Zimmer enPulsPro User manual

Category
Skin care appliances
Type
User manual
User Manual
enPulsPro
USA
Illustrations
Front of the device
Fig. 1
1 Control unit
2 Pulse energy controller
3 Display
4 Frequency controller
5 Slot for SD card
6 Power switch
7 Handpiece
8 Vent, front
9 Vent with fan, rear
10 Holder for hand piece
11 Footswitch
12 Rotatable swivel base
2
5
7
8
4
3
10
11
1
9
12
6
Illustrations
Rear of the device
Fig. 2
13 Socket for mains cable
14 Mains fuse
15 Socket for handpiece channel I
16 Socket for handpiece channel II
17 Socket for footswitch
18 Identification plate
16
15
17
13
18
14
Illustrations
Screens / Displays
Fig. 3
19 On-screen buttons
20 Status bar
21 Navigation bar
22 VAS- Pre scale
23 VAS- Post scale
Fig. 4
Navigation bar
Description of the
functions
(A) Back
Takes you back one step
(B) Therapy
Switches to the therapy screen
(C) History
(D) Protocols
Switches to the VAS evaluation
Switches to protocols
(E) Programs
Switches to the program list
(F) Favorites
Switches to the favorites area
21
19
20
F
E
D
C
B
A
22
23
Illustrations
Applicator heads and accessories
Fig. 5
24 Applicator head, 25 mm
25 Applicator head, 15 mm
26 Applicator head, 6 mm
27 Silicone protection cap
24
25
26
27
Explanation of Symbols
In the instructions for use, this symbol indicates danger.
Caution!
In the instructions for use, this symbol indicates a "Caution" with regard to
possible damage of the device.
Socket for handpiece
Socket for footswitch
User Manual
Follow user manual
Serial number
Article number
Manufacturer
Date of manufacture
Applied part type BF
Value of accessible fuses
Interval operation - Follow user manual
Contents
Illustrations
Front of the device
Rear of the device
Screens / Displays
Applicator heads and accessories
Explanation of Symbols
Page
1
Indications / Contraindications
1
2
Side Effects
2
3
Application Information
3
4
Warnings
4
5
enPulsPro in brief
6
6
Device Set-Up
7
7
Settings
8
8
Operating Instructions
8.1 Device Description
8.2 Notes on Operation
8.3 Performing the Treatment
8.4 Display and Buttons
8.5 SD card
8.6 Favorites List Retrieving Programs, Editing Lists
8.7 VAS Visual Analogue Scale
10
12
13
14
16
17
20
9
Technical Information
9.1 Info-Chart Duration of Treatment
24
26
10
Cleaning / Disinfection
27
Contents
11
Scope of Delivery and Accessories
31
12
Device Combination
32
0
13
Safety and Maintenance
13.1 Safety
13.2 Maintenance
33
34
14
Functional Test
35
32
15
Safety Check / Metrological Control
36
16
Error Messages / Troubleshooting / Disposal
37
17
Manufacturer´s EMC Declaration
39
Valid for the device enPulsPro (NG).
This user manual is an integral part of the device.
It must be stored with the device and kept accessible at all times for anyone authorized to
operate this device. Full observation of these instructions is a requirement for the correct
application and operation of the equipment and for consequent safety for patient and
operator.
The user manual is valid as of September 2019.
U.S.A. Federal Law restricts this device to sale by or on the order of a practitioner
licensed by the law of the State in which he/she practices to use or order the use of
the device.
Indications / Contraindications
1
Page 1
Indications
For relief of minor muscle aches and pains and for temporary increase in
local blood circulation.
Contraindications
vascular diseases present in or near the treatment area
local infections in the treatment area
around malignant or benign tumors
directly on cartilage surfaces or near the small facet joints of the spinal
column
directly over implanted electronic devices such as pacemakers, analgesic
pumps, etc.
in areas, in which mechanical energy in the form of vibrations may lead to
tissue damage such as metal implants after a fracture
In general we advise against treatments
if blood clotting disorders are present or the patient is receiving treatment that
results in a change in the blood clotting behavior
during pregnancy
on patients with neurological diseases resulting in impairment of the
vasomotor function in the treatment area
over air-filled cavities such as treatment on the thoracic spine, etc.
on children, particularly around the epiphyseal plates
Care is required for patients
with impaired sensibility
with severe autonomic disorders
under the influence of drugs and/or alcohol
as circulatory stresses and inadequate treatment responses cannot be
excluded.
Side Effects
2
Page 2
Side Effects
Treatments with enPulsPro may occasionally cause irritation, petechial,
bruising, swelling or pain.
Application Information
3
Page 3
Before using the device on a patient, the user should become acquainted with
the operating instructions and individual treatment methods as well as the
indications / contraindications, warnings and application information. Additional
sources of information about types of therapy should be consulted.
Caution!
Before use, ensure that the device is powered via a properly earthed plug with
a grounded outlet (electrical installation according to DIN VDE 0100 Part 710).
The device must only be operated with the supplied power cord. The power
cord must be protected against mechanical stress.
Caution!
Operation of this equipment in the vicinity of strong electromagnetic fields (e.g.
tomography, x-ray or diathermy equipment) may interfere with the operation of
the device. Please keep a safe distance of several meters.
enPulsPro is not suitable for use in areas with an explosive, flammable or
combustive environment.
When in operation, the device must be positioned in such a way that direct
access to the device's central power supply is possible, so that it can be
disconnected from the mains at any time.
To avoid the risk of electric shock, the device must be disconnected from the
mains supply before performing any maintenance or cleaning activities.
Inspect the device prior to use. If damaged, it must not be used.
Caution!
Only accessories provided by Zimmer MedizinSysteme GmbH are to be used.
Caution!
The device may cause malfunctions or may interfere with the operation of
equipment in the vicinity by electromagnetic effects. It may be necessary to
take appropriate remedial action, such as realignment, re-arrangement of the
device or the electromagnetic shielding.
Caution!
The handpiece of the device is not designed for continuous operation. After
max. 6,000 shocks, a break of 15 min. must be observed.
Caution!
If enPulsPro is not mounted on the designated system cart, make sure that
enPulsPro is placed on a stable surface.
Caution!
To avoid heat accumulation in the handpiece, it is essential to ensure that your
hand or any other object does not block the ventilation openings on the top and
especially on the bottom of the handpiece.
Warnings
4
Page 4
Caution!
In the extreme case, the maximum treatment time is limited to 4 min followed
by a break of at least 15 min.
If the treatment time is exceeded, the handpiece may overheat.
Caution!
Users of the enPulsPro radial pulse therapy device must be instructed in the
proper use of the system and must possess the appropriate skills.
Caution!
The device is intended to be used exclusively by medical professionals.
Caution!
Any treatment instructions regarding treatment location, duration and intensity
require medical knowledge and should be given by authorised physicians,
therapists and health paraprofessionals. It is imperative that these instructions
are followed.
The patient must not be left unattended during therapy.
Persons undergoing simultaneous treatment involving the reduction and / or
alteration of blood clotting or the prolongation of the blood clotting time (e.g.
acetylsalicylic acid) should consult their therapist regarding a possible
suspension of this therapy, as the use of radial pulse in such patients may
cause bleeding and bruising in varying degrees.
Radial pulse become significantly scattered in air-filled areas and generate
reflections that can have negative effects.
Therefore, do not perform treatments directly above the lungs (intercostal
space) or the gastrointestinal area.
The use in wet areas is not permitted and may in case of non-compliance lead
to considerable damage to the device and endanger both the patient and the
user.
Dispose of the packaging material properly. Make sure that it is not accessible
to children.
On patients with implants or implanted electronic device do not conduct
treatment before the risk has been assessed and found negligible.
Patients may not be connected to a high-frequency surgical device at the same
time. This can lead to burns.
This device should not be used over swollen or inflamed areas or skin
eruptions. Do not use in presence of unexplained calf pain.
Consult a physician.
Warnings
4
Page 5
In worst case condition maximum therapy time is limited to 4 minutes treatment
time and a break time of 5 min. If you will exceed this duty cycle the handpiece
can be in overtemperature.
enPulsPro - in brief
5
Page 6
What is enPulsPro?
A state of the art therapeutic massager.
Radial pulse therapy
Radial Pulse Therapy is a procedure for relief of minor muscle aches and
pains and for temporary increase in local blood circulation.
What does
enPulsPro do?
The handpiece contains a projectile that is accelerated through the
electromagnetic transfer of kinetic energy. This kinetic energy is transferred
into impact energy in the applicator head. The impact energy delivered from
the applicator head results in radial pulses developed in the target tissue.
With enPulsPro, a maximum penetration depth of about 35 mm into human
tissue can be achieved.
How does the
enPulsPro generate
radial pulses?
An electromagnetic field is generated at the rear end of the handpiece by
means of a coil.
A projectile is accelerated through the field crashing into the applicator head
at the front of the handpiece generating radial pulses that spread radially into
the tissue.
Why enPulsPro?
The innovative technology allows for a compact design without a
compressor.
The modern multi-color display, showing all therapy-relevant parameters, the
modern touch operation and the ability to simultaneously connect 2
handpieces.
Customizable program startup settings and a clear, simple menu offer
maximum convenience for the user.
Adjustable frequencies and various applicators enable the therapy to be
tailored to the respective status of the patient.
What else does
enPulsPro offer?
An integrated VAS scale provides an orienting overview of the chronological
progression and success of the therapy.
Intended Use
enPulsPro is an electromagnetic therapy system for the generation and
application of radial pulses in orthopedics and physiotherapy.
Note:
The application of the device is reserved to medical professionals (such as
physicians, therapists and health paraprofessionals).
enPulsPro has been constructed and designed solely for the application on
superficial skin problems in humans.
The device is intended for use mechanical massage devices.
Device Set-Up
6
Page 7
Note:
If enPulsPro is not mounted on the designated system trolley, please make
sure that it is standing on a stable surface. We recommend using the optional
swivel base.
Note:
Make sure that the power switch on the device is set to "0".
Connecting the
mains cable
Connect the mains cable to the designated socket (13) on the device and
connect the cable to the mains.
Note:
The device must only be connected to power outlets with an earthing contact.
Connecting the
handpiece
Connect the handpiece to one of the designated sockets Channel I (15) or
Channel II (16) and set it down.
Note:
Make sure that an applicator head is placed inserted to handpiece and that it
is correctly and completely screwed in.
Connecting the
footswitch
Connect the footswitch to the designated socket (17) and place it on the floor.
Switching the device
on
Switch the device on using the power switch (6).
Switching the device
off
The device is switched off using the power switch (6).
In order to completely disconnect the device (all-phase) from the mains,
remove the mains cable.
Caution!
All cables must be protected against jamming or other mechanical damage.
Settings
7
Page 8
Note:
Changes to the basic settings can only be made from the start up screen.
Start up screen
Once the device has been switched on and the self-test performed, the home
start up screen opens.
Note:
Pressing the "Therapy" button (1) immediately switches to the therapy screen
Configuration menu
The factory settings can be changed and adjusted individually using the
configuration menu.
Selecting
Configuration
Pressing the "Settings" button (2) opens the "Settings" screen.
The settings options are described below.
By default, the factory settings are pre-programmed as shown on the screen.
1
2
3
4
5
6
7
2
1
8
9
10
9
Settings
7
Page 9
(1) Startup settings
1. Start menu:
Individual startup settings options.
The selection is made directly in the corresponding row.
(2) Language
Selecting the language.
The selection is made directly in the corresponding row.
(3) Welcome
Pressing the "Welcome" field opens a window with an alphabetic keyboard to
enter a welcome message for the start-up screen.
Pressing the “OK” key saves the text that has been entered.
Pressing the “X” key returns you to the configuration menu.
(4) OK
Pressing the button switches to the home screen.
(5) Version
Pressing the "Version" button opens a window with information regarding the
current software version.
(6) Load default
Pressing the "Load default" button restores the default factory settings.
(7) Color scheme
Pressing the key switches between the two display settings. A bright or dark
screen can be selected.
(8) Brightness
The Adjustment of the brightness is made using the two arrow keys.
(9) Volume
The Adjustment of the volume is made using the two arrow keys.
(10) Handpiece
counter status
In this display panel, the meter reading of any connected handpiece(s) is
shown.
(11) Maintenance
After pressing the (1) key in the maintenance menu, the maintenance
message, which appears after 1.8 and 2 million collisions, can be turned off.
Deactivation results in a one-time maintenance message after 1.8 and 2
million shocks.
The maintenance message can only be deactivated by customer service.
The following points are not relevant for the user. They are used only by
customer service.
1
Operating Instructions
8.1 Device Description
8
Page 10
Handpiece
The handpiece (7) contains the radial pulse generator, a cooling fan for heat
dissipation and the socket for the various applicator heads. It is connected to
the control device (1).
Note:
The pulse generator in the handpiece is a wear part and must be replaced
after a certain period of operation, as its ability to function decreases with time.
Zimmer MedizinSysteme GmbH guarantees the unrestricted use of at least
two million shocks per radial pulse generator.
Depending on power and frequency, the generator can produce far in excess
of 2 million radial pulses.
Further information regarding the need to replace the pulse generator can be
found in Chapter 16.
In order to use the handpiece on the patient, an applicator head must be
completely and securely screwed onto the handpiece.
Interval operation
The handpiece is not suitable for continuous operation. The heat resulting
from friction increases the temperature of the applicator head with increasing
duration of treatment, to such a degree that thermal damage to the patient
cannot be ruled out.
Please refer to the maximum values for the treatments in the figure in Chapter
9.1. After reaching the specified values, the treatment must be interrupted and
can only be resumed once the applicator head has cooled down to room
temperature.
Note:
Please note that the temperature monitoring of the handpiece (described on
page 1)1 only demonstrates a technical shutdown to protect the mechanical
parts inside the handpiece. It provides no information regarding the
temperature of the applicator head.
Recommendation:
To avoid interruption to ongoing therapy in treatments with more than 6,000
pulses, the connection of a second hand piece is recommended.
During the cooling phase for one handpiece, you can switch to the second
hand piece and immediately continue the treatment.
Note:
When switching from one handpiece to another during ongoing therapy, the
pulse frequency and the number of shocks remain unchanged. The pulse
energy returns to 0 and must be reset.
Operating Instructions
8.1 Device Description
8
Page 11
Handpiece
temperature
monitoring /
regulation
The generation of mechanical shock energy causes substantial heat
generation within the handpiece. A thermal safety switch is integrated so that
the life of the handpiece is not affected. In the event of overheating, an internal
shut-off switch ensures the cooling down of the handpiece.
In addition to temperature monitoring, enPulsPro features temperature
regulation by means of a temperature sensor in the handpiece. The fan in the
handpiece starts upon activation using the footswitch and automatically stops
when a certain temperature is reached.
When the temperature reaches a critical level, a cooling phase is initiated.
This is indicated in the display by the following message:
"Overtemperature in the applicator. Please allow the applicator to cool."
The emission of pulses is no longer possible.
After confirming the message with "OK", the treatment screen comes to the
foreground. In the selection window of the handpiece, the handpiece is shown
a deactivated. The time required for the applicator to cool down is displayed.
Applicator heads
There are 3 different applicator heads available for therapy.
Changing the
applicator heads
To change the different applicator heads, hold the handpiece with one hand
and unscrew the applicator head from the handpiece with the other hand,
turning in an anti-clockwise direction. Then screw in the desired head, turning
clockwise until it stops.
Note:
The applicator heads are subject to wear and must be replaced after a certain
period of operation (see Chapter 13.2 Maintenance).
Footswitch
Position the corresponding foot switch in such a way that is can be easily
reached during therapy. The control element of the switch is not direction-
dependent, meaning that the precise alignment of the footswitch is not
necessary.
To avoid damage, ensure that only light pressure is applied to the switch. The
switch should be operated with the front of the foot, not the heel.
The switch has no locking mechanism, i.e. it only remains activated for as long
as pressure is applied to the switch.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52

Zimmer enPulsPro User manual

Category
Skin care appliances
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI