Velleman DVM52IT Specification

Category
Multimeters
Type
Specification
DVM52IT VELLEMAN
1
DVM52IT – DIGITAL INSULATION TESTER
1. Introduction & Safety Prescriptions
To all residents of the European Union
Important environmental information about this product
This symbol on the device or the package indicates that disposal of the device after its lifecycle could harm
the environment.
Do not dispose of the unit (or batteries) as unsorted municipal waste; it should be taken to a specialized
company for recycling.
This device should be returned to your distributor or to a local recycling service.
Respect the local environmental rules.
If in doubt, contact your local waste disposal authorities.
Thank you for buying the DVM52IT! Please read the manual carefully before bringing this device into service.
The meter was designed according to IEC 348 and according to IEC-1010 concerning the safety requirements for
electronic measuring instruments with an overvoltage category (CAT II) and pollution 2. This manual contains a
number of safety prescriptions that must be followed to the letter. Read these instructions before using the device!!
Injury or death can occur even with low voltages and low current. Consequently it is extremely important that you
read these safety instructions before using your device. Follow all safety and operating procedures as described in
this manual.
Protect yourself against electroshocks.
Do not use this device for any other application than those described in this manual.
Make sure the device was not damaged in transit.
Make sure the insulation of the test leads is not damaged and/or the wire itself is not exposed.
Full compliance with safety standards can only be guaranteed if the device is used with the supplied test leads. If
necessary, they should be replaced with identical leads or leads with identical electric ratings. All test leads
should be in good working order.
Never exceed the specified limit values for the various measurement ranges.
Do not touch unused terminals when the meter is connected to a circuit.
Do not measure voltages > 1000V above earth ground.
Exercise extreme caution when working with voltages in excess of 60VDC or 30Vrms AC. Keep your fingers
behind the probe barriers while using the device.
Do not connect the leads to a voltage source while the function switch is in one of the following modes: insulation
resistance, resistance or continuity.
Never perform measurements on live circuits.
Disconnect all test leads from the circuit to be tested prior to selecting a different function or range.
Have the device checked by a qualified technician in case of malfunction.
Never use the meter if the back panel is not in place and firmly fixed.
Do not use or store the device in areas exposed to direct sunlight, high temperatures or high humidity.
2. Symbols
Important information with reference to safety, consult the manual!
Double insulation (Protection class II)
DVM52IT VELLEMAN
2
3. Maintenance
Do not remove the back cover while a voltage is being applied. Only a qualified technician should handle repairs.
Always disconnect the test leads from all current sources before opening the meter.
Always place the function and range switches in the OFF position when the device is not in use.
Remove the batteries to keep them from leaking if the device is to be stored for a prolonged period of time.
Use a damp cloth and a mild detergent to clean the device. Never apply abrasives or solvents to the meter
4. Description
General
This digital insulation tester is equipped with two alarm functions. A humming sound is produced every two
seconds to prevent damage from improper operation when the function or range switch is in the wrong position.
Press the test button and the alarm system will produce a humming sound every two seconds. The high voltage
output indicator will flash red, warning the operator of the risk of electroshocks because of the high output voltage.
The device is protected against overload and a low-battery indication is provided.
The device also has a data-hold function.
Two rotary switches are used to select the functions and ranges.
The DVM52IT has a test button that can be locked. Releasing the test button automatically discharges the
capacitance of the device.
Front & Back Panel
1. High voltage output indicator
2. LCD display
3. Front panel
4. Test button
5. Data-hold switch
6. Function switch
7. Range switch
8. HIGH jack
9. LOW jack
10. Battery cover
11. Case
5. Specifications
Max. accuracy is achieved during a one-year period after calibration. Ideal circumstances require a temperature of 18
to 28°C (64 to 82°F) and a max. relative humidity of 75%.
Earth ground
Low battery
AC (alternating current)
DC (direct current)
Continuity buzzer
DVM52IT VELLEMAN
3
5.1 General Specifications
Max. voltage between terminals and earth 1000V dc or ac
Measuring Method Dual-slope integration A/D converter
Sampling Time ±0.4sec.
LCD Height 22mm
Max. Display 1999 points (3 ½ digits)
Polarity Indication ‘-‘ indicates negative polarity
Overrange Indication ‘1’ or ‘-1’ is displayed
Displayed Measuring Unit Unit of electrical capacity
Power Supply 6 AA-batteries of 1.5V (alkaline batteries are recommended)
Power Consumption ±5mA (100Vdc, 750Vac, 200, )
±30mA (200M, 250V)
±50mA (200M/500V)
±100mA (2000M/1000V)
Battery-Low Indication ”-symbol is displayed
Operating Temperature 0 to 40°C (32 to 104°F)
Operating Humidity < 85%RH
Storage Temperature -10 to +50°C (10 to 122°F)
Dimensions 192 x 122 x 55mm
Weight ± 545g (with batteries)
5.2 Electrical Specifications
5.2.1. Insulation Resistance
Range
200M
/ 250V 200M
/ 500V 2000M
/ 1000V
Test Voltage 250Vdc ± 10% 500Vdc ± 10% 1000Vdc ± 10%
0 ~2000M
Measuring Ranges
0 ~200M 0 ~200M
0 ~1000M 1000 ~2000M
Accuracy ± 3% rdg ± 5 digits ± 5% rdg ± 5 digits
Output Voltage on Open Circuit 250V ± 10% 500V ± 10% 1000V ± 10%
Min. Output Voltage
225V at 0.25M 450V at 0.5M 900V at 1M
Test Current (approx.)
1mA at 0.25M 1mA at 0.5M 1mA at 1M
Output Short-Circuit Current
2.5mA
5.2.2. AC Voltage
Range Resolution Accuracy
700V 1V ±1.2% of rdg + 5 digits
Input Impedance 10M
Max. Input Voltage 700V rms AC or 1000V DC
Frequency Range 40 to 400Hz
Response average, calibrated in rms of a sine wave
DVM52IT VELLEMAN
4
5.2.3. DC Voltage
Range Resolution Accuracy
1000V 1V ±0.8% of rdg + 3 digits
Input Impedance 10M
Max. Input Voltage 1000V DC or 700V rms AC
5.2.4. Resistance
Range Resolution Accuracy
200 0.1
±1.0% of rdg + 3 digits
5.2.5. Continuity
Range Function
Built-in buzzer sounds if continuity exists (resistance < 50)
Open Circuit Voltage ±2.5V
Overload Protection 250V DC or rms AC
6. Operating Instructions
6.1. Selecting the Appropriate Function and Range
When measuring AC voltage, DC voltage or resistance or when performing a continuity test, you should place the
function switch in the 200 , 700V~, 1000V position. Place the range switch in the 200 , 700V~ or
1000V position.
When measuring insulation resistance, you should place the function switch in the MANU., LOCK 1min., LOCK 2min.
or LOCK 4min. position. The device will produce a humming sound every 2 seconds and the display will show a
random reading if either the range or the function switch is in the wrong position.
6.2. Sonorous and Visual Alarm Signals
The device will emit a humming sound every two seconds and the red high voltage output indicator in the top right-
hand corner of the LCD will flash if the test button is pressed while the insulation range is selected.
6.3. Data-Hold Function
Simply toggle the hold switch to the right if you want to hold a particular value while measuring. The data-hold symbol
appears on the LCD. Slide the switch to the left again to leave the data-hold mode.
DVM52IT VELLEMAN
5
6.4. Manual Operation and Using the Test Button
When measuring insulation, you can rotate the function switch to the “MANU.” position. The meter is now operated
manually. The measurement begins when you press the test button. For prolonged measurements it can be very
practical to lock the test button : just press the button and turn it anticlockwise. You can place the function switch in
the LOCK 1min., LOCK 2min. or LOCK 4min. position to perform measurements of the indicated duration. Simply
place the function switch in the MANU. position again if you want to interrupt your measurements before the 1, 2 or 4
minutes are up.
6.5. Preparing for Measurements
When the symbol is printed beside on of the jacks, this means that the indicated limit values must be respected in
order to protect the inner circuits from damage.
The low-battery sign indicates that the batteries should be replaced. This symbol is displayed when the battery
voltage drops below 7V.
6.6. Measuring Insulation Resistance
Insulation tests should only be conducted on deactivated circuits. Connect the DVM52IT and the
circuit in question with test leads and press the test button to execute the test.
1. Select the required test mode (MANU., LOCK 1min., LOCK 2min., LOCK 4min.) with the function switch.
2. Select the required range (200M/250V, 200M/500V, 2000M/1000V) with the range selector.
3. Connect the black test lead with the LOW jack and the red test lead with the HIGH jack.
4. Connect the test leads with the circuit to be tested.
5. Press the test button. In the manual mode, you can press and then rotate the test button anticlockwise to lock it.
6. Read the measured value from the LCD.
Never touch the circuit under test during your measurements. Do not rotate the range switch while
the test button is in the depressed position.
When your measurements are completed you must release the test button before the test leads are
disconnected. This is because the system is charged up and must be allowed to discharge through
the tester’s internal discharge resistor.
6.7. Measuring AC Voltage
1. Connect the black test lead with the LOW jack and the red test lead with the HIGH jack.
2. Put the function switch in the 200 , 700V~ or 1000V position.
3. Put the range switch in the 700V~position.
4. Connect the test leads with the source or load to be tested.
5. Read the measured value form the LCD.
The symbol indicates that the limit values of 700V rms AC and 1000V DC have to be respected. It may be
possible to measure higher voltages, but this may destroy the inner circuit. Be careful to avoid electroshocks when
performing measurements.
DVM52IT VELLEMAN
6
6.8. Measuring DC Voltage
1. Connect the black test lead with the LOW jack and the red test lead with the HIGH jack.
2. Put the function switch in the 200 , 1000V . or 700V~ position.
3. Put the range switch in the 1000V position.
4. Connect the test leads with the source or load to be tested.
5. Read the measured voltage and the polarity of the red lead connection from the LCD.
The symbol indicates that the limit values of 700V rms AC and 1000V DC have to be respected. It may be
possible to measure higher voltages, but this may destroy the inner circuit. Be careful to avoid electroshocks when
performing measurements.
6.9. Measuring Resistance
Disconnect the circuit to be tested and make sure that all capacitors are fully discharged before measuring the in-
circuit resistance.
1. Connect the black test lead with the LOW jack and the red test lead with the HIGH jack.
2. Put the function switch in the 200 , 1000V . or 700V~ position.
3. Place the range switch in the 200 position.
3. Connect the test leads with the resistance under measurement.
4. Read the value from the LCD.
REMARKS
The LCD displays the overrange indication “1” if the measured value is too high for the range you are using.
Select a higher range.
The overrange indication “1” is displayed when the input is not connected e.g. in case of an open circuit.
6.10. Continuity Test
1. Connect the black test lead with the LOW jack and the red test lead with the HIGH jack.
2. Put the function switch in the 200 , 1000V . or 700V~ position.
3. Place the range switch in the position.
4. Connect the test leads with two points of the circuit to be tested.
5. The built-in buzzer will sound if continuity exists (resistance < 50).
7. Maintenance
7.1. Replacing the Batteries
Disconnect the test leads from the circuit to be tested before removing the battery cover. This is
necessary to avoid the risk of electroshocks.
1. The batteries need to be replaced when the sign appears on the LCD.
2. Open the battery cover with the included screwdriver.
3. Replace the old batteries with new ones.
4. Replace the battery cover.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24