GHM G 1111-VAC Owner's manual

  • Hello! I'm a chat assistant that has analyzed the provided document about the GREISINGER G 1111-xx-VAC vacuum meter. This document details the supplementary instructions for the device, highlighting features like its additional pressure-compensating port and suitability for industrial packaging. I'm ready to answer your questions about the device and its specifications.
  • What is the purpose of the additional pressure port on the G 1111-xx-VAC?
    What is the pore width of the pressure-compensating element?
    What type of pressure port is used in the G 1111-xx-VAC?
    In what applications is the G 1111-xx-VAC suitable for?
H88.0.16.6AC-01
Supplementary Instructions
for the G 1111 Vacuum-Variant G 1111-xx-VAC
The standard G 1111 is especially suitable to measure rough vacuum via
the pressure port very fast. Due to the water tight construction a complete
evacuation of the instrument may degrade water protection: The
keypad/display screen may break loose! If the instrument should be used in
this application, we recommend removing the sealing O-ring of the battery
compartment (p.r.t. figure).
Attention: The Instrument then is no more water protected.
HINWEIS
Alternatively, there are the specialized variants G 1111-xx-VAC available,
which are described here.
The G 1111-xx-VAC do have an additional pressure port besides the
pressure-connection-port, which works as a pressure-compensating
element between the outer and the inner pressure of the instrument.
Therefore no dangerous pressure difference can build up inside at the one
side, and at the other particles are kept from ingression into the instrument
during the vacuum relaxation. As the compensating element works as a fine
porous filter.
The extreme small pores as well give a certain water protection.
Therefore, the instrument is suitable for rough use in industrial packaging
machines as well, where e.g. > 1000 packaging processes per hours are
completed.
The instrument variant G 1111-xx-VAC can substitute fully encapsulated
instruments very reasonable.
Pore width of pressure-compensating element (equals art. GDZ-34): 40 - 50
µm.
Due to the universal G1/8 pressure port, other filter elements can be
equipped as well.
GHM GROUP - Greisinger
GHM Messtechnik GmbH
Hans-Sachs-Str. 26
93128 Regenstauf | GERMANY
Pressure-compensating element
Pressure connection
Order Code
1
.
Pressure connection
-UT
1/8 inch port incl. connection for tube 6x1 mm (4 mm
inner) or 8x1 mm (6 mm inner)
-QC6
1/8 inch port incl. quick connector for tube with outer-Ø 6
mm
-ST6
1/8 inch port incl. screw connector for tube with outer-Ø 6
mm
-MCM
1/8 inch port incl. mini quick coupling plug
2
.
Option Anschluss
-VAC
Add. housing vent for fast complete evacuation
3
.
Set-Option
Device only
-SET
Measuring device, incl. 1 m tube 6x1 PVC, suitcase GKK
1002
-WPD
Measuring device, incl. 1 m tube 6x1 PVC, suitcase GKK
1002, WPD5 calibration certificate
-SETVAC
Measuring device, incl. 1 m tube 6x1 PVC, GDZ-34, GKK
1002, WPD5-VAC calibration cert. + adj. < 1 mbar
/