Cannon Densi-H User manual

Category
Measuring & layout tools
Type
User manual
Densi-H
CAL
Densi-H
Handheld Density/Specific Gravity Meter
Operator’s Manual
Revision 2.0, February 5, 2016
30312854
Table of Contents
1 Introduction
3
2 Safety information
3
2.1 Definitions of signal words and warning symbols 3
2.2 Product specific safety notes 3
3 Description of the instrument
6
3.1 Overview 6
3.2 Display 7
3.3 Keys 7
4 Tutorial
8
4.1 Inserting batteries 8
4.2 Configuration 9
4.3 Test measurement 10
4.4 Adjustment 10
4.5 Cleaning 11
4.6 Switching off and on 11
5 Menu (menu)
11
5.1 Using the menu 11
5.2 Measurement Unit (Measure Unit) 12
5.3 Temperature Unit (Temp. Unit) 15
5.4 Measure Mode 15
5.5 Adjustment Mode (Calib. Mode) 16
5.6 Interface 16
5.7 Beep 16
5.8 Background Lightning and LCD-Contrast (LCD) 17
5.9 Automatic Switch-off (Power) 17
5.10 Software Version (Version No.) 17
5.11 Date and Time (Date&Time) 17
5.12 Identification 17
6 Measurement (meas)
17
6.1 Procedure for proper measurement 17
6.2 Filling the measuring cell using the built-in sample pump 18
6.3 Filling the measuring cell using an external syringe 18
6.4 Measuring 19
6.5 Saving the results 19
6.6 Displaying and marking saved results 20
6.7 Printing and transferring results 20
6.8 Deleting results 21
7 Adjustment (cal)
21
7.1 Adjusting the measuring cell with water 22
7.2 Adjusting the measuring cell with a density standard 22
7.3 Adjusting the measuring cell with air 23
7.4 Adjusting the temperature display 24
8 Interface
25
8.1 RS-P26 printer settings 25
8.2 Data transfer to PC 25
8.3 Data format 25
Densi-H 1Revision 2.0, February 5, 2016
9 Error messages and malfunctions
26
10 Cleaning and maintenance
27
10.1 Cleaning the measuring cell 27
10.2 Cleaning the housing 27
10.3 Replacing the sample tube 28
10.4 Replacing the sample pump 28
11 Standard and optional equipment
28
11.1 Standard equipment 28
11.2 Optional equipment 29
12 Technical data
30
13 Appendix
32
13.1 Density of pure water (0…40 °C) 32
13.2 Temperature-compensation coefficients α 32
Index
35
Densi-H2 Revision 2.0, February 5, 2016
1 Introduction
The CANNON Densi-H is a portable measuring instrument suited for
determining the density of liquids. The instrument uses the oscillating body
method. To fill the measuring cell the built-in sample pump or a syringe
may be used. The results are automatically calculated into one of the
following units: density, specific gravity, API degrees, Brix, % Alcohol, %
H2SO4, °Baumé, °Plato, Proof or a user-defined unit. The value is then
shown on the backlit display. For exact measurements, it is imperative to
correct the temperature’s influence on the density. Depending on the
selected unit the Densi-H carries out this result correction automatically.
For this task, the instrument uses either internally- stored tables or one of
the 10 temperature-compensation coefficients entered by the user.
The results, along with the sample identification, temperature, temperature-
coefficient, date and time can be saved. Together with the instrument
identification they can then, via the integrated infrared interface, be
transferred to the computer or printed out on a printer.
2 Safety information
Read and understand the instructions in this manual before you use
the instrument.
Keep this manual for future reference.
Include this manual if you pass on the instrument to other parties.
If the instrument is not used according to the instructions in this manual or
if it is modified, the safety of the instrument may be impaired and CANNON
Instrument Company® assumes no liability.
2.1 Definitions of signal words and warning symbols
Safety notes are marked with signal words and warning symbols. These
show safety issues and warnings. Ignoring the safety notes may lead to
personal injury, damage to the instrument, malfunctions and false results.
WARNING
for a hazardous situation with medium risk, possibly resulting
in severe injuries or death if not avoided.
CAUTION
for a hazardous situation with low risk, resulting in minor or
medium injuries if not avoided.
NOTICE
for a hazardous situation with low risk, resulting in damage to
the device or the property or in loss of data.
Note
(no symbol)
for useful information about the product.
Warning symbols
Harmful
substance
Acid / Corrosion
2.2 Product specific safety notes
The instrument has been tested for the experiments and intended purposes
documented in the appropriate manual. However, this does not absolve
you from the responsibility of performing your own tests of the products
supplied by us regarding their suitability for the methods and purposes
you intend to use them for.
Densi-H 3Revision 2.0, February 5, 2016
Intended use
The measuring instrument is intended for performing measurements for
determining the density of liquids.
Any other type of use and operation beyond the limits of technical specifi-
cations without written consent from CANNON Instrument Company®, is
considered as not intended.
Staff qualification
Incorrect use of the instrument or the chemicals used in the analysis can
lead to death or injury. The following qualifications are needed for
operating the instrument.
Knowledge and experience in working with toxic and caustic
substances.
Knowledge and experience in working with standard laboratory
equipment.
Knowledge and experience in working in accordance with general lab
safety rules.
Protective Clothing
Wear protective clothing in the laboratory when working with hazardous or
toxic substances.
Wear suitable eye protection such as goggles.
Use appropriate gloves when handling chemicals or hazardous
substances and check their integrity before use.
Wear a lab coat.
Safety notes
WARNING
Danger of injuries and death due to harmful substances!
Chemicals can cause injuries if they come in contact with
bare skin or are inhaled.
1 When using chemicals and solvents, comply with the
instructions of the producer and the general lab safety
rules.
2 Set up the instrument in a well-ventilated location.
3 Clean any spills immediately.
4 If you measure substances which form toxic gases,
place the instrument in a fume hood.
Densi-H4 Revision 2.0, February 5, 2016
WARNING
Risk of chemical burns from acids and alkalis!
The contact with acids and alkalis can cause severe
Poisoning, burns, skin irritation and cause eye damage.
1 Observe and comply the Specifications of the material
safety data sheets of the used testing medium.
2 When working with acids and alkalis always wear
protective clothing, chemical protective gloves und
protective goggles.
3 Do not breathe and swallow acids and alkalis.
4 Avoid skin contact with acids and alkalis.
5 Rinse eye immediately with clear water after eye contact.
Seek immediate medical help.
NOTICE
Danger of material damage due to non-intended use of
the device!
Non-intended use can damage the device.
1 Never press on the measuring cell window. This can
influence the oscillation characteristics of the measuring
cell.
2 Do not clean the measuring cell with concentrated NaOH
(causic soda) or HF (hydrogen fluoride). Both
substances chemically corrode the measuring cell.
3 Use batteries of the specified type only. Otherwise,
proper operation cannot be guaranteed.
4 Do not submerge the housing in liquid. The instrument
is only resistant to splashed water.
Site requirements
The instrument has been developed for operation in a well-ventilated area.
Avoid the following environmental influences:
Conditions outside of the ambient conditions specified in the technical
data
Powerful vibrations
Direct sunlight
high humidity present
Corrosive gas atmosphere
Explosive atmosphere of gases, steam, fog, dust and flammable dust
Powerful electric or magnetic fields
only ambient temperatures between -20 °C and 70 °C
Densi-H 5Revision 2.0, February 5, 2016
3 Description of the instrument
3.1 Overview
1
2
3
5
6
7
8
12 1113
10
4
9
Densi-H
CAL
1 Backlit Display
2 Keypad
3 Drain button
4 Sample pump
5 Cover of sample pump
6 Connector for syringe adapter
7 Fixing screw for sample tube
8 Sample tube
9 Measuring cell
10 Infrared interface
11 Lock button
12 Fill button
13 Battery compartment cover
Densi-H6 Revision 2.0, February 5, 2016
3.2 Display
1 2 3 4
5 6 7 8 9 10
11
(20)
(20)
1 Selected unit of measurement
2 Result
3 Sample identification (a…z or space)
4 Sample number, or error number if an error has occurred
5
Appears if Stability is set to Auto
6
Appears if Memory in is set to Auto
7
Appears if Memory out is set to Auto. If a printer or PC is connected,
the data are transferred automatically
8 Battery-power indicator
9 Temperature (°C / °F)
10 Appears if delete mode is activated
11 Mark for results
For the identification of invalid or incorrect results or for marking a
sample change
Δ
Reference temperature (for the units Comp. Density and SG only)
3.3 Keys
3
1
2
4
5
6
Red symbols:
Press key longer than 2 seconds.
Blue symbols:
Press key briefly.
Densi-H 7Revision 2.0, February 5, 2016
No. Symbol Short key-press Long key-press
1 Move marker to the left
Mark saved results
Delete saved results
2 Move marker upward
Select sample number
(ascends)
Switch between Yes and
No setting
Select temperature-
compensation
coefficient α
3 Move marker to the right
Display saved results
Transfer one saved
result to a printer/PC
Transfer a series of
saved results to a
printer/PC
4 Move marker downward
Select sample number
(descends)
Switch between Yes and
No setting
Call up adjustment
mode
5
esc
Exit the menu Switch instrument on or
off
6
ok/meas.
Start measurement
Confirm input
Confirm data delete
Confirm data transfer
While pressing and
holding key 5: enter the
menu
Only the arrow symbols are used to represent keys 1 – 4 in the
following operating instructions.
4 Tutorial
4.1 Inserting batteries
1 Open the battery compartment cover on the back of the Densi-H.
2 Remove battery compartment.
3 Insert batteries into the battery compartment, observing correct polarity.
4 Insert battery compartment, ensuring that the contacts of the battery
compartment face the instrument.
Densi-H8 Revision 2.0, February 5, 2016
5 Close battery compartment cover.
ð The instrument switches on automatically and is immediately ready for
operation. With the backlighting switched off the battery lasts about 90
hours.
Note
§ If nothing appears in the display.
Check polarity of the batteries and battery compartment.
4.2 Configuration
Configuring the instrument for test measurement.
Entering the menu
Press the keys ok and esc simultaneously.
ð The instrument switches to the FUNCTION menu. Measure Unit
appears in reverse display (white on black).
Selecting setting
1 Press the ok key.
ð The Measure Unit menu is confirmed.
2 Press the key repeatedly until SG (t/t) is marked.
3 Press the ok key.
ð SG (t/t) is confirmed.
Setting date and time
1 Press the key repeatedly until Date&Time is marked.
2 Press the ok key.
ð The Date&Time menu is confirmed.
Date is marked.
3 Press the ok key.
4 Use the side arrows to select the number to change, and use the up
and down arrows to change the value.
5 Press the ok key.
ð The Date is set.
6 Use the key to switch to Time and confirm with ok.
7 Set the time as described above.
ð The Time is set.
Exiting the menu
Press the esc key.
ð The instrument is configured for test measurement.
Densi-H 9Revision 2.0, February 5, 2016
4.3 Test measurement
12
3
1 Push drain button (3) downward completely.
2 Ensure that the sample tube is located in the accompanying density
standard or in distilled water.
3 Press the fill button (12) slowly to fill the measuring cell.
4 Ensure that no air bubbles are contained within the measuring cell.
5 Press the meas. key.
ð The instrument automatically executes a measurement and the result
appears in reverse display.
The result for water must be 1.000 in the selected measurement unit
SG (t/t).
Note
§ If the deviation is ≤ 0.0005
Adjustment is ok, the instrument is ready for measurement.
§ If the deviation is > 0.0005
Adjust instrument, see [Adjustment}10].
4.4 Adjustment
Note
You must be in the "expert" mode to perform a water calibration
1 Press and hold the key until CALIB (Water) appears on the display.
ð The instrument adjusts automatically (duration: approx. 1 minute).
After adjustment is completed, the measured deviation from the
theoretical value and Execute? ( No ) appears.
2 Press the or key.
ð Execute? ( Yes ) appears.
3 Press the ok key.
ð The adjustment is confirmed.
Densi-H10 Revision 2.0, February 5, 2016
4.5 Cleaning
§ Clean the measuring cell daily with a suitable cleaning liquid. The
built-in sample pump or an external syringe can be used to fill the
measuring cell with cleaning liquid.
1 Press the drain button downward completely to empty the measuring
cell.
2 Press the fill button to fill the measuring cell with cleaning liquid.
3 For very dirty measuring cells, let the inside of the cell soak in the
cleaning liquid.
4 Repeat cleaning if necessary.
5 If the housing is soiled, clean with a cleaning tissue.
4.6 Switching off and on
Switching Off
Press and hold the esc key until the display is turned off.
ð The instrument is now turned off.
Switching On
Press and hold the esc key until the display appears.
ð The instrument is ready for operation.
5 Menu (menu)
The menu of the Densi-H offers the following functions:
[Measurement Unit (Measure Unit)}12]
[Temperature Unit (Temp. Unit)}15]
[Measure Mode}15]
[Adjustment Mode (Calib. Mode)}16]
[Interface}16]
[Beep}16]
[Background Lightning and LCD-Contrast (LCD)}17]
[Automatic Switch-off (Power)}17]
[Software Version (Version No.)}17]
[Date and Time (Date&Time)}17]
[Identification}17]
5.1 Using the menu
Entering the menu
Press the keys ok and esc simultaneously.
ð The instrument switches to the menu.
Densi-H 11Revision 2.0, February 5, 2016
Selecting functions
1 Press the and keys repeatedly until the desired function is
marked.
2 Press the ok key to activate the marked function.
ð The instrument switches to the corresponding submenu or activates the
desired function.
Numerical input
1 Use the and keys to select the decimal places.
2 Use the and keys to change the value.
3 Press the ok key to confirm the value.
Exiting the menu
Press the esc key.
5.2 Measurement Unit (Measure Unit)
The following units are available for selection:
Density
Available density units:
g/cm3
lb/gal (US)
1 g/cm³ = 8.3454 lb/gal
lb/gal (IP)
1 g/cm³ = 10.0224 lb/gal
Comp. Density (Temperature-compensated density)
Density measurement in g/cm
3
at a reference temperature.
All results at the same reference temperature (T
0
, e.g. 20 °C), regardless
of the measuring temperature (T).
Compensated density = measured density * (1 + α * (T – T
0
))
10 temperature-compensation coefficients can be saved.
The following input is required:
Comp. No.
Number of the temperature-compensation coefficient
(0…9)
Comp. Temp.
Reference temperature (T
0
)
a*1000
Temperature-compensation coefficient
For typical temperature-compensation coefficients see [Temperature-
compensation coefficients α}32] and for calculation of α see
[Measurement Unit (Measure Unit)}14].
Densi-H12 Revision 2.0, February 5, 2016
SG (t/t) (specific gravity)
T Measuring temperatures
ρ
s
Density of the sample at T
ρ
w
Density of water at T
SG (specific gravity, temperature-compensated)
ρ
s
(T
0
) Density of the sample, temperature-compensated for T
0
ρ
w
(T
1
) Density of water at T
1
10 temperature-compensation coefficients can be saved.
The following input is required:
Comp. No.
Number of the temperature-compensation coefficient
(0…9)
Comp. T.
T
0
/T
1
°C (enter temperatures T
0
and T
1
)
a*1000
Temperature-compensation coefficient
For typical temperature-compensation coefficients see [Temperature-
compensation coefficients α}32] and for calculation of α see
[Measurement Unit (Measure Unit)}14].
API (American Petroleum Institute)
The measured values are converted to a reference temperature of 15 °C or
60 °F.
This conversion is based on API tables:
product group A:
crude oil
product group B:
fuel
petroleum products
product group D:
lubricants
The result can be expressed as density (e.g. API A (Density)), or directly
in API degrees (e.g. API A (degrees)).
Brix (Sucrose)
Measurement of the sucrose concentration, percent weight at 20 °C.
Alcohol (Ethanol)
Measurement of the ethanol concentration in water, percent weight (Wt%)
or percent volume (Vol%) at 20 °C.
H2SO4 (Sulphuric acid concentration)
Measurement of the sulphuric acid concentration, percent weight at 20 °C.
Densi-H 13Revision 2.0, February 5, 2016
Baume (°Baumé scale)
Measurement in °Baumé (heavy Baumé for d>1, light Baumé for d<1)
calculated to a reference temperature.
10 temperature-compensation coefficients can be saved.
Input of temperature-compensation coefficients via Comp. Density.
For typical temperature-compensation coefficients see [Temperature-
compensation coefficients α}32] and for calculation of α see
[Measurement Unit (Measure Unit)}14].
Plato
Measurement of the °Plato at 20 °C.
Proof
Measurement of the Proof degree at 60 °F.
The Proof degree is a unit used to determine the ethanol content.
Proof (US):
Measurement with US unit (100 v/v % = 200 US Proof ).
Proof (IP):
Measurement with IP unit (100 v/v % = 175 British
Proof ).
Conc. (Concentrations)
Measurement of the concentration via the input of the desired concen-
tration-conversion formula y = a + bx at a reference temperature.
y concentration in % or without units
a, b sample-dependent coefficients
x measured density
1/compensated density - 1
compensated density - 1
compensated density
1/specific gravity - 1
specific gravity - 1
specific gravity
10 temperature-compensation coefficients can be saved.
Input of temperature-compensation coefficients via Comp. Density.
For typical temperature-compensation coefficients see [Temperature-
compensation coefficients α}32] and for calculation of α see
[Measurement Unit (Measure Unit)}14].
Calculation of the temperature-compensation coefficient α
1 Measure density at a temperature (T
1
) above the normal measuring
temperature.
2 Measure density at a temperature (T
2
) below the normal measuring
temperature.
3 Calculate α according to the formula:
density at T
2
= density at T
1
* (1 + α * (T
1
– T
2
)) or
4 Enter value a x 1000 into the instrument.
Densi-H14 Revision 2.0, February 5, 2016
Note
The temperatures T
1
and T
2
have to be entered in the selected unit °C or °F,
see [Temperature Unit (Temp. Unit)}15].
Example
(measured) density at 26 °C (T
1
) = 0.7844 g/cm
3
(measured) density at 15 °C (T
2
) = 0.7937 g/cm
3
Enter this value into the instrument.
10 temperature-compensation coefficients can be saved.
Input of temperature-compensation coefficients via Comp. Density.
For typical temperature-compensation coefficients, see [Temperature-
compensation coefficients α}32].
5.3 Temperature Unit (Temp. Unit)
Data in °C or °F (selectable).
5.4 Measure Mode
Configuration of sample identification, stability mode and method of data
storage.
Sample Name
Sample identification. A letter (a…z or space) can be set for the identi-
fication of samples.
Stability
Stability control.
Auto
The result is accepted automatically when the display is stable.
Manu
Result accepted by pressing the ok key.
Mode
Method of data storage.
Labo
By pressing the ok key the result is saved and transferred (printer,
PC).
Field
By pressing the ok key the result is saved.
Custom
User-defined setting.
Memory in: Save results.
Auto:
Save result automatically.
Manu:
Result saved by pressing meas. key.
Densi-H 15Revision 2.0, February 5, 2016
Memory out: Transfer result to PC or printer.
Auto:
Transfer result automatically.
Manu:
Transfer result by pressing the key.
5.5 Adjustment Mode (Calib. Mode)
Off
Adjust measuring cell with accompanying density standard or distilled
water.
On
Adjust measuring cell with desired density standard.
The following input is required:
Density
Density of the density standard (g/cm³) at the reference
temperature.
Temp.
Reference temperature
a*1000
Temperature-compensation coefficient of the density standard *
1000
5.6 Interface
PRN
Printer interface.
Printer with serial interface and connected infrared adapter. Results are
formatted for output to a printer.
Transfer rate (baud rate), parity and stop and data bits must be configured
according to the peripheral device.
The following settings are required for the RS-P26 printer:
Baud rate: 9600
Parity: none
Stop bits: 1
Data bits: 8
RS
Serial interface.
The Excel macro “PortableCapt” for data transfer to the computer via the
infrared adapter is located on the CD-ROM.
IrDA
Data transfer to the PC with the integrated IrDA interface in accordance
with protocol 1.20.
5.7 Beep
Off
Beep off.
On
Beep on.
Densi-H16 Revision 2.0, February 5, 2016
5.8 Background Lightning and LCD-Contrast (LCD)
Light
The backlighting is turned off automatically 5 seconds after the last time a
key has been pressed (Auto off), or it is always off (Always off).
Contrast
Display contrast is adjustable to one of 9 levels with the keys.
5.9 Automatic Switch-off (Power)
Off
Automatic switch-off off. The instrument must be switched off manually.
On
The instrument switches off automatically if not operated for 10 minutes.
5.10 Software Version (Version No.)
The software version is displayed.
5.11 Date and Time (Date&Time)
The settings for date and time can be edited here. The date is displayed in
the format year/month/day (e.g. 2003/03/04 for March 4, 2003). Both
date and time are included in the data transfer to a printer or computer.
5.12 Identification
An identification consisting of 10 letters and numbers (instrument name,
user, etc.) may be entered here. This identification will be included in the
data transfer to a printer or computer.
6 Measurement (meas)
6.1 Procedure for proper measurement
Procedure for proper measurement
1 Test the instrument for accuracy with the accompanying density
standard or distilled water before beginning any measurements, see
[Test measurement}10].
2 Insufficient cleaning leads to remaining residue in the measuring cell
and, therefore, incorrect results. Ensure that the measuring cell is clean
before each measurement.
3 Check the sample for chemical resistance of the instrument materials.
Instrument materials
Sample tube: PTFE (polytetrafluorethylene)
Measuring cell: Borosilicate glass
Measuring cell holder: PPS (polyphenylene sulphide)
Sample pump: PP (polypropylene)
4 Ensure that the samples to be measured are liquid enough to be
sucked up or injected.
5 Ensure that the samples to be measured can be dissolved with a
solvent suitable for cleaning the measuring cell.
Densi-H 17Revision 2.0, February 5, 2016
6 Ensure that the samples to be measured are homogeneous (no
emulsions or suspended particles, no air bubbles).
7 Ensure that the samples to be measured have reached ambient
temperature in the measuring cell.
8 For samples which are 20 °C colder than the ambient temperature,
heat the sample before beginning the sampling procedure.
ð The unit display flashes when the difference in temperature between
the sample and the ambient air is greater than ±5 °C.
9 For samples of high viscosity (> 2000 mPa*s), use an external
syringe to fill the measuring cell.
6.2 Filling the measuring cell using the built-in sample pump
1 Ensure that the sample tube is in the sample.
2 Press the fill button (12) slowly to fill the measuring cell.
3 Ensure that there are no air bubbles in the measuring cell.
6.3 Filling the measuring cell using an external syringe
11
3
6
Densi-H
CAL
1 Press the drain button (3).
2 Slide up the lock button (11).
3 Remove the screw stopper (6) with a coin.
4 Screw in the adapter for the external syringe.
Densi-H18 Revision 2.0, February 5, 2016
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44

Cannon Densi-H User manual

Category
Measuring & layout tools
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI