Patterson Medical Midland 4512EC Operation And Service Manual

Type
Operation And Service Manual
MODEL 4512E & 4512EC:
Midland
®
Stand-In Table
Operation and Service Guide
Midland
®
by
Sammons Preston
4 Sammons Court
Bolingbrook, IL 60440-4989
(630) 226-1300
TABLE
OF CONTENTS
Purpose of Stand-In Table
Page 1
Setup
Pagel
Method of Operation
Page 2
Cleaning
Page 2
Preventative Maintenance
Page 2
Trouble Shooting Guide
Page 3
Warranty
Page 3
Service
Page 3
Technical Specifications and Replacement Parts
Page 4
PURPOSE OF HI-LO TREATMENT TABLE
The purpose of the Stand-In Table is to assist in moving the patient from a sitting position in
a wheelchair to a standing position.
SET UP
Uncrate the table and check to ensure that all parts listed are included in your shipment.
Inspect the power cord, plug, controls, frame and upholstery for damage.
Model 4512E is completely assembled and ready to use. If any part is missing (see
illustration drawings and parts list on page 5) or damaged, please call customer service at
Sammons Preston at
1-800-323-5547.
Set up the table in a location that allows ample clearance around, above, and below the table,
and has access to the needed power supply. Securely lock all four casters for stability. The
unit is shipped with a 3-pole male hospital grade 15 amp plug (110V models). To maintain
adequate grounding, the unit must be plugged into a 3-pole receptacle (110V models)
(preferably a hospital grade) that is correctly wired to the building power system. 3 pole
male plugs are provided for 220V models due to the variation of receptacles. Illustration A
shows the correct color connections of a hospital grade cord to a 3 pole plug (220V models).
1
METHOD OF OPERATION
The Clinician must fully lower the unit to the floor before placing wheelchair with
patient up to stand in table. Adjust the width of the Lifting Arms to the size of the
patient. Position the harness under the patient buttocks and attach the hooks in the
desired holes. By means of the pneumatic hand-held switch, activate the motor gear
drive assembly raising the patient from a sitting position to a standing position. Once
in the full standing position the Back Support Pad should be locked into position and
the Chest Pad can be adjusted to desired height. To lower the patient back into the
wheel chair you would reverse the procedure.
WARNING
Never roll patient into lifting position before fully lowering unit to the floor.
CAUTIONS
H
Never leave a patient unattended.
H
Always inform the patient of the procedure before beginning.
H
Make sure the patient is clear of moving parts.
CLEANING
"UNPLUG UNIT BEFORE ATTEMPTING ANY CLEANING AND/OR
DISINFECTING"
Infective material should be removed by scrubbing with an approved germicidal
solution. After scrubbing, rinse with a wet cloth and allow to dry.
PREVENTATIVE
MAINTENANCE
HJoints use oil impregnated bronze bushings and should never need to be lubricated.
H
Frequently inspect the power cord and plug for frayed wires and/or damaged
insulation.
2
TROUBLE SHOOTING GUIDE
Should the table fail to operate:
1.
Check to ensure that the wall outlet is working.
2.
Check to ensure that the AC plug is inserted fully.
3.
Check to ensure that the pneumatic switch assemblies are connected correctly. If the
line to the hand control is pinched, the control will not work properly.
4.
The motor is protected by a thermal overload device. Should this safety device trip or
engage, let the motor cool down for 15 minutes before operating the table.
5.
If the table is still inoperable, see the service section for instructions on contacting
Sammons Preston.
WARRANTY
The Stand-In Table is guaranteed to be free from defects in material and/or workmanship
under normal use and service for a period of 5 years from date of purchase. The above
guarantee is null and void if unauthorized repair and/or modification is performed on
the
unit.
SERVICE
If a defect is detected or if the table requires service, contact the Sammons Preston customer
service department at
1-800-323-5547.
Please provide the Customer Service Representative
with the following information:
H
Sammons Preston item number (if known)
HTable serial number
H
Specific problems and/or parts involved (see part list)
HFacility name, contact name, shipping address and phone number
If the table is still under warranty, Sammons Preston will work with you to correct the defect
(either by sending replacement parts or by bringing the table back for repair if necessary). If
the table is out of warranty, repair parts can be purchased and sent to you.
WARNING
An electrical shock hazard may exist when the motor covers and control boxes have been
removed. In addition, the mechanical action of the table may subject the repair person to
significant shearing and crushing forces. Therefore, repair
must
be performed by technically
qualified personnel who proceed with care and use proper servicing techniques.
Note:
Unauthorized servicing may void the remainder of the warranty. Check with
Sammons Preston to determine the warranty status of a particular unit.
3
TECHNICAL SPECIFICATIONS AND REPLACEMENT PARTS
Model 4512E Specifications
Power requirements:
115 Volts
Voltage
110 VAC
Current
3.3 Amps
Frequency
60 HZ
Phase
Single Phase
60 HZ Leakage Currents:
Normal
1 ua or less
Max. Leakage
100 ua or less (TYP. 18 ua)
Grounding Characteristics:
Resistance from ground pin on plug and exposed metal surfaces -- 0.1 Ohm or less.
(TYP. 0.08 OHMS).
Table Dimensions:
Base Dimension:
LENGTH:
42" (107 cm)
WIDTH:
38" (97 cm)
Top Size:
LENGTH:
32 3/4" (83 cm)
WIDTH:
24 1/4" (62 cm)
Table Height: (from floor to top surface)
Minimum:
42 1/2" (108 cm)
Maximum:
52 1/2" (133 cm)
Table Weight:
210 lb.
(uncrated)
295 lb. (crated)
Maximum Lifting Capacity:
(patient weight) 300 lb. or 135 kg
Maximum Top Capacity:
100 lb. or 45 kg
220 Volts
220 Volts
1.06 Amps
50 HZ
Single Phase
4
Replacement Parts list
No.
Part No.
1
MT2038
or
MT2085
2
MT2056
or
MT2052
3
MT2080
4
MT2042
5
4512E-13-00
6
4512E-14-00
7
4512E-16-00
8
4512E-11-00
9
4512E-12-00
10
4512E-18-00
11
HW3423
12
HW3482
13
HW325 1
14
4512E-10-16P
15
4512E-19-01
16
4512E-20-01
Description
Qty.
Hanning Motor 115v KL-88
1
Hanning Motor 220v
1
Oriental Motor 115v 3RK15GN/AUL
1
Oriental Motor 220v 3RK15GN/CE
1
Gear Head 3RK15GN
1
Double Control Switch Hanning
1
Upholstered Chest Pad
1
Upholstered back Support Pad
1
Upholstered Knee Pad
1
Upholstered Side Pad
2
Upholstered Center Pad
1
Lifting Harness
1
Knob
1
/2-13x1
1
Knob 3/8-16x1
4
Caster 2"
4
Link Bar
2
Collar #1
3
Collar #2
2
5
Replacement Parts list for 4512E
Part No.
MT2038
MT2085
MT2056
MT2052
MT2080
MT2042
4512E-13-00
4512E-14-00
4512E-16-00
4512E-11-00
4512E-12-00
4512E-18-00
HW3423
HW3482
HW3251
4512E-10-16P
4512E-19-01
4512E-20-01
4512E -u
p
s.co
p
'4512E-15-co
p
Description
Hanning Motor 115v KL-88
Hanning Motor 220v
Oriental Motor 115v 3RK15GN/AUL
Oriental Motor 220v 3RK15GN/CE
Gear Head 3RK15GN
Double Control Switch Hanning
Upholstered Chest Pad
Upholstered back Support Pad
Upholstered Knee Pad 1S"
A
I "
J
W
Upholstered Side Pad
Upholstered Center Pad
Lifting Harness
Knob Y2-13x1
Knob 3/8-16x1
Caster 2"
Link Bar
Collar #1
Collar #2
Moroa.1j
FT
Pa
z
"
ts,
Arts J
6
Replacement Parts list or 4512EC
No.
Part
No.
Description .
Qty.
1
MT203 8
Hanning Motor 115v KL-88
1
or
MT2085
Hanning Motor 220v
1
2
MT2056
Oriental Motor 115v 3RK15GN/AUL
1
or
MT2052
Oriental Motor 220v 3RK15GN/CE
1
3
MT2080 Gear Head 3RK15GN
1
4
MT2042
Double Control Switch Hanning
1
5
4512E-13-00 Upholstered Chest Pad
1
6
4512E-14-00 Upholstered back Support Pad
1
7
4512E-16-00
Upholstered Knee Pad
1
8
4512E-11-00 Upholstered Side Pad
2
9
4512E-12-00 Upholstered Center Pad
1
10
4512E-18-00
Lifting Harness
1
11
HW3423
Knob % - 13 x 1
1
12
HW3482
Knob3/8-16x1
4
13
HW3251
Caster 2"
4
14
.
4512E-10-16P Link Bar
2
15
4512E-19-01
Collar #1
3
16
4512E-20-01 Collar #2
2
17
4512EC-18-00 T Weldment (Stainless Steel)
1
18
4512EC-15-00P Shoe Support
2
19
4512EC-22-00 Wooden Shoe assembly
3
20
HW3422
Knob
l/4 -
20
2
L^
^j € -
2-Op
1
ZI
Z2
7
JAN-26-2007 11:48 FROM:HAUSMANN IND
:aAP:/25 2OQ7 /THU 03:09 PM -TIE SPRINGS
12017676302
TO:12623878748
,. ,
:
PAK N
'e:
'
^192730168
Page 6
Received Time Jan•26.
9:58AM
i$
JAN-26-2007 11:48 FROM:HAUSMANN IND
12017676302
TO:12623878748
IM/25i2007/THU ,03 : D9 PM ^'lil; SPRINGS
^` ' °
'FAX
Na„ 5182730168
P.003/003
004 ;
Received Time Jan,26.
9:58AM
I
REVISIONS
LTR
DESCRIPTION
NATE
' APPROVALS
SCALES
DIM ZL
El/26197
DA
APPROVALS
HD.
AssEM
HA
REEL
8119!93
DRIB ZL
DRIP. ZL
10118/94
TgLE
OIR9/96
1
RICKER LEVEL
Warp.
UM
in
REV.
4512E-00-00
451 2E-
)
J. 0 F -D OR
SIZE
LOUD
FT
SOT. 1
OF
I
RD.
REEL
1
451 EE-l 0-03
a
ri
j.p 2E- 112
M.M[
aa
W
4
4512 -1 D- J
F
5
45 zE-1 D-0'
2:
ri
451
-IQ-1C
Ufa
111911
I
1IM MAC EWM111M
.-
M
IM
M gO
PRODUCT NO.
4512E
REAR LIFT ASSEMBLY- .
REVISIONS
mu.
ASSN NIL
POI
4512E-10A-0OP
3
2
451 2E-I 01:1-,O OP
3
4512c-10C-OOP
4
4512E-10G-00P
5
1-12E-10-OUP
6
4512E-10F-OOP
IT
4512E-4 -16P
..
4512E-10-17P
1
9
10
11
1a
DATE
APPROVALS
DESCRMT ON
PRODUCT
Kfl.
nRII.
I-IC
19 93
EE.
URIC
ZL
afra/94 TE RE
LAYOUT OF LEFT PARTS
DRIC
ZL
REV.
DRK.
JLEVU.
MR- 151
II IL NO.
s
[2
4512E-0D-00
E
451 EE-IG
.
1
:00
LAYER:
L
_
ASSEM
stir.
OF
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13

Patterson Medical Midland 4512EC Operation And Service Manual

Type
Operation And Service Manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI