Important Information
2
PROOF (LR #3557), 2002-11-25 «5118-51A Cozmo User Manual»
☞
Make sure that all caregivers, including for instance
school personnel, daycare providers, etc., are familiar
with the features of the pump, and can use and trouble-
shoot the pump. Make sure they are familiar with any in-
formation given to you by your healthcare provider.
☞
Deltec infusion sets are recommended for use with this
pump. Pump operation and accuracy may not be assured
if you use other manufacturers’ products.
☞
If there are any system problems, the pump will stop
the tests and display an alarm screen letting you know
there is a problem. If this happens, do not use the pump.
☞
The time and date must be set correctly, since basal pat-
terns are time-based, and all of the history is stored based
on time and date. Occasionally you will need to reset the
time, for example, during daylight savings time or to adjust
for a different time zone.
☞
When stopping delivery, remember that since the pump uses
only short-acting insulin, your insulin level will start to fall
and blood glucose level to rise very quickly (in as little as 90
minutes). You must test your blood glucose when stop-
ping delivery for any length of time.
☞
Once the pump is locked, you will not be able to enter the
menu without the proper code. Touch Bolus is also not al-
lowed when the pump is locked. Make sure there is some-
one around who knows the proper code.
☞
Always use the pump’s Load feature when starting a new
cartridge. This will assure that the cartridge is properly
loaded and the tubing and access device are filled with in-
sulin.
☞
Whenever you load a new cartridge and attach an infu-
sion set, you must test your BG level within 1 to 2 hours
to make sure that insulin is being properly delivered. If
you do not test within this time-frame, and insulin delivery
is delayed or blocked, you may develop hyperglycemia. If
desired, you can use the Glucose Reminder alert to remind
you to test your BG level (see Personalizing Alerts, page 28).
☞
Always use “aseptic” technique, particularly when work-
ing with the cartridge, infusion set, catheter and access
site. An increased risk of skin infection is associated with
pump use, as the access device remains in place for 2 or
more days (see “Aseptic Technique,” page 9).
☞
Always read the instructions for use provided with the
insulin, catheter, infusion set, and any other accessory used
with the Deltec Cozmo™ Insulin Pump.
☞
To avoid accidentally infusing insulin, disconnect the
infusion set tubing from the access device (or site) be-
fore removing a used cartridge or replacing an infusion
set. Never use the pump’s “load cartridge” or “fill tubing”
features while an infusion set is connected to your body.
You could deliver an unintended dose of insulin, resulting
in low blood glucose (hypoglycemia).
☞
If not properly tightened, insulin could leak from the
cartridge and infusion set connections and disrupt de-
livery (the insulin that leaks out is not being delivered into
your body like it should be and, depending on how much
insulin is leaking, could cause your BG level to rise). Signs