  • Hello! I am an AI chatbot trained to assist you with the Midmark Multi-parameter Monitors Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
  • How do I activate the Multigas parameter?
    How do I enable the AG module?
    How do I display Multigas parameters?
    What happens if Multigas is not attached?
    How do I restart the monitor?
English-1
TP201 Rev. A
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | www.midmark.com
003-2747-00 Rev. AA1 (3/4/2020)
Multigas Kit Instructions [002-1745-00 & 002-1746-00]
B
A
Style G
Airway adapters and sampling lines are
intended for single patient use.
To reorder visit www.midmarkserviceparts.com,
or call 1-800-MIDMARK.
Attention
Prior to running this procedure be aware
that this will require the monitor to restart,
upon restart, ALL current patient alarm
data will be purged.
Step 1: Settings
A) Press power button on side of monitor to
turn on power.
B) Press the “Settings” Touch Screen button.
C) Press the “Module Setup” button.
C
Step 2: Module Setup menu
A) Press the “Off” Touch Screen button next to AG.
B) Press the “Masimo” Touch Screen button
A
B
Applies to Models:
8019 Multiparameter Monitor
English-2
TP201 Rev. A
003-2747-00
Step 3: Module Setup menus
A) Press “X” on top of Module Setup menu.
B) Press the “Ok” button in Warning menu to restart.
C) Count down starts and shuts down monitor.
Note: The options you have selected requires the monitor to restart,
Upon restart, all current patient alarm data will be purged.
B
C
A
Step 4: Display
A) Press power button on side of monitor to turn on power.
B) Press “Displays” on Touch Screen until display with AA
parameter is visible.
A
B
AA
Step 5: N
2
O Display - if Desired
A) Press “AA parameter” to turn on N
2
O display.
B) Press the “Off” on Touch Screen beside N2O Display.
A
B
Step 6: N
2
O Display - if Desired
A) Press “On” to turn on N
2
O display.
B) Press the “X” to close AA Setup menu.
A
B
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | www.midmark.com
English-3
TP201 Rev. A
Multigas Kit Instructions [002-1745-00 & 002-1746-00]
Step 1: Multigas parameter setup.
A) Turn on power button for monitor.
B) Press the “NEXT” Touch Screen Quick Access button.
C) Press the “Main Menu” Touch Screen Quick Access button.
D) Scroll down and press “Maintenance”
E) Enter “2016” and press “ENTER” on the keypad.
F) Scroll down and press COM_C MOD LOCK and select AG.
G) Click OK and press power button to turn off monitor.
H) Press power button to turn on monitor.
E
D
C
B
A
G
F
Scroll down
H
Airway adapters and sampling lines are
intended for single patient use.
To reorder visit www.midmarkserviceparts.com,
or call 1-800-MIDMARK.
Style G
003-2747-00
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | www.midmark.com
Applies to Models:
8013 Cardell Touch
English-4
TP201 Rev. A
Step 3: Display Mode setup.
Step 3 is required if Multigas parameters do not display
A) Press the “PREVIOUS” Touch Screen Quick Access button.
B) Press the “Display Modes” Touch Screen Quick Access button.
C) Press the “AG Screen” button.
D) Refer to Cardell Touch User’s Guide for Multi-gas ISA
(sidestream) and IRMA (mainstream) connection procedures
and parameter setup and usage instructions.
E
D
C
B
A
F
Scroll
down
A
B
C
Step 2: Monitor setup.
A) Press the “NEXT” Touch Screen Quick Access button.
B) Press the “Main Menu” Touch Screen Quick Access button.
C) Press the “Monitor Setup” button.
D) Press the “Module Setup” button.
E) Scroll down and press “AG” to show the side menu which
displays “ON” and “OFF.
F) Press “ON” to turn AG module on. The side menu will
automatically close once ON or OFF buttons are pressed.
Note: Alarm will sound if Multigas is not attached.
003-2747-00