WIKA CPU6000-M tag:model:CPU6000-S tag:model:CPU6000-W Operating instructions

Category
Measuring & layout tools
Type
Operating instructions
Operating instructions
Betriebsanleitung
Manuale d'uso
EN
DE
IT
CalibratorUnit, CPU6000 series
CalibratorUnit, Serie CPU6000
CalibratorUnit, serie CPU6000
CPU6000 series with iPad-App
®
CPB-CAL
2
14097659.04 03/2016 EN/DE/IT
WIKA operating instructions CalibratorUnit, CPU6000 series
EN
DE
IT
Operating instructions model CPU6000 Page 3 - 44
Betriebsanleitung Typ CPU6000 Seite 45 - 86
Manuale d’uso serie CPU6000 Pagina 87 - 127
Further languages can be found at www.wika.com.
© 2014 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA
®
is a registered trademark in various countries.
WIKA
®
ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!
Prima di iniziare ad utilizzare lo strumento, leggere il manuale d’uso!
Conservare per future consultazioni!
3WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
Contents
Contents
1. General information 5
2. Safety 7
2.1 Intended use. . . . . . . . . . . . . . . . . . . . . . . 7
2.2 Personnel qualification . . . . . . . . . . . . . . . . . . . 8
2.3 Special hazards. . . . . . . . . . . . . . . . . . . . . . 8
2.4 Safety instructions for the power supply unit . . . . . . . . . . . . 9
2.5 Labelling, safety marks . . . . . . . . . . . . . . . . . . . 10
2.5.1 Product labels. . . . . . . . . . . . . . . . . . . . . 10
2.5.2 Explanation of symbols. . . . . . . . . . . . . . . . . . 11
3. Specifications 12
3.1 Specifications of the CPU6000 series CalibratorUnit . . . . . . . . . 12
3.2 Dimensions in mm / in . . . . . . . . . . . . . . . . . . . 15
3.2.1 Weather station, model CPU6000-W . . . . . . . . . . . . . 15
3.2.2 Pressure balance sensor box, model CPU6000-S . . . . . . . . 16
3.2.3 Digital multimeter, model CPU6000-M . . . . . . . . . . . . 17
3.3 Pin assignments . . . . . . . . . . . . . . . . . . . . . 18
3.3.1 Weather station, model CPU6000-W . . . . . . . . . . . . . 18
3.3.2 Model CPU6000-S pressure balance sensor box or
model CPU6000-M digital multimeter . . . . . . . . . . . . .19
3.4 Dimensions with wall mounting in mm / in . . . . . . . . . . . . .20
3.4.1 Model CPU6000-W weather station or
model CPU6000-M digital multimeter . . . . . . . . . . . . .20
3.4.2 Isometric view . . . . . . . . . . . . . . . . . . . . . 20
3.4.3 Pressure balance sensor box, model CPU6000-S . . . . . . . . 21
3.4.4 Isometric view . . . . . . . . . . . . . . . . . . . . . 21
4. Design and function 22
4.1 Description . . . . . . . . . . . . . . . . . . . . . . . 22
4.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . 22
4.3 Weather station, model CPU6000-W . . . . . . . . . . . . . . 23
4.4 Pressure balance sensor box, model CPU6000-S . . . . . . . . . . 24
4.5 Digital multimeter, model CPU6000-M . . . . . . . . . . . . . . 25
4.6 Sensor connection to model CPU6000-S pressure balance sensor box. . . 26
4.6.1 Piston temperature sensor, built-in version . . . . . . . . . . . 26
4.6.2 Float position sensor, built-in version . . . . . . . . . . . . . 26
4.6.3 Cable version - Standard - Piston temperature sensor . . . . . . . 27
4.6.4 Cable version - Standard - Floating position sensor . . . . . . . . 27
4.7 Connection of pressure transmitter to model CPU6000-M digital multimeter . 27
4 WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
Contents
Declarations of conformity can be found online at www.wika.com.
5. Transport, packaging and storage 28
5.1 Transport . . . . . . . . . . . . . . . . . . . . . . . .28
5.2 Packaging . . . . . . . . . . . . . . . . . . . . . . . 28
5.3 Storage . . . . . . . . . . . . . . . . . . . . . . . . 28
6. Commissioning, operation 29
6.1 Connecting the CPU6000 series . . . . . . . . . . . . . . . .29
6.2 Setting up the models CPU6000 . . . . . . . . . . . . . . . .30
6.2.1 Setting up model CPU6000-W . . . . . . . . . . . . . . . 30
6.2.2 Setting up model CPU6000-S . . . . . . . . . . . . . . . 31
6.2.3 Setting up model CPU6000-M . . . . . . . . . . . . . . . 33
6.3 Mass calculation . . . . . . . . . . . . . . . . . . . . . 34
7. Remote instruction set 35
8. Maintenance, cleaning and calibration 38
8.1 Maintenance. . . . . . . . . . . . . . . . . . . . . . . 38
8.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . 38
8.3 Recalibration. . . . . . . . . . . . . . . . . . . . . . . 38
9. Faults 39
10. Dismounting, return and disposal 40
10.1 Dismounting . . . . . . . . . . . . . . . . . . . . . . .40
10.2 Return . . . . . . . . . . . . . . . . . . . . . . . . .40
10.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . 41
11. Accessories 42
Appendix: EC declaration of conformity 43
5WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
1. General information
The model CPU6000 CalibratorUnits described in these operating instructions have
been designed and manufactured using state-of-the-art technology. All components
are subject to stringent quality and environmental criteria during production. Our
management systems are certified to ISO 9001 and ISO 14001.
These operating instructions contain important information on handling the instrument.
Working safely requires that all safety instructions and work instructions are observed.
Observe the relevant local accident prevention regulations and general safety
regulations for the instrument’s range of use.
The operating instructions are part of the product and must be kept in the immediate
vicinity of the instrument and readily accessible to skilled personnel at any time. Pass
the operating instructions onto the next operator or owner of the instrument.
Skilled personnel must have carefully read and understood the operating instructions
prior to beginning any work.
The general terms and conditions contained in the sales documentation shall apply.
Subject to technical modifications.
Further information:
- Internet address: www.wika.de / www.wika.com
- Relevant data sheet: CT 35.02
- Application consultant:
Tel.: +49 9372 132-0
Fax: +49 9372 132-5049
1. General information
6 WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
1. General information
Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious injury
or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries
or damage to equipment or the environment, if not avoided.
DANGER!
... identifies hazards caused by electrical power. Should the safety
instructions not be observed, there is a risk of serious or fatal injury.
Information
... points out useful tips, recommendations and information for efficient and
trouble-free operation.
Abbreviations
2-wire The two connection lines are used for the voltage supply.
The measurement signal also provides the supply current.
3-wire Two connection lines are used for the voltage supply.
One connection line is used for the measurement signal.
UB Positive power supply terminal
S+ Positive output terminal
7WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
2. Safety
2. Safety
WARNING!
Before installation, commissioning and operation, ensure that the
appropriate CalibratorUnit has been selected in terms of measuring range,
design and specific measuring conditions.
Non-observance can result in serious injury and/or damage to the
equipment.
Further important safety instructions can be found in the individual chapters
of these operating instructions.
2.1 Intended use
The CPU6000 series is made up of three instruments: the model CPU6000-W weather
station, the model CPU6000-S pressure balance sensor box and the model CPU6000-M
digital multimeter. All of these products serve for the acquisition of measured value data,
relevant for calibration certificates. The new calibration concept simplifies the generation
of certificates and increases the productivity and quality of the certificate generation
process.
The instrument has been designed and built solely for the intended use described here,
and may only be used accordingly.
The technical specifications contained in these operating instructions must be observed.
Improper handling or operation of the instrument outside of its technical specifications
requires the instrument to be taken out of service immediately and inspected by an
authorised WIKA service engineer.
Handle electronic precision measuring instruments with the required care (protect from
humidity, impacts, strong magnetic fields, static electricity and extreme temperatures,
do not insert any objects into the instrument or its openings). Plugs and sockets must be
protected from contamination.
If the instrument is transported from a cold into a warm environment, the formation of
condensation may result in instrument malfunction. Before putting it back into operation,
wait for the instrument temperature and the room temperature to equalise.
The manufacturer shall not be liable for claims of any type based on operation contrary to
the intended use.
8 WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
2. Safety
2.2 Personnel qualification
WARNING!
Risk of injury should qualification be insufficient!
Improper handling can result in considerable injury and damage to
equipment.
The activities described in these operating instructions may only be carried
out by skilled personnel who have the qualifications described below.
Special operating conditions require further appropriate knowledge, e.g. of aggressive
media.
Skilled electrical personnel
Skilled electrical personnel are understood to be personnel who, based on their technical
training, know-how and experience as well as their knowledge of country-specific
regulations, current standards and directives, are capable of carrying out work on
electrical systems and independently recognising and avoiding potential hazards.
The skilled electrical personnel have been specifically trained for the work environment
they are working in and know the relevant standards and regulations.
The skilled electrical personnel must comply with current legal accident prevention
regulations.
2.3 Special hazards
WARNING!
For hazardous media such as oxygen, acetylene, flammable or toxic gases
or liquids, and refrigeration plants, compressors, etc., in addition to all
standard regulations, the appropriate existing codes or regulations must
also be followed.
Residual media on the test item and/or sensors of the CalibratorUnit can
result in a risk to personnel, the environment and equipment. Take sufficient
precautionary measures.
9WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
2. Safety
2.4 Safety instructions for the power supply unit
DANGER!
Danger of death caused by electric current
Upon contact with live parts, there is a direct danger of death.
Operation using a defective power supply unit (e.g. short-circuit from
the mains voltage to the output voltage) can result in life-threatening
voltages at the instrument!
If there is any visible damage to the case or the wiring, do not use the
power supply unit!
Never install nor store the power supply unit in the following locations, as
this can lead to a failure in operation:
- places where there is strong humidity or condensation
- places which have specific environmental influences
- places which are subject to constant vibration
- places which are subject to strong fluctuations in temperature
- outdoors
During thunderstorms or when it won’t be used for long periods,
disconnect the power supply unit from the mains supply.
Do not pull on the cable in order to disconnect the power supply unit from
the mains supply.
The power supply unit is maintenance-free. It must not be opened (risk of
electrical shock).
No modification of the power supply unit is permitted (voids the
warranty).
The power supply unit may only be repaired by authorised skilled
personnel.
Before cleaning, disconnect the power supply unit from the mains supply.
Do not clean with chemical cleaning fluids. Only clean with a dry cloth.
Dispose of the power supply unit correctly (Waste Electrical and
Electronic Equipment in accordance with local regulations).
The power supply unit should only be operated in an ambient temperature of 0 ... 40 °C
(32 ... 104 °F) (humidity: to 90 % relative humidity, non-condensing).
10 WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
2.5 Labelling, safety marks
2.5.1 Product labels
Weather station, model CPU6000-W
Pressure balance sensor box, model CPU6000-S
Digital multimeter, model CPU6000-M
8
1
7
2
6
3
5
4
Serial no. and date of manufacture
Ambient temperature
Relative humidity
Atmospheric pressure
WIKA-Wireless antenna
Interface connector
Power supply
Model
1
2
3
4
5
6
7
8
7
1
6
2
5
3
4
Serial no. and date of manufacture
Piston temperature
Floating position of the masses
WIKA-Wireless antenna
Interface connector
Power supply
Model
1
2
3
4
5
6
7
6
1
5
2
4
3
Serial no. and date of manufacture
Interface
WIKA-Wireless antenna
Interface connector
Power supply
Model
1
2
3
4
5
6
2. Safety
11WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
2.5.2 Explanation of symbols
Before mounting and commissioning the instrument, ensure you read the
operating instructions!
CE, Communauté Européenne
Instruments bearing this mark comply with the relevant European directives.
This marking on the instruments indicates that they must not be disposed of
in domestic waste. The disposal is carried out by return to the manufacturer
or by the corresponding municipal authorities (see EU directive 2012/19/EU).
2. Safety
12 WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
3. Specifications
3.1 Specifications of the CPU6000 series CalibratorUnit
Weather station, model CPU6000-W
Symbol Description Measuring range Accuracy
Ambient temperature
-40 ... +123.8 °C
(-40 ... +254.84 °F)
±0.5 °C (0.9 °F)
1)
Relative humidity 0 … 100 % r. h. ±5.0 % r. h.
2)
Atmospheric
pressure
550 … 1,150 mbar abs.
(7.85 ... 16.68 psi abs.)
(opt. 551 … 1,172 mbar abs.)
(8 ... 17 psi abs.))
0.05 % FS
(opt. 0.01 % of reading)
1) The calibration is performed for the points 18 °C (64.4 °F), 23 °C (73.4 °F) and 25 °C (77 °F) following a warm-up time of
120 minutes.
2) The calibration is performed for the points 30 % r. h., 50 % r. h. and 80 % r. h. following a warm-up time of 120 minutes.
Pressure balance sensor box, model CPU6000-S
Symbol Description Measuring range Accuracy
Piston temperature
-50 ... +250 °C
(-58 ... +482 °F)
±0.1 °C at 20 °C
(±0.18 °F at 68 °F)
1)
Floating position of
the masses
4 … 20 mA
20 … 200 mm
(0.79 ... 7.87 in)
±0.5 mm (±0.02 in)
1) The calibration is performed for the points 18 °C (64.4 °F), 23 °C (73.4 °F) and 25 °C (77 °F) following a warm-up time of
120 minutes.
3. Specifications
13WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
It is recommended to install the sensor box for pressure balances of the
CPB5000, CPB5000HP and CPB5800 series.
Digital multimeter, model CPU6000-M
Symbol Description
+ mA -
Current input
Measuring range 0 … 20 mA
Accuracy ±5 µA
+ V -
Voltage input
Measuring range 0 … 10 V
Accuracy ±1 mV
+ Out -
Voltage supply DC 24 V (load: max. 50 mA; min. 20 mA)
Interface
Model D-10
Models CPT6100 and CPT6180
RS-232 interface, 9600 Baud, 8N1
Base instrument Model
CPU6000-W
Model
CPU6000-S
Model
CPU6000-M
Voltage supply
Power supply DC 24 V, 625 mA
Permissible ambient conditions
Operating temperature 15 ... 45 °C (59 ... 113 °F)
Storage temperature 0 ... 70 °C (32 ... 158 °F)
Communication
Interface USB with PC, WIKA-Wireless 4.0 with iPad
®
3.0 or higher
WIKA-Wireless
Frequency range 2.402 ... 2.480 GHz
HF output power max. 2 dBm (+ 2 dBi)
3. Specifications
14 WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
Base instrument Model
CPU6000-W
Model
CPU6000-S
Model
CPU6000-M
Case
Dimensions
(W x H x D)
190.4 x 58 x 110 mm
(7.50 x 2.28 x 4.33 in)
190.4 x 58 x 80 mm
(7.50 x 2.28 x 3.15 in)
190.4 x 58 x 110 mm
(7.50 x 2.28 x 4.33 in)
Weight 1.44 kg (3.18 lbs) 1.18 kg (2.60 lbs) 1.29 kg (2.84 lbs)
Approvals
Logo Description Country
EC declaration of conformity
EMC directive 2004/108/EC
2004/108/EC, EN 61326-1 Emission (group 1, class B) and
interference immunity (“basic environment”: Commercial,
laboratories, service centres or workshops)
R&TTE directive 1999/5/EC
EN 300 228, harmonised frequency range
2,400 … 2,500 MHz is used
The instrument may be used without limitations in the EU
and also CH, NO and LI.
European Community
EAC
Electromagnetic compatibility
Low voltage directive
Eurasian Economic
Community
Certificates
Certificate
Calibration Standard: 3.1 calibration certificate per DIN EN 10204
Option: DKD/DAkkS calibration certificate
Recommended
recalibration interval
1 year (dependent on conditions of use)
Approvals and certificates, see website
For further specifications see WIKA data sheet CT 35.02 and the order documentation.
3. Specifications
15WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
3. Specifications
3.2 Dimensions in mm / in
3.2.1 Weather station, model CPU6000-W
16 WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
3. Specifications
3.2.2 Pressure balance sensor box, model CPU6000-S
1
Piston temperature sensor connector
2
Floating position sensor connector
1 2
17WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
3.2.3 Digital multimeter, model CPU6000-M
1
RS-232 interface
2
Current input
3
Voltage input
4
Voltage supply for test item
1 2 3 4
3. Specifications
18 WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
3.3 Pin assignments
3.3.1 Weather station, model CPU6000-W
1
On/Off switch
2
Power connection DC 24 V
3
USB interface
4
4 mm hose connection for internal barometric reference
5
WIKA-Wireless 4.0 antenna
6
Humidity-temperature sensor
1 2 3 4
6 5
3. Specifications
19WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
3.3.2 Model CPU6000-S pressure balance sensor box or
model CPU6000-M digital multimeter
1
On/Off switch
2
Power connection DC 24 V
3
USB interface
4
WIKA-Wireless 4.0 antenna
1 2 3 4
3. Specifications
20 WIKA operating instructions CalibratorUnit, CPU6000 series
14097659.04 03/2016 EN/DE/IT
EN
3.4 Dimensions with wall mounting in mm / in
3.4.1 Model CPU6000-W weather station or model CPU6000-M digital multimeter
3.4.2 Isometric view
1
1
Screw hole closed with decorative
strips
3. Specifications
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128

WIKA CPU6000-M tag:model:CPU6000-S tag:model:CPU6000-W Operating instructions

Category
Measuring & layout tools
Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI