BD TPCA™ Pain Management Pump Quick start guide

Type
Quick start guide
TPCA
TM
Pain Management Pump
Quick User Guide
Users must have undertaken training before operang this device.
Refer to Operang Manual for full operang instrucons.
Screen informaon/sequences may vary with dierent soware versions.
The guidance given is based on pump default sengs; the pump set-up for your hospital/department may
vary if addional features/set-up opons have been added.
Always follow screen prompts and before pressing keys to proceed ensure selecons made correspond
with what is required.
Doc Ref: QUG24/TPCA Pain Mgt/August 2015/v1
While the infusion is running, repeatedly press INFO to display:
Press 1 - Infusion summary (VTBI/VI).
Press 2 - Boluses summary (aempted/given).
Press 3 - Clinician bolus summary.
Press 4 - Shi total.
Press 5 - Baery level.
Press 6 - Basal rate protocol review.
Press 7 - Paent bolus protocol review.
Press 8 - Date and me/occlusion pressure.
Accessing Informaon
Remember to remove and re-apply keypad lock.
1. To pause current infusion, press the STOP/NO.
2. Press the FF key.
3. Enter access code, press START/YES.
4. Use +/- keys to enter bolus, press START/YES.
5. Observe the screen showing Clinician Acvated Bolus delivery.
On compleon of bolus delivery the current infusion resumes.
Clinician Acvated Bolus (Designated and Authorised Personnel Only)
Notes
When an ALERT is acvated:
1. The infusion connues and the LED indicator remains green.
2. Three beeps are heard approximately every three/four minutes.
3. A screen message alternates with the infusion running screen unl the end of the baery/infusion.
When an ALARM is acvated:
1. The infusion stops.
2. The LED indicator light turns from green to red.
3. The alarm sound connuously unl either the pump is paused or the problem is reced.
4. A message appears on the display indicang the alarm cause.
Screen Prompt Result and Acons
Alert and Alarm Acvaon
Troubleshoong
Keypad Locked
Pump Stopped
Press YES to Resume
Press YES to Resume,
NO for New Therapy
Near End
Low Baery
Pump Paused
Too Long
Syringe Empty,
Remove Syringe
End Baery
Syringe displaced,
Check Syringe
Occlusion/Empty
Syringe, Check Line
System Error. Press and Hold INFO for
Details. If problem persists send pump
for service.
ERROR. Startup MotMov Fail. If problem
persists send pump for service.
Only the STOP, START and INFO keys are accessible.
Deacvate keypad lock if further access required.
The pump has been powered o and powered on again.
Press YES to resume the current protocol. Press NO to start a new Therapy (protocol).
The infusion has been stopped.
Press YES to Resume the infusion or press NO to connue stopped state.
Alert: Protocol is about to end/syringe is almost empty.
Prepare to change syringe or disconnue pump use.
Alert: Baery is almost depleted.
Prepare to charge baery.
Alarm: The pump has been stopped/paused for more than 2 minutes with no intervenons or key presses.
Either press START/YES to resume, press STOP/NO to connue pause or turn the power o.
Alarm: Current infusion has completed and the syringe is empty.
Prepare to change syringe or disconnue pump use.
Alarm: Baery will fail imminently.
Charge baery.
Alarm: One or more of the syringe detecon sensors is not detecng.
Check the syringe and re-seat as necessary. Check screen messages for assistance.
Alarm: Clamped line, occluded or kinked. Actuator has reached the minimum travel posion.
Release the clamp, ush/replace the access device or clear the occlusion.
A technical or system error has occurred.
The pump alarms if system fault has been detected and the unit will be inoperave.
The user may be prompted to power o and restart, which may recfy the error.
If the problem cannot be reced: power o, remove from paent use.
Follow local policy and contact service department.
The Event Log will record the error/alarm event.
Tel: 01253 206700 | Fax: 01253 896648 | Email: customersuppor[email protected] | cmemedical.co.uk
Remember to de-acvate and acvate keypad lock if necessary
1. With infusion in progress, use +/- keys to select “Basal Rate Titraon” opon, press START/YES.
2. Enter the relevant access code using +/- keys, press START/YES.
3. Enter the mL/hour rate using +/- keys, press START/YES.
4. Check that the new basal mL/hour rate displays on the infusion running screen.
Remember to remove and re-apply keypad lock.
1. With infusion in progress, use +/- keys to select either “Bolus Dose Titraon” or “Lockout Time Titraon
opon, press START/YES.
2. Enter the relevant access code using +/- keys, press START/YES.
3. Enter the dose or me required using +/- keys, press START/YES.
4. Check that the new dose or lockout me displays on the infusion running screen.
To change syringe when end alarm has acvated:
1. To conrm the end of infusion alarm, press START/YES.
2. Press STOP/NO then power o by pressing ON/OFF.
3. Clamp line.
4. Remove empty syringe and aach new syringe to administraon set.
5. Power on by pressing ON/OFF.
6. Follow screen prompts and when “Press Yes to Resume, No for New Therapy” screen displays press
START/YES (to retain infusion history).
7. When “Load Syringe” screen displays, load the new syringe, re-check protocol parameters and start the
infusion.
Basal Rate Titraon During Delivery
Changing Paent Bolus Dose and/or Lockout Time
Syringe Change
Starng a Protocol for a New Paent
Re-check Protocol Parameters
Start Protocol/Infusion
Power On
Select and Check the Protocol
Review Syringe Sizes Enabled for the Protocol
Observe Pre-Loading (Automac Acuator Movement)
Syringe Loading, Detecon and Conrmaon
Purge System
Conrm Deliverable Volume
PAIN MANAGER
SYRINGE PUMP
SV: xxxxxxx
ID: xxxxxxxx
Enter Level 1 Code
0
Set +/- and Press YES
SELECT PROTOCOL
A:
B: Morphine 2mg
C:
D:
E:
F:
Select +/-, Press YES
B: Morphine
Rate 0 ml/h
Bolus 2 mg
Lock out 5 mins
Last bolus - - - mins
50ml BD Plaspak
B: Morphine 2 mg
Rate 2 ml/hr
Min. 0 ml/hr
Max. 20 ml/hr
Press YES to Conrm
26/11/2008 16:45
Occlusion 540mmHG
Titraon Enabled
Baery Status 79%
ID: xxxxxxxx
B: Morphine 2 mg
Rate 2 ml/hr
Min. 0 ml/hr
Max. 20 ml/hr
Press YES to Conrm
Bolus Dose 2 mg
(0 - 20mg)
Lock Out 5 mins
Bolus Rate 300 ml/hr
Press YES to Conrm
Bolus Dose 2 mg
(0 - 20mg)
Lock Out 5 mins
Bolus Rate 300 ml/hr
Press YES to Conrm
Syringe Sizes Enabled
No Preference
Press YES to Conrm
Pre-Loading
PRESS NO TO STOP
Purge System Slack?
Start Protocol?
50ml BD Plaspak
Press YES to Conrm
or +/- for opons
50ml BD Plaspak
Volume 49.3ml
Press YES to Conrm
Load Syringe
Enter the access code.
Press START/YES.
Use the +/- keys to scroll up/down the menu. When protocol required is
highlighted, press START/YES.
Check basal rate
parameters, press
START/YES.
Check bolus
parameters, press
START/YES.
Press START/YES.
Wait for the next screen to display.
If required, manually adjust the actuator posion then load syringe into the 3 x
sensors. Load the syringe into the 3 sensors.
Check that the syringe inserted into the pump matches the syringe displaying,
when they match, press START/YES.
If not required, press STOP/NO.
If required, press START/YES and follow screen prompts.
Check that the volume displaying matches the volume in the syringe, press
START/YES.
If the prompt Press YES to Resume, NO for New Program displays, press STOP/NO.
If the screen prompt “Protocol Lock, displays, select OFF and press START/YES.
Press ON/OFF key to power up.
The last line alternates between “Pump Operang” and the syringe size/brand
conrmed for the protocol.
Connect set/cannula to paent.
To commence the protocol, press START/YES.
Check protocol basal
rate parameters, press
START/YES.
Check protocol bolus
parameters, press
START/YES.
  • Page 1 1
  • Page 2 2

BD TPCA™ Pain Management Pump Quick start guide

Type
Quick start guide