.0
8.2 Using a Wi-Fi connection
Your ReliaSENS is natively configured as an Access Point that generates a Wi-Fi network with the following
features:
SSID: ReliaSENS_serialnumber
Password: Enter the Password provided to you by your Eurotech sales representative
Where serialnumber is the serial number for your ReliaSENS.
Using the Wi-Fi network you can directly connect to the ReliaSENS unit and have access to Eurotech Gateway
Admin interface, the local administration user interface of the unit.
Alternatively, you can access the Eurotech Gateway Admin interface from the wired Ethernet connection, as
described in the previous section.
From the Eurotech Gateway Admin you can reconfigure the ReliaSENS unit to operate in “station mode” for
connecting to a local Wi-Fi network.
Follow these steps:
1. Configure the Ethernet interface of your PC in DHCP
2. Connect the Ethernet cable of the ReliaSENS unit to your PC. Use the cable included in the cable kit for this
purpose.
3. After a few minutes your PC will receive an IP address from the ReliaSENS unit in the class 172.16.0.X
4. Browse the Eurotech Gateway Admin interface to: http://172.16.0.1/esf
5. Enter the Username and Password provided to you by your Eurotech sales representative
6. Select Network in the System menu
7. Select the ppp0 interface
8. Select the TCP/IP tab, and:
a) Disable the ppp0 interface by setting Status to “Disabled”
b) Press Apply
NOTE: You must only have 1 interface configured in WAN mode
9. Select the wlan0 interface
10. Select the TCP/IP tab, and:
a) Set Status to “Enabled for WAN”
b) Set Configure to “Manually” or ”Using DHCP” depending on your network features
c) Complete the configuration according to your network features
d) Press Apply
Establish a network connection ReliaSENS 18-12-00 User manual
20
RESENS-18-12-
00_UsrMan_EN_1.0
11. Select the Wireless tab, and:
a) Set Wireless mode to “Station”
b) Complete the configuration according to your network features
c) Press Apply
The unit connects to the local Wi-Fi and, within a few minutes, should get connected to the Cloud.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34

Eurotech ReliaSENS 18-12 Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI