Thermo Fisher Scientific FAIMS Pro System User guide

  • Hello! I am an AI chatbot trained to assist you with the Thermo Fisher Scientific FAIMS Pro System User guide. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
FAIMS Pro System
User Guide
High-Field Asymmetric Waveform Ion Mobility
Spectrometry System
98100-97000 Revision A January 2019
© 2019 Thermo Fisher Scientific Inc. All rights reserved.
EASY-Spray NG, FAIMS Pro, Nanospray Flex NG, and TSQ are trademarks; Unity is a registered service
mark; and Orbitrap, Pierce, Thermo Scientific, and Tribrid are registered trademarks of Thermo Fisher
Scientific Inc. in the United States.
The following are registered trademarks in the United States and other countries: Liquinox is a registered
trademark of Alconox, Inc. Microsoft and Windows are registered trademarks of Microsoft Corporation.
The following are registered trademarks in the United States and possibly other countries: Tygon is a registered
trademark of Saint-Gobain Performance Plastics Company. Viton is a registered trademark of DuPont
Performance Elastomers LLC.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Release history: Rev A, January 2019
Software version: (Thermo) Foundation 3.1 SP5 and later, Xcalibur 4.2 and later, Tune 3.1 and later
For Research Use Only. Not for use in diagnostic procedures.
Regulatory Compliance
Thermo Fisher Scientific performs complete testing and evaluation of its products to ensure full compliance with
applicable North American and European regulations. Your system meets the applicable requirements in the
electromagnetic compatibility (EMC) and product safety standards described in this section.
Unauthorized changes that you make to your system will void regulatory compliance and may defeat the built-in
protections for your instrument. Some examples of unauthorized changes include using replacement parts or adding
components, options, or peripherals that Thermo Fisher Scientific has not qualified and authorized. Unauthorized
changes can also result in bodily injury and/or damage to your system and laboratory.
Ensure continued compliance with regulatory standards:
Follow all installation instructions provided in the documentation that comes with your system.
Order replacement parts (as specified in the instrument manual) and additional components, options, and periph-
erals directly from Thermo Fisher Scientific or an authorized representative.
FCC Compliance Statement
Low Voltage Directive 2014/35/EU
This device complies with Low Voltage Directive 2014/35/EU and the safety standards EN 61010-1:2010, UL
61010–1:2012, CAN/CSA 22.2 61010-1-12; EN 61010-2-010:2014, UL 61010-010:2015, CAN/CSA 22.2 61010-
2-010:2015.
EMC Directive 2014/30/EU
This device was tested by TÜV Rheinland of North America and complies with the following EMC standards:
47 CFR 15, Subpart B, Class A: 2015 EN 55011:2009 + A1 EN 61000-4-4: 2012
CISPR11: 2009 + A1 IEC/EN 61326-1: 2013 EN 61000-4-5: 2014
ICES-003 Issue 6: 2016 IEC 61000-3-2: 2014 EN 61000-4-6: 2014
AS/NZS CISPR22: 2009 IEC/EN 61000-3-3: 2013 IEC/EN 61000-4-11: 2004 /
ISH1:2010
AS/NZS CISPR32: 2015 IEC/EN 61000-4-2
VCCI CISPR 32: 2015 IEC 61000-4-3: 2006 + A2
THIS DEVICE COMPLIES WITH PART 15 OF THE FCC RULES. OPERATION IS SUBJECT TO
THE FOLLOWING TWO CONDITIONS: (1) THIS DEVICE MAY NOT CAUSE HARMFUL
INTERFERENCE, AND (2) THIS DEVICE MUST ACCEPT ANY INTERFERENCE RECEIVED,
INCLUDING INTERFERENCE THAT MAY CAUSE UNDESIRED OPERATION.
Notice on Lifting and Handling of
Thermo Scientific Instruments
For your safety, and in compliance with international regulations, the physical handling of this Thermo Fisher Scientific
instrument requires a team effort to lift and/or move the instrument. This instrument is too heavy and/or bulky for one
person alone to handle safely.
Notice on the Proper Use of
Thermo Scientific Instruments
In compliance with international regulations: This instrument must be used in the manner specified by Thermo Fisher
Scientific to ensure protections provided by the instrument are not impaired. Deviations from specified instructions on
the proper use of the instrument include changes to the system and part replacement. Accordingly, order replacement
parts from Thermo Fisher Scientific or one of its authorized representatives.
CAUTION Read and understand the various precautionary notes, signs, and symbols contained inside
this manual pertaining to the safe use and operation of this product before using the device.
WEEE Directive
2012/19/EU
Thermo Fisher Scientific is registered with B2B Compliance (B2Bcompliance.org.uk) in the UK and with the
European Recycling Platform (ERP-recycling.org) in all other countries of the European Union and in Norway.
If this product is located in Europe and you want to participate in the Thermo Fisher Scientific Business-to-Business
(B2B) Recycling Program, send an email request to weee.recycle@thermofisher.com with the following information:
WEEE product class
Name of the manufacturer or distributor (where you purchased the product)
Number of product pieces, and the estimated total weight and volume
Pick-up address and contact person (include contact information)
Appropriate pick-up time
Declaration of decontamination, stating that all hazardous fluids or material have been removed from the product
For additional information about the Restriction on Hazardous Substances (RoHS) Directive for the European Union,
search for RoHS on the Thermo Fisher Scientific European language websites.
IMPORTANT This recycling program is not for biological hazard products or for products that have been medically
contaminated. You must treat these types of products as biohazard waste and dispose of them in accordance with
your local regulations.
Thermo Scientific FAIMS Pro System User Guide vii
C
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiii
Accessing Documentation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiii
Viewing the Product Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiv
Accessing the Help Menu Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiv
Viewing Online User Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiv
Providing Documentation Feedback. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
Compatible Mass Spectrometers and API Sources. . . . . . . . . . . . . . . . . . . . . . . xv
Special Notices, Symbols, and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xvi
Safety Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xvii
Contacting Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xix
Chapter 1 FAIMS Theory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
What is FAIMS Technology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
FAIMS Value Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Supported FAIMS Platforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Standard-Resolution and High-Resolution Modes . . . . . . . . . . . . . . . . . . . . . . . 3
Standard Resolution Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
High Resolution Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Additional Available Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Chapter 2 FAIMS Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Selection of Ions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
FAIMS Pro Instrument-Control Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Dispersion Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Compensation Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Electrode Temperatures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Carrier Gas Flow Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Readback Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Configuring the System Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Calibration Solutions and Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
FAIMS Component On/Off Status Under Varying Power Conditions . . . . . . . 15
Chapter 3 Functional Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Main Controller Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Housing Adapter and RF Transformer Box. . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Contents
Contents
viii FAIMS Pro System User Guide Thermo Scientific
Electrode Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Electrode Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Gas Delivery to the FAIMS Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Chapter 4 Installing the FAIMS Pro System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Removing the MS Ion Sweep Cone. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Installing the Electrode Assembly to the Housing Adapter . . . . . . . . . . . . . . . . 27
Connecting the Nitrogen Gas and Electrical Cables to the MCB . . . . . . . . . . . 28
Installing the FAIMS Pro System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Installing the EASY-Spray NG Ion Source on the FAIMS Pro System . . . . . . . 30
Installing the Nanospray Flex NG Ion Source on the FAIMS Pro System . . . . . 31
Chapter 5 Generating a FAIMS Diagnostics Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Running the FAIMS User Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Diagnostics Results of the Tune DV RF Process . . . . . . . . . . . . . . . . . . . . . . . . 37
Diagnostics Results of Check DV RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Example Diagnostics Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Chapter 6 Identifying Optimal CV Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Configuring the CV Scan Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Defining the Scan Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Saving the Optimal CV Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Chapter 7 Conducting an Example FAIMS Experiment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
Using Tune to Acquire Sample Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Using Method Editor Templates to Acquire Sample Data. . . . . . . . . . . . . . . . . 62
Using FreeStyle to View Previously Acquired Sample Data . . . . . . . . . . . . . . . . 63
Chapter 8 Maintaining the FAIMS Pro System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65
Maintenance Schedule. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Tools and Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Shutting Down the LC/MS/FAIMS Pro System . . . . . . . . . . . . . . . . . . . . . . . . 68
Flushing the Inlet Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Removing Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Removing the Ion Source from the FAIMS Pro System. . . . . . . . . . . . . . . . . 70
Removing the Electrode Assembly from the Housing Adapter. . . . . . . . . . . . 70
Removing the Electrode Set from the Entrance Plate. . . . . . . . . . . . . . . . . . . 71
Removing the FAIMS Pro System from the MS . . . . . . . . . . . . . . . . . . . . . . 72
Cleaning Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Cleaning the Entrance Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Cleaning the Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Cleaning the RF Transformer Box’s Air Filter . . . . . . . . . . . . . . . . . . . . . . . . 75
Cleaning the Housing Adapter’s Drain Adapter. . . . . . . . . . . . . . . . . . . . . . . 75
Fixing a Gas Leak Between the MS and the Electrodes . . . . . . . . . . . . . . . . . . . 77
Contents
Thermo Scientific FAIMS Pro System User Guide ix
Reassemble Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Chapter 9 Replaceable Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
Additional Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Consumable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Power Cord. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Chapter 10 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87
Contents
xFAIMS Pro System User Guide Thermo Scientific
Thermo Scientific FAIMS Pro System User Guide xi
F
Figure 1. FAIMS Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Figure 2. FAIMS Pro and Obritrap Lumos . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Figure 3. Standard Resolution Mode Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Figure 4. High Resolution Mode Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 5. Animation: FAIMS effect showing ion drift (parallel plate example) . . . . . . . . . . 8
Figure 6. Animation: FAIMS effect with CV showing ion transmission (parallel plate
example) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Figure 7. FAIMS-MS/MS selection of drug and metabolite . . . . . . . . . . . . . . . . . . . . . . . . 9
Figure 8. Gas flow with ions in the cylindrical electrodes . . . . . . . . . . . . . . . . . . . . . . . . . 10
Figure 9. Example ideal FAIMS waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Figure 10. LEDs on the MCB front . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Figure 11. Interconnect on the MCB rear . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Figure 12. Adapter housing and RF transformer box (front and back with partial cable) . . . 20
Figure 13. Electrode assembly with unlocked levers (front and right side) . . . . . . . . . . . . . . 21
Figure 14. Entrance plate with and without the electrode assembly (back) . . . . . . . . . . . . . 21
Figure 15. Electrode assembly with unlocked levers (back and bottom) . . . . . . . . . . . . . . . 21
Figure 16. Electrode set removed from the entrance plate . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Figure 17. Partial exploded view of the electrode set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Figure 18. Heater endcap with heater, bushing assembly, and outer electrode . . . . . . . . . . 23
Figure 19. Schematic of the gas delivery to the FAIMS system and MS . . . . . . . . . . . . . . . 23
Figure 20. FAIMS Diagnostics pane - Tune DV RF in progress . . . . . . . . . . . . . . . . . . . . . 34
Figure 21. FAIMS Diagnostics pane - Tune DV RF complete . . . . . . . . . . . . . . . . . . . . . . 35
Figure 22. FAIMS Diagnostics pane - Check DV RF in progress . . . . . . . . . . . . . . . . . . . . 36
Figure 23. FAIMS Diagnostics pane - Check DV RF completed . . . . . . . . . . . . . . . . . . . . 37
Figure 24. High field RF coarse tuning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Figure 25. High field RF fine tuning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Figure 26. Low field RF coarse tuning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Figure 27. Low field RF fine tuning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Figure 28. Wait DV modulation DC to settle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Figure 29. HF & LF modulation DAC adjust . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Figure 30. DV RF ramp check HF amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Figure 31. DV ramp check of LF Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Figure 32. Wait DV modulation DC to settle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Figure 33. Check DV RF Diagnostic Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Figure 33. Check DV RF Diagnostic Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Figure 34. Tune DV RF Diagnostic Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Figure 34. Tune DV RF Diagnostic Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Figures
Figures
xii FAIMS Pro System User Guide Thermo Scientific
Figure 34. Tune DV RF Diagnostic Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Figure 34. Tune DV RF Diagnostic Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Figure 35. Results of FAIMS CV Scan in Tune . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Figure 36. Ion Source pane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Figure 37. Select FAIMS mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Figure 38. Setup Scan Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Figure 39. Activate FAIMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Figure 40. Setup CV Scan Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Figure 41. Report Generation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Figure 42. CV Scan Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Figure 43. CV Scan Results Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Figure 44. Entrance plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Figure 45. Separate the inner electrode and the outer electrode from electrode set assembly 74
Figure 46. Air filter on the bottom of the RF transformer box . . . . . . . . . . . . . . . . . . . . . . 75
Figure 47. Drain adapter on the housing adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Figure 48. O-ring seals for electrode bushing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Figure 49. O-ring seals for the entrance plate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Figure 50. Assembled electrode set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Figure 51. Electrode set inserted into entrance plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Figure 52. FAIMS assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Figure 53. FAIMS assembly attached to MS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Thermo Scientific FAIMS Pro System User Guide xiii
P
Preface
The FAIMS Pro System User Guide contains the installation instructions, descriptions of the
primary components, and maintenance procedures for the Thermo Scientific™ FAIMS Pro™
system, which is a high-field asymmetric waveform ion mobility spectrometry (FAIMS)
system. To ensure a compatible setup, see Compatible Mass Spectrometers and API Sources.
Also, because this guide uses drawings of various connections and component parts to help
illustrate procedures, be sure to start from , no matter where it appears.
Contents
Accessing Documentation
Providing Documentation Feedback
Compatible Mass Spectrometers and API Sources
Special Notices, Symbols, and Cautions
Safety Precautions
•Contacting Us
Accessing Documentation
The FAIMS Pro system includes complete documentation.
Viewing the Product Manuals
Accessing the Help Menu Options
Viewing Online User Documentation
For mass spectrometer (MS) requirements, refer to the release notes on the software DVD.
Preface
xiv FAIMS Pro System User Guide Thermo Scientific
Viewing the Product Manuals
The Thermo Fisher Scientific™ service engineer installs the instrument control applications
and the instrument manuals on the data system computer.
To view the product manuals
From the Microsoft™ Windows™ taskbar, choose Start > All Apps (Windows 10) or
All Programs (Windows 7) > Thermo Instruments > model x.x, and then open the
applicable PDF file.
Accessing the Help Menu Options
Follow this procedure to view the Help systems for the instrument-control applications.
To view the Help
Do the following as applicable:
Thermo Tune instrument-control application: Click the Options icon, , and
choose Tu n e H e l p.
Thermo Xcalibur™ Method Editor application: Choose an option from the Help
menu (or press the F1 key).
Viewing Online User Documentation
Visit the Thermo Fisher Scientific website for product manuals and more.
To view user documentation from the Thermo Fisher Scientific website
1. Go to thermofisher.com.
2. Point to Services & Support, and click Manuals on the left.
3. In the Refine Your Search box, search by the product name.
4. From the results list, click the title to open the document in your web browser, save it, or
print it.
To return to the document list, click the browser Back button.
Preface
Thermo Scientific FAIMS Pro System User Guide xv
Providing Documentation Feedback
To suggest changes to the documentation or to the Help
Complete a brief survey about this document by clicking the button below.
Thank you in advance for your help.
Compatible Mass Spectrometers and API Sources
Use the FAIMS Pro system with the appropriate Thermo Scientific MS and ion source for
nanoelectrospray (commonly referred to as nanoES or nanospray™) analysis. Table 1 lists
several compatible MSs. For information about your specific system, refer to the product
manuals.
CAUTION If you connect the FAIMS Pro system to another type of MS, you might
impair the safety protection provided by the equipment.
Table 1. Compatible Thermo Scientific ion sources and mass spectrometers
Ion source models Mass spectrometer models
•EASY-SprayNG
Nanospray Flex NG™
Orbitrap™ Tribrid™ Seriesa
•TSQSeriesII
aIn 2019, the name “Orbitrap Fusion™ Series” changed to “Orbitrap Tribrid Series.”
Preface
xvi FAIMS Pro System User Guide Thermo Scientific
Special Notices, Symbols, and Cautions
Make sure you understand the special notices, symbols, and caution labels in this guide. Most
of the special notices and cautions appear in boxes; those pertaining to safety also have
corresponding symbols. Some symbols are also marked on the instrument itself and can
appear in color or in black and white. For complete definitions, see Table 2.
Table 2. Notices, symbols, labels, and their meanings (Sheet 1 of 2)
Notice, symbol, or label Meaning
IMPORTANT Highlights information necessary to prevent damage to
software, loss of data, or invalid test results; or might contain
information that is critical for optimal performance of the
product.
Note Highlights information of general interest.
Tip Highlights helpful information that can make a task easier.
Caution: Read the cautionary information associated with this
task.
Chemical hazard: Observe safe laboratory practices and
procedures when handling chemicals. Only work with volatile
chemicals under a fume or exhaust hood. Wear gloves and other
protective equipment, as appropriate, when handling toxic,
carcinogenic, mutagenic, corrosive, or irritant chemicals. Use
approved containers and proper procedures to dispose of waste
oil and when handling wetted parts of the instrument.
Hot surface: Allow any heated components to cool before
touching them.
Risk of electric shock: This instrument uses voltages that can
cause electric shock and personal injury. Before servicing the
instrument, shut it down and disconnect it from line power.
While operating the instrument, keep covers on.
Risk of eye injury: Eye injury can occur from splattered
chemicals, airborne particles, or sharp objects. Wear safety
glasses when handling chemicals or servicing the instrument.
Preface
Thermo Scientific FAIMS Pro System User Guide xvii
Safety Precautions
Read and understand the following cautions that are specific to the shutdown of the MS
system or to the removal of parts for cleaning.
Sharp object: Avoid handling the tip of the syringe needle.
Trip obstacle: Be aware of cords, hoses, or other objects located
on the floor.
Table 2. Notices, symbols, labels, and their meanings (Sheet 2 of 2)
Notice, symbol, or label Meaning
CAUTION If you must turn off the MS in an emergency, turn off the main power switch
located on the right-side power panel. This switch turns off all power to the MS,
including the forepump, without harming components within the system. However, do
not use this method as part of the standard shutdown procedure. Instead, see Shutting
Down the LC/MS/FAIMS Pro System. To turn off the LC, autosampler, and data system
computer in an emergency, use their respective on/off switch or button.
CAUTION To avoid an electrical shock, be sure to follow the instructions in Shutting
Down the LC/MS/FAIMS Pro System.
CAUTION Do not turn the MS on if you suspect that it has incurred any kind of electrical
damage. Instead, disconnect the power cord and contact Thermo Fisher Scientific
technical support for a product evaluation. Do not attempt to use the MS until it has been
evaluated. (Electrical damage might have occurred if the system shows visible signs of
damage, or has been transported under severe stress.)
CAUTION The FAIMS Pro system must connect to a certified Thermo Scientific MS,
which supplies high voltage capable of delivering a maximum of 8 kV and 100 μA. If you
connect the FAIMS Pro system to another type of MS, you might impair the protection
provided by the equipment.
CAUTION Do not place any objects (for example, containers with liquids) on top of the
MS, unless instructed to in the documentation. Leaking liquids might contact the
electronic components and cause an electrical short circuit.
Preface
xviii FAIMS Pro System User Guide Thermo Scientific
CAUTION Do not perform any servicing other than that contained in this manual. To
avoid personal injury or damage to the FAIMS Pro system, do not perform any servicing
other than that contained in this manual or related manuals unless you are qualified to do
so.
CAUTION Do not disconnect the power cords at the MS or MCB while the other ends are
still plugged into the electrical outlet.
Risk of electric shock: The FAIMS Pro system must be properly earthed through a supply
outlet that has an earthing connection. If you are not sure, check with a qualified electrician.
CAUTION Allow heated components to cool to room temperature (approximately
20 minutes) before servicing them.
Preface
Thermo Scientific FAIMS Pro System User Guide xix
Contacting Us
Contact Email Telephone QR Codea
aYou can use your smartphone to scan a QR Code, which opens your email application or browser.
U.S. Technical Support us.techsupport.analyze@thermofisher.com (U.S.) 1 (800) 532-4752
U.S. Customer Service
and Sales
us.customer-support.analyze@thermofisher.com (U.S.) 1 (800) 532-4752
Global Support To find global contact information or customize your request
1. Go to thermofisher.com.
2. Click Contact Us, select the country, and then select the type of support
you need.
3. At the prompt, type the product name.
4. Use the phone number or complete the online form.
To find product support, knowledge bases, and resources
Go to thermofisher.com/us/en/home/technical-resources.
To find product information
Go to thermofisher.com/us/en/home/brands/thermo-scientific.
Note To provide feedback for this document, go to surveymonkey.com/s/PQM6P62 or send an email message to
Technical Publications (techpubs-lcms@thermofisher.com).
Preface
xx FAIMS Pro System User Guide Thermo Scientific
/