8 - 2 SONOACE PICO Operation Manual
Probes
The probe is a device that sends and receives ultrasound for acquiring image data. It is also
called a Transducer or Scanhead.
The system limits patient contact temperature to 43 degrees Celsius, and acoustic power values
to their respective U.S. FDA limits. A power protection fuse circuit protects against over-current
conditions. If the power monitor protection circuit senses an over-current condition, then the
drive current to the probe is shut off immediately, preventing overheating of the probe surfaces
and limiting acoustic output. Validation of the power protection fuse circuit is performed under
normal system operation. For invasive probes, additional protections are designed to keep
patient contact surface temperature under 43 degrees Celsius in the event of a single fault
failure.
The ultrasonographic image scanner uses the image data received by the probe to display the
images of each organ on the screen. For best quality images, select the appropriate probe for
the each application.
Product Applications
C2-4ES General, Abdomen, Cardiac
C2-5ET General, Abdomen, Fetal Heart, Gynecology, OB, Renal
C3-7ED General, Abdomen, Fetal Heart, Gynecology, OB, Renal
C3-7EP General, Abdomen, Fetal Heart, Gynecology, OB, Renal
C4-7ED General, Abdomen, Pediatric, Appendix
C4-9ED General, Neonatal, Pediatric, Vascular
EC4-9ED General, Gynecology, OB, Urology
EC4-9ES General, Gynecology, OB, Urology
L4-7EL General, Breast, Small Part, Vascular, Musculoskeletal
L5-9EC General, Breast, Small Part, Vascular, Musculoskeletal
L5-9EE General, Breast, Small Part, Vascular, Musculoskeletal
L5-12/50EP General, Breast, Small Parts, Vascular, Musculoskeletal
HC2-5ED General, Abdomen, Fetal Heart, Gynecology, OB, Renal
HL5-9ED General, Breast, Small Part, Vascular, Musculoskeletal
HL5-12ED General, Breast, Small Part, Vascular, Musculoskeletal