1 ©AdamEquipmentCompany2017
AdamEquipment
MUWHEALTHANDFITNESSSCALE
USERMANUAL
(P.N.7.00.6.6.0229RevisionB–Aug2017)
2 ©AdamEquipmentCompany2017
CONTENTS
CONTENTS ......................................................................................................................... 2
1.0 INTRODUCTION ....................................................................................................... 3
2.0 SET UP ..................................................................................................................... 3
2.1 UNPACKING AND SETTING UP YOUR MUW ..................................................... 3
3.0 LOCATING AND PROTECTING YOUR SCALE ....................................................... 4
4.0 KEYPAD / DISPLAY FUNCTION .............................................................................. 5
4.1 REAR PANEL ........................................................................................................ 6
4.2 INTERNAL RECHARGEABLE BATTERY ............................................................. 6
5.0 BASIC OPERATION ................................................................................................. 7
5.1 SWITCHING THE BALANCE ON .......................................................................... 7
5.2 ZEROING .............................................................................................................. 7
5.3 WEIGHING UNITS ................................................................................................ 7
5.4 HEIGHT UNITS ..................................................................................................... 7
5.5 WEIGHING/HEIGHT/BMI ...................................................................................... 8
5.6 PRINTING AND OUTPUTTING THE RESULTS ................................................... 9
5.7 CLEARING A PREVIOUS RESULT .................................................................... 10
6.0 PARAMETERS ....................................................................................................... 10
6.1 SETTING THE BACKLIGHT OPERATION F1 EL .................................................. 10
6.2 SETTING AUTO OFF F2 OFF ................................................................................ 11
6.3 SETTING THE PRINTING CONTROLS F3 SEr .................................................... 11
6.4 PRINTING DETAIL CONTROL F4 PtSEL .............................................................. 1 2
6.5 ULTRASONIC HEIGHT CALIBRATION F5 ULtrA ................................................. 13
6.6 SETTING THE HOLD FUNCTION F6 HOLD .......................................................... 13
6.7 SETTING DATE AND TIME F7 d-t ........................................................................ 14
6.8 SET UP F8 SET UP.................................................................................................. 14
6.9 TECHNICAL PARAMETERS F9 TECH .................................................................. 15
7.0 USER CALIBRATION ............................................................................................. 15
8.0 SERIAL INTERFACE SPECIFICATIONS ............................................................... 16
9.0 TROUBLE SHOOTING ........................................................................................... 17
9.1 ERROR MESSAGES .......................................................................................... 17
10.0 TECHNICAL SPECIFICATIONS ............................................................................. 17
11.0 WARRANTY INFORMATION ................................................................................. 18
3 ©AdamEquipmentCompany2017
1.0 INTRODUCTION
The MUW Health and Fitness scale provides the weight and height of a patient
beforecalculatingtheBMIbasedontheresults.Eachresultwillbedisplayedonthe
Indicator and held until the result is cleared, the BMI result can be used to help
evaluate some ones overall health. Units of weighing are Kg, Lbs, and Lbs and
Ounces. Height measurements are displayed in cm, feet and inches, and inches.
There is an internal rechargeable battery as well as wheels fitted to the base that
maketheMUWtotallyportable.TheresultscanbesenttoaprinterorPCviathe
RS232communicationport,eithermanuallyorautomatically.
2.0 SETUP
2.1 UNPACKINGANDSETTINGUPYOURMUW
TheMUWhasbeenpackedinsuchawaytomakeitquickandeasyfortheUSERto
assemble,pleasefollowtheinstructionsasbelow.
1) Removethebaseandpillarsectionsfromthepackaging.
2) Liftthelowerpillarsectionandplaceitwiththescrewholesattherearonto
thebackofthebaseallowingforthecabletohangfreelyintheopenarea.
3) Fitandtightenthe4xM5x15boltsandwashersthroughthefixingplateof
thelowerpillarsectionandintothebasethenslidethepillarcoverdownso
that it locates securely. Fix the pillar cover at the rear using 2 x M4 x 10
screws.
4) Liftthecentralpillarsectionintopositionandslideitwiththescrewholesat
therearontothefixinginsertensuringthecablehangsfreelyinsidethepillar.
5) Fitandtighten4xM4x10screwsthroughthepillarandintotheinsert.
6) Liftthetoppillarsectionintopositionandslideitontothefixinginsertwith
thesensorhousingfacingthefront,againensuringthecablehangs freely
insidethepillar.
7) Fitandtighten4xM4x10screwsthroughthepillarandintotheinsert.
4 ©AdamEquipmentCompany2017
TheMUWcomeswith4xwallfixingbracketsthatcanbesecuredtothepillar
beforefixingto the wall.Ifthe wallfixingbrackets arebeingusedattach these
whenyoufixtheM4x10screwstothepillarfixinginserts.Fastensecurelytothe
wallonceyouhaveselectedthepositiontheMUWwillbeusedin.
3.0 LOCATINGANDPROTECTINGYOURSCALE
Avoid extremes of temperature. Do not place in
directsunlightornearairconditioningvents.
Avoidunstablepowersources.Donotusenearlarge
users of electricity such as welding equipment or
largemotors.Donotletthebalancebatterygoflat.
Ifyouarenotusingitforalongtimeyoushould
chargethebatteryperiodicallytomakesurethe
batterydoesnotloseitscharge.
Keepfreefromvibration.Donotplacenearheavyor
vibratingmachinery.
Avoidhighhumiditythatmightcausecondensation.
Keep away from direct contact with water. Do not
sprayorimmersethebalanceinwater.
Donotplacenearopenwindows,airconditioning
ventsorfansthatmaycauseadraftandunstable
readings.
Keepthescaleclean.Donotstackmaterialonthe
toppanwhenitisnotinuse.
5 ©AdamEquipmentCompany2017
4.0 KEYPAD/DISPLAYFUNCTION
Thekeypadanddisplayhaveanumberoffeaturesasshownandexplainedbelow:
ChargingLED
BatterySymbol
ZeroIndicator
StableIndicator
TareSymbol
UltrasonicSensor
Symbol
6 ©AdamEquipmentCompany2017

4.1 REAR
PANEL
Therearpanelhasaconnectorinputfortheexternalpoweradaptor (12 volts DC
@800ma).
4.2 INTERNAL
RECHARGEABLEBATTERY
The scale can be operated from the internal rechargeable batteryorusingan
adapter.Thebatterylifeisapproximately24hourswiththebacklightoff.Depending
ontheusageofthebacklightthelifetimeofthebatterywillreduce.
Thedisplaywillshowanindicationwhenthebatteryneedstobecharged.Tocharge
the battery plug the adapter intothePOWERinputsocketinthe rear panel and
connectittothemainspowersupply.Thechargingindicatorinthelowerleftcorner
oftheLCDwillbeontoshowitischargingandwillturngreenwhenthebatteryis
fullycharged.
KEYS PRIMARYFUNCTION SECONDARYFUNCTION
ToturnthescaleOnorOff.
Toclearthereadingsonthedisplay
readyforthenextweighing
Tochangethedisplayedheight
readingbetweencm,feetorfeetand
inches
Tosaveandescapebacktothe
parametersettings
Tosendacommandtoprintthe
currentdisplayedreadings
TochangethesettingwheninaUSER
parameter
Tochangethedisplayedweight
readingbetweenKg,Lb’sorLb’sand
Ounces
TochangetheUSERparameter
ToZerothescalebeforeweighing
ToTareanyitemtoobtainnetweight
Toentertheselectedsettingwhenina
USERparameter
7 ©AdamEquipmentCompany2017
5.0 BASICOPERATION
5.1 SWITCHINGTHEBALANCEON
Plug in the unit using the adapter or use the internal rechargeablebattery.Itis
recommendedthatyouchargethebatteryforatleast8hoursbeforefirstuse.
1) Toturnonpressthe[On/Off] keyonceandrelease.Thescalewillshowthe
softwarerevision,thebatteryvoltage,andthenaself‐testbeforeshowingthe
stablesignandzeroweightonthedisplay.
2) Thescalecannowbeused.
3) Toturnthescaleoffafterusepressthe[On/Off]keyagain.Thereisanauto
power‐off function that will automatically turn the unit off ifnotusedfora
periodoftime.Thiscanbesetintheparameterssection6.0.
5.2 ZEROING
Youcanpressthe[Zero]keytosetanewzeropointandshowazeroreading.This
may be necessary if the weight reading is not indicating zero with nothing on the
platform,andshouldbedoneifrequiredbeforeeachnewweighingtakesplace.The
zero indicator will show up in the upper left corner of the LCD when the scale
displaysZero.
5.3 WEIGHING
UNITS
Tochangethedisplayed weighing unit, pressthe[kg/lb]keywhichwillthencycle
youthroughtheavailableoptionsofweighingunits.Thiscanonlybedonebeforeor
afteraweighingtakesplace.
5.4 HEIGHT
UNITS
Tochangethedisplayedheightmeasuringunit,pressthe[cm/ft]keywhichwillthen
cycleyouthroughtheavailableoptionsofheightmeasuringunits.Thiscanonlybe
donebeforeorafteraweighingtakesplace.
8 ©AdamEquipmentCompany2017
5.5 WEIGHING/HEIGHT/BMI
To determine a person’s weight, height and BMI the patient should stand on the
platformwithinthemarkedareaandlookstraightahead.Theultrasonicsensorwill
detect that a person is on the platform signified by the symbol in the upper right
part of the display and start to determine the height. When a stable reading of
weightisobtainedthethreereadingswillbedisplayed.
Theresultswillremainonthedisplayuntilthe[Clear]keyispressed.
ResultsfromaBMImeasurementcanbedeterminedasthetablebelow.
BMI WeightStatus
Below18.5 Underweight
18.5–24.9 Normal
25.0–29.9 Overweight
30.0andAbove Obese
9 ©AdamEquipmentCompany2017
Mother and Baby weighing function
TheMDWscaleallowsuserstoweighthemotherandthenweighthe baby being
heldinthemothersarms.Tooperatethefunction,havethemother step on the
weighingpanandwaitfortheweighttobestable.
ThenpresstheTarebutton.ThedisplayshouldreadZero.
Havethemotherholdthebabyinherarmswhilstshestandsonthe scale. The
weightdisplayedthenistheweightofthebaby.
Themothercannowgetoffthescale.PresstheTarebuttontoclearthemother’s
weightfromthememory.
5.6 PRINTING
ANDOUTPUTTINGTHERESULTS
If a printer or other data collection device is being used the results will be sent
automatically or manually depending on the printing control setup.Thedata
transmissionspeedmustbesetcorrectlytoallowtheresultstobereceived.
TheDoctorandpatientsidentitynumbercanbeprogrammedintotheIndicatorto
giveyouatraceablerecordwhichalsorecordsdateandtime.
ToentertheDoctorsIdpressandholdthe[Print]keyfor2seconds.Thedisplaywill
showasbelow, usethe [Print]and[Clear]keystomovetheflashingdigitandthe
[kg/lb]and[cm/ft]keystoentertheIdnumberyourequireandpress[Zero]to
save.
ThedisplaywillnowshowthePatientsIdscreen,usethekeysasabovetosettheId
numberandpress[Zero]tosaveandexitbacktonormalweighing.

Seesection6.3and6.4forprintingcontrolsanddatatransmissionspeedoptions
10 ©AdamEquipmentCompany2017
5.7 CLEARINGAPREVIOUSRESULT
Oncetheresultshavebeencalculatedanddisplayedtheywillremaindisplayeduntil
the[Clear] keyispressed.Whenthe patienthassteppedofftheplatformandthe
results have been recorded or printed, the [Clear] key can then be pressed which
willreturnthedisplaytozeroreadyforthenextpatienttobeweighed.
6.0 PARAMETERS
Thebalancehas8parameterswhereoptionscanbesetbytheuser.
FUNCTION SECTION DESCRIPTION
F1
EL
Seesection6.1 Setsthebacklightoperation
DefaultisAUTO
F2
OFF
Seesection6.2 Autoofffunction
Defaultis10
F3
SE
R
Seesection6.3 Setsthe printingcontroland datatransmissionspeed
Defaultis9600
F4
P
T
SEL
Seesection6.4 Printingcontrols
F5
UL
TR
A
Seesection6.5 Calibrationoftheheightsensor
F6
HOLD
Seesection6.6 SetstheHOLDfunctiontoONorOFF
F7
D
-
T
Seesection6.7 Settingthedateandtime
F8
SET
UP
Seesection6.8 Setupparameters
F9
TECH
Serviceonly
6.1 SETTINGTHEBACKLIGHTOPERATION
F1
EL
Thebacklightfunctioncanbesetbytheuser.
AUTO
Setsthebacklighttooperateautomaticallywhenaweightis
placedontheplatformorakeyispressed.
OFF
Setsthebacklighttobeonatalltimes.
ON
Setsthebacklighttobeoffatalltimes.
1) Switchonthescalethenpressthe[kg/lb]keyduringtheself‐checkingtestof
thedisplay.
2) Thedisplaywillshowthefirstfunction
F1
EL.

11 ©AdamEquipmentCompany2017
3) Press the [Zero] key to select
F1
EL
and show the current setting, default is
AUTO.
4) Pressthe[Print]keytochangethesettingforthebacklightandpress[Zero]
tosave,then[kg/lb]toselectthenextparameter.
6.2 SETTING
AUTOOFF
F2
OFF
AnautomaticpowerdowntimecanbesetsothattheMUWturnsoffafteraperiod
wherenoactivityisdetected.Thispreservesthelifetimeofthebatterywhenused
inthismode.
SettingsareOFF,5,10,20,and30minutes.
1) Switchonthescalethenpressthe[kg/lb]keyduringtheself‐checkingtestof
thedisplay.
2) Thedisplaywillshowthefirstfunction
F1
EL.

3) Pressthe[kg/lb]keytoselect
F2
OFF
andthen[Zero]toenterand showthe
currentsetting,defaultis10.
4) Pressthe[Print]keytochangethesettingforthebacklightandpress[Zero]
tosave,then[kg/lb]toselectthenextparameter.
6.3 SETTING
THEPRINTINGCONTROLS
F3
SE
R
TheMUW canbesettoautomaticallyprint“AUTO”,or printondemand“PrN”by
pressingthe[Print]button.
Thedatatransmissionspeed“Bps”oftheMUWmustmatchthatofthePrinteror
otherdevicebeingusedtocollectdatainformationsuccessfully.
Theprintingcontrolandspeedcanbeselectedasfollows.
1) Switchonthescalethenpressthe[kg/lb]keyduringtheself‐checkingtestof
thedisplay.
2) Thedisplaywillshowthefirstfunction
F1
EL.

3) Pressthe[kg/lb]keytoselect
F3
S
ER
andthen[Zero]toenterandshowthe
currentprintingcontrolsetting“UArt”.
12 ©AdamEquipmentCompany2017
4) Eitherprintingondemand(PrN),pressingthe[Print]keyfollowingaweighing,
or printing automatically (AUTO), automatically prints out whenastable
readingandBMIcalculationisdisplayedcanbeselected.
5) Pressthe[Print]keytochangethesettingfortheprintingcontrolandpress
[Zero]tosave.
6) Thecurrentsettingfordatatransmissionspeed“Bps”willnowbedisplayed.
600,1200,2400,4800,9600areavailable.Defaultis9600.
Pressthe[Print]keytochangethesettingandpress[Zero]tosave.
Testwillbedisplayed,pressthe[Print]keytosendatestprintcommand.
7) Press[Zero]tosave,[cm/ft]toexit,then[kg/lb]toselectthenextparameter.
6.4 PRINTINGDETAILCONTROL
F4
P
T
SEL
The MUW can be programmed to provide a customised print out to give you the
readingsyourequire.Doctor,patient,DiSP1,DiSP2,DiSP3detailscanallbeenabled
ordisabledtosuittheinformationyouneedtosee.
1) Switchonthescalethenpressthe[kg/lb]keyduringtheself‐checkingtestof
thedisplay.
2) Thedisplaywillshowthefirstfunction
F1
EL.

3) Pressthe[kg/lb]keytoselect
F4
P
T
SEL
andthen [Zero]toenterandshowthe
currentsettings.
SETTING DESCRIPTION SELECTIONS
dr.Id
DoctorsIdnumber
ON/OFF
PAt.Id
PatientIdnumber
ON/OFF
DiSP 1
Display1showstheweightofthepatient
ON/OFF
DiSP 2
Display2showstheheightofthepatient
ON/OFF
DiSP 3
Display3showstheBMIofthepatient
ON/OFF
CAL
Printoutfollowingcalibration
ON/OFF
d-t
Dateandtime
ON/OFF
13 ©AdamEquipmentCompany2017
4) Pressthe[Print]keytochangethesettingfortheprintingdetailcontroland
press[Zero]tosave.
5) Pressthe[kg/lb]keytoexitandcontinue.
6.5
ULTRASONICHEIGHTCALIBRATION
F5
UL
TR
A
TheUltrasonicheightsensorhasbeenfactorycalibratedandshould not require
frequentcalibration.Ifthereisaproblemwithheightmeasuringproceedasbelow.
1) Switchonthescalethenpressthe[kg/lb]keyduringtheself‐checkingtestof
thedisplay.
2) Thedisplaywillshowthefirstfunction
F1
EL.

3) Press the [kg/lb] key to select
F5
U
LTRA
and then [Zero] to enter the sensor
calibrationsection“AdJUST”XX(XXbeingafactorysetfigure)
4) Pressthe[Zero]keyandthedisplaywillshow“base‐H220”,press[Zero]to
calibratethe“Zero”heightreading.
5) Thedisplaywillshow“test”anda4digitnumber.Pressthe[Zero]keyfroma
positionwheretheheightsensorcannotdetectyou.
6) Whencalibrationiscompletethedisplaywillshow“AdJUSTXX”.
Ifaftercalibrationtheheightsensorisnotmeasuringcorrectly, press [Print] or
[Clear] at point 3 to change XX. A lower number will decrease the height
measurementandahighernumberwillincreaseit.Thencontinuefrompoint4on.
*Aprintoutwithdateandtimecanbeissuedfollowingeverycalibration.
6.6 SETTING
THEHOLDFUNCTION
F6
HOLD
TheHoldfunctioncanbesetsothattheheightandBMIreadingsarevisibleasthey
are being calculated (OFF), or remain as dashes until the weight reading is stable
uponwheretheywillbedisplayed(ON).
1) Switchonthescalethenpressthe[kg/lb]keyduringtheself‐checkingtestof
thedisplay.
2) Thedisplaywillshowthefirstfunction
F1
EL.

14 ©AdamEquipmentCompany2017
3) Pressthe[kg/lb]keytoselect
F6
HOLD
and then[Zero]toenterandshowthe
currentsetting,defaultisON.
4) Pressthe[Print]keytochangethesettingforHold,press[Zero]tosave,then
[kg/lb]toselectthenextparameter.
6.7 SETTING
DATEANDTIME
F7
D
-
T
1) Switchonthescalethenpressthe[kg/lb]keyduringtheself‐checkingtestof
thedisplay.
2) Thedisplaywillshowthefirstfunction
F1
EL
..
3) Pressthe[kg/lb]keytoselect
F7
D
-
T
andthen [Zero]toenterthedatesetting
display.
4) Thedisplaywillshowtheday,monthandyear.
5) Tochangeanysettingpressthe[kg/lb]keytoincrementtheflashingdigitand
eitherthe[Print]or[Clear]keytomovetheflashingdigittotheleftorright.
6) Pressthe[Zero]keytosavethedateandenterthetimesettingparameter
7) Tochangethetimepressthe[kg/lb]keytoincrementtheflashingdigitand
eitherthe[Print]or[Clear]keytomovetheflashingdigittotheleftorright.
8) Press[Zero]tosave,[cm/ft]toexitthen[kg/lb]toselectthenextparameter.
6.8 SET
UP
F8
SET
UP

Four performance parameters, Filter, Auto Zero, Stability and weight tracking are
availabletotheUSER.
Theseparametersinfluencethespeedofthedisplayupdateandhavebeenfactory
set,thedefaultsettingsshouldnormallybeusedtooptimiseperformance.
Availableparametersandsettingsareasbelow.
PARAMETER DESCRIPTION DEFAULTVALUE
FILT
Filtersettingsof1‐6areavailable
1isthefastestfilter,3istheslowest.
6
ZERO
AutomaticRe‐zerorange
Settingsof1to6areavailable
2
STA
Stabilitysymbolrange
6
15 ©AdamEquipmentCompany2017
Settingsof1to6areavailable
1isthesmalleststabilityrangeand6isthelargest.
ZTR
WeightTrackingRange
Settingsof1to6areavailable
4
1) Switchonthescalethenpressthe[kg/lb]keyduringtheself‐checkingtestof
thedisplay.
2) Thedisplaywillshowthefirstfunction
F1
EL.

3) Pressthe[kg/lb]keytoselect
F8
SET
UP
and then[Zero]toenterandshowthe
firstparameter.
4) Pressthe[Print]keytochangethesettingand[Zero]tosaveandmovetothe
nextparameter,press[cm/ft]toexit.
ToreturntonormaloperationturnthescaleOFFandthenONagain.
6.9
TECHNICALPARAMETERS
F9
TECH
Thisisamenuavailableonlytoqualifiedservicepersonnelforfactoryadjustments.
7.0 USERCALIBRATION
1) Switchonthescaleandwhilstthedisplaycountsdownpressthe[Clear]key.
2) Thedisplaywillshowkgorlb,press[Print]toselecttheunitofweightbeing
usedforthecalibrationandthen[Zero]toconfirm.
3) ThedisplaywillshowCALXX(XXbeingthelastcalibrationweighttobeused).
4) Pressthe[Print]keytochangethevalueifrequiredand[Zero]toconfirm.
5) ThedisplaywillshowtheZeroA‐Dwhichshouldbestable.
6) Pressthe[Zero]keytoconfirmtheZeropointandthedisplaywillshowLOd
XX(XXbeingthecalibrationweightvaluetobeloadedontothetoppan).
7) Loadthetoppanwiththecalibrationweightandpressthe[Zero]key.
8) Thedisplaywillshow“PASSforasuccessfulcalibrationandreturntonormal
weighingdisplayingthecalibrationweightvaluecorrectly.
16 ©AdamEquipmentCompany2017
9) RemovetheweightfromthepanandthescalewillreturntoZero.
*Aprintoutwithdateandtimecanbeissuedfollowingeverycalibration.
If an error code was displayed when calibrating please see the error code
tableinsection9.1
8.0 SERIALINTERFACESPECIFICATIONS
ThebalancecomesequippedwithanRS‐232SerialInterface.
TheconnectorisaDE‐9Pfemalefittingwiththefollowingconnections.
Pin2 TXDTransmitteddata
Pin3 RXD Receiveddata
Pin5GND Signalground
PrintingOutput:

Thelineswillincludeaheadingforeachdetailthenthevalue.Atypicaloutputwhen
weighingisshownbelow.6linesofdataareprintedifallareenabled.
Alllinesendwithacarriagereturn<cr>andalinefeed<lf>,(0dHand0aHinASCII).
Inputcommandformat:
ThebalancecanbecontrolledbyaPCwiththefollowingcommands.Thecommands
mustbesentinuppercaseletters,i.e.“T”not“t”.
Z<cr><lf> WillZerothescale,performsthesamefunctionsaspressingthe[Zero] key.
P<cr><lf>
WillsendaPrintcommandtoaPCorPrinter. Performsthesamefunctionsas
pressingthe[Print]key.
C<cr><lf> Will clear the last recorded reading from the display.Performsthesame
DoctorId:00001
PatientId:0000000123
Weight:80.00kg
Height:183cm
BMI: 23.9
D&T 30
/
07
/
201211.20
17 ©AdamEquipmentCompany2017
functionsaspressingthe[Clear] key
W<cr><lf> Will changethe Weight measuring unit being usedbetweenkg,lbsandlbs
andounces.Performsthesamefunctionsaspressingthe[kg/lb]key.
H<cr><lf> Will changethe Height measuringunit being used betweencm,inchesand
feetandinches.Performsthesamefunctionsaspressingthe[cm/ft]key.
9.0 TROUBLESHOOTING
9.1 ERRORMESSAGES
Ifanerrormessageisshown,repeatthestepthatcausedthemessage.Iftheerror
messageisstillshownthencontactyourdealerforsupport.
DISPLAYEDMESSAGE DESCRIPTION
Err
Measuredheightisabove
210cm/82.5inches
UNDER
Measuredheightisbelow
80cm/31.5inches
Err H
Calibrationweightvalue
usedtoohigh
Err L
Calibrationweightvalue
usedtoolow
- - - - - -
Weightontoppanexceeds
300kg/660Lbs
10.0 TECHNICALSPECIFICATIONS
Model MUW300L
MaximumCapacity 300kg/660Lbs
Readability 0.05kg/0.1Lbs
Repeatability(s.d.) 0.1kg/0.2Lbs
Linearity± 0.1kg/0.2Lbs
UnitsofHeight Cm,inches,feetandinches
Heightrange 80cm
210cm/31.5‐ 82.5inches
UnitsofWeight Kg,lb’s,lb’sandounces
PowerSupply 12vDC@800maorInternal
rechargeable6vDCbattery
OverallDimensions
wxdxh
0.38mx0.6mx2.3m
15x23.6x90.2inches
18 ©AdamEquipmentCompany2017
11.0 WARRANTYINFORMATION
AdamEquipmentoffersLimitedWarranty(PartsandLabour)forthecomponentsfailedduetodefects
inmaterialsorworkmanship.Warrantystartsfromthedateofdelivery.
Duringthewarrantyperiod,shouldanyrepairsbenecessary,thepurchasermustinformitssupplieror
AdamEquipmentCompany.Thecompanyor itsauthorisedTechnicianreservestherighttorepairor
replacethecomponentsatanyofitsworkshopsdependingontheseverityoftheproblems.However,
anyfreight involvedin sendingthe faultyunitsorpartstotheservicecentreshouldbe bornebythe
purchaser.
Thewarrantywillceasetooperateiftheequipmentisnotreturnedintheoriginalpackagingandwith
correct documentation for a claim to be processed. All claims are at the sole discretion of Adam
Equipment.
This warranty does not cover equipment where defects or poor performance is due to misuse,
accidental damage, exposure to radioactive or corrosive materials, negligence, faulty installation,
unauthorised modifications or attempted repair or failure to observe the requirements and
recommendations as given in this User Manual.Additionally rechargeable batteries (where supplied)
arenotcoveredunderwarranty.

Repairs carried out under thewarranty does notextendthe warranty period. Components removed
duringthewarrantyrepairsbecomethecompanyproperty.

The statutory right of the purchaser is not affected by this warranty.Thetermsofthiswarrantyis
governedbytheUKlaw.ForcompletedetailsonWarrantyInformation,seethetermsandconditions
ofsaleavailableonourweb‐site.
©AdamEquipmentCompany2017
WEEE2012/19/EU
Thisdevicemaynotbedisposedofindomesticwaste.Thisalsoappliestocountriesoutsidethe
EU,pertheirspecificrequirements.Disposalofbatteries(iffitted)mustconformtolocallaws
andrestrictions.
Cetappareilnepeutêtreéliminéaveclesdéchetsménagers.L’éliminationdelabatteriedoit
êtreeffectuéeconformémentauxloisetrestrictionslocales.
DiesesGerätnichtmitdemHausmüllentsorgt.
Dispositivonopuedeserdesechadojuntoconlosresiduosdomésticos
Dispositivononpuòesseresmaltitoneirifiutidomestici.
FCC/ICCLASSADIGITALDEVICEEMCVERIFICATIONSTATEMENT
NOTE:ThisequipmenthasbeentestedandfoundtocomplywiththelimitsforaClassAdigital
device,pursuanttoPart15oftheFCCrulesandCanadianICES‐003/NMB‐003regulation.These
limits are designed to provide reasonable protection against harmful interference when the
equipmentisoperatedinacommercialenvironment.Thisequipmentgenerates,usesandcan
radiateradiofrequencyenergyand,ifnotinstalledandusedinaccordancewiththeinstruction
manual, may cause harmful interference to radio communications.Operationofthis
equipmentinaresidentialareaislikelytocauseharmfulinterference inwhichcase theuser
willberequiredtocorrecttheinterferenceathisownexpense.
CALIFORNIAPROPOSITION65‐MANDATORYSTATEMENT
WARNING:Thisproductincludesasealedlead‐acidbatterywhichcontainschemicalsknownto
theStateofCaliforniatocausecancerandbirthdefectsorotherreproductiveharm.
Adam Equipment products have been tested with, and are always supplied
with mains power adaptors which meet all legal requirements forthe
intended country or region of operation, including electrical safety,
interference and energy efficiency. As we often update adaptor products to
meetchanginglegislationitisnotpossibletorefertotheexactmodelinthis
manual.Pleasecontactusifyouneedspecificationsorsafetyinformationfor
yourparticularitem.Donotattempttoconnectoruseanadaptor not
suppliedbyus.

©AdamEquipmentCompany2017
ADAMEQUIPMENTisanISO9001:2008certifiedglobalcompanywithmorethan40years’experienceinthe
productionandsaleofelectronicweighingequipment.
AdamproductsarepredominantlydesignedfortheLaboratory,Educational,HealthandFitness,Retailand
IndustrialSegments.Theproductrangecanbedescribedasfollows:
‐AnalyticalandPrecisionLaboratoryBalances
‐CompactandPortableBalances
‐HighCapacityBalances
‐Moistureanalysers/balances
‐MechanicalScales
‐CountingScales
‐DigitalWeighing/Check‐weighingScales
‐HighperformancePlatformScales
‐Cranescales
‐MechanicalandDigitalElectronicHealthandFitnessScales
‐RetailScalesforPricecomputing
ForacompletelistingofallAdamproductsvisitourwebsiteat
www.adamequipment.com
AdamEquipmentCo.Ltd.
MaidstoneRoad,Kingston
MiltonKeynes
MK100BD
UK
Phone:+44(0)1908274545
Fax:+44(0)1908641339
e‐mail:
AdamEquipmentInc.
1,FoxHollowRd.
Oxford,CT
06478
USA
Phone:+12037904774
Fax:+12037923406
e‐mail:
AEAdamGmbH.
Instenkamp4
D‐24242Felde
Germany
Phone+49(0)4340403000
Fax:+49(0)43404030020
e‐mail:
AdamEquipmentS.A.(Pty)Ltd.
7MegawattRoad,
SpartanEXT22
KemptonPark,
Johannesburg,
RepublicofSouthAfrica
Phone+27(0)119749745
Fax:+27(0)113922587
e‐mail:
AdamEquipment(S.E.ASIA)PTY
Ltd
2/71TacomaCircuit
CANNINGVALE6155
Perth
WesternAustralia
Phone:+61(0)864616236
Fax+61(0)894564462
e‐mail:
AdamEquipment(Wuhan)Co.Ltd.
ABuildingEastJianhua
PrivateIndustrialPark
ZhuanyangAvenue
WuhanEconomic&Technological
DevelopmentZone
430056Wuhan
P.R.China
Phone:+86(27)59420391
Fax+86(27)59420388
e‐mail:
©CopyrightbyAdamEquipmentCo.Allrightsreserved.Nopartofthispublicationmaybereprintedortranslatedin
anyformorbyanymeanswithoutthepriorpermissionofAdamEquipment.
Adam Equipment reserves the right to make changes to the technology, features, specificationsand design ofthe
equipmentwithoutnotice.
Allinformationcontainedwithinthispublicationistothebestofourknowledgetimely,completeandaccuratewhen
issued.However,wearenotresponsibleformisinterpretationswhichmayresultfromthereadingofthismaterial.
ThelatestversionofthispublicationcanbefoundonourWebsite.
www.adamequipment.com
/