4 Important Safeguards
Vessel Safeguards
• Only use vessels and vessel components purchased directly from CEM or through its authorized dealer network.
• All vessel components must be dry prior to use
• Always condition new liners and plugs
• Do not batch old liners with new liners unless new liners have been conditioned with at least 3 runs using the
customer’s sample
• Ensure vessels are assembled properly
• Plugs should not freely fall into liners. Some resistance should be felt. Group liners and plugs with similar
resistance.
• Align tab on load disc in center notch of liner
• Look for vent hole through the frame
• Align liner and sleeve properly on insulator disc at bottom – liner and sleeve ush with tab
• Use torque block
• Invert sleeves after every run to extend lifetime and promote better sealing & even heating
• Slide liner into the end of the sleeve with most resistance felt if unsure which side was previously used
• CEM recommends using a high temperature pen (P/N 302070) to mark sleeves for a visual reference.
• Try to group liners with similar sample history and usage
• Segregate liners for HNO3 and non-HNO3 applications
• Do not use a new liner with a set of liners that has been run consistently until the new liner has been
conditioned with that sample type
• Pull ones that run 20 ºC hotter than average and run them together
• Bake out liners and plugs to remove residual HNO3 & NOx gases
• Use Bake Out procedure for iPrep vessels (210 °C for 6 hrs – overnight)
• Bake out liners if TempGuardTM is triggered
Microwave Safeguards
• Always use the TempGuard feature in the method for software versions 1.43 - 1.48. TempGuard is preset and
non-adjustable in version 1.49 (or greater).
• Set the TempGuard limit at 30 °C > control temperature if control temperature is < 250 °C.
• Set the TempGuard limit at 20 °C > control temperature if control temperature is > 250 °C
• Limit TempGuard to 270 °C for digestion of organics with HNO3 (primarily)
• Limit TempGuard to 300 °C for H3PO4/H2SO4 (primarily) applications
• Limit control temperature of organics with HNO3 to 250 °C
• Choose Sample Type in method with software version 1.45 or higher as follows:
• Standard = Any acid mixture other than HCl or HF dominant
• HCl = aqua regia or acid mixtures that are ≥ 75% HCl
• Always use the Max Power Monitor and set the alarm time for 10 minutes
• Maximum programmable temperature is 300 °C
• Maximum temperature varies per acid. See "Operating Parameters" section for temperatures
• Maximum programmable Hold Time is 45 minutes
• If a run is stopped or paused for more than 30 seconds during the digestion, do not restart the method with hot
vessels
• Allow vessels to cool < 80 °C
• Vent vessels and reseal
• Restart method from the beginning
• Allow vessels to cool below 80 ºC before opening the microwave door.