Roche cobas z 480 User manual

Type
User manual

This manual is also suitable for

1
cobas
®
4800 System
User Assistance version 2.0
Software version 2.2
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
2
Publication information
Edition notice This publication is intended for operators of the
cobas x 480 instrument and/or the cobas z 480
analyzer.
Every effort has been made to ensure that all the
information contained in this publication is correct at the
time of publishing. However, the manufacturer of this
product may need to update the publication information
as output of product surveillance activities, leading to a
new version of this publication.
Where to find information The User Assistance contains all information about the
product, including the following:
• Routine operation
• Maintenance
• Safety
• Troubleshooting information
• A software reference
• Configuration information
• Background information
The Safety chapter contains important safety
information. You must read the Safety chapter before
operating the system.
Publication
version
System software
version
Revision date Change description
1.0 2.2 May 2015 First version. New manual structure and design. Software
update from 2.1 to 2.2
2.0 2.2 April 2017 o Added RoHS directive 2011/65/EU to Approvals section.
o Compliance marks updated to include ‘CSA’ and
‘Laboratory Equipment’.
o Symbols used on product updated to include ‘Cont.’,
‘Content’ and ‘A/C only’.
o Transferred general safety messages to Safety chapter and
revised safety messages throughout manual.
o In chapter Introduction to the user interface, added About
the sample editor section and About the results work area
section.
o Revised cleaning procedure of the instrument.
o Updated images of thermal cycler door (analyzer) in
Exchanging fuses section.
o Added new Operation chapter with available workflows.
o Added information on new integrated heater/shaker unit.
o Revised description for sealing the microwell plate.
o Changed name of deliverable from cobas
®
4800 System
System Manual to cobas
®
4800 System User Assistance.
o Added new Safety chapter. Information of previous Safety
manual is now integrated in this manual as Safety chapter.
y Revision history
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
3
The test-specific Instructions for Use provides
additional information such as instructions on storage
and handling of reagents, samples, and controls. The
Instructions for Use also provides information on manual
PCR setup of the microwell plate.
Before using the system, it is important that the operator
reads these publications.
!
General attention
To avoid serious or fatal injury, ensure that you are
familiar with the instructions and safety information
before you use the instrument or analyzer.
r Pay particular attention to all safety precautions.
r Always follow the instructions in this publication.
r Do not use the instrument or analyzer in a way that is
not described in this publication.
r Store all publications in a safe and easily retrievable
place.
Privacy notice
When you use User Assistance on line, viewing events
(topics viewed and searches performed) and IP addresses
are logged.
The data collected is for Roche internal use only and is
never forwarded to third parties. It is anonymized, and
after one year it is automatically deleted.
Viewing events are analyzed to improve User Assistance
content and search functionality. IP addresses are used to
classify regional behavior.
You can configure how data is collected.
u Configuring User Assistance data collection (26)
Training Do not carry out operation tasks or maintenance actions
unless you have received training from Roche
Diagnostics. Leave tasks that are not described in the
user documentation to trained Roche Service
representatives.
Images The screenshots and hardware images in this publication
have been added exclusively for illustration purposes.
Configurable and variable data in screenshots, such as
tests, results, or path names visible therein must not be
used for laboratory purposes.
Warranty Any customer modification to the system renders the
warranty or service agreement null and void.
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
4
For conditions of warranty, contact your local sales
representative or refer to your warranty contract partner.
Always leave software updates to a Roche Service
representative or perform such updates with their
assistance.
Copyright © 2009-2017 F. Hoffmann-La Roche Ltd. All rights
reserved.
License information
cobas
®
4800 software is protected by contract law,
copyright law, and international treaties. cobas
®
4800
software contains a user license between F. Hoffmann-La
Roche Ltd. and a license holder, and only authorized
users may access the software and use it. Unauthorized
use and distribution may result in civil and criminal
penalties.
Trademarks The following trademarks are acknowledged:
COBAS, COBAS X, COBAS Z, and LIFE NEEDS ANSWERS
are trademarks of Roche.
All other trademarks are the property of their respective
owners.
Feedback Every effort has been made to ensure that this publication
fulfills the intended use. All feedback on any aspect of
this publication is welcome and is considered during
updates. Contact your Roche representative, should you
have any such feedback.
Approvals The instrument and the analyzer meet the protection
requirements laid down in:
Directive 98/79/EC of the European Parliament and of the
Council of 27 October 1998 on in vitro diagnostic medical
devices.
Directive 2011/65/EU of the European Parliament and of
the Council of 8 June 2011 on the restriction of the use of
certain hazardous substances in electrical and electronic
equipment.
The analyzer is manufactured and tested according to the
following international standards:
• IEC 61010-1
• IEC 61010-2-101
• UL 61010-1
• CAN/CSA C22.2 No. 61010-1
Compliance with the applicable directive(s) is provided by
means of the Declaration of Conformity.
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
5
The following marks demonstrate compliance:
Contact addresses
Complies with the IVD directive 98/79/EC on in vitro
diagnostic medical devices.
For in vitro diagnostic use.
Issued by Underwriters Laboratories, Inc. (UL) for
Canada and the US.
Issued by CSA Group for Canada and the US.
'Laboratory Equipment' is the product identifier as
shown on the name plate.
CUS
cobas z 480 analyzer and cobas
®
4800 software
Roche Molecular Systems Inc.
1080 US Highway 202 South
Branchburg, NJ 08876
USA
Made in Switzerland
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
cobas x 480 instrument
HAMILTON Bonaduz AG
Via Crusch 8
CH-7402 Bonaduz
Switzerland
Made in Switzerland
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
6
Distributed in the United States by:
Roche Diagnostics
9115 Hague Road
Indianapolis
Indiana
USA
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
7
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
8
Table of contents 9
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
Table of contents
Publication information 2
Contact addresses 5
Table of contents 9
Preface 11
Intended use 12
Symbols and abbreviations 12
What is new in publication version 2.0 15
What is new in publication version 1.0 16
User Assistance
1 User Assistance
Using User Assistance 23
Safety
2 General safety information
Introduction 31
Safety classifications 32
Safety precautions 33
Warning messages 36
Caution messages 50
Notices 53
Safety labels on the system 55
Safety information for disposal 61
System description
3 Overview
System overview 67
4 Hardware
Instrument overview 77
Analyzer overview 98
Consumables overview 103
Control unit overview 108
About the hand-held barcode reader 109
Barcode configuration specifications 110
5 Introduction to the user interface
About the screen design 113
About the sample editor 123
About the results work area 130
About the database 145
About the audit trail 147
Operation
6 Operation
About workflows 153
Full workflow overview 155
Recovery workflow overview 190
PCR Only workflow overview 202
Configuration
7 Configuration
About configuration 225
About managing users 230
Exporting support data 238
About result counters 239
Maintenance
8 Maintenance
Instrument maintenance 245
Analyzer maintenance 276
Troubleshooting
9 Troubleshooting
Troubleshooting information in the user
interface 293
About the messages work area 295
Reviewing and confirming new alarms 297
Deleting alarms and messages 298
Troubleshooting software 299
Troubleshooting hardware 302
List of error messages and system flags
10 Error messages and system flags
Error messages 315
List of error messages 316
System flags 350
List of system flags 351
Glossary
11 Glossary
Index
Index 363
10 Table of contents
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
11
Preface
The system integrates fully automated total nucleic acid
isolation directly from primary and secondary tubes,
automated PCR setup, and real time PCR.
This manual covers the complete system:
• cobas
®
4800 software
• cobas x 480 instrument used for sample preparation
• cobas z 480 analyzer used for amplification and
detection using real time PCR
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
12
Intended use
System
(instrument and analyzer)
The cobas
®
4800 System is intended to be used as a
diagnostic or screening system providing sample
preparation, amplification and detection of specific
targets from human samples. It integrates fully automated
total nucleic acid isolation directly from primary and
secondary tubes, automated PCR setup, and real time
PCR.
Analyzer The cobas z 480 analyzer including dedicated software
for IVD use is a plate-based automated amplification and
detection system for nucleic acids. The analyzer is
intended to be used as an in-vitro diagnostic or screening
system providing real time PCR amplification and
detection for human samples within the system.
The cobas x 480 instrument and cobas z 480 analyzer
are to be used by laboratory professionals trained in
laboratory techniques and by instruction on the use of the
system and analyzer.
Symbols and abbreviations
Product names Except where the context clearly indicates otherwise, the
following product names and abbreviations are used:
Product name Descriptor
cobas
®
4800 software
software
cobas x 480 instrument instrument
cobas z 480 analyzer analyzer
cobas
®
4800 System
system
cobas
®
4800 System Sample
Preparation Kit
sample preparation kit
Extraction plate 1.6ml deepwell plate
Extraction plate 2.0ml deepwell plate
AD-plate 0.3ml microwell plate
Reagent reservoir 50ml 50 mL reagent reservoir
Reagent reservoir 200ml 200 mL reagent reservoir
Core Tips with Filter 1ml pipette tip
PreservCyt
®
carrier
collection medium carrier
(test-specific)
y Product names
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
13
Symbols used in the publication
Symbols used on product
Symbol Explanation
o List item
u Related topics containing further information.
q Tip. Extra information on correct use or useful
hints.
r Start of a task
I Extra information within a task
f Result of a user action within a task
c Frequency of a task
n Duration of a task
d Materials that are required for a task
j Prerequisites of a task
u Topic. Used in cross-references to topics.
p Task. Used in cross-references to tasks.
w Figure. Used in figure titles and cross-
references to figures.
y Table. Used in table titles and cross-references
to tables.
z Equation. Used in cross-references to
equations.
k Code example. Used in code titles and cross-
references to codes.
Print. Used for printable documents.
Safety alert
Electrical and electronic equipment marked
with this symbol are covered by the European
Directive 2002/96/EC on waste electrical and
electronic equipment (WEEE).
The symbol denotes that the equipment must
not be disposed of in the municipal waste
system.
y Symbols used in the publication
Symbol Explanation
Global Trade Item Number.
y Symbols used on product
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
14
Abbreviations The following abbreviations are used:
Quantity contained in the package.
Quantity contained in the package.
Indicates that the equipment is suitable for
alternating current only; to identify relevant
terminals.
Symbol Explanation
y Symbols used on product
Abbreviation Definition
AD Amplification and detection
ANSI American National Standards Institute
cc Cubic centimeter
cLLD Capacitive liquid level detection
CSA Canadian Standards Association
dBA Decibel weighted against the A-frequency
response curve. This curve approximates
the audible range of the human ear.
DWP Deepwell plate
EC European Community
e.g. Exempli gratia – for example
EN European standard
i.e. Id est – that is to say
IEC International Electrical Commission
iSWAP Internal Swivel Arm Plate handler
IVD In-vitro diagnostic
LED Light emitting diode
LIS Laboratory information system
MWP Microwell plate
n/a Not applicable
PCR Polymerase chain reaction
UL Underwriters Laboratories Inc.
UPS Uninterruptible power supply
WEEE Waste Electrical and Electronic Equipment
XML Extensible Markup Language
y Abbreviations
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
15
What is new in publication version 2.0
Short description of all major changes in this version.
User Assistance chapter Added new chapter on using the User Assistance.
u Using User Assistance (23)
Safety information chapter Added new general safety information chapter.
u Safety classifications (32)
Integrated heater/shaker unit Added information on new integrated heater/shaker unit
in the instrument.
u Heater/shaker unit (94)
Sample editor section Added section on sample editor for handling work order
files.
u About the sample editor (123)
Results work area section Added section on the results work area.
u About the results work area (130)
New Operation chapter Added new Operation chapter with possible workflows
for performing a run.
u About workflows (153)
Cleaning procedure The cleaning solution for the instrument has been revised.
u About cleaning the instrument (247)
New decontamination section Added new section on instrument decontamination.
u Decontamination (264)
Thermal cycler door The thermal cycler door of the analyzer has been
modified. The door is now fitted with a metal hinge
covering the secondary fuses. To replace a secondary
fuse, you must now open the thermal cycler door.
u To exchange a secondary fuse (285)
Error messages and flags chapter Added new chapter on error messages and system flags.
u Error messages (315)
u System flags (350)
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
16
What is new in publication version 1.0
Short description of all major changes in publication
version 1.0.
Solid waste door New hardware. The solid waste door allows you to access
and replace the small solid waste bag or waste chute
easily.
u About the waste station setup (84)
Enhancement to pipette tip pickup To minimize generating multiple partial tip racks after a
run, the system uses pipette tips from partial tip racks
before pipette tips from full tip racks.
u About the pipette tips (104)
Supported character formats for barcodes / is a supported barcode character.
u Barcode configuration specifications (110)
Sorting results function Results can be sorted by quantity.
u About sorting by column (117)
Result filter by sample Filtering options for controls. You can now filter results
based on sample types.
u Creating result filters (141)
Sorting of reports The laboratory manager can configure how the results
are sorted in the PDF results report.
u Reporting overview (226)
Button name The Extract support data button is renamed to Export
support data button.
u Exporting support data (238)
Weekly maintenance Additional alignment check during weekly maintenance
to ensure that instrument is within alignment
(approximately 7 extra minutes for weekly maintenance).
u Weekly maintenance (259)
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
17
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
18
Roche Diagnostics
cobas® 4800 System · Software version 2.2 · User Assistance · version 2.0
User Assistance
1 User Assistance ...................................................................................................... 21
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272
  • Page 273 273
  • Page 274 274
  • Page 275 275
  • Page 276 276
  • Page 277 277
  • Page 278 278
  • Page 279 279
  • Page 280 280
  • Page 281 281
  • Page 282 282
  • Page 283 283
  • Page 284 284
  • Page 285 285
  • Page 286 286
  • Page 287 287
  • Page 288 288
  • Page 289 289
  • Page 290 290
  • Page 291 291
  • Page 292 292
  • Page 293 293
  • Page 294 294
  • Page 295 295
  • Page 296 296
  • Page 297 297
  • Page 298 298
  • Page 299 299
  • Page 300 300
  • Page 301 301
  • Page 302 302
  • Page 303 303
  • Page 304 304
  • Page 305 305
  • Page 306 306
  • Page 307 307
  • Page 308 308
  • Page 309 309
  • Page 310 310
  • Page 311 311
  • Page 312 312
  • Page 313 313
  • Page 314 314
  • Page 315 315
  • Page 316 316
  • Page 317 317
  • Page 318 318
  • Page 319 319
  • Page 320 320
  • Page 321 321
  • Page 322 322
  • Page 323 323
  • Page 324 324
  • Page 325 325
  • Page 326 326
  • Page 327 327
  • Page 328 328
  • Page 329 329
  • Page 330 330
  • Page 331 331
  • Page 332 332
  • Page 333 333
  • Page 334 334
  • Page 335 335
  • Page 336 336
  • Page 337 337
  • Page 338 338
  • Page 339 339
  • Page 340 340
  • Page 341 341
  • Page 342 342
  • Page 343 343
  • Page 344 344
  • Page 345 345
  • Page 346 346
  • Page 347 347
  • Page 348 348
  • Page 349 349
  • Page 350 350
  • Page 351 351
  • Page 352 352
  • Page 353 353
  • Page 354 354
  • Page 355 355
  • Page 356 356
  • Page 357 357
  • Page 358 358
  • Page 359 359
  • Page 360 360
  • Page 361 361
  • Page 362 362
  • Page 363 363
  • Page 364 364
  • Page 365 365
  • Page 366 366
  • Page 367 367
  • Page 368 368

Roche cobas z 480 User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI