TPCA
TM
Pain Management Pump
Quick User Guide
• Users must have undertaken training before operang this device.
• Refer to Operang Manual for full operang instrucons.
• Screen informaon/sequences may vary with dierent soware versions.
• The guidance given is based on pump default sengs; the pump set-up for your hospital/department may
vary if addional features/set-up opons have been added.
• Always follow screen prompts and before pressing keys to proceed ensure selecons made correspond
with what is required.
Doc Ref: QUG24/TPCA Pain Mgt/August 2015/v1
While the infusion is running, repeatedly press INFO to display:
Press 1 - Infusion summary (VTBI/VI).
Press 2 - Boluses summary (aempted/given).
Press 3 - Clinician bolus summary.
Press 4 - Shi total.
Press 5 - Baery level.
Press 6 - Basal rate protocol review.
Press 7 - Paent bolus protocol review.
Press 8 - Date and me/occlusion pressure.
Accessing Informaon
Remember to remove and re-apply keypad lock.
1. To pause current infusion, press the STOP/NO.
2. Press the FF key.
3. Enter access code, press START/YES.
4. Use +/- keys to enter bolus, press START/YES.
5. Observe the screen showing Clinician Acvated Bolus delivery.
On compleon of bolus delivery the current infusion resumes.
Clinician Acvated Bolus (Designated and Authorised Personnel Only)
Notes
When an ALERT is acvated:
1. The infusion connues and the LED indicator remains green.
2. Three beeps are heard approximately every three/four minutes.
3. A screen message alternates with the infusion running screen unl the end of the baery/infusion.
When an ALARM is acvated:
1. The infusion stops.
2. The LED indicator light turns from green to red.
3. The alarm sound connuously unl either the pump is paused or the problem is reced.
4. A message appears on the display indicang the alarm cause.
Screen Prompt Result and Acons
Alert and Alarm Acvaon
Troubleshoong
Keypad Locked
Pump Stopped
Press YES to Resume
Press YES to Resume,
NO for New Therapy
Near End
Low Baery
Pump Paused
Too Long
Syringe Empty,
Remove Syringe
End Baery
Syringe displaced,
Check Syringe
Occlusion/Empty
Syringe, Check Line
System Error. Press and Hold INFO for
Details. If problem persists send pump
for service.
ERROR. Startup MotMov Fail. If problem
persists send pump for service.
Only the STOP, START and INFO keys are accessible.
Deacvate keypad lock if further access required.
The pump has been powered o and powered on again.
Press YES to resume the current protocol. Press NO to start a new Therapy (protocol).
The infusion has been stopped.
Press YES to Resume the infusion or press NO to connue stopped state.
Alert: Protocol is about to end/syringe is almost empty.
Prepare to change syringe or disconnue pump use.
Alert: Baery is almost depleted.
Prepare to charge baery.
Alarm: The pump has been stopped/paused for more than 2 minutes with no intervenons or key presses.
Either press START/YES to resume, press STOP/NO to connue pause or turn the power o.
Alarm: Current infusion has completed and the syringe is empty.
Prepare to change syringe or disconnue pump use.
Alarm: Baery will fail imminently.
Charge baery.
Alarm: One or more of the syringe detecon sensors is not detecng.
Check the syringe and re-seat as necessary. Check screen messages for assistance.
Alarm: Clamped line, occluded or kinked. Actuator has reached the minimum travel posion.
Release the clamp, ush/replace the access device or clear the occlusion.
A technical or system error has occurred.
• The pump alarms if system fault has been detected and the unit will be inoperave.
• The user may be prompted to power o and restart, which may recfy the error.
• If the problem cannot be reced: power o, remove from paent use.
• Follow local policy and contact service department.
• The Event Log will record the error/alarm event.