Rosemount 340 Trace Moisture Analyzer-Rev R Owner's manual

Type
Owner's manual
Instruction Manual
081854-R
May 2002
http://www.processanalytic.com
Model 340
Trace Moisture Analyzer
Emerson Process Management
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434
http://www.processanalytic.com
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and
international standards. Because these instruments are sophisticated technical products, you
MUST properly install, use, and maintain them to ensure they continue to operate within their
normal specifications. The following instructions MUST be adhered to and integrated into your
safety program when installing, using, and maintaining Rosemount Analytical products. Failure to
follow the proper instructions may cause any one of the following situations to occur: Loss of life;
personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Rosemount Analytical representative
for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of
the product.
• Install your equipment as specified in the Installation Instructions of the appropriate
Instruction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and
maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by
Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe
operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result
in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified persons, to prevent electrical shock and personal
injury.
The information contained in this document is subject to change without notice.
Teflon® is a registered trademark of E.I. duPont de Nemours and Co., Inc.
Viton® is a registered trademark of E.I. duPont de Nemours and Co., Inc.
Freon12® is a registered trademark of E.I. duPont de Nemours and Co., Inc.
SNOOP
®
is a registered trademark of NUPRO Co.
Instruction Manual
081854-R
May 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i
Model 340
TABLE OF CONTENTS
PREFACE...........................................................................................................................................P-1
Definitions ...........................................................................................................................................P-1
Safety Summary .................................................................................................................................P-2
General Precautions For Handling And Storing High Pressure Gas Cylinders .................................P-4
Documentation....................................................................................................................................P-5
1-0 DESCRIPTION AND SPECIFICATIONS..............................................................................1-1
1-1 Overview................................................................................................................................1-1
1-2 Instrument Configurations .....................................................................................................1-1
1-3 Sample Gases .......................................................................................................................1-4
a. Suitable Sample Gases ..................................................................................................1-4
b. Unsuitable Sample Gases ..............................................................................................1-4
1-4 Specifications ........................................................................................................................1-5
a. Performance....................................................................................................................1-5
b. Alarm Option (panel mount analyzers only)....................................................................1-5
c. Sample ............................................................................................................................1-5
2-0 INSTALLATION ....................................................................................................................2-1
2-1 Facility Preparation................................................................................................................2-1
a. Outline and Mounting Dimensions ..................................................................................2-1
b. Interconnection Diagram (Explosion Proof Analyzers Only)...........................................2-1
c. Location...........................................................................................................................2-1
d. Utility Specifications ........................................................................................................2-1
2-2 Unpacking..............................................................................................................................2-2
2-3 Electrical Connections ...........................................................................................................2-2
a. Output Selection and Cable Connections for Potentometric Recorder ..........................2-5
b. Output Selection and Cable Connections for Current Recorder (AC Analyzers Only)...2-6
c. Alarm Output Connection and Alarm Function Selection (Optional, for Panel
Mounted Analyzers Only)................................................................................................2-7
d. Setting the Deadband .....................................................................................................2-8
e. Electrical Interconnection for Explosion Proof Analyzer .................................................2-8
f. Electrical Power Connection ...........................................................................................2-8
2-4 Sample Connections and Sample Handling recommendations ............................................2-9
2-5 Purge Connections and Requirements .................................................................................2-10
3-0 OPERATING CONTROLS AND INDICATORS....................................................................3-1
3-1 Range Selector Switch and Meter.........................................................................................3-1
3-2 Sample Flow Control Valve and Sample Flowmeter.............................................................3-1
3-3 Bypass Flow Control Valve and Bypass Flowmeter..............................................................3-1
3-4 Controls of Alarm Setpoint Accessory (Panel Mount Analyzers Only).................................3-1
Instruction Manual
081854-R
May 2002
ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model 340
4-0 STARTUP..............................................................................................................................4-1
4-1 Systems Utilizing Pressurized Gas Sample ..........................................................................4-1
a. Initial Dry-Down...............................................................................................................4-1
b. System Leak Check ........................................................................................................4-3
c. Instrument Calibration.....................................................................................................4-4
d. Computation of Sample Flowmeter Settings ..................................................................4-4
e. Pressure (Elevation) Corrections To Computed Flowmeter Values ...............................4-6
f. Experimental Calibration of Sample Flowmeter..............................................................4-6
g. Temperature Corrections (Portable Analyzers Only)......................................................4-7
4-2 systems utilizing the low pressure sampling accessory ........................................................4-8
a. Calibration Procedure for Sample Flowmeter.................................................................4-9
b. Operating Parameter Selection.......................................................................................4-10
Vacuum Reading.............................................................................................................4-10
Sample Flowmeter Setting..............................................................................................4-10
c. Setup for Normal Operation ............................................................................................4-11
5-0 OPERATION .........................................................................................................................5-1
5-1 Recommended Calibration Frequency..................................................................................5-1
5-2 Shutdown...............................................................................................................................5-1
6-0 THEORY................................................................................................................................6-1
6-1 Principle of Operation ............................................................................................................6-1
6-2 Flow System ..........................................................................................................................6-1
6-3 Electronic Circuitry.................................................................................................................6-3
a. Electrolytic Call and Switch Assembly (All Analyzers)....................................................6-3
b. Amplifier Circuit Board (All Analyzers)............................................................................6-3
c. Current Output Board (Optional for AC Analyzers Only) ................................................6-4
d. ±15 Volt Power Supply (AC Analyzers Only)..................................................................6-5
e. Alarm Setpoint Accessory and Universal Alarm Board (Optional, for Panel Mount
Analyzers Only)...............................................................................................................6-5
7-0 MAINTENANCE AND SERVICE ..........................................................................................7-1
7-1 Maintenance ..........................................................................................................................7-1
a. Care of the Electrolytic Cell.............................................................................................7-1
b. Replacing Electrolytic Cell ..............................................................................................7-1
c. Cleaning and Re-sensitizing Electrolytic Cell, Using PN 642257 Kit..............................7-1
7-2 Service...................................................................................................................................7-3
a. Subnormal or Zero Meter Reading .................................................................................7-3
b. Off-Scale Meter Reading ................................................................................................7-3
c. Erratic Meter Reading .....................................................................................................7-3
8-0 REPLACEMENT PARTS ......................................................................................................8-1
8-1 Matrix .....................................................................................................................................8-1
8-2 Circuit Board Replacement Policy .........................................................................................8-2
8-3 Selected Replacement Parts.................................................................................................8-2
a. Door Assembly – Panel Mount Instruments ...................................................................8-3
b. Chassis Assembly...........................................................................................................8-4
c. 193005 Portable AC Trace Moisture Analyzer ...............................................................8-6
d. 194772 Flowmeter Accessory.........................................................................................8-8
Instruction Manual
081854-R
May 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Model 340
9-0 RETURN OF MATERIAL ......................................................................................................9-1
9-1 Return Of Material .................................................................................................................9-1
9-2 Customer Service ..................................................................................................................9-1
9-3 Training..................................................................................................................................9-1
LIST OF ILLUSTRATIONS
Figure 1-1. Panel Mounted Trace Moisture Analyzer .............................................................. 1-2
Figure 1-2. Explosion-Proof Trace Moisture Analyzer ............................................................. 1-2
Figure 1-3. Portable Trace Moisture Analyzer and Flowmeter Accessory............................... 1-3
Figure 2-1. Interior of Panel Mount Analyzer ........................................................................... 2-2
Figure 2-2. Interior of Explosion Proof Analyzer ...................................................................... 2-3
Figure 2-3. Amplifier Board – Location of Selector Plugs ........................................................ 2-3
Figure 2-4. Current Output Board - Location of Selector Plugs ............................................... 2-4
Figure 2-5. Universal Alarm Board (PN 193913) - Locations of Alarm Selector Plugs............ 2-4
Figure 2-6. Universal Alarm Board (PN 620695) – Location of Alarm Select Jumpers and
Deadband Adjustment Potentiometer.................................................................... 2-5
Figure 2-7. Connections for Potentiometric Recorder with Intermediate Span........................ 2-6
Figure 2-8. Installation of Purge Kit........................................................................................ 2-10
Figure 3-1. Operating Controls of Model 340 Panel Mount Analyzer ...................................... 3-2
Figure 3-2. Operating Controls of Explosion Proof Analyzer ................................................... 3-3
Figure 3-3. Operating Controls of the Portable Analyzer and Flowmeter Accessory .............. 3-3
Figure 4-1. Interconnection of Low Pressure Analysis System................................................ 4-8
Figure 5-1. Ice Point vs. Parts-Per-Million H
2
O by Volume...................................................... 5-3
Figure 6-1. Schematic Diagram of Internal Flow System......................................................... 6-2
Figure 8-1. 194782 Door Assembly – Panel Mount Instrument............................................... 8-3
Figure 8-2. Chassis Assembly ................................................................................................. 8-5
Figure 8-3. 193005 Portable AC Trace Moisture Analyzer ...................................................... 8-7
Figure 8-4. 194772 Flowmeter Accessory ............................................................................... 8-8
LIST OF TABLES
Table 2-1. Accessory Devices for Sample Pressure Ranges............................................... 2-11
Table 4-1. Typical Settings for Sample Flowmeter ................................................................. 4-4
Table 4-2. Normal Barometric Pressures for Various Elevations ........................................... 4-5
Table 5-1. Conversion Factors for Water Vapor Concentrations............................................ 5-2
Instruction Manual
081854-R
May 2002
iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model 340
LIST OF DRAWINGS
193136 Installation Drawing - Model 340 Panel Mount
194745 Pictorial Wiring Diagram - Trace Moisture Analyzer, Portable AC
194749 Schematic Diagram - Portable DC
194754 Schematic Diagram - Trace Moisture Analyzer (sheet 1 only)
194757 Schematic Diagram - Trace Moisture Analyzer, Portable AC
194759 Interconnect Diagram - Trace Moisture Analyzer, Explosion Proof
194760 Schematic Diagram, Alarm Setpoint Assembly (sheet 3 only)
194761 Pictorial Wiring Diagram - Trace Moisture Analyzer, Panel Mount
194771 Board Assembly, Battery Pack
194775 Installation Drawing - Model 340 Portable
194789 Installation Drawing - Model 340 Explosion Proof
619710 Schematic Diagram - ±15V Power Supply
620696 Schematic Diagram - Universal Alarm
624265 Board Assembly, Amplifier (sheet 2 only)
654443 Schematic Diagram, Current Output Board
780213 Schematic Diagram, 230 VAC Operation
780796 Pictorial Wiring Diagram, 230 VAC Operation
Instruction Manual
081854-R
May 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Model 340
PREFACE
The purpose of this manual is to provide information concerning the components,
functions, installation and maintenance of the Model 340.
Some sections may describe equipment not used in your configuration. The user should
become thoroughly familiar with the operation of this module before operating it.
Some sections pertain to models that are no longer available. This information is included
for those instruments still in use.
Read and understand this instruction manual completely.
DEFINITIONS
The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout
this publication.
DANGER .
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial
property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure,
condition or statement.
Instruction Manual
081854-R
May 2002
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 340
SAFETY SUMMARY
If this equipment is used in a manner not specified in these instructions, protective systems
may be impaired.
AUTHORIZED PERSONNEL
To avoid explosion, loss of life, personal injury and damage to this equipment and on-site
property, all personnel authorized to install, operate and service the this equipment should be
thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE
INSTRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can
cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a properly grounded
three-wire source of power.
Alarm switching relay contacts wired to a separate power source must be disconnected before
servicing.
WARNING.
POSSIBLE EXPLOSION HAZARD
This analyzer is of a type capable of analysis of sample gases which may be flammable. If used for
analysis of such gases, the detection section of the analyzer must be either in an explosion proof
enclosure suitable for the hazard classification of the gas, or protected by a continuous dilution
purge system in accordance with Standard ANSI/NFPA-496-1986 (Chapter 8) or IEC Publication 79-
2-1983 (Section Three).
If explosive gases are introduced into this analyzer, the sample containment system must be care-
fully leak checked upon installation and before initial startup, during routine maintenance and any
time the integrity of the sample containment system is broken, to ensure that the system is in leak
proof condition. Leak check instructions are provided in Section 4-1b, page 4-3.
Internal leaks resulting from failure to observe these precautions could result in an explosion,
causing death, personal injury or property damage.
Instruction Manual
081854-R
May 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Model 340
WARNING.
HIGH PRESSURE GAS CYLINDERS
Fuel, air and calibration gas cylinders are under pressure. Mishandling of gas cylinders could re-
sult in death, injury or property damage. See General Precautions for Handling and Storing High
Pressure Cylinders, page P-4.
CAUTION
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product.
Use only factory documented components for repair.
.
Instruction Manual
081854-R
May 2002
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 340
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed
Gases" published in 1981
Compressed Gas Association
1235 Jefferson Davis Highway
Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of
weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the
shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or
bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a
suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is
attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125
°
F (52
°
C). A flame should
never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding,
precautions must be taken to prevent striking an arc against the cylinder.
Instruction Manual
081854-R
May 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Model 340
DOCUMENTATION
The following Model 340 instruction materials are available. Contact Customer Service Center or the local
representative to order.
081854 Instruction Manual (this document)
Instruction Manual
081854-R
May 2002
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 340
Instruction Manual
081854-R
May 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Model 340
SECTION 1
DESCRIPTION AND SPECIFICATIONS
1-1 OVERVIEW
The Model 340 Trace Moisture Analyzer
automatically and continuously measures water
vapor concentrations, up to a maximum of 1000
ppm, in a gaseous sample stream. The
determination is based on the simultaneous
absorption and electrolysis of water. The
instrument has a wide range of applications, in
monitoring many gases used in manufacturing
processes. (Suitable and unsuitable sample
gases are listed in Section 1-3, page 1-4.)
Permissible sample pressure range for the
standard instrument is 10 to 100 psig. Optional
sampling accessories permit monitoring gas
streams at atmospheric or sub-atmospheric
pressures.
The analyzer provides direct readout on a front
panel meter and a selectable output for an
accessory potentiometric recorder. With all AC
operated versions of the analyzer, a selectable
output for a current type recorder is obtainable
through use of an optional plug in the circuit
board.
1-2 INSTRUMENT CONFIGURATIONS
The Model 340 Analyzer is made in the
following configurations:
1. Panel Mounted Analyzer, Figure 1-1 on
page 1-2, with detector, electronic circuitry,
and operating controls housed in a single
purgeable case. Available with internal flow
system of either stainless steel (193000
Analyzer) or brass (193001 Analyzer).
2. The 193004 Explosion Proof Analyzer,
Figure 1-2 on page 1-2.
NOTE
The Model 340 TMA Explosion Proof (PN
193004) is no longer available.
Designed for use in the chemical,
petrochemical, and petroleum
industries, in applications where the
sample stream contains flammable
gases, or where explosive vapors may
be present at the installation site.
Control section is similar to that of the
Panel Mounted Analyzer. Detector
section is contained in an explosion
proof housing that meets the
requirements for installation under
hazardous conditions specified as Class
1, Group D, Division 1, in the National
Electrical Code. Flow system is of
stainless steel.
3. Portable Analyzer, Figure 1-3 on page
1-3.
NOTE
The Model 340 TMA Portable (PN 193005)
is no longer available.
Available for operation on either 115
VAC, 50/60 Hz (193005 Analyzer) or
±15 VDC from a self contained battery
pack (193006 Analyzer).
Except where specifically stated otherwise,
information in this manual applies to all
versions of the instrument.
Instruction Manual
081854-R
May 2002
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 340
Note: Illustration applicable to part numbers 193000 and 193001 Analyzers
Figure 1-1. Panel Mounted Trace Moisture Analyzer
Note: This instrument is no longer available – Consult Factory
Figure 1-2. Explosion-Proof Trace Moisture Analyzer
Alarm Setpoint
Accessory
Detector Section Control Section
Alarm Setpoint
Accessory
Instruction Manual
081854-R
May 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Model 340
Notes: Illustration applicable to part numbers 193005 and 193006 Analyzers
This instrument is no longer available – Consult Factory
Figure 1-3. Portable Trace Moisture Analyzer and Flowmeter Accessory
Instruction Manual
081854-R
May 2002
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 340
1-3 SAMPLE GASES
WARNING
POSSIBLE EXPLOSION HAZARD
This analyzer is of the type capable of
analysis of sample gases which may be
flammable. If used for analysis of such
gases, the detection section of the ana-
lyzer must be either in an explosion proof
enclosure suitable for the hazardous clas-
sification of the gas , or, protected by a
continuous dilution purge system in ac-
cordance with Standard ANSI /N FPA-496
-1986 (Chapter 8) or IEC Publication
79-2-1983 (Section Three).
If explosive gases are introduced into this
analyzer, the sample containment system
must be carefully leak checked upon in-
stallation and before initial startup, during
routine maintenance and any time the in-
tegrity of the sample containment system
is broken, to ensure that the system is in
leak proof condition. Leak check instruc-
tions are provided in Section 4-1b on page
4-3.
Internal leaks resulting from failure to ob-
serve these precautions could result in an
explosion causing death, personal injury
or property damage.
Determination of whether a sample stream of
a particular composition is suitable for
monitoring depends on its compatibility with
the construction materials in a) the detector
cell, and b) the instrument flow system. In all
instruments, the detector cell utilizes a thin
film of phosphorous pentoxide (P
2
O
6
) on
rhodium electrodes. Depending on the
intended application of a given instrument
version, its internal flow system is constructed
of either stainless steel (for corrosion
resistance) or brass (for non-corrosive sample
gases only).
a. Suitable Sample Gases
Elemental Gases
Argon, Helium, Neon, Nitrogen, Oxygen,
Hydrogen.
Inorganic Gaseous Mixtures and
Compounds
Air, Carbon Dioxide, Carbon Monoxide,
Sulphur Dioxide, Sulphur Hexafluoride.
Organic Gaseous Compounds
Butane, Ethane, Freon 12 , Methane,
Propane, Halogenated Hydrocarbons.
b. Unsuitable Sample Gases
Gases that react with P206 to produce
additional water
Example: alcohols, HF.
Gases that react with construction
materials of the instrument
Gases that react with P
2
O
6
to alter
required absorption characteristics of
the P
2
O
6
film
Examples: ammonia, amines.
Gases that polymerize to form a solid
or liquid phase (they gradually
desensitize the detector cell by coating
or clogging)
Example: Unsaturated hydrocarbons -
alkynes, alkadienes and alkenes.
Gases that contain particulate solid or
liquid materials such as dust and dirt
found in furnace atmosphere gases.
These must be avoided or filtered out
upstream; oil mist or dust from some
types of dryers can clog the detector cell
or desensitize the P
2
O
6
film by forming a
layer over the film.
Instruction Manual
081854-R
May 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Model 340
1-4 SPECIFICATIONS
a. Performance
Ranges .......................................... 0 to 10, 0 to 50, 0 to 100, 0 to 500, 0 to 1000 ppm H
2
O by volume
Accuracy........................................ ±5% of fullscale (not applicable to 0 to 10 ppm range or to
hydrogen or oxygen sample stream containing less than 25 ppm
H
2O
Sensitivity ...................................... Less than 1 ppm
Bypass Flow .................................. Adjustable of 0 to 2 cubic feet per hour (940 cc/min.) is standard in
non-portable instruments, and is obtainable for portable
instruments through use of optional flowmeter accessory
Ambient Temperature.................... 0°F to 120°F (-18°C to 49°C)
Recorder Potentiometric Output.... All analyzers provide selectable output of 0 to 10 mV, 0 to 100 mV,
0 to 1 V, or 0 to 5 V
Recorder Current Output Option (for AC power analyzers only)
Plug-in circuit board provides 4 selectable outputs:
Output (mA) Maximum Permissible Load (ohms)
0 to 5 800
1 to 5 8000
4 to 20 2000
10 to 50 700
b. Alarm Option (panel mount analyzers only)
Setpoint Accuracy ......................... ±1/2 of fullscale, or 25 mV
Repeatability.................................. 1% of fullscale
Setpoint Range.............................. 0 to 100% or 0 to 5 VDC, displayed on front panel meter
Hysteresis...................................... 2% of fullscale is standard, adjustable by changing resistor on
circuit board
Output............................................ (1) isolated (2) 190 VDC or VAC maximum, (3) 1.5 amperes AC or
DC maximum
c. Sample
Sample Flow Rate ......................... 100 cc/minute
Sample Pressure........................... Standard range 10 to 100 psig (69 to 690 kPa).
Low Pressure Sampling Accessory provides range of 10 inches
mercury vacuum to +10 psig.
Sample Temperature..................... 32°F to 120°F (0°C to 80°C)
Sample Inlet/Outlet Connections... 1/8 inch bulkhead tube fittings (Double ferrule compression type)
Instruction Manual
081854-R
May 2002
1-6 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 340
Instruction Manual
081854-R
May 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Model 340
SECTION 2
INSTALLATION
Sections 2-1a through 2-1d provide information that
may be required prior to installation.
2-1 FACILITY PREPARATION
a. Outline and Mounting Dimensions
For significant dimensions of the
instrument, refer to the appropriate
Drawing at the back of the manual.
b. Interconnection Diagram (Explosion
Proof Analyzers Only)
Drawing 194759 shows electrical
interconnection for the 193004 Explosion
Proof Analyzer.
NOTE
Separate conduits should be used for
the power cable and the interconnec-
tion cable.
c. Location
Ambient temperature range for all
analyzers is 0°F to 120°F (-18°C to 49°C).
Additional requirements, specific to the
various analyzer configurations, are given
in the following:
193000 and 193001 Panel Mounted
Analyzers
Install in a clean area, not subject to
excessive vibration or extreme
temperature variations. Preferably, the
instrument should be mounted near the
sample stream, to minimize transport
time.
193004 Explosion Proof Analyzer
Detector Section: Criteria for installation
site are proximity to sample point,
protection from environment, and
accessibility for servicing. Protect the unit
adequately against shock and extreme
vibration.
Control Section: Principal criteria for the
installation site is that it must be outside
the hazardous area. Hazardous locations
are defined in Article 500 of the National
Electrical Code. An additional
consideration is convenience in taking
readings and servicing the unit.
d. Utility Specifications
Electrical power requirements are listed in
the following table:
193000 and 193001 Panel Mounted Analyzers
193004 Explosion Proof Analyzer
107 to 127 VAC
50/60 Hz, 80 watts
193005 Portable Analyzer 107 to 127 VAC, 50/60 Hz, 20 watts
193006 Portable Analyzer ± 15 VDC from self contained battery pack
Instruction Manual
081854-R
May 2002
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 340
±15V Power Supply
Universal Alarm Board
Amplifier Board
Cell Holder
Bypass Flowmeter
Sample Flowmeter
TB1 TB2
Partial view of inside door
Alarm Setpoint Accessory
J5 (shorting plug shown)
2-2 UNPACKING
Examine the shipping carton and contents
carefully for any signs of damage. Save the
carton and packing material until the analyzer
is operational. If carton or contents damage
(either external or concealed) is discovered,
notify the carrier immediately.
2-3 ELECTRICAL CONNECTIONS
Depending on the particular options used,
electrical setup may entail insertion of various
selector plugs into appropriate positions in the
associated circuit boards. Locations of circuit
boards and other components within the
several analyzer configurations are shown in
Figure 2-1 below and Figure 2-2 on page 2-3.
Locations of selector plugs on the individual
board are shown in Figure 2-3 on page 2-3,
Figure 2-4 on page 2-4, Figure 2-5 on page 2-
4, and Figure 2-6 on page 2-5. Make
electrical connections in the following
sequence:
1. If a recorder is to be used, select the
particular output required and make the
appropriate cable connections as
explained in Section 2-3a on page 2-5
(potentiometric recorder) or Section 2-3b
on page 2-6 (current recorder). All
analyzers provide potentiometric output.
Current output is obtainable from AC
operated instruments only, through use of
the optional current output circuit board.
2. If an alarm system is to be used, select
the desired function and connect the
output as explained in Section 2-3c on
page 2-7. Alarm output is obtainable from
panel mounted instruments only, through
use of the Alarm Setpoint Accessory and
Universal Alarm Board in combination.
3. With Explosion Proof Analyzer,
interconnect detector and control modules
per Section 2-3e on page 2-8.
4. Supply electrical power to analyzer per
Section 2-3f on page 2-8.
Figure 2-1. Interior of Panel Mount Analyzer
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90