Midmark 646 Podiatry Procedures Chair User guide

Type
User guide
User Guide
TP202 20-42-FO-00014 Rev A2 C2169
Podiatry
Procedures Chair
For Models:
646
Style P
003-10160-99 Rev. AA3 (8/9/19)
English
EspaƱol
FranƧais
English - 2
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
MA10683i
Product Information
(The information below is required when calling for service.)
Model / Serial Number:
Date of Purchase:
Midmark Authorized Service
Company:
Model / Serial
Number Label
Dealer :
English - 3
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
Important Information
Safety Symbols ...........................................................................................................4
Transportation / Storage Conditions ...........................................................................5
Disposal of Equipment ................................................................................................5
Intended Use ..............................................................................................................6
Electrical Requirements ..............................................................................................6
Electromagnetic Interference ......................................................................................6
EMC - Manufacturerā€™s Declaration and Guidance ...........................................................7
Power to the Chair.................................................................................................................8
Connecting the Foot Control ................................................................................................9
Touch Pad Location ............................................................................................................10
Operation
Back / Tilt Operation ...........................................................................................................11
QuickChair Function ...........................................................................................................12
Foot Section / Debris Tray ..................................................................................................13
Chair Arms ............................................................................................................................ 14
Maintenance
Calling for Service ....................................................................................................15
Cleaning ...................................................................................................................15
Preventative Maintenance..................................................................................................15
Troubleshooting
Powered Function(s) Not Working Properly ....................................................................16
Crash Avoidance System ...................................................................................................16
Speciī‚æcations
Range of Motion / Speciī‚æcations Table ............................................................................17
Dimensions ........................................................................................................................... 18
Model Identiī‚æcation / Compliance Chart ..........................................................................19
Warranty Information
Warranty Information ..........................................................................................................20
Table of Contents
English - 4
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
8A
250 V
F1
F2
015-3495-00
015-10312-00
F1
250 V
8A
250mA
F2
6.3A
250 V
250 V
Keep dry
Safety Symbols
WARNING
Indicates a hazardous situation which could result in serious injury if not avoided.
CAUTION
Indicates a potentially hazardous situation which could result in minor or moderate injury if
not avoided. It may also be used to alert against unsafe practices.
Equipment Alert
Indicates a potentially hazardous
situation which could result in equipment
damage if not avoided.
Note
ī€ƒ Ampliī‚æesī€ƒaī€ƒprocedure,ī€ƒpractice,ī€ƒorī€ƒcondition.
Protective earth ground
Duty Cycle (motor run time):
30ī€ƒsecondsī€ƒON,ī€ƒ5ī€ƒminutesī€ƒOFF
Fuse replacement label
Typeī€ƒB,ī€ƒAppliedī€ƒPart
Proper Shipping
Maximum stacking height
(palletted units)
Fragile
Important Information
Emergency Stop
Refer to manual
Manufacturer
Catalogue Number
Serial Number
Symbol Glossary
English - 5
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
Transportation / Storage Conditions
Ambient Temperature Range: .................................................................+41Ā°Fī€ƒtoī€ƒ100Ā°Fī€ƒ(+5Ā°Cī€ƒtoī€ƒ+38Ā°C)
Relative Humidity....................................................................................10% to 90% (non condensing)
Atmospheric Pressure .............................................................500hPaī€ƒtoī€ƒ1060hPaī€ƒ(0.49atmī€ƒtoī€ƒ1.05atm)
Disposal of Equipment
Atī€ƒtheī€ƒendī€ƒofī€ƒproductī€ƒlife,ī€ƒtheī€ƒchair,ī€ƒaccessories,ī€ƒandī€ƒotherī€ƒconsumableī€ƒgoodsī€ƒmayī€ƒbecomeī€ƒ
contaminatedī€ƒfromī€ƒnormalī€ƒuse.ī€ƒConsultī€ƒlocalī€ƒcodesī€ƒandī€ƒordinancesī€ƒforī€ƒproperī€ƒdisposalī€ƒofī€ƒequipment,ī€ƒ
and other consumable goods.
English - 6
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
Intended Use
This product is intended to be used as a chair / table to provide positioning and support for patients during
examinationsī€ƒandī€ƒproceduresī€ƒconductedī€ƒbyī€ƒqualiī‚æedī€ƒmedicalī€ƒprofessionalsī€ƒ(primarily podiatrists).
CAUTION
Be sure that all persons and equipment are clear of the table before activating any
function. Failure to do so could result in personal injury.
Electrical Requirements
CAUTION
When using high frequency surgical devices or endocardial catheters:
ā€¢ Use non-conductive material to insulate patient from metal portions of the chair.
ā€¢ Consult operating instructions for the device before using in conjunction with drawer
or upholstery heater.
Failure to comply may result in electric shock or burns to patient.
Electromagnetic Interference
Thisī€ƒproductī€ƒisī€ƒdesignedī€ƒandī€ƒbuiltī€ƒtoī€ƒminimizeī€ƒelectromagneticī€ƒinterferenceī€ƒwithī€ƒotherī€ƒdevices.ī€ƒHowever,ī€ƒifī€ƒ
interference is noticed between another device and this product:
ī€ƒ ā€¢ī€ƒ ī€ƒRemoveī€ƒinterferingī€ƒdeviceī€ƒfromī€ƒroom
ī€ƒ ā€¢ī€ƒ ī€ƒPlugī€ƒchairī€ƒintoī€ƒisolatedī€ƒcircuit
ī€ƒ ā€¢ī€ƒ ī€ƒIncreaseī€ƒseparationī€ƒbetweenī€ƒchairī€ƒandī€ƒinterferingī€ƒdevice
ī€ƒ ā€¢ī€ƒ ī€ƒContactī€ƒMidmarkī€ƒifī€ƒinterferenceī€ƒpersists
Equipment Alert
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the
Exam Chair including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
Warning
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
Warning
No modification of this equipment is allowed.
English - 7
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
Immunity Test Immunity Test Level
Electrostatic Discharge Ā±ī€ƒ8ī€ƒkVī€ƒContact,ī€ƒĀ±15ī€ƒkVī€ƒAir
Radiated RF EM Fields 3V/m
80MHz-2.7 GHz 80% AM at 1kHz
Rated Power Frequency Magnetic Fields 30 A/m
Electrical Fast Transients Ā±2 kV at 100kHz repetition frequency
Surges Line to Line: Ā±1 kV
Line to Ground: Ā±2 kV
Conducted RF 3V:ī€ƒ0.15MHzī€ƒ-80MHz
6Vī€ƒinī€ƒISMī€ƒBandsī€ƒbetweenī€ƒ0.15MHzī€ƒ-ī€ƒ80MHz
80% AM at 1kHz
Voltage Dips 0ī€ƒ%ī€ƒUT;ī€ƒ0,5ī€ƒcycle
Atī€ƒ0Ā°,ī€ƒ45Ā°,ī€ƒ90Ā°,ī€ƒ135Ā°,ī€ƒ180Ā°,ī€ƒ225Ā°,ī€ƒ270Ā°ī€ƒandī€ƒ315Ā°
0 % UT; 1 cycle
and
70ī€ƒ%ī€ƒUT;ī€ƒ25/30ī€ƒcycles
Single phase: at 0Ā°
Voltage Interruptions 0ī€ƒ%ī€ƒUT;ī€ƒ250/300ī€ƒcycle
Note
The emissions characteristics of this equipment make it suitable for use in industrial areas and
hospitals. The exam chair is not intended to be used in a residential environment.
EMC - Manufacturerā€™s Declaration and Guidance
Radiated Test Radiated Test Level
Conducted and Radiated RF Emissions CISPR 11 Group 1 Class A
English - 8
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
MA10684i
5
A
M
P
5
A
M
P
Power to the Chair
The power indicator light is illuminated when power is properly connected to the chair.
Ifī€ƒtheī€ƒchairī€ƒhasī€ƒnoī€ƒpower,ī€ƒallī€ƒfunctionsī€ƒwillī€ƒbeī€ƒinoperable.
Power Indicator
Light
Note
The chair has a dual power cord. Both cords
must be connected for proper operation.
Circuit Breakers
Tripped
(no power)
Reset
(power)
To reset circuit breaker(s)...
Push the circuit breaker toggle switch.
Note: The circuit breakers supply voltage to receptacles only.
English - 9
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
MA10685i
To connect the foot control cord...
A) Slide narrow portion of cord connector into bracket.
B) Connect cord at desired inlet (see Note).
C) Tighten locking ring.
Note: The cord connector is ā€œkeyedā€ and must be aligned with inlet.
Connecting the Foot Control
The foot control cord inlet is located on the base column at the foot end of the chair.
Locking Ring
Cord Inlet
Note
Chair must be unplugged before installing foot control.
English - 10
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
MA10686i
Touch Pad Location
Chair functions can be activated from either of the touch pads located on the foot section.
Equipment Alert
The foot section is not a seat. Patients should
sit only on the seat section of the chair. Sitting on the foot
section may result in damage or inoperable equipment.
English - 11
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
Back / Tilt Operation
To operate any of the chairā€™s powered functions...
Press & hold the Up or Down arrow button for the desired function.
Note: In the event of a malfunction, press the Stop button to terminate movement.
Back Up / Down
Tilt Up / Down
Stop
Operation
Back Up / Down
Tilt Up / Down
Stop
CAUTION
When the Back function is in full down
position, and the Tilt function is in full up position,
there is less than 2 in (5 cm) of clearance between
the back section and the īƒ€oor. Keep all objects &
personnel clear to prevent damage / injury.
English - 12
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
QuickChair Function
The QuickChairā„¢ function moves the table into the ā€œchair positionā€ with the touch of a single button.
Functions activated during QuickChair Operation
Back Up / Tilt Down
Stop
QuickChair
QuickChair
Stop
To activate the QuickChair function...
A) Press & hold the QuickChair button.
B) Release button when table reaches the chair position.
Note: In the event of a malfunction, press the Stop button to terminate movement.
English - 13
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
MA10688i
MA10687i
Foot Section / Debris Tray
The foot section can be extended up to 9 in. (23 cm) to accommodate taller patients.
The retractable debris tray is located directly under the foot section.
To access the debris tray...
Pull tray outward.
To extend / retract the foot section...
A) Squeeze handle on either side of foot section.
B) Move foot section in / out to desired position.
C) Release handle(s) to lock in position.
English - 14
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
MA10689i
To rotate the chair arm(s)...
A) Press the pivot lever.
B) Rotate the chair arm to desired position.
C) Release pivot lever to lock in position.
Pivot Lever
Chair Arms
The chair arms can be rotated 90Ā° to allow the patient to mount / dismount the chair.
English - 15
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
Maintenance
Calling for Service
Ifī€ƒserviceī€ƒisī€ƒrequired,ī€ƒcontactī€ƒyourī€ƒauthorizedī€ƒMidmarkī€ƒdealer.
To contact Midmark directly:
ī€ƒ 1-800-MIDMARKī€ƒ(1-800-643-6275)
ī€ƒ 8:00ī€ƒamī€ƒuntilī€ƒ5:00ī€ƒpmī€ƒMondayī€ƒ-ī€ƒFridayī€ƒ(EST)ī€ƒ[excludingī€ƒstandardī€ƒU.S.ī€ƒholidays]
Note
Model / serial number information is required when calling for service.
Preventative Maintenance
Periodically inspect the following areas:
ī€ƒ ā€¢ī€ƒī€ƒPowerī€ƒcordsī€ƒshouldī€ƒbeī€ƒfreeī€ƒofī€ƒcutsī€ƒorī€ƒotherī€ƒvisibleī€ƒdamage.
ī€ƒ ā€¢ī€ƒī€ƒAllī€ƒfastenersī€ƒshouldī€ƒbeī€ƒinī€ƒplaceī€ƒandī€ƒtightenedī€ƒsecurely.
ī€ƒ ā€¢ī€ƒī€ƒAllī€ƒmechanicalī€ƒfunctionsī€ƒshouldī€ƒoperateī€ƒproperly.
Periodicallyī€ƒlubricateī€ƒtheī€ƒbackī€ƒhingeī€ƒtoī€ƒmaintainī€ƒquiet,ī€ƒsmooth,ī€ƒoperationī€ƒ(useī€ƒlightī€ƒmachineī€ƒoil).
Haveī€ƒaī€ƒqualiī‚æedī€ƒserviceī€ƒtechnicianī€ƒinspectī€ƒyourī€ƒequipmentī€ƒeveryī€ƒtwelveī€ƒmonths.
Cleaning
Upholstery
Washī€ƒyourī€ƒupholsteryī€ƒweeklyī€ƒwithī€ƒaī€ƒmildī€ƒliquidī€ƒsoapī€ƒandī€ƒwaterī€ƒmixture,ī€ƒrinseī€ƒwithī€ƒclearī€ƒwater
and dry completely to remove disinfection cleaner build-up.
Disinfect your upholstery with a solution of standard bleach and water mixed 1 in 10 (10%) or chlorine based
cleaners
.
Follow this with a clear water rinse and thorough drying of material. See current CDC Guideline for
Disinfection & Sterilization in Healthcare Facilities.
Toī€ƒminimizeī€ƒdamageī€ƒcausedī€ƒbyī€ƒdisinfectantī€ƒcleanerī€ƒresidueī€ƒbuild-up,ī€ƒdoī€ƒnotī€ƒallowī€ƒdisinfectantsī€ƒtoī€ƒpoolī€ƒonī€ƒ
theī€ƒupholsteryī€ƒsurface.ī€ƒOnceī€ƒtheī€ƒapprovedī€ƒcontactī€ƒtimeī€ƒhasī€ƒbeenī€ƒobtained,ī€ƒremoveī€ƒandī€ƒdryī€ƒexcessī€ƒliquidī€ƒ
remaining on the surface.
Detailed care and maintenance instructions are included with your product. This information is also available
on www.midmark.com in the Technical library under the User Information tab for your product.
Painted Metal / Plastic Surfaces
Cleanī€ƒtheī€ƒpaintedī€ƒmetalī€ƒandī€ƒplasticī€ƒsurfacesī€ƒweeklyī€ƒusingī€ƒaī€ƒcleanī€ƒsoftī€ƒcloth,ī€ƒandī€ƒmildī€ƒcleaner.ī€ƒ
Equipment Alert
The upholstery is resistant to most medicinal-type stains, but may be damaged
by solvents and dyes. Immediately remove any fluids spilled on the upholstery.
English - 16
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
Powered Function(s) Not Working Properly
Function Problem Correction
Back Up No movement
(ā€œbeepsā€)
Contact your authorized service provider.
Back Down No movement
(ā€œbeepsā€)
Contact your authorized service provider.
Tilt Up No movement
(ā€œbeepsā€)
Contact your authorized service provider.
Tilt Down No movement
(ā€œbeepsā€)
* Remove any obstructions under the foot section.
Activateī€ƒtheī€ƒTiltī€ƒUp,ī€ƒthenī€ƒretryī€ƒTiltī€ƒDown.
Ifī€ƒproblemī€ƒpersists,ī€ƒcontactī€ƒyourī€ƒauthorizedī€ƒserviceī€ƒprovider.
* Crash Avoidance System
This chair is equipped with Midmarkā€™s Crash Avoidance System. The system is designed to prevent
injury and equipment damage by disabling certain functions if the chair is in a potential ā€œcrash positionā€.
The chair is considered to be in a potential ā€œcrash positionā€ when the systemā€™s sensors and switches
indicateī€ƒthatī€ƒtheī€ƒfootī€ƒsectionī€ƒisī€ƒcloseī€ƒtoī€ƒcollidingī€ƒwithī€ƒtheī€ƒīƒ€oor,ī€ƒorī€ƒanotherī€ƒforeignī€ƒobject.
Theī€ƒCrashī€ƒAvoidanceī€ƒSystemī€ƒmayī€ƒaļ‚‡ectī€ƒtheī€ƒTiltī€ƒDownī€ƒfunctionī€ƒonly.
Troubleshooting
English - 17
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
MA10690i
Range of Motion / Speciī‚æcations Table
Back Function
Full Down: 0Ā° +1/2Ā°
Full Up: 80Ā° +3Ā°
Patient Weight (max): 450ī€ƒlbsī€ƒ(204ī€ƒkg)
Weight of Chair: 290 lbs (131 kg)
Power Cord Length: 8 ft (244 cm)
Foot Control Voltage: 10ī€ƒVAC,ī€ƒSELVī€ƒ(Safetyī€ƒExtraī€ƒLowī€ƒVoltage)
Chair Receptacle
Maximum Load:
115ī€ƒVAC,ī€ƒ3ī€ƒamps,ī€ƒ50ī€ƒ/ī€ƒ60ī€ƒHz
Duty Cycle
(Motor Run Time):
Intermittent Operation
(30ī€ƒsecondsī€ƒONī€ƒ-ī€ƒ5ī€ƒminutesī€ƒOFF)
Protection against
ingress of īƒ€uids:
Ordinary Equipment (IPX0)
Foot control only: IPX1
Classiī‚æcations: Classī€ƒ1,ī€ƒTypeī€ƒB,ī€ƒAppliedī€ƒPart
Electrical Requirements: See model identiī‚æcation chart
Regulatory Compliance: See model identiī‚æcation chart
Equipment not suitable for use in the presence of a īƒ€ammable anesthetic mixture with air, oxygen, or nitrous oxide.
Speciī‚æcations
Tilt Function
Full Down: 0Ā° +1/2Ā°
Full Up: 30Ā° +2Ā°
English - 18
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
MA10691i
Dimensions
22 in
(56 cm)
24 in
(61 cm)
38.5 in
(98 cm)
Base Casting
58 in
(147 cm)
17 in
(43 cm)
29 in
(74 cm)
26 in
(66 cm)
Upholstered Top
Extends Up to:
9 in (23 cm)
19 in. (48 cm)
English - 19
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
Model Identiī‚æcation / Compliance Chart
Model Description
Complies To: Electrical Ratings:
UL
60601-1
CAN / CSA
22.2,
#601.1-M90
IEC
60601-1-2
(EMC)
VAC
+/- 10%
Amps
Cycles
(Hz)
646-001 Two-Function Chair
(Back & Tilt)
Non-Programmable
w/Receptacles
ā€¢ ā€¢
115 6.0 60
English - 20
003-10160-99 Ā© Midmark Corporation 2018
TP202 20-42-FO-00014 Rev A1 C2169
SCOPEī€ƒOFī€ƒWARRANTYī€ƒMidmarkī€ƒCorporationī€ƒ(ā€œMidmarkā€)ī€ƒwarrantsī€ƒtoī€ƒtheī€ƒoriginalī€ƒretailī€ƒpurchaserī€ƒthatī€ƒitī€ƒwill,ī€ƒatī€ƒMidmarkā€™sī€ƒoption,ī€ƒrepairī€ƒorī€ƒ
replace components of the domestic and international medical products manufactured by Midmark (except for components not warranted
under ā€œExclusionsā€) that are defective in material or workmanship under normal use and service. The sole remedy under this limited warran-
tyī€ƒisī€ƒtheī€ƒrepairī€ƒorī€ƒreplacement,ī€ƒatī€ƒMidmarkā€™sī€ƒoption,ī€ƒofī€ƒtheī€ƒapplicableī€ƒcomponents.ī€ƒThisī€ƒlimitedī€ƒwarrantyī€ƒshallī€ƒonlyī€ƒapplyī€ƒtoī€ƒdefectsī€ƒthatī€ƒareī€ƒ
reportedī€ƒtoī€ƒMidmarkī€ƒwithinī€ƒtheī€ƒapplicableī€ƒwarrantyī€ƒperiodī€ƒandī€ƒwhichī€ƒareī€ƒdeterminedī€ƒtoī€ƒexistī€ƒuponī€ƒexaminationī€ƒbyī€ƒMidmark,.ī€ƒThisī€ƒwarrantyī€ƒ
extends only to the original retail purchaser of a product and is not transferable or assignable. Replacement components or products may
beī€ƒusedī€ƒand/orī€ƒrefurbishedī€ƒcomponentsī€ƒorī€ƒproducts,ī€ƒprovidedī€ƒtheyī€ƒareī€ƒofī€ƒlikeī€ƒqualityī€ƒandī€ƒspeciī‚æī€ƒcationsī€ƒasī€ƒnewī€ƒcomponentsī€ƒorī€ƒproducts.
Midmark warrants to the original retail purchaser that during the applicable warranty period it will repair or replace software contained
within the products manufactured by Midmark (except for those not warranted under ā€œExclusionsā€) if: (1) the media on which the software is
furnished exhibits defects in material or workmanship under normal use; or (2) the software does not substantially conform to its published
speciī‚æī€ƒcations.
APPLICABLEī€ƒWARRANTYī€ƒPERIODī€ƒTheī€ƒapplicableī€ƒwarrantyī€ƒperiod,ī€ƒmeasuredī€ƒfromī€ƒtheī€ƒdateī€ƒofī€ƒinvoiceī€ƒtoī€ƒtheī€ƒoriginalī€ƒretailī€ƒpurchaserī€ƒofī€ƒ
the product and shall be one (1) year for all warranted products and components.
OBTAINING WARRANTY SERVICE Warranty service must be obtained through either Midmark or an authorized dealer in the Midmark
product line for which warranty service is requested. Midmark may be contacted for warranty service inquiries or issues via email at www.
midmark.com;ī€ƒbyī€ƒphoneī€ƒatī€ƒ1-800-MIDMARKī€ƒorī€ƒbyī€ƒmailī€ƒtoī€ƒMidmarkī€ƒCorporation,ī€ƒ60ī€ƒVistaī€ƒDrive,ī€ƒVersailles,ī€ƒOhioī€ƒ45380.ī€ƒItī€ƒisī€ƒtheī€ƒretailī€ƒpur-
chaserā€™sī€ƒobligationī€ƒtoī€ƒarrangeī€ƒforī€ƒdeliveryī€ƒofī€ƒaī€ƒproductī€ƒtoī€ƒMidmarkī€ƒorī€ƒoneī€ƒofī€ƒitsī€ƒauthorizedī€ƒdealersī€ƒforī€ƒwarrantyī€ƒservice,ī€ƒwhichī€ƒdeliveryī€ƒshallī€ƒ
be at retail purchaserā€™s expense. It is also the retail purchaserā€™s obligation to comply with the warranty service instructions provided either
by Midmark or its authorized dealer. The retail purchaser must provide Midmark with completed warranty registration information within
thirtyī€ƒ(30)ī€ƒdaysī€ƒafterī€ƒpurchaseī€ƒinī€ƒorderī€ƒtoī€ƒobtainī€ƒtheī€ƒbeneī‚æī€ƒtsī€ƒofī€ƒthisī€ƒlimitedī€ƒwarranty.
EXCLUSIONS: This limited warranty does not cover and Midmark shall not be liable for the following:
(1)ī€ƒDefects,ī€ƒdamageī€ƒorī€ƒotherī€ƒconditionsī€ƒcaused,ī€ƒinī€ƒwholeī€ƒorī€ƒinī€ƒpart,ī€ƒbyī€ƒmisuse,ī€ƒabuse,ī€ƒnegligence,ī€ƒalteration,ī€ƒaccident,ī€ƒfreightī€ƒdamage,ī€ƒ
negligentī€ƒstorage,ī€ƒtamperingī€ƒorī€ƒfailureī€ƒtoī€ƒseekī€ƒandī€ƒobtainī€ƒrepairī€ƒorī€ƒreplacementī€ƒinī€ƒaī€ƒtimelyī€ƒmanner;
(2)ī€ƒProductsī€ƒwhichī€ƒareī€ƒnotī€ƒinstalled,ī€ƒused,ī€ƒandī€ƒproperlyī€ƒcleanedī€ƒandī€ƒmaintainedī€ƒasī€ƒrequiredī€ƒorī€ƒrecommendedī€ƒinī€ƒtheī€ƒMidmarkī€ƒā€œInstallationā€ī€ƒ
and/orī€ƒā€œInstallation/Operationī€ƒManualā€ī€ƒforī€ƒtheī€ƒapplicableī€ƒproduct,ī€ƒincludingī€ƒtheī€ƒspeciī‚æī€ƒedī€ƒstructuralī€ƒandī€ƒoperationalī€ƒenvironmentalī€ƒconditionsī€ƒ
and electrical requirements;
(3) Products considered to be of a consumable or sterile nature;
(4) Accessories or parts not manufactured by Midmark;
(5)ī€ƒChargesī€ƒbyī€ƒanyoneī€ƒforī€ƒadjustments,ī€ƒrepairs,ī€ƒreplacementī€ƒparts,ī€ƒinstallationī€ƒorī€ƒotherī€ƒworkī€ƒperformedī€ƒuponī€ƒorī€ƒinī€ƒconnectionī€ƒwithī€ƒsuchī€ƒ
products which are not expressly authorized in writing in advance by Midmark;
(6) Costs and expenses of routine maintenance and cleaning; and
(7) Representations and warranties made by any person or entity other than Midmark.
(8)ī€ƒMatchingī€ƒofī€ƒcolor,ī€ƒgrainī€ƒorī€ƒtextureī€ƒexceptī€ƒtoī€ƒcommerciallyī€ƒacceptableī€ƒstandards;
(9)ī€ƒChangesī€ƒinī€ƒcolorī€ƒcausedī€ƒbyī€ƒnaturalī€ƒorī€ƒartiī‚æī€ƒcialī€ƒlight;
(10) Custom manufactured products;
(11)ī€ƒAlterationsī€ƒorī€ƒmodiī‚æī€ƒcationsī€ƒtoī€ƒtheī€ƒproductī€ƒbyī€ƒanyī€ƒpersonī€ƒorī€ƒentityī€ƒotherī€ƒthanī€ƒMidmark;ī€ƒand
(12)ī€ƒProductsī€ƒthatī€ƒwouldī€ƒotherwiseī€ƒbyī€ƒcoveredī€ƒunderī€ƒthisī€ƒlimitedī€ƒwarranty,ī€ƒbutī€ƒareī€ƒacquired:ī€ƒ(i)ī€ƒfromī€ƒaī€ƒpersonī€ƒorī€ƒentityī€ƒthatī€ƒisī€ƒnotī€ƒMidmarkī€ƒ
or one of its authorized dealers; or (ii) from a Midmark dealer that is not authorized to sell the product at issue in the geographic territory
whereī€ƒtheī€ƒpurchaserī€ƒisī€ƒlocated,ī€ƒorī€ƒisī€ƒnotī€ƒauthorizedī€ƒtoī€ƒsellī€ƒtheī€ƒproductī€ƒatī€ƒissueī€ƒwithinī€ƒtheī€ƒmedicalī€ƒmarket.
SOFTWARE;ī€ƒWITHī€ƒRESPECTī€ƒTOī€ƒSOFTWAREī€ƒTHATī€ƒISī€ƒAī€ƒPRODUCTī€ƒORī€ƒCOMPONENTī€ƒTHEREOF,ī€ƒMIDMARKī€ƒDOESī€ƒNOTī€ƒWARRANTī€ƒ
THAT THE SOFTWARE: (1) IS ERROR FREE; (2) CAN BE USED WITHOUT PROBLEMS OR INTERRUPTIONS; OR (3) IS FREE FROM
VULNERABILITY TO INTRUSION OR ATTACK BY VIRUSES OR OTHER METHODS.
EXCLUSIVE REMEDY; CONSEQUENTIAL DAMAGES DISCLAIMER MIDMARKā€™S ONLY OBLIGATION UNDER THIS LIMITED WARRAN-
TY IS THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS. MIDMARK SHALL NOT BE LIABLE FOR AND HEREBY DISCLAIMS
ANYī€ƒDIRECT,ī€ƒSPECIAL,ī€ƒINDIRECT,ī€ƒINCIDENTAL,ī€ƒEXEMPLARYī€ƒORī€ƒCONSEQUENTIALī€ƒDAMAGESī€ƒORī€ƒDELAYS,ī€ƒINCLUDING,ī€ƒBUTī€ƒ
NOTī€ƒLIMITEDī€ƒTO,ī€ƒDAMAGESī€ƒFORī€ƒLOSSī€ƒOFī€ƒPROFITSī€ƒORī€ƒINCOME,ī€ƒLOSSī€ƒOFī€ƒUSE,ī€ƒLOSSī€ƒOFī€ƒDATA,ī€ƒDOWNTIME,ī€ƒCOVERī€ƒANDī€ƒ
EMPLOYEEī€ƒORī€ƒINDEPENDENTī€ƒCONTRACTORī€ƒWAGES,ī€ƒPAYMENTSī€ƒANDī€ƒBENEFITS.ī€ƒTHISī€ƒDISCLAIMERī€ƒSHALLī€ƒSURVIVEī€ƒANYī€ƒ
FAILURE OR ASSERTED FAILURE OF THE ESSENTIAL PURPOSE OF THIS LIMITED WARRANTY OR ITS REMEDIES SPECIFIED
HEREIN. WARRANTY DISCLAIMER THIS WARRANTY IS MIDMARKā€™S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRAN-
TIES,ī€ƒEXPRESSī€ƒORī€ƒIMPLIED.ī€ƒMIDMARKī€ƒMAKESī€ƒNOī€ƒIMPLIEDī€ƒWARRANTIESī€ƒOFī€ƒANYī€ƒKINDī€ƒINCLUDINGī€ƒANYī€ƒIMPLIEDī€ƒWARRANTIESī€ƒ
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACE-
MENT OF DEFECTIVE PARTS.
STATUTEī€ƒOFī€ƒLIMITATIONSī€ƒNoī€ƒactionī€ƒmayī€ƒbeī€ƒbroughtī€ƒagainstī€ƒMidmarkī€ƒforī€ƒbreachī€ƒofī€ƒthisī€ƒlimitedī€ƒwarranty,ī€ƒanī€ƒimpliedī€ƒwarranty,ī€ƒifī€ƒany,ī€ƒorī€ƒforī€ƒ
anyī€ƒotherī€ƒclaimī€ƒarisingī€ƒoutī€ƒofī€ƒorī€ƒrelatingī€ƒtoī€ƒtheī€ƒproducts,ī€ƒmoreī€ƒthanī€ƒninetyī€ƒ(90)ī€ƒdaysī€ƒfollowingī€ƒexpirationī€ƒofī€ƒtheī€ƒlimitedī€ƒwarrantyī€ƒperiod.
NOī€ƒAUTHORIZATIONī€ƒNoī€ƒpersonī€ƒorī€ƒī‚æī€ƒrmī€ƒisī€ƒauthorizedī€ƒtoī€ƒcreateī€ƒorī€ƒapproveī€ƒforī€ƒMidmarkī€ƒanyī€ƒotherī€ƒobligationī€ƒorī€ƒliabilityī€ƒinī€ƒconnectionī€ƒwithī€ƒ
the products.
Warranty Information
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68

Midmark 646 Podiatry Procedures Chair User guide

Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI